The optimization of signal detectability in digital fluoroscopy.

Abstract:

:An optimization analysis of digital fluoroscopy has been carried out using (signal/noise)2/dose as a figure of merit. Three signal detection tasks have been considered: (i) small iodine-containing blood vessels; (ii) regions of fat; and (iii) microcalcifications. The effects of x-ray beam filtration and kilovoltage on the figure of merit have been parameterized and the associated x-ray tube power loadings have been defined. Results are presented in a fashion that facilitates the choice of parameters (amount of aluminum filtration, kilovoltage, and x-ray tube power loading) such that an acceptable balance can be achieved between signal detection sensitivity, patient dose, and x-ray tube power loading.

journal_name

Med Phys

journal_title

Medical physics

authors

Zamenhof RG

doi

10.1118/1.595130

subject

Has Abstract

pub_date

1982-09-01 00:00:00

pages

688-94

issue

5

eissn

0094-2405

issn

2473-4209

journal_volume

9

pub_type

杂志文章
  • SU-E-T-535: Proton Dose Calculations in Homogeneous Media.

    abstract:PURPOSE:To develop a pencil beam dose calculation algorithm for scanned proton beams that improves modeling of scatter events. METHODS:Our pencil beam algorithm (PBA) was developed for calculating dose from monoenergetic, parallel proton beams in homogeneous media. Fermi-Eyges theory was implemented for pencil beam tr...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735624

    authors: Chapman J,Fontenot J,Newhauser W,Hogstrom K

    更新日期:2012-06-01 00:00:00

  • Comparative assessment of linear least-squares, nonlinear least-squares, and Patlak graphical method for regional and local quantitative tracer kinetic modeling in cerebral dynamic 18 F-FDG PET.

    abstract:PURPOSE:Dynamic 18 F-FDG PET allows quantitative estimation of cerebral glucose metabolism both at the regional and local (voxel) level. Although sensitive to noise and highly computationally expensive, nonlinear least-squares (NLS) optimization stands as the reference approach for the estimation of the kinetic model p...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13366

    authors: Ben Bouallègue F,Vauchot F,Mariano-Goulart D

    更新日期:2019-03-01 00:00:00

  • Comparison of three strategies in management of independent movement of the prostate and pelvic lymph nodes.

    abstract:PURPOSE:Concurrent irradiation of the prostate and pelvic lymph nodes is technically challenging due to treating one moving target and one immobile target. The purposes of this article are to propose a new management strategy and to compare this strategy to the conventional isocenter shift method and the previously pro...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3480505

    authors: Xia P,Qi P,Hwang A,Kinsey E,Pouliot J,Roach M 3rd

    更新日期:2010-09-01 00:00:00

  • Investigation of gamma knife image registration errors resulting from misalignment between the patient and the imaging axis.

    abstract::The ability of Leksell GammaPlan to perform stereotactic space localizations with image sets where there is misalignment of the patient's head (stereotactic frame and fiducial apparatus) relative to the computed tomography (CT) scanner coordinate system was studied. Misalignment is sometimes necessary for patient comf...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2179751

    authors: Cernica G,Wang Z,Malhotra H,de Boer S,Podgorsak MB

    更新日期:2006-04-01 00:00:00

  • Measurement of surface dose in an MR-Linac with optically stimulated luminescence dosimeters for IMRT beam geometries.

    abstract:PURPOSE:This study aims to measure the surface dose on an anthropomorphic phantom for intensity-modulated radiation therapy (IMRT) plans treated in a 1.5 T magnetic resonance (MR)-Linac. Previous studies have used Monte Carlo programs to simulate surface dose and have recognized high surface dose as a potential limitin...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14185

    authors: Lim-Reinders S,Keller BM,Sahgal A,Chugh B,Kim A

    更新日期:2020-07-01 00:00:00

  • SU-E-T-07: Edema Induced Changes in Tumor Cell Survival Fraction and Tumor Control Probability in 131Cs Permanent Prostate Implant Patients.

