Calibration of crushable foam plasticity models for synthetic bone material for use in finite element analysis of acetabular cup deformation and primary stability.

Abstract:

:Polyurethane (PU) foam is a material often used in biomechanical experiments and demands for the definition of crushable foam plasticity (CFP) in numerical simulations of the primary stability and deformation of implants, to describe the crushing behaviour appropriately. Material data of PU foams with five different densities (10-40 pounds per cubic foot were ascertained experimentally in uniaxial compression test and used to calibrate CFP models for finite element modelling. Additionally, experimental and numerical deformation, push-out and lever-out tests of press-fit acetabular cups were carried out to assess the influence of the chosen material definition (linear elastic and CFP) on the numerical results. Comparison of the experimentally and numerically determined force-displacement curves of the uniaxial compression test showed a mean deviation of less than 3%. In primary stability testing, the deviation between the experimental and numerical results was in a range of 0%-27% for CFP modelling and 64%-341% for the linear elastic model. The material definition selected, highly influenced the numerical results in the current study. The use of a CFP model is recommended for further numerical simulations, when a deformation of the foam beyond the yield strength is likely to occur.

authors

Schulze C,Vogel D,Sander M,Bader R

doi

10.1080/10255842.2018.1524884

subject

Has Abstract

pub_date

2019-01-01 00:00:00

pages

25-37

issue

1

eissn

1025-5842

issn

1476-8259

journal_volume

22

pub_type

杂志文章
  • Quantification of soft tissue balance in total knee arthroplasty using finite element analysis.

    abstract::Unbalanced contact force on the tibial component has been considered a factor leading to loosening of the implant and increased wear of the bearing surface in total knee arthroplasty. Because it has been reported that good alignment cannot guarantee successful clinical outcomes, the soft tissue balance should be check...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2013.765409

    authors: Oh KJ,Park WM,Kim K,Kim YH

    更新日期:2014-01-01 00:00:00

  • Comparison of dynamic response of three TLIF techniques on the fused and adjacent segments under vibration.

    abstract::To explore which TLIF techniques are advantageous in reducing the risk of complications and conducive to bone fusion under the vibration. The L1-L5 finite element lumbar model was modified to simulate three different TLIF techniques (a unilateral standard cage, a crescent-shaped cage, and bilateral standard cages). Th...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2020.1829604

    authors: Wang QD,Guo LX

    更新日期:2020-10-13 00:00:00

  • Influence of fluid-flow direction on effective permeability of the vertebral end plate: an analytical model.

    abstract::Convective transports in the vertebral end plate (VEP) play a significant role in the homeostasis of the spine. A few studies hypothesised that the hydraulic resistance or effective permeability of the VEP could be dependant upon fluid-flow direction. Results were influenced by species, region of interest within the e...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2010.518960

    authors: Swider P,Accadbled F,Laffosse JM,Sales de Gauzy J

    更新日期:2012-01-01 00:00:00

  • Simulation of swallowing dysfunction and mechanical ventilation after a Montgomery T-tube insertion.

    abstract::The Montgomery T-tube is used as a combined tracheal stent and airway after laryngotracheoplasty, to keep the lumen open and prevent mucosal laceration from scarring. It is valuable in the management of upper and mid-tracheal lesions, while invaluable in long and multisegmental stenting lesions. Numerical simulations ...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2014.930448

    authors: Trabelsi O,Malvè M,Mena Tobar A,Doblaré M

    更新日期:2015-01-01 00:00:00

  • Mechanical vulnerability of lower second premolar utilising visco-elastic dynamic stress analysis.

    abstract::Stress analysis determines vulnerability of dental tissues to external loads. Stress values depend on loading conditions, mechanical properties and constrains of structural components. The critical stress levels lead to tissue damage. The aim of this study is to analyse dynamic stress distribution of lower second prem...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255840902788579

    authors: Khani MM,Tafazzoli-Shadpour M,Aghajani F,Naderi P

    更新日期:2009-10-01 00:00:00

  • Quantitative underwater 3D motion analysis using submerged video cameras: accuracy analysis and trajectory reconstruction.

