The influence of cannulation technique on blood flow to the brain in rats undergoing cardiopulmonary bypass: a cautionary "tail".

Abstract:

:Recently, there has been an increase in the use of rat models of cardiopulmonary bypass (CPB) for research purposes. Much of this work has focused on cerebral injury associated with CPB. Many of these studies employ a peripheral cannulation approach, often utilizing the caudal artery and internal or external jugular vein. The aim of the present study was to establish whether there is any alteration in blood flow to the brain associated with the use of different cannulation routes. Twenty-four adult male Sprague Dawley rats were allocated to one of three study groups: Group 1- caudal artery return, Group 2 -open-chest aortic return, and Group 3- nonbypass control group. Colored microspheres were injected into all animals at four time points (postinduction, initiation of bypass, midbypass, and end bypass). After the termination of each experiment, the brains were excised, the tissue was digested, the microspheres were harvested, and the global blood flow to the brain was assessed using the reference flow method. There was a significant reduction in blood flow to the brain between both bypass groups and the control group. Additionally, cerebral blood flow was significantly lower in the caudal return group than in the aortic return group. There is a significant drop in blood flow to the brain associated with the initiation and continuation of CPB when compared to non-CPB controls. These results also confirm a considerable cerebral hypoperfusion associated with the peripheral cannulation technique, and suggest that peripheral bypass may exaggerate the influence CPB has on cerebral injury. This technique must therefore be employed with caution.

journal_name

Artif Organs

journal_title

Artificial organs

authors

Gourlay T,Modine T

doi

10.1111/j.1525-1594.2009.00917.x

subject

Has Abstract

pub_date

2010-06-01 00:00:00

pages

498-503

issue

6

eissn

0160-564X

issn

1525-1594

pii

AOR917

journal_volume

34

pub_type

杂志文章
  • Analysis of left ventricular impedance in comparison with ultrasound images.

    abstract::Cardiac monitoring of ventricular assist devices (VADs) is important for detecting heart failure risks, such as critical arrhythmia and ventricular fibrillation, and for supplying data that are useful for hemodynamic control. Specifically, impedance cardiograms (ICGs) are especially beneficial because they have no eff...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2011.01381.x

    authors: Choi SW,Park SM

    更新日期:2012-05-01 00:00:00

  • In vivo characterization of a decellularized dermis-polymer complex for use in percutaneous devices.

    abstract::To develop a method for making percutaneous devices that have high biocompatibility and do not induce downgrowth of epidermal cells, we prepared a partial decellularized dermis (DD)/poly(methyl methacrylate) (PMMA) complex (PDPC) with a PMMA rod firmly stabilized inside. The porcine decellularized tissue was chosen be...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12330

    authors: Nam K,Matsushima R,Kimura T,Fujisato T,Kishida A

    更新日期:2014-12-01 00:00:00

  • Long-term performance of homograft and autograft valves.

    abstract::Five hundred and eighty homograft valves have been inserted into the aortic position as isolated valve replacements, and followed up for 7--14 years. After ten years 90% of frozen homografts, 80% of autografts and 70% of freeze-dried homografts were free of valve-related death. There was only one embolism reported giv...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1980.tb03894.x

    authors: Bodnar E,Wain WH,Martelli V,Ross DN

    更新日期:1980-02-01 00:00:00

  • Vascular access for continuous arteriovenous hemofiltration in infants and young children.

    abstract::The blood flow through the hemofilter device is the key point for ultrafiltrate production during continuous arteriovenous hemofiltration. It mainly depends on arteriovenous pressure gradient and vascular access, which is the major problem for arteriovenous hemofiltration in infants and small children. In infants, we ...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1988.tb01518.x

    authors: Zobel G,Trop M,Beitzke A,Ring E

    更新日期:1988-02-01 00:00:00

  • Performance of baffled bubble blood oxygenators.

    abstract::Bubble blood oxygenators equipped with baffles of various types in the oxygenating column were studied. The rate of hemolysis, the volumetric coefficient for oxygen absorption into blood, and the fractional gas holdup were found to be affected mainly by the superficial gas velocity. When compared with the conventional...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1979.tb01032.x

    authors: Ohshima N,Yoshido F

    更新日期:1979-05-01 00:00:00

  • Optimal continuous retrograde coronary perfusion flow rate with beating heart in normal temperature.

    abstract:PURPOSE:We were led to the premise that with retrograde oxygenated blood perfusion, it would be possible to perform operations on the beating heart. METHODS:The experiment consisted of two stages: procedure 1 and 2. In procedure 1, six pigs were used. Retrograde perfusion flow was controlled to be in the range of 5.0-...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2004.47279.x

    authors: Tsukioka T,Tomita S,Watanabe G,Takemura H

    更新日期:2004-03-01 00:00:00

  • In vivo test of pressure head and flow rate estimation in a continuous-flow artificial heart.

