X-ray beam filtration, dosimetry phantom size and CT patient dose conversion factors.

Abstract:

:We examine how the choice of CT x-ray beam filtration and phantom size influences patient dose (D) to computed tomography dose index (CTDI) conversion factors (i.e. D/CTDI). The ratio of head to body phantom CTDI(w) for a defined scan technique is alpha, and the ratio of organ dose when the body filter is changed to the head filter is beta. CTDI and organ doses were obtained using the ImPACT CT patient dosimetry calculator, and values of alpha and beta were determined for 39 CT scanners. The average value of alpha for the 39 CT scanners covering a 20 year period was 1.99 +/- 0.23, but 30% of scanners had alpha values that differed by more than 10% from the average. For GE, the value of alpha has been approximately constant at approximately 2.0. Both Philips and Siemens show a definite upward trend from values well below 2.0 in the early 1990s to well over 2.0 for their latest models. The data for Toshiba show no overall trend with time with half the data points below 2.0 and the remainder above this value. The average value of beta was 1.09 +/- 0.25. All vendors showed a downward trend in the beta parameter, and where the most recent scanners from each vendor had a beta value close to unity. Our results show that average D/CTDI conversion factors for a body phantom/filter combination are typically double those appropriate for a head phantom/filter combination.

journal_name

Phys Med Biol

authors

Huda W,Sterzik A,Tipnis S

doi

10.1088/0031-9155/55/2/014

subject

Has Abstract

pub_date

2010-01-21 00:00:00

pages

551-61

issue

2

eissn

0031-9155

issn

1361-6560

pii

S0031-9155(10)29446-0

journal_volume

55

pub_type

杂志文章
  • Positron range estimations with PeneloPET.

    abstract::Technical advances towards high resolution PET imaging try to overcome the inherent physical limitations to spatial resolution. Positrons travel in tissue until they annihilate into the two gamma photons detected. This range is the main detector-independent contribution to PET imaging blurring. To a large extent, it c...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/58/15/5127

    authors: Cal-González J,Herraiz JL,España S,Corzo PM,Vaquero JJ,Desco M,Udias JM

    更新日期:2013-08-07 00:00:00

  • A numerical study on the influence of vulnerable plaque composition on intravascular thermography measurements.

    abstract::Intracoronary thermography is a technique that measures lumen wall temperatures for vulnerable plaque detection. In this paper the influence of vulnerable plaque composition on lumen wall temperatures was studied numerically. Concerning the vulnerable plaque heat generation, the location of the heat source and its hea...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/51/22/010

    authors: ten Have AG,Gijsen FJ,Wentzel JJ,Slager CJ,Serruys PW,van der Steen AF

    更新日期:2006-11-21 00:00:00

  • A multiresolution image based approach for correction of partial volume effects in emission tomography.

    abstract::Partial volume effects (PVEs) are consequences of the limited spatial resolution in emission tomography. They lead to a loss of signal in tissues of size similar to the point spread function and induce activity spillover between regions. Although PVE can be corrected for by using algorithms that provide the correct ra...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/51/7/016

    authors: Boussion N,Hatt M,Lamare F,Bizais Y,Turzo A,Cheze-Le Rest C,Visvikis D

    更新日期:2006-04-07 00:00:00

  • Measurements of angular distributions of Rayleigh and Mie scattering events in biological models.

    abstract::Remission spectroscopy in living tissues can only be performed because scattering processes induce a pronounced amount of backscattered light. New types of scattering chambers were constructed in order to investigate the angular distribution of light intensities caused by scattering events typical for living tissues. ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/34/12/012

    authors: Frank KH,Kessler M,Appelbaum K,Albrecht HP,Mauch ED

    更新日期:1989-12-01 00:00:00

  • An estimation of noise levels in HMPAO RCBF SPECT images using simulation and phantom data; comparison with results obtained from repeated normal controls.

    abstract::The purpose of this work was to determine noise levels in HMPAO RCBF SPECT images. Eight simulated images of a uniform sphere of activity were made at each of three different count levels. Three images of the Amersham brain phantom were obtained at each of three count levels, roughly corresponding to the simulation le...

    journal_title:Physics in medicine and biology

    pub_type: 临床试验,杂志文章

    doi:10.1088/0031-9155/39/5/006

    authors: Houston AS,Kemp PM,Griffiths PT,MacLeod MA

    更新日期:1994-05-01 00:00:00

  • On the correction, perturbation and modification of small field detectors in relative dosimetry.

    abstract::The purpose of this study was to derive a complete set of correction and perturbation factors for output factors (OF) and dose profiles. Modern small field detectors were investigated including a plastic scintillator (Exradin W1, SI), a liquid ionization chamber (microLion 31018, PTW), an unshielded diode (Exradin D1V...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/59/19/5937

    authors: Papaconstadopoulos P,Tessier F,Seuntjens J

    更新日期:2014-10-07 00:00:00

  • The calibration of CT Hounsfield units for radiotherapy treatment planning.