    abstract::In this study we aim at investigating the applicability of underwater 3D motion capture based on submerged video cameras in terms of 3D accuracy analysis and trajectory reconstruction. Static points with classical direct linear transform (DLT) solution, a moving wand with bundle adjustment and a moving 2D plate with Z...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2012.664637

    authors: Silvatti AP,Cerveri P,Telles T,Dias FA,Baroni G,Barros RM

    更新日期:2013-01-01 00:00:00

  • Subject-specific musculoskeletal modelling in patients before and after total hip arthroplasty.

    abstract::The goal of this study was to define the effect on hip contact forces of including subject-specific moment generating capacity in the musculoskeletal model by scaling isometric muscle strength and by including geometrical information in control subjects, hip osteoarthritis and total hip arthroplasty patients. Scaling ...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2016.1181174

    authors: Wesseling M,De Groote F,Meyer C,Corten K,Simon JP,Desloovere K,Jonkers I

    更新日期:2016-11-01 00:00:00

  • Real-time technique for conversion of skin temperature into skin blood flow: human skin as a low-pass filter for thermal waves.

    abstract::Monitoring of skin blood flow oscillations related with mechanical activity of vessels is a very useful modality during diagnosis of peripheral hemodynamic disorders. In this study, we developed a new model and technique for real-time conversion of skin temperature into skin blood flow oscillations, and vice versa. Th...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2019.1615058

    authors: Sagaidachnyi A,Fomin A,Usanov D,Skripal A

    更新日期:2019-09-01 00:00:00

  • Determining the location of hip joint centre: application of a conchoid's shape to the acetabular cartilage surface of magnetic resonance images.

    abstract::Preoperative planning, or intraoperative navigation of hip surgery, including joint-preserving procedures such as osteotomy or joint-replacing procedures such as total arthroplasty, needs to be performed with a high degree of accuracy to ensure a successful outcome. The ability to precisely localise the hip joint rota...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2010.495064

    authors: Kang MJ,Sadri H,Stern R,Magnenat-Thalmann N,Hoffmeyer P,Ji HS

    更新日期:2011-01-01 00:00:00

  • Validation of a finite element model with six-year-old child anatomical characteristics as specified in Euro NCAP Pedestrian Human Model Certification (TB024).

    abstract::Accident statistics show that more than 80% of car-to-pedestrian collisions (CPC) occur when pedestrians cross the road. It is very important to establish a finite element model with natural walking posture to study the kinematics and injury mechanism of pedestrians. In this study, a finite element model of six-year-o...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2020.1810677

    authors: Li H,Li K,Huang Y,Lv W,Cui S,He L,Ruan JS,Wang C

    更新日期:2020-09-02 00:00:00

  • A constitutive model for the periodontal ligament as a compressible transversely isotropic visco-hyperelastic tissue.

    abstract::This study is devoted to the development of a non-linear anisotropic model for the human periodontal ligament (PDL). A thorough knowledge of the behaviour of the PDL is vital in understanding the mechanics of orthodontic tooth mobility, soft tissue response and proposed treatment plans. There is considerable evidence ...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/13639080701314894

    authors: Zhurov AI,Limbert G,Aeschlimann DP,Middleton J

    更新日期:2007-06-01 00:00:00

  • Comparing parametric solid modelling/reconfiguration, global shape modelling and free-form deformation for the generation of 3D digital models of femurs from X-ray images.

    abstract::At present, computer assisted surgery systems help orthopaedic surgeons both plan and perform surgical procedures. To enable these systems to function, it is crucial to have at one's disposal 3D models of anatomical structures, surgical tools and prostheses (if required). This paper analyses and compares three methods...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255840903077329

    authors: Filippi S,Motyl B,Bandera C

    更新日期:2009-02-01 00:00:00

  • Finite element analysis of peri-implant bone volume affected by stresses around Morse taper implants: effects of implant positioning to the bone crest.

    abstract:OBJECTIVES:The purpose of the present study was to evaluate the distribution and magnitude of stresses through the bone tissue surrounding Morse taper dental implants at different positioning relative to the bone crest. MATERIALS AND METHODS:A mandibular bone model was obtained from a computed tomography scan. A three...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2018.1507025

    authors: Macedo JP,Pereira J,Faria J,Souza JCM,Alves JL,López-López J,Henriques B