    abstract::To avoid using sensors with low biocompatibility and low durability in implantable total artificial heart (TAH) systems, the authors previously proposed a new method for estimating instantaneous values of flow rate and pressure head on the basis of voltage, current, and rotational speed in a motor driven centrifugal p...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.2003.07175.x

    authors: Tanaka A,Yoshizawa M,Abe K,Takeda H,Yambe T,Nitta S

    更新日期:2003-01-01 00:00:00

  • A clinical monitoring system for centrifugal blood pumps.

    abstract::In clinical application of rotary blood pumps, flow obstruction as a result of suction of the inflow cannula, kinking of tubing, or thrombus formation occurs quite frequently. Early detection of such problems is essential to avoid hemolysis, tissue degradation, or release of thrombi to the patient. A program was devel...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1995.tb02409.x

    authors: Holzer S,Scherer R,Schmidt C,Schwendenwein I,Wieselthaler G,Noisser R,Schima H

    更新日期:1995-07-01 00:00:00

  • Cadaver fitting study of the DexAide right ventricular assist device.

    abstract::The DexAide right ventricular assist device (RVAD) has been developed to provide an implantable RVAD option to surgeons. The aim of this study was to determine the optimal cannula design and optimal implantation location of the DexAide RVAD in preparation for its clinical use. Separately, a HeartMate XVE left ventricu...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2007.00438.x

    authors: Ootaki Y,Ootaki C,Kamohara K,Akiyama M,Cingoz F,Kopcak MW Jr,Dessoffy R,Liu J,Massiello A,Horvath D,Chen JF,Zhou Q,Benefit S,Golding LA,Fukamachi K

    更新日期:2007-08-01 00:00:00

  • Chaotic dynamics in circulation with Tohoku University vibrating flow pump.

    abstract::For the development of a totally implantable ventricular assist system (VAS), we have been developing the vibrating flow pump (VFP), which can generate oscillated blood flow with a relative high frequency (10-50 Hz) for a totally implantable system. In this study, the effects of left ventricular assistance with this u...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.1999.06288.x

    authors: Nitta S,Yambe T,Kobayashi S,Hashimoto H,Yoshizawa M,Mastuki H,Tabayashi K,Takeda H

    更新日期:1999-01-01 00:00:00

  • Short hemodialysis: long-term mortality and morbidity.

    abstract::Individualized short hemodialysis treatment schedules, approximately three 3-h treatments per week, were introduced in a nephrology unit in 1976. In May 1985, 259 patients had been treated, which corresponds to approximately 60 new patients per million inhabitants per year. Ages at start of therapy ranged between 17 a...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1986.tb02541.x

    authors: Wauters JP,Bercini-Pansiot S,Gilliard N,Stauffer JC

    更新日期:1986-06-01 00:00:00

  • The compliance problem: a major obstacle in the development of implantable blood pumps.

    abstract::The compliance problem represents a major obstacle in the development of implantable blood pumps. The motor-driven pumps are enclosed in an airtight casing with gas in the motor space. The trapped gas in the motor space either increases the load on the motor during systole or impedes blood sac filling. Potential solut...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1984.tb04248.x

    authors: Lee S,Rosenberg G,Donachy JH,Wisman CB,Pierce WS

    更新日期:1984-02-01 00:00:00

  • Ferritin: a reliable indicator of iron supplementation in patients on chronic hemodialysis/hemofiltration treatment?

    abstract::Four hundred and forty-two serum ferritin determinations were performed in 144 patients on chronic intermittent hemodialysis treatment with intravenous iron substitution and/or oral iron substitution. Iron substitution should be done individually according to regular serum ferritin determinations. Intravenous iron sub...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1981.tb03980.x

    authors: Müller HA,Schneider H,Hövelborn U,Streicher E

    更新日期:1981-05-01 00:00:00

  • Pre- and postoperative magnetic resonance imaging in neonatal arterial switch operation using warm perfusion.

    abstract::Neurological morbidity is a major concern in pediatric cardiac surgery. Cardiopulmonary bypass is one of the few modifiable factors affecting neurodevelopmental outcome. This study aimed to measure the incidence of abnormalities apparent by magnetic resonance imaging (MRI) after neonatal arterial switch operation usin...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2011.01325.x

    authors: Durandy Y,Rubatti M,Couturier R,Rohnean A

    更新日期:2011-11-01 00:00:00

  • The roles of vascular smooth muscle cells in the aortic wall thinness under prolonged continuous flow left heart bypass.

