Abstract:
PURPOSE:In this study, a dedicated device for ion chamber measurements of absorbed dose-to-water for a Nucletron microSelectron-v2 HDR (192)Ir brachytherapy source is presented. The device uses two ionization chambers in a so-called sandwich assembly. Using this setup and by taking the average reading of the two chambers, any dose error due to difficulties in absolute positioning (centering) of the source in between the chambers is cancelled to first order. The method's accuracy was examined by comparing measurements with absorbed dose-to-water determination based on the AAPM TG-43 protocol. METHODS:The optimal source-to-chamber distance (SCD) for (192)Ir dosimetry was determined from ion chamber measurements in a water phantom. The (192)Ir source was sandwiched between two Exradin A1SL chambers (0.057 cm(3)) at the optimal SCD separation. The measured ionization was converted to the absorbed dose-to-water using a (60)Co calibration factor and a Monte Carlo-calculated beam quality conversion factor, kQ, for (60)Co to (192)Ir. An uncertainty estimate of the proposed method was determined based on reproducibility of measurements at different institutions for the same type of source. RESULTS:The optimal distance for the A1SL chamber measurements was determined to be 5 cm from the (192)Ir source center, considering the depth dependency of kQ for (60)Co to (192)Ir and the chamber positioning. The absorbed dose to water measured at (5 cm, 90°) on the transverse axis was 1.3% lower than TG-43 values and its reproducibility and overall uncertainty were 0.8% and 1.7%, respectively. The measurement doses at anisotropic points agreed within 1.5% with TG-43 values. CONCLUSIONS:The ion chamber measurement of absorbed dose-to-water with a sandwich method for the (192)Ir source provides a more accurate, direct, and reference dose compared to the dose-to-water determination based on air-kerma strength in the TG-43 protocol. Due to the simple but accurate assembly, the sandwich measurement method is useful for daily dose management of (192)Ir sources.
journal_name
Med Physjournal_title
Medical physicsauthors
Araki F,Kouno T,Ohno T,Kakei K,Yoshiyama F,Kawamura Sdoi
10.1118/1.4816673subject
Has Abstractpub_date
2013-09-01 00:00:00pages
092101issue
9eissn
0094-2405issn
2473-4209journal_volume
40pub_type
杂志文章相关文献
MEDICAL PHYSICS文献大全abstract::The calibration (in rad/nC) of an air-filled tissue-equivalent ionization chamber for neutrons of 15-MeV average energy was determined by measuring absolute fluences and calculating the kerma per unit fluence. The neutron calibration determined from the gamma-ray calibration and application of the Bragg-Gray relation ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.594310
更新日期:1977-03-01 00:00:00
abstract::Simultaneous rest 99mTc-Sestamibi/ 123I-BMIPP cardiac SPECT imaging has the potential to replace current clinical 99mTc-Sestamibi rest/stress imaging and therefore has great potential in the case of patients with chest pain presenting to the emergency department. Separation of images of these two radionuclides is diff...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3063544
更新日期:2009-02-01 00:00:00
abstract::Several updated Monte Carlo (MC) codes are available to perform calculations of voxel S values for radionuclide targeted therapy. The aim of this work is to analyze the differences in the calculations obtained by different MC codes and their impact on absorbed dose evaluations performed by voxel dosimetry. Voxel S val...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3103401
更新日期:2009-05-01 00:00:00
abstract:PURPOSE:Intensity-modulated proton therapy (IMPT) is known to be sensitive to patient setup and range uncertainty issues. Multiple robust optimization methods have been developed to mitigate the impact of these uncertainties. Here, we propose a new robust optimization method, which provides an alternative way of robust...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13276
更新日期:2019-01-01 00:00:00
abstract::Intrafraction organ motion can limit the advantage of highly conformal dose techniques such as intensity modulated radiation therapy (IMRT) due to target position uncertainty. To ensure high accuracy in beam targeting, real-time knowledge of the target location is highly desired throughout the beam delivery process. T...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2905225
更新日期:2008-05-01 00:00:00
abstract:PURPOSE:Contrast-enhanced imaging of the breast is frequently used in breast MRI and has recently become more common in mammography. The purpose of this study was to make single-scan contrast-enhanced imaging feasible for photon-counting breast CT (pcBCT) and to assess the spectral performance of a pcBCT scanner by eva...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14133
更新日期:2020-07-01 00:00:00
abstract:PURPOSE:To develop a new biological equivalent dose (BED) formulism that includes repair, repopulation and resensitization (redistribution and reoxygenation), and can be used to calculate prescribed dose for permanent implant with Cs-131. METHODS:Because of lack of sufficient clinical data, the prescribed dose with ne...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4735551