    更新日期:2018-09-01 00:00:00

  • Accuracy of femur reconstruction from sparse geometric data using a statistical shape model.

    abstract::Sparse geometric information from limited field-of-view medical images is often used to reconstruct the femur in biomechanical models of the hip and knee. However, the full femur geometry is needed to establish boundary conditions such as muscle attachment sites and joint axes which define the orientation of joint loa...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2016.1263301

    authors: Zhang J,Besier TF

    更新日期:2017-04-01 00:00:00

  • Cluster analysis of pressure pain threshold maps from the trapezius muscle.

    abstract::The aim of this study was to investigate and present a new mapping method to describe muscle pain sensitivity based on the assessment of pressure pain threshold (PPT) over the trapezius muscle. PPT data were recorded from 36 points in 20 healthy males using a standardised grid. Points were clustered using the K-means ...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255840903446979

    authors: Binderup AT,Arendt-Nielsen L,Madeleine P

    更新日期:2010-12-01 00:00:00

  • Optimisation of solute transport in dialysers using a three-dimensional finite volume model.

    abstract::Dialyser manufacturers only provide limited information about mass removal under well-defined flow and solute conditions in commercially available dialysers for hemodialysis. This computational study aimed at assessing the solute transport efficiency in a dialyser for different geometries (fiber lengths and diameters)...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255840601002728

    authors: Eloot S,Vierendeels J,Verdonck P

    更新日期:2006-12-01 00:00:00

  • Simulation of the behaviour of the L1 vertebra for different material properties and loading conditions.

    abstract::Three-dimensional finite element models of the thoracolumbar junction (T12-L2) and isolated L1 vertebra were developed to investigate the role of material properties and loading conditions on vertebral stresses and strains to predict fracture risk. The geometry of the vertebrae was obtained from computed tomography im...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2011.636741

    authors: Erdem I,Truumees E,van der Meulen MC

    更新日期:2013-01-01 00:00:00

  • Flexural and creep properties of human jaw compact bone for FEA studies.

    abstract::The aim of this work was to improve the constitutive model of the human mandible and dentition system by taking into account the non-linear material properties of the structural boney matrix that forms the human jaw bone or mandible. Due to the specific structure of the jaw bone the time dependence of the mechanical p...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255840310001637257

    authors: Vitins V,Dobelis M,Middleton J,Limbert G,Knets I

    更新日期:2003-10-01 00:00:00

  • The accuracy of active shape modelling and end-plate measurements for characterising the shape of the lumbar spine in the sagittal plane.

    abstract::The 2D shape of the lumbar spine in the sagittal plane can be determined from lordosis angles measured between the corresponding end-plates of the vertebral bodies or by using an active shape model (ASM) of the vertebral body outline. The ASM was previously shown to be a more efficient and reliable method, but its acc...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2010.518962

    authors: Ali AH,Cowan AB,Gregory JS,Aspden RM,Meakin JR

    更新日期:2012-01-01 00:00:00

  • Stress distribution in cylindrical and conical implants under rotational micromovement with different boundary conditions and bone properties: 3-D FEA.

    abstract::Factors related to micromovements at bone-implant interface have been studied because they are considered adverse to osseointegration. Simplifications are commonly observed in these FEA evaluations. The aim of this study was to clarify the influence of FEA parameters (boundary conditions and bone properties) on the st...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2017.1309394

    authors: Dos Santos MBF,Meloto GO,Bacchi A,Correr-Sobrinho L

    更新日期:2017-06-01 00:00:00

  • A computational study of the EN 1078 impact test for bicycle helmets using a realistic subject-specific finite element head model.

    abstract::In the present study, the free fall impact test in accordance with the EN1078 standard for certification of bicycle helmets is replicated using numerical simulations. The impact scenario is simulated using an experimentally validated, patient-specific head model equipped with and without a bicycle helmet. Head acceler...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2018.1511775

    authors: Sandberg M,Tse KM,Tan LB,Lee HP

    更新日期:2018-09-01 00:00:00

  • New head models extracted from thermal infrared (IR) images for dosimetry computations.