    abstract::Aortic wall thinness was one of the most characteristic changes observed in experimental animals under prolonged continuous flow left heart bypass. The goal of this study was to determine the roles of smooth muscle cells in the vascular remodeling process in cases demonstrating aortic wall thinness under prolonged con...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.2003.00022.x

    authors: Mizuno T,Nishinaka T,Ohnishi H,Tatsumi E,Tsukiya T,Oshikawa M,Shioya K,Takewa Y,Homma A,Takano H,Kitamura S,Taenaka Y

    更新日期:2003-10-01 00:00:00

  • Heart Transplant Waiting List Implications of Increased Ventricular Assist Device Use as a Bridge Strategy: A National Analysis.

    abstract:BACKGROUND:The use of ventricular assist devices (VAD) as a bridge to heart transplant (HT) is increasing, while HT volume remains stagnant. This may portend longer waiting times and an otherwise more competitive environment for all patients on the HT waiting list. METHODS:A retrospective analysis of patients who were...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.13833

    authors: Han JJ,Elzayn H,Duda MM,Iyengar A,Acker AM,Patrick WL,Helmers M,Birati EY,Atluri P

    更新日期:2020-10-01 00:00:00

  • Multicenter Experience With Mechanical Circulatory Support Using a New Diagonal Pump in 233 Children.

    abstract::Technological innovations in pediatric extracorporeal life support circuits can reduce system-related complications and may improve patients' outcome. The Deltastream DP3 (Medos Medizintechnik AG, Stolberg, Germany) is a novel rotational pump with a diagonally streamed impeller that can be used over a broad range of f...

    journal_title:Artificial organs

    pub_type: 杂志文章,多中心研究

    doi:10.1111/aor.13016

    authors: Stiller B,Houmes RJ,Rüffer A,Kumpf M,Müller A,Kipfmüller F,Köditz H,Herber Jonat S,Schmoor C,Benk C,Tibboel D,Fleck T

    更新日期:2018-04-01 00:00:00

  • Immunologic studies of hollow-fiber dialyzer extracts.

    abstract::Testing failed to detect free isocyanates in dialyzer rinses or significant levels of immunoglobulin E antibodies to isocyanate-protein conjugates in the serum of individuals reacting during hemodialysis with dialyzers containing Cuprophan membranes. Extended immunization of rabbits with Limulus lysate-reactive materi...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1984.tb04298.x

    authors: Butcher BT,Reed MA,O'Neil CE,Leech S,Pearson FC

    更新日期:1984-08-01 00:00:00

  • Tissue reactions to various percutaneous materials with different surface properties and structures.

    abstract::Tissue responses to various percutaneous materials with different surface properties and structures were investigated. Dense hydroxyapatite (HA), tricalcium phosphate (TCP), glassy carbon (GC), and 2 types of porous HA were used. Cutaneous and subcutaneous tissues were tightly attached to the surface of the penetratin...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1997.tb00514.x

    authors: Shin Y,Akao M

    更新日期:1997-09-01 00:00:00

  • The impact of pump speed and inlet cannulation site on left ventricular unloading with a rotary blood pump.

    abstract:BACKGROUND:Ventricular assist devices are gaining ground in the therapeutic treatment of chronic heart failure. These devices are sometimes used as a bridge to recovery by unloading the left ventricle (LV) and restoring its function. It is therefore important to preserve the heart muscle and apply less invasive implant...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2004.07374.x

    authors: Vandenberghe S,Nishida T,Segers P,Meyns B,Verdonck P

    更新日期:2004-07-01 00:00:00

  • Cell investigation into the biocompatibility of adult human dermal fibroblasts with PCL nanofibers/TiO2 nanotubes on the surface of Ti-30Ta alloy for biomedical applications.

    abstract::The polymer poly(ε-caprolactone) (PCL) has been used in the biomaterial field for its relatively inexpensive price and suitability for modification. Also, its chemical and biological properties are desirable for biomedical applications. The electrospinning process has been used for producing polymer fibers of PCL due ...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.13713

    authors: Capellato P,Silva G,Popat K,Simon-Walker R,Alves Claro AP,Zavaglia C

    更新日期:2020-08-01 00:00:00

  • Usability and safety of ventricular assist devices: human factors and design aspects.

    abstract::The purpose of this study was the investigation of the usability and ergonomics of ventricular assist devices (VADs) in everyday usage. Patients with four different VAD types were observed. After implantation, instruction, and discharge from the hospital, the patients returned on a regular basis to the outpatient clin...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2009.00844.x

    authors: Geidl L,Zrunek P,Deckert Z,Zimpfer D,Sandner S,Wieselthaler G,Schima H

    更新日期:2009-09-01 00:00:00

  • Significance of percutaneous cardiopulmonary bypass support for volume reduction surgery with severe hypercapnia.