更新日期:2012-06-01 00:00:00
abstract:PURPOSE:Partial scan reconstruction (PSR) artifacts are present in myocardial perfusion imaging using dynamic multidetector computed tomography (MDCT). PSR artifacts appear as temporal CT number variations due to inconsistencies in the angular data range used to reconstruct images and compromise the quantitative value ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3665767
更新日期:2012-01-01 00:00:00
abstract:PURPOSE:Brachytherapy with 106 Ru/106 Rh plaques offers good outcomes for small-to-medium choroidal melanomas and retinoblastomas. The dose measurement of the plaques is challenging, due to the small range of the emitted beta particles and steep dose gradients involved. The scarce publications on film dosimetry of 106 ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.12777
更新日期:2018-04-01 00:00:00
abstract::Absorbed doses from main charged-particle beams and charged-particle fragments have been measured with high accuracy for particle therapy, but there are few reports for doses from neutron components produced as fragments. This study describes the measurements on neutron doses produced by carbon beams; microdosimetric ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2767933
更新日期:2007-09-01 00:00:00
abstract:PURPOSE:Upgrade and benchmarking of a research 4D treatment planning system (4DTPS) suitable for realistic patient treatment planning and treatment simulations taking into account specific requirements for scanned ion beam therapy, i.e., modeling of dose heterogeneities due to interplay effects and range changes caused...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4800802
更新日期:2013-05-01 00:00:00
abstract::Generating an optimized radiation treatment plan requires understanding the factors affecting tumour control. Mathematical models of tumour dynamics may help in future studies of factors predicting tumor sensitivity to radiotherapy. In this study, a time-dependent differential model, incorporating biological cancer ma...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2965923
更新日期:2008-07-01 00:00:00
abstract::A method for the in vivo determination of the isocenter dose, Diso, and mid-plane dose, Dm, using the transmitted signal St measured by 25 central pixels of an aSi-based EPID is here reported. The method has been applied to check the conformal radiotherapy of pelvic tumors and supplies accurate in vivo dosimetry avoid...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2360014
更新日期:2006-11-01 00:00:00
abstract::The number of intensity modulated radiation therapy (IMRT) procedures is continuously growing worldwide and it is necessary to develop tools for patient specific quality assurance (QA) that avoid using machine time that could be employed in treating additional patients. One way of achieving this goal is to perform a m...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2815359
更新日期:2008-01-01 00:00:00
abstract::This study compared air exposures during conventional dental and maxillofacial radiography and similar views using free-focus radiography with conventional image receptors. The results show that periapical type surveys on nonscreen film placed extraorally or in the buccal fold may be carried out with an exposure to th...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.594690
更新日期:1980-07-01 00:00:00
abstract:PURPOSE:To use EGSnrc Monte Carlo simulations for magnetic field dosimetry to determine optimal measurement orientations, calculate beam quality conversion factors for 32 cylindrical and three parallel-plate (PP) ion chambers, evaluate the beam quality and angular dependence of these factors, and examine the magnetic f...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.12716
更新日期:2018-02-01 00:00:00
abstract:PURPOSE:This study aims to measure the surface dose on an anthropomorphic phantom for intensity-modulated radiation therapy (IMRT) plans treated in a 1.5 T magnetic resonance (MR)-Linac. Previous studies have used Monte Carlo programs to simulate surface dose and have recognized high surface dose as a potential limitin...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14185
更新日期:2020-07-01 00:00:00
abstract:PURPOSE:To characterize the modulation transfer function (MTF) of proton/carbon radiography using Monte Carlo simulations. To assess the spatial resolution of proton/carbon radiographic imaging. METHODS:A phantom was specifically modeled with inserts composed of two materials with three different densities of bone and...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4819816
更新日期:2013-09-01 00:00:00
abstract:PURPOSE:To obtain on-treatment volumetric patient anatomy during respiratory gated volumetric modulated arc therapy (VMAT). METHODS:On-board imaging device integrated with Linacs offers a viable tool for obtaining patient anatomy during radiation treatment delivery. In this study, the authors acquired beam-level kV im...