    abstract::Computer tomographic (CT) scans are used to correct for tissue inhomogeneities in radiotherapy treatment planning. In order to guarantee a precise treatment, it is important to obtain the relationship between CT Hounsfield units and electron densities (or proton stopping powers for proton radiotherapy), which is the b...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/41/1/009

    authors: Schneider U,Pedroni E,Lomax A

    更新日期:1996-01-01 00:00:00

  • Examination of errors caused by random summing in in vivo prompt gamma ray analysis.

    abstract::A technique is described for examining the mode of operation of radiation detectors and estimating the losses and gains caused by random summing effects in spectra acquired at high counting rates. The technique is applied to estimating the errors caused by random summing when measuring the amounts of carbon and nitrog...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/34/8/010

    authors: Yankuba S,Oldroyd B,Metcalfe SC,Godfrey JE,Oxby CB

    更新日期:1989-08-01 00:00:00

  • High-resolution numerical model of the middle and inner ear for a detailed analysis of radio frequency absorption.

    abstract::In order to enable a detailed analysis of radio frequency (RF) absorption in the human middle and inner ear organs, a numerical model of these organs was developed at a spatial resolution of 0.1 mm, based on a real human tissue sample. The dielectric properties of the liquids (perilymph and endolymph) inside the bony ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/52/7/001

    authors: Schmid G,Uberbacher R,Samaras T,Jappel A,Baumgartner WD,Tschabitscher M,Mazal PR

    更新日期:2007-04-07 00:00:00

  • Single and multiple cross sections for ionizing processes of biological molecules by protons and alpha-particle impact: a classical Monte Carlo approach.

    abstract::A classical trajectory model has been used to predict total cross sections of single and double ionizing processes (including capture processes) for several ion-biological molecule collisional systems in the intermediate and high energy range. In this work, the systems studied are water, adenine or cytosine targets io...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/53/4/N03

    authors: Abbas I,Champion C,Zarour B,Lasri B,Hanssen J

    更新日期:2008-02-21 00:00:00

  • Fast online Monte Carlo-based IMRT planning for the MRI linear accelerator.

    abstract::The MRI accelerator, a combination of a 6 MV linear accelerator with a 1.5 T MRI, facilitates continuous patient anatomy updates regarding translations, rotations and deformations of targets and organs at risk. Accounting for these demands high speed, online intensity-modulated radiotherapy (IMRT) re-optimization. In ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/57/5/1375

    authors: Bol GH,Hissoiny S,Lagendijk JJ,Raaymakers BW

    更新日期:2012-03-07 00:00:00

  • Commissioning of pencil beam and Monte Carlo dose engines for non-isocentric treatments in scanned proton beam therapy.

    abstract::This work describes the dosimetric commissioning of the treatment planning system (TPS) RayStation v6.1 from RaySearch Laboratories (Stockholm, Sweden) for a synchrotron-based scanned proton beam delivery with isocentric and non-isocentric setups at MedAustron. Focus was on the comparison of the pencil beam (PBv4.1) a...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab3557

    authors: Carlino A,Böhlen T,Vatnitsky S,Grevillot L,Osorio J,Dreindl R,Palmans H,Stock M,Kragl G

    更新日期:2019-08-28 00:00:00

  • Design of applicators for a 27 MHz multielectrode current source interstitial hyperthermia system; impedance matching and effective power.

    abstract::In interstitial heating one of the main requirements for achieving a certain elevated temperature in a tumour is that the effective power per applicator (Peff), i.e. the power which is actually deposited in the tissue, is sufficiently high. In this paper this requirement is discussed for the applicators of the 27 MHz ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/42/6/007

    authors: Kaatee RS,Crezee J,Kanis AP,Lagendijk JJ,Levendag PC,Visser AG

    更新日期:1997-06-01 00:00:00

  • Calculation of 10 MV x-ray spectra emitted by a medical linear accelerator using the BFGS quasi-Newton method.

    abstract::To calculate photon spectra for a 10 MV x-ray beam emitted by a medical linear accelerator, we performed numerical analysis using the aluminium transmission data obtained along the central axis of the beam under the narrow beam condition corresponding to a 3x3 cm2 field at a 100 cm distance from the source. We used th...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/52/2/014

    authors: Shimozato T,Tabushi K,Kitoh S,Shiota Y,Hirayama C,Suzuki S

    更新日期:2007-01-21 00:00:00

  • Performance of a multi-axis ionization chamber array in a 1.5 T magnetic field.

    abstract::At the UMC Utrecht a prototype MR-linac has been installed. The system consists of an 8 MV Elekta linear accelerator and a 1.5 T Philips MRI system. This paper investigates the performance of the IC PROFILER™, a multi-axis ionization chamber array, in a 1.5 T magnetic field. The influence of the magnetic field on the ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/59/7/1845

    authors: Smit K,Kok JG,Lagendijk JJ,Raaymakers BW

    更新日期:2014-04-07 00:00:00

  • Dual energy x-ray laser measurement of calcaneal bone mineral density.