    abstract::In patients with reduced respiratory function, lung resection is associated with high risk because separate ventilation is generally needed for safe management. For patients with end-stage emphysema, intraoperative respiratory management is important and particularly difficult because neither incomplete oxygenation no...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.2000.06339.x

    authors: Tsunezuka Y,Sato H,Tsubota M,Seki M

    更新日期:2000-01-01 00:00:00

  • Testosterone and hemoglobin in hemodialysis male and female patients.

    abstract::It has been speculated that testosterone stimulates erythropoiesis. We hypothesized that hemoglobin levels in hemodialysis (HD) patients are associated with serum testosterone concentrations. Testosterone, hemoglobin, and other biochemical parameters were measured in a representative sample of 98 chronic HD patients (...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12218

    authors: Ekart R,Taskovska M,Hojs N,Bevc S,Hojs R

    更新日期:2014-07-01 00:00:00

  • Influence of different hemodialysis membranes on red blood cell susceptibility to oxidative stress.

    abstract::Oxidative stress is crucial in red blood cell (RBC) damage induced by activated neutrophils in in vitro experiments. The aim of the study was to evaluate whether the bioincompatibility phenomena occurring during hemodialysis (HD) (where neutrophil activation with increased free radical production is well documented) m...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.2000.06432.x

    authors: Lucchi L,Bergamini S,Botti B,Rapanà R,Ciuffreda A,Ruggiero P,Ballestri M,Tomasi A,Albertazzi A

    更新日期:2000-01-01 00:00:00

  • Design of a symmetry controller for cycling induced by electrical stimulation: preliminary results on post-acute stroke patients.

    abstract::This study deals with the design of a controller for cycling induced by functional electrical stimulation. The controller will be exploitable in the rehabilitation of hemiparetic patients who need to recover motor symmetry. It uses the pulse width as the control variable in the stimulation of the two legs in order to ...

    journal_title:Artificial organs

    pub_type: 临床试验,杂志文章

    doi:10.1111/j.1525-1594.2009.00941.x

    authors: Ambrosini E,Ferrante S,Schauer T,Ferrigno G,Molteni F,Pedrocchi A

    更新日期:2010-08-01 00:00:00

  • Comparison of rapid in situ, regular in situ, and ex vivo flushing on hepatic function.

    abstract::The efficacy of three flushing techniques on subsequent liver function was assessed using the in vivo isolated liver perfusion model (ILPM). Livers from brain-dead mongrel dogs were flushed with cold Euro-Collins as follows: Group I, rapid in situ flushing (10 min); Group II, regular in situ flushing (45 min); Group I...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1988.tb02760.x

    authors: Castellanos J,Toledo-Pereyra LH,Finkelstein I,Kestenberg W,Chapman M,Choudhury S

    更新日期:1988-06-01 00:00:00

  • Hepatocyte growth factor-transfected skeletal myoblasts to limit the development of postinfarction heart failure.

    abstract::Stem cells transplanted to an injured heart affect the host myocardium indirectly. The cytokine hepatocyte growth factor (HGF) may play a key role in this paracrine activity. We hypothesized that HGF-overexpressing stem cells would restore cardiac function after myocardial infarction (MI). Because there is a high rate...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2011.01328.x

    authors: Poppe A,Golsong P,Blumenthal B,von Wattenwyl R,Blanke P,Beyersdorf F,Schlensak C,Siepe M

    更新日期:2012-03-01 00:00:00

  • Assessment of total hip arthroplasty by means of computed tomography 3D models and fracture risk evaluation.

    abstract::Total hip arthroplasty (THA) can be achieved by using a cemented or noncemented prosthesis. Besides patient's age, weight, and other clinical signs, the evaluation of the quality of the bones is a crucial parameter on which orthopedic surgeons base the choice between cemented and noncemented THA. Although bone density...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12033

    authors: Gargiulo P,Pétursson T,Magnússon B,Bifulco P,Cesarelli M,Izzo GM,Magnúsdóttir G,Halldórsson G,Ludvigsdóttir GK,Tribel J,Jónsson H Jr

    更新日期:2013-06-01 00:00:00

  • Evaluation of double filtration plasmapheresis, thermofiltration, and low-density lipoprotein adsorptive methods by crossover test in the treatment of familial hypercholesterolemia patients.

    abstract::A comparative assessment has been made regarding efficacy and safety of the double filtration plasmapheresis (DFPP), thermofiltration (TFPP), and low-density lipoprotein (LDL) adsorptive (PA) methods by making a crossover test on heterozygous familial hypercholesterolemia patients. Treatments by DFPP, TFPP (secondary ...

    journal_title:Artificial organs

    pub_type: 临床试验,杂志文章

    doi:10.1111/j.1525-1594.1996.tb04448.x

    authors: Suzuki M,Yamane S,Matsugane T,Nobuto T,Azuma N,Nishide T,Shinomiya M,Saito K,Sasaki N,Nosé Y

    更新日期:1996-04-01 00:00:00