    abstract:PURPOSE:The purpose of this study is to estimate the effect of edema, developed during implant procedure, on tumor cell surviving fraction(SF) and tumor control probability(TCP) in the patients of prostate cancer who underwent 131 Cs permanent seed implants. METHODS:The impact of edema on SF and TCP, was calculated us...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735061

    authors: Kehwar TS,Huq MS,Smith RP

    更新日期:2012-06-01 00:00:00

  • The application of sparse arrays in high frequency transcranial focused ultrasound therapy: a simulation study.

    abstract:PURPOSE:Transcranial focused ultrasound is an emerging therapeutic modality that can be used to perform noninvasive neurosurgical procedures. The current clinical transcranial phased array operates at 650 kHz, however the development of a higher frequency array would enable more precision, while reducing the risk of st...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4829510

    authors: Pajek D,Hynynen K

    更新日期:2013-12-01 00:00:00

  • SU-F-BRCD-09: Total Variation (TV) Based Fast Convergent Iterative CBCT Reconstruction with GPU Acceleration.

    abstract:PURPOSE:To improve image quality and reduce imaging dose in CBCT for radiation therapy applications and to realize near real-time image reconstruction based on use of a fast convergence iterative algorithm and acceleration by multi-GPUs. METHODS:An iterative image reconstruction that sought to minimize a weighted leas...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735747

    authors: Xu Q,Yang D,Tan J,Anastasio M

    更新日期:2012-06-01 00:00:00

  • Potential of in vivo MRI-based nonlinear finite-element analysis for the assessment of trabecular bone post-yield properties.

    abstract:PURPOSE:Bone strength is the key factor impacting fracture risk. Assessment of bone strength from high-resolution (HR) images have largely relied on linear micro-finite element analysis (μFEA) even though failure always occurs beyond the yield point, which is outside the linear regime. Nonlinear μFEA may therefore be m...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4802085

    authors: Zhang N,Magland JF,Rajapakse CS,Bhagat YA,Wehrli FW

    更新日期:2013-05-01 00:00:00

  • The effect of pitch in multislice spiral/helical CT.

    abstract::The purpose of this study is to understand the effect of pitch on raw data interpolation in multislice spiral/helical computed tomography (CT) and provide guidelines for scanner design and protocol optimization. Multislice spiral CT is mainly characterized by the three parameters: the number of detector arrays, the de...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598804

    authors: Wang G,Vannier MW

    更新日期:1999-12-01 00:00:00

  • Transrectal electrical impedance tomography of the prostate: spatially coregistered pathological findings for prostate cancer detection.

    abstract:PURPOSE:Prostate cancer ranks as one of the most common malignancies and currently represents the second leading cancer-specific cause of death in men. The current use of single modality transrectal ultrasound (TRUS) for biopsy guidance has a limited sensitivity and specificity for accurately identifying cancerous lesi...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4803498

    authors: Wan Y,Borsic A,Heaney J,Seigne J,Schned A,Baker M,Wason S,Hartov A,Halter R

    更新日期:2013-06-01 00:00:00

  • Traceable micro-CT scaling accuracy phantom for applications requiring exact measurement of distances or volumes.

    abstract:PURPOSE:Volumetric x-ray microcomputed tomography (CT) can be employed in a variety of quantitative research applications such as image-guided interventions or characterization of medical devices. To ensure the highest geometric fidelity of images for these applications, a phantom and image processing algorithm have be...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4752083

    authors: Waring CS,Bax JS,Samarabandu A,Holdsworth DW,Fenster A,Lacefield JC

    更新日期:2012-10-01 00:00:00

  • A scanning system for chest radiography with regional exposure control: theoretical considerations.

    abstract::Conventional chest radiography is limited by the presence of scattered radiation and the small useful exposure range of radiographic film. A computer-assisted scanning system to minimize these two effects is outlined. The system uses a small beam of radiation swept over the patient's chest in a raster pattern to expos...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595327

    authors: Plewes DB

    更新日期:1983-09-01 00:00:00

  • Characterizing the modulation transfer function (MTF) of proton/carbon radiography using Monte Carlo simulations.

    abstract:PURPOSE:To characterize the modulation transfer function (MTF) of proton/carbon radiography using Monte Carlo simulations. To assess the spatial resolution of proton/carbon radiographic imaging. METHODS:A phantom was specifically modeled with inserts composed of two materials with three different densities of bone and...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4819816

    authors: Seco J,Oumano M,Depauw N,Dias MF,Teixeira RP,Spadea MF

    更新日期:2013-09-01 00:00:00

  • SU-E-J-186: CT Textures Can be Predictive for Tumor Shrinkage.