journal_title:Medical physics
pub_type: 信件
doi:10.1118/1.4794925
更新日期:2013-04-01 00:00:00
abstract::A comprehensive performance testing program is an essential ingredient of high-quality single-photon emission computed tomography (SPECT). Many of the procedures previously published are complicated, time consuming, or require a special testing environment. This Task Group developed a protocol for evaluating SPECT ima...
journal_title:Medical physics
pub_type: 指南,杂志文章
doi:10.1118/1.597605
更新日期:1995-04-01 00:00:00
abstract:PURPOSE:Stationary computed tomography (s-CT) conceptually offers several advantages over existing rotating gantry-based CT. Over the last 40 yr, s-CT has been investigated using different technological approaches. We are developing a s-CT system specifically for head/brain imaging using carbon nanotube (CNT)-based fie...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14696
更新日期:2020-12-31 00:00:00
abstract::In order to understand the role of water desorption, i.e., the opposite of water adsorption, in optical clearing of biological tissue created by the application of hyperosmotic agents, dynamics of water loss in fresh porcine muscle tissue administrated with glycerol or ethylene glycol was investigated with the near-in...
journal_title:Medical physics
pub_type:
doi:10.1118/1.1576228
更新日期:2003-06-01 00:00:00
abstract:PURPOSE:Cone-beam CT (CBCT) images contain severe shading artifacts mostly due to scatter. Many algorithms have been proposed to alleviate this problem by data correction on projections. Sophisticated methods have also been designed to further improve the image quality when prior patient information is available. In th...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4736218
更新日期:2012-06-01 00:00:00
abstract:PURPOSE:Two types of tumor hypoxia most likely exist in human cancers: Chronic hypoxia due to the paucity of blood capillaries and acute hypoxia due to temporary shutdoWn of microvasculatures or fluctuation in the red cell flux. In a recent hypoxia imaging study using 18F-FMISO PET, the authors observed variation in tr...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3213092
更新日期:2009-10-01 00:00:00
abstract::Our purpose in this paper is to present an in vivo dosimetry program designed both for measuring the rectal dose and for avoiding misadministrations in gynecological intracavitary implants. A device containing an energy compensated diode was specially designed for these measurements. Our calibration procedure as well ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.598598
更新日期:1999-05-01 00:00:00
abstract::An energy discriminating detector for dual-energy radiography can be configured as a two-layer sandwich, where the mean energy of photons detected by the two layers differs. To characterize the quantum noise of such a detector, the noise covariance between the two layers must be known in addition to the noise variance...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.596311
更新日期:1989-11-01 00:00:00
abstract::The visual observer performance for detecting low-contrast patterns by employing both the receiver operating characteristic (ROC) and the 9-alternative forced choice (9-AFC) method is investigated. For each observer and each experimental technique, the results were reproducible with respect to a repeated observation. ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.596986
更新日期:1993-05-01 00:00:00
abstract:PURPOSE:Obtaining an integral depth-dose (IDD) curve using a recently developed acrylic-disk radiation sensor (ADRS) is time-consuming because its single structure requires point-by-point measurements in a water phantom. The goal of this study was to verify the ability of a newly designed multilayer ADRS, composed of 2...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14326
更新日期:2020-09-01 00:00:00
abstract::Radiation therapy treatment modalities continue to develop and have become increasingly complex. With this, dose verification and quality assurance (QA) is of great importance to ensure that a prescribed dose is accurately and precisely delivered to a patient. Radiochromic film dosimetry has been adopted as a convenie...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4740126
更新日期:2012-07-01 00:00:00
abstract::For the case of eye tumor therapy with protons, improvements are introduced compared to the standard dose calculation which implies straight-line optics and the constant-density assumption for the eye and its surrounding. The progress consists of (i) taking account of the lateral scattering of the protons in tissue by...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2168067
更新日期:2006-03-01 00:00:00