Abstract:
:Two side-by-side energy windows, one at the photopeak and one at lower energy, are sometimes employed in quantitative SPECT studies. We measured the count-rate losses at moderately high activities of 131I for two multihead Anger cameras in such a dual-window-acquisition mode by imaging a decaying source composed of two hot spheres within a warm cylinder successively over a total of 23 days. The window locations were kept fixed and the paralyzable model was assumed. In addition, for the Picker Prism 3000 XP camera, the source was viewed from three different angles separated by 120 degrees and the final results are from an average over these three angles. For the Picker camera, the fits to the data from the individual windows are good (the mean of the squared correlation coefficient equals 0.98) while for the Siemens Multispect camera fits to the data from head 1 and from the lower-energy, monitor window are relatively poor. Therefore, with the Siemens camera the data from the two windows are combined for deadtime computation. Repeated autopeaking might improve the fits. At the maximum count rate, corresponding to a total activity of 740 MBq (20 mCi) in the phantom, the multiplicative deadtime correction factor is considerably larger for the Picker than for the Siemens camera. For the Picker camera, it is 1.11, 1.12, and 1.12 for heads 1-3 with the photopeak window and 1.10 for all heads with the lower-energy monitor window. For the Siemens camera, the combined-window deadtime correction factor is 1.02 for head 1 and 1.03 for head 2. Differences between the deadtime correction factor for focal activity and for the total activity do not support the hypothesis of count misplacement between foci of activity at these count rates. Therefore, the total-image dead time correction is recommended for any and all parts of the image.
journal_name
Med Physjournal_title
Medical physicsauthors
Koral KF,Zasadny KR,Ackermann RJ,Ficaro EPdoi
10.1118/1.598162subject
Has Abstractpub_date
1998-01-01 00:00:00pages
85-91issue
1eissn
0094-2405issn
2473-4209journal_volume
25pub_type
杂志文章相关文献
MEDICAL PHYSICS文献大全abstract:PURPOSE:A modular phantom for dosimetry and imaging performance evaluation in cone-beam computed tomography (CBCT) is reported, providing a tool for quantitative technical assessment that can be adapted to a broad variety of CBCT imaging configurations and clinical applications. METHODS:The phantom presents a set of m...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13952
更新日期:2020-02-01 00:00:00
abstract::Dose to the total body from induced radiation resulting from primary exposure to radiotherapeutic beams is not detailed in routine treatment planning though this information is potentially important for better estimates of health risks including secondary cancers. This information can also allow better management of p...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1612947
更新日期:2003-10-01 00:00:00
abstract:PURPOSE:This paper presents the design of a micro-CT guided small animal robotic needle positioning system. In order to simplify the robotic design and maintain a small targeting error, a novel implementation of the remote center of motion is used in the system. The system has been developed with the objective of achie...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4771958
更新日期:2013-01-01 00:00:00
abstract:PURPOSE:The quality of a treatment plan for stereotactic body radiotherapy (SBRT) depends on an experience of each treatment planner. Therefore, the treatment plans are subjectively determined by comparison of several treatment plans developed by time consuming iterative manners, while considering the benefit to a tumo...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4735950
更新日期:2012-06-01 00:00:00
abstract:PURPOSE:Accurate, patient-specific radiation dosimetry for CT scanning is critical to optimize radiation doses and balance dose against image quality. While Monte Carlo (MC) simulation is often used to estimate doses from CT, comparison of estimates to experimentally measured values is lacking for advanced CT scanners ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13780
更新日期:2019-11-01 00:00:00
abstract::Intensity modulated radiation therapy (IMRT) treatment planning typically considers beam optimization and beam delivery as separate tasks. Following optimization, a multi-leaf collimator (MLC) or other beam delivery device is used to generate fluence patterns for patient treatment delivery. Due to limitations and char...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1477230
更新日期:2002-06-01 00:00:00
abstract:PURPOSE:This work introduces a rigid registration framework for patient positioning in radiotherapy, based on real-time surface acquisition by a time-of-flight (ToF) camera. Dynamic properties of the system are also investigated for future gating/tracking strategies. METHODS:A novel preregistration algorithm, based on...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3664006