    abstract:PURPOSE:To determine if NSCLC GTV CT textures can be predictive for tumor shrinkage after proton therapy with concurrent chemotherapy. METHODS:Simulation and weekly 4DCTs were obtained from 25 patients with locally advanced NSCLC treated by proton therapy at a 74 Gy dose level with concurrent chemotherapy. In-house de...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735027

    authors: Hunter L,Chen Y,Zhang L,Liao Z,Dong L,Court L

    更新日期:2012-06-01 00:00:00

  • Modeling the dosimetry of organ-at-risk in head and neck IMRT planning: an intertechnique and interinstitutional study.

    abstract:PURPOSE:To build a statistical model to quantitatively correlate the anatomic features of structures and the corresponding dose-volume histogram (DVH) of head and neck (HN) Tomotherapy (Tomo) plans. To study if the model built upon one intensity modulated radiation therapy (IMRT) technique (such as conventional Linac) ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4828788

    authors: Lian J,Yuan L,Ge Y,Chera BS,Yoo DP,Chang S,Yin F,Wu QJ

    更新日期:2013-12-01 00:00:00

  • Deadtime correction for two multihead Anger cameras in 131I dual-energy-window-acquisition mode.

    abstract::Two side-by-side energy windows, one at the photopeak and one at lower energy, are sometimes employed in quantitative SPECT studies. We measured the count-rate losses at moderately high activities of 131I for two multihead Anger cameras in such a dual-window-acquisition mode by imaging a decaying source composed of tw...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598162

    authors: Koral KF,Zasadny KR,Ackermann RJ,Ficaro EP

    更新日期:1998-01-01 00:00:00

  • Inter- and intra-software reproducibility of computed tomography lung density measurements.

    abstract:PURPOSE:Multiple commercial, open-source, and academic software tools exist for objective quantification of lung density in computed tomography (CT) images. The purpose of this study was to evaluate the intersoftware reproducibility of CT lung density measurements. METHODS:Computed tomography images from 50 participan...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14130

    authors: Kirby M,Hatt C,Obuchowski N,Humphries SM,Sieren J,Lynch DA,Fain SB,QIBA Lung Density Committee.

    更新日期:2020-07-01 00:00:00

  • On-board SPECT for localizing functional targets: a simulation study.

    abstract::Single photon emission computed tomography (SPECT) was investigated for imaging on-board radiation therapy machines in order to localize functional and molecular targets. A computer-simulated female NCAT phantom was positioned supine on a flat-top treatment couch. Twenty tumor locations were defined in the upper torso...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3113902

    authors: Roper J,Bowsher J,Yin FF

    更新日期:2009-05-01 00:00:00

  • A novel approach to evaluate spatial resolution of MRI clinical images for optimization and standardization of breast screening protocols.

    abstract:PURPOSE:Stringent quality assurance is required in MRI breast screening to ensure that different scanners and imaging protocols reach similar diagnostic performance. The authors propose a methodology, based on power spectrum analysis (PSA), to evaluate spatial resolution in clinical images. To demonstrate this approach...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4966704

    authors: Borri M,Scurr ED,Richardson C,Usher M,Leach MO,Schmidt MA

    更新日期:2016-12-01 00:00:00

  • Radiofrequency penetration and absorption in the human body: limitations to high-field whole-body nuclear magnetic resonance imaging.

    abstract::This study presents experimental results about the effective depth of penetration and about the radiofrequency (rf) power absorption in humans as a function of frequency. The frequency range investigated covers 10 up to 220 MHz. For the main part, the results were derived from bench measurements of the quality factor ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595995

    authors: Röschmann P

    更新日期:1987-11-01 00:00:00

  • A method for characterizing and matching CT image quality across CT scanners from different manufacturers.