    abstract::In electromagnetic dosimetry, anatomical human models are commonly obtained by segmentation of magnetic resonance imaging or computed tomography scans. In this paper, a human head model extracted from thermal infrared images is examined in terms of its applicability to specific absorption rate (SAR) calculations. Sinc...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2011.563738

    authors: Gasmelseed A

    更新日期:2011-07-01 00:00:00

  • Fibre Orientation in Human Fetal Heart and Ventricular Mechanics : A Small Perturbation Analysis.

    abstract::The study of the topological organisation of myocardial cells is a basic requirement for understanding the mechanical design of the normal and pathological heart. Anatomical observations show that cardiac muscle tissue has a highly specialized architecture. We have made new quantitative measurements of fibre orientati...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255849908907980

    authors: Ohayon J,Usson Y,Jouk PS,Cai H

    更新日期:1999-01-01 00:00:00

  • Finite element analysis of a hemi-pelvis: the effect of inclusion of cartilage layer on acetabular stresses and strain.

    abstract::An appropriate method of application of the hip-joint force and stress analysis of the pelvic bone, in particular the acetabulum, is necessary to investigate the changes in load transfer due to implantation and to calculate the reference stimulus for bone remodelling simulations. The purpose of the study is to develop...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2013.843674

    authors: Ghosh R,Pal B,Ghosh D,Gupta S

    更新日期:2015-01-01 00:00:00

  • Non-Newtonian flow-induced deformation from pressurized cavities in absorbing porous tissues.

    abstract::We investigate the behavior of a spherical cavity in a soft biological tissue modeled as a deformable porous material during an injection of non-Newtonian fluid that follows a power law model. Fluid flows into the neighboring tissue due to high cavity pressure where it is absorbed by capillaries and lymphatics at a ra...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2017.1376323

    authors: Ahmed A,Siddique JI,Mahmood A

    更新日期:2017-10-01 00:00:00

  • An efficient algorithm for retinal blood vessel segmentation using h-maxima transform and multilevel thresholding.

    abstract::Retinal blood vessel detection and analysis play vital roles in early diagnosis and prevention of several diseases, such as hypertension, diabetes, arteriosclerosis, cardiovascular disease and stroke. This paper presents an automated algorithm for retinal blood vessel segmentation. The proposed algorithm takes advanta...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2010.545949

    authors: Saleh MD,Eswaran C

    更新日期:2012-01-01 00:00:00

  • A computational prediction for the effective drug and stem cell treatment of human airway burns.

    abstract::Burns in the airway from inhaling hot gases lead to one of the most common causes of death in the United States. In order to navigate tissues with large burn areas, the velocity, temperature, and heat flux distributions throughout the human airway system are computed for the inhalation of hot air using the finite-elem...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2015.1105966

    authors: Park S

    更新日期:2016-01-01 00:00:00

  • Parametric study of anterior percutaneous endoscopic cervical discectomy (APECD).

    abstract::Anterior percutaneous endoscopic cervical discectomy (APECD) is a common treatment for cervical spondylotic radiculopathy (CSR). In this study, the effects of various channel diameters and approach angles on cervical vertebrae on postoperative outcomes in APECD surgery were explored. A finite element model of intact c...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2020.1846186

    authors: Sun MS,Yuchi CX,Cai XY,Du CF,Mo ZJ

    更新日期:2020-12-01 00:00:00

  • Biomechanical assessment and clinical analysis of different intramedullary nailing systems for oblique fractures.

    abstract::The aim of this study is to evaluate the fracture union or non-union for a specific patient that presented oblique fractures in tibia and fibula, using a mechanistic-based bone healing model. Normally, this kind of fractures can be treated through an intramedullary nail using two possible configurations that depends o...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2015.1125473

    authors: Alierta JA,Pérez MA,Seral B,García-Aznar JM

    更新日期:2016-09-01 00:00:00

  • A two population model of prion transport through a tunnelling nanotube.

    abstract::This article develops a two prion population model that simulates prion trafficking between an infected dendritic cell and a neuron. The situation when the two cells are connected by a tunnelling nanotube (TNT) is simulated. Two mechanisms of prion transport are considered: lateral diffusion in the TNT membrane and ac...

    journal_title:Computer methods in biomechanics and biomedical engineering

    pub_type: 杂志文章

    doi:10.1080/10255842.2013.763938

    authors: Kuznetsov IA,Kuznetsov AV