更新日期:2012-01-01 00:00:00
abstract::We have developed two prototypes of portable gamma cameras for medical applications based on a previous prototype designed and tested by our group. These cameras use a CsI(Na) continuous scintillation crystal coupled to the new flat-panel-type multianode position-sensitive photomultiplier tube, H8500 from Hamamatsu Ph...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2358199
更新日期:2006-11-01 00:00:00
abstract:PURPOSE:A detailed analysis of 2728 intensity modulated proton therapy (IMPT) fields that were clinically delivered to patients between 2007 and 2013 at Paul Scherrer Institute (PSI) was performed. The aim of this study was to analyze the results of patient specific dosimetric verifications and to assess possible corre...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4964449
更新日期:2016-11-01 00:00:00
abstract:PURPOSE:The objective of this study is to evaluate requirements for radionuclide-based solid tumor therapy by assessing the radial dose distribution of beta-particle-emitting and alpha-particle-emitting molecules localized either solely within endothelial cells of tumor vasculature or diffusing from the vasculature thr...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3431999
更新日期:2010-06-01 00:00:00
abstract::In order to understand the role of water desorption, i.e., the opposite of water adsorption, in optical clearing of biological tissue created by the application of hyperosmotic agents, dynamics of water loss in fresh porcine muscle tissue administrated with glycerol or ethylene glycol was investigated with the near-in...
journal_title:Medical physics
pub_type:
doi:10.1118/1.1576228
更新日期:2003-06-01 00:00:00
abstract:PURPOSE:To evaluate the energy dependence of Gafchromic EBT3 film for establishing a quality assurance method of bolus electron conformal radiotherapy. METHODS:We irradiated electron beam to EBT3 films, which were set in the water tank. The linear accelerator used was Varian Clinac 21EX. The energy of electron beams w...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4735200
更新日期:2012-06-01 00:00:00
abstract:PURPOSE:For cone-beam computed tomography (CBCT) systems, we propose a sphere phantom based method to estimate the full three-dimensional (3D) modulation transfer function (MTF). METHODS:The FDK reconstruction of CBCT system in a local region was modeled by a triple convolution operator. Afterward, we calculated the d...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14147
更新日期:2020-07-01 00:00:00
abstract::The availability of digital radiographic imaging on simulators has led to the investigation of a number of new imaging possibilities, including digital linear tomography (tomosynthesis). It has been shown that a single set of projections in this case is sufficient to reconstruct images from multiple planes, including ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.598006
更新日期:1997-06-01 00:00:00
abstract:PURPOSE:Hip fracture is a serious health problem and textural methods are being developed to assess bone quality. The authors aimed to perform textural analysis at femur on high-resolution digital radiographs compared to three-dimensional (3D) microarchitecture comparatively to bone mineral density. METHODS:Sixteen ca...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3528125
更新日期:2011-01-01 00:00:00
abstract:PURPOSE:Dosimetry with ionization chambers in clinical ion beams for radiation therapy requires correction for recombination effects. However, common radiation protocols discriminate between initial and general recombination and provide no universal correction method for the presence of both recombination types in ion ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4962483
更新日期:2016-10-01 00:00:00
abstract:PURPOSE:Dual-energy computed tomography and dual-energy cone-beam computed tomography (DE-CBCT) are promising modalities for applications ranging from vascular to breast, renal, hepatic, and musculoskeletal imaging. Accordingly, the optimization of imaging techniques for such applications would benefit significantly fr...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4736420
更新日期:2012-08-01 00:00:00
abstract:PURPOSE:Accurate detection of pulmonary nodules remains a technical challenge in computer-aided diagnosis systems because some nodules may adhere to the blood vessels or the lung wall, which have low contrast compared to the surrounding tissues. In this paper, the analysis of typical shape features of candidate nodules...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4907961
更新日期:2015-03-01 00:00:00