    abstract:PURPOSE:The purpose of this study was to quantitatively characterize the fundamental aspects of image quality (IQ) associated with different computed tomography (CT) reconstruction algorithms, the resolution, noise texture, noise magnitude per dose, and use those data to devise a methodology to match IQ between differe...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12554

    authors: Winslow J,Zhang Y,Samei E

    更新日期:2017-11-01 00:00:00

  • Correcting magnetic resonance k-space data for in-plane motion using an optical position tracking system.

    abstract:PURPOSE:Motion is a major confound of image quality in MRI. A method of retrospectively correcting the effects of rotations and translations on the acquired k-space data is presented. METHODS:In two phantom experiments of well-controlled translation and rotation, two MRI-compatible infrared cameras recorded motion dat...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3254189

    authors: Marxen M,Marmurek J,Baker N,Graham SJ

    更新日期:2009-12-01 00:00:00

  • A restricted angular scattering model for electron penetration in dense media.

    abstract::A restricted angular scattering model for electron penetration in dense media is presented. In the model, the Fermi-Eyges transport equation is modified through the addition of an extra term which may be interpreted as representing an apparent force opposing the scattering of electrons into wider angles. The introduct...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596747

    authors: McLellan J,Sandison GA,Papiez L,Huda W

    更新日期:1991-01-01 00:00:00

  • Analyzing the impact of intrafraction motion: correlation of different dose metrics with changes in target D95%.

    abstract:PURPOSE:A number of techniques are available to determine the dosimetric impact of intrafraction motion during intensity modulated radiation therapy (IMRT). Motion-induced dose perturbations can be determined both computationally and experimentally using a number of different dosimetric metrics. However, these measures...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3605633

    authors: Waghorn BJ,Meeks SL,Langen KM

    更新日期:2011-08-01 00:00:00

  • Stereomammography: evaluation of depth perception using a virtual 3D cursor.

    abstract::We are evaluating the usefulness of stereomammography in improving breast cancer diagnosis. One area that we are investigating is whether the improved depth perception associated with stereomammography might be significantly enhanced with the use of a virtual 3D cursor. A study was performed to evaluate the accuracy o...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.599008

    authors: Goodsitt MM,Chan HP,Hadjiiski L

    更新日期:2000-06-01 00:00:00

  • Effects of tumor motion in GRID therapy.

    abstract::Clinical and biological evidence suggest that the success of GRID therapy in debulking large tumors depends on the high peak-to-valley contrast in the dose distribution. In this study, we show that the peaks and valleys can be significantly blurred out by respiration-induced tumor motion, possibly affecting the clinic...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2977538

    authors: Naqvi SA,Mohiuddin MM,Ha JK,Regine WF

    更新日期:2008-10-01 00:00:00

  • Experimental verification of a technique for predicting scattered radiation transfer: application to low photon energies.

    abstract::It is shown that predictions can be made of scatter-to-primary ratios for a variety of mammographic configurations. The different configurations can be produced by changes in source-detector distance; source-phantom distance; air gap; photon energy; phantom composition, thickness, and cross-sectional area. A detailed ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595807

    authors: Bernstein H,Muntz EP

    更新日期:1986-11-01 00:00:00

  • SU-E-I-99: An Ultra-High Resolution Small Field-Of-View Solid State X-Ray Imaging Detector Based on an Electron Multiplying CCD.

    abstract:PURPOSE:To demonstrate the ultra-high resolution capability of a small field-of-view (FOV) solid state x-ray imaging detector based on an EMCCD sensor. METHODS:A micro-solid state x-ray image intensifier (micro-SSXII) was developed to serve as an ultra-high resolution region-of-interest (ROI) imaging detector. It is b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734816

    authors: Singh V,Loughran B,Jain A,Sharma P,Bednarek D,Rudin S

    更新日期:2012-06-01 00:00:00

  • Specific recommendations for accurate and direct use of PET-CT in PET guided radiotherapy for head and neck sites.

    abstract:PURPOSE:To provide specific experience-based guidance and recommendations for centers wishing to develop, validate, and implement an accurate and efficient process for directly using positron emission tomography-computed tomography (PET-CT) for the radiotherapy planning of head and neck cancer patients. METHODS:A PET-...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4867856

    authors: Thomas CM,Pike LC,Hartill CE,Baker S,Woods E,Convery DJ,Greener AG

    更新日期:2014-04-01 00:00:00