abstract:PURPOSE:As hypoxic cells are more resistant to photon radiation, it is desirable to obtain information about the oxygen distribution in tumors prior to the radiation treatment. Noninvasive techniques are currently not able to provide reliable oxygenation maps with sufficient spatial resolution; therefore mathematical m...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4812431
更新日期:2013-08-01 00:00:00
abstract::A beam intensity monitor was tested in a 230-MeV proton beam at the Loma Linda Proton Therapy Accelerator during its commissioning at Fermi National Accelerator Laboratory. The intensity monitor was designed to regulate the beam intensity extracted from the proton synchrotron. The proton beam is tunable between 70 and...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.596636
更新日期:1991-07-01 00:00:00
abstract:PURPOSE:To investigate the effect of energy (kVp) and filters (no filter, half Bowtie, and full Bowtie) on the dose response curves of the Gafchromic XRQA2 film and nanoDot optical stimulated luminescence dosimeters (OSLDs) in CBCT dose fields. To measure surface and internal doses received during x-ray volume imager (...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4803466
更新日期:2013-06-01 00:00:00
abstract:PURPOSE:A series of examples are presented in which potential errors in the delivery of radiation therapy were prevented through use of incident learning. These examples underscore the value of reporting near miss incidents. METHODS:Using a departmental incident learning system, eight incidents were noted over a two-y...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4760774
更新日期:2012-11-01 00:00:00
abstract:PURPOSE:A long-standing conundrum is why photoacoustic tomography (PAT) possesses the unique ability to produce images devoid of speckle artifacts while all other coherent imaging technologies do not. METHODS:In this paper, we explain the inherent mechanism that suppresses speckle in PAT, and the analysis was validate...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3187231
更新日期:2009-09-01 00:00:00
abstract::Respiration can cause tumors in the thorax or abdomen to move by as much as 3 cm; this movement can adversely affect the planning and delivery of radiation treatment. Several techniques have been used to compensate for respiratory motion, but all have shortcomings. Manufacturers of computed tomography (CT) equipment h...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2044427
更新日期:2006-01-01 00:00:00
abstract:PURPOSE:To present a system for robotic 4D ultrasound (US) imaging concurrent with radiotherapy beam delivery and estimate the proportion of liver stereotactic ablative body radiotherapy (SABR) cases in which robotic US image guidance can be deployed without interfering with clinically used VMAT beam configurations. M...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4964454
更新日期:2016-11-01 00:00:00
abstract:PURPOSE:Transcranial magnetic resonance-guided focused ultrasound surgery is a noninvasive technique for causing selective tissue necrosis. Variations in density, thickness, and shape of the skull cause aberrations in the location and shape of the focal zone. In this paper, the authors propose a hybrid simulation-MR-AR...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4865778
更新日期:2014-03-01 00:00:00
abstract::A new scheme, called direct aperture deformation (DAD), for online correction of interfraction geometric uncertainties under volumetric imaging guidance is presented. Using deformable image registration, the three-dimensional geometric transformation matrix can be derived that associates the planning image set and the...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2374675
更新日期:2006-12-01 00:00:00
abstract:PURPOSE:To use the Monte Carlo code PENELOPE to study attenuation and tissue equivalence properties of a-Al2O3:C for OSL dosimetry. METHODS:Mass attenuation coefficients of α-Al2O3 and α-Al2O3:C with carbon percent weight concentrations from 1% to 150% were simulated with PENELOPE Monte Carlo code and compared to mass...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4735237
更新日期:2012-06-01 00:00:00
abstract::Radiation therapy treatment modalities continue to develop and have become increasingly complex. With this, dose verification and quality assurance (QA) is of great importance to ensure that a prescribed dose is accurately and precisely delivered to a patient. Radiochromic film dosimetry has been adopted as a convenie...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4740126
更新日期:2012-07-01 00:00:00
abstract:PURPOSE:To correct for the deviations of the detector response when typical radiation detectors are used under non-reference conditions, factor kNR was calculated from the known energy dependence of the detector response at photon energies from 10 keV upwards and from clinical photon spectra within a large water phanto...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4735184
更新日期:2012-06-01 00:00:00