    abstract::In dual energy x-ray absorptiometry (DXA) the photon attenuation is assumed to be similar in soft tissue overlying, adjacent to and inside the measured bone. In the calcaneal dual energy x-ray laser (DXL) technique, this assumption is not needed as attenuation by soft tissues at the local bone site is determined by co...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/48/12/305

    authors: Hakulinen MA,Saarakkala S,Töyräs J,Kröger H,Jurvelin JS

    更新日期:2003-06-21 00:00:00

  • CT reconstruction algorithm for a dental panoramic x-ray unit.

    abstract::A variable Jacobian and weighted backprojection algorithm, used for medical CT, was adapted to perform CT reconstructions on data obtained with a dental panoramic x-ray unit. A detector array, fitted to the unit for the purpose of acquiring digital panoramic radiographs, was used to collect the data. Compensations wer...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/37/12/001

    authors: Lee MD,Waggener RG,McDavid WD,Dove SB,Welander U,Tronje G

    更新日期:1992-12-01 00:00:00

  • An iodine-125 thyroid measurement method.

    abstract::A two-probe technique has been developed to measure 125I in the thyroid. In this technique simultaneous emissions from a source are detected as coincidence events between two probes rather than as a sum peak in a single probe (the sum peak method). The background count rate when counting coincidence events between pro...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/25/3/004

    authors: Burns PA,Peggie JR

    更新日期:1980-05-01 00:00:00

  • Fluence correction factors in plastic phantoms for clinical proton beams.

    abstract::In recent codes of practice for reference dosimetry in clinical proton beams using ionization chambers, it is recommended to perform the measurement in a water phantom. However, in situations where the positioning accuracy is very critical, it could be more convenient to perform the measurement in a plastic phantom. I...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/47/17/302

    authors: Palmans H,Symons JE,Denis JM,de Kock EA,Jones DT,Vynckier S

    更新日期:2002-09-07 00:00:00

  • Monte Carlo simulations of a nozzle for the treatment of ocular tumours with high-energy proton beams.

    abstract::By the end of 2002, 33 398 patients worldwide had been treated with proton radiotherapy, 10 829 for eye diseases. The dose prediction algorithms used today for ocular proton therapy treatment planning rely on parameterizations of measured proton dose distributions, i.e., broad-beam and pencil-beam techniques, whose pr...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/50/22/002

    authors: Newhauser W,Koch N,Hummel S,Ziegler M,Titt U

    更新日期:2005-11-21 00:00:00

  • Analytic treatment of resolution precision in electronically collimated SPECT imaging involving multiple-interaction gamma rays.

    abstract::The widely applied single-interaction analytic expression characterizing the energy resolution component of the angular resolution precision for an electronically collimated point source is extended to include multiple-interaction Compton scatter sequences as well as sequences terminated by photoelectric absorption. T...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/43/10/021

    authors: Durkee JW Jr,Antich PP,Tsyganov EN,Constantinescu A,Kulkarni PV,Smith BJ,Arbique GM,Lewis MA,Nguyen T,Raheja A,Thambi G,Parkey RW

    更新日期:1998-10-01 00:00:00

  • Learning image based surrogate relevance criterion for atlas selection in segmentation.

    abstract::Picking geometrically relevant atlases from the whole training set is crucial to multi-atlas based image segmentation, especially with extensive data of heterogeneous quality in the Big Data era. Unfortunately, there is very limited understanding of how currently used image similarity criteria reveal geometric relevan...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/61/11/4223

    authors: Zhao T,Ruan D

    更新日期:2016-06-07 00:00:00

  • The assessment of the non-equilibrium effect in the 'Patlak analysis' of Fdopa PET studies.

    abstract::'Patlak analysis' is a common approach used in Fdopa PET studies to calculate the uptake constant (Ki) of the tracer. It is assumed in the Patlak analysis that the reversible compartment of tissue radioactivity used is in effective equilibrium with the tracer in plasma. Therefore, using the data prior to equilibrium i...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/40/7/007

    authors: Yu DC,Huang SC,Barrio JR,Phelps ME

    更新日期:1995-07-01 00:00:00

  • Comparison of risk of radiogenic second cancer following photon and proton craniospinal irradiation for a pediatric medulloblastoma patient.

    abstract::Pediatric patients who received radiation therapy are at risk of developing side effects such as radiogenic second cancer. We compared proton and photon therapies in terms of the predicted risk of second cancers for a 4 year old medulloblastoma patient receiving craniospinal irradiation (CSI). Two CSI treatment plans ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/58/4/807

    authors: Zhang R,Howell RM,Giebeler A,Taddei PJ,Mahajan A,Newhauser WD

    更新日期:2013-02-21 00:00:00

  • Improving SAR estimations in MRI using subject-specific models.

    abstract::To monitor and strategically control energy deposition in magnetic resonance imaging (MRI), measured as a specific absorption rate (SAR), numerical methods using generic human models have been employed to estimate worst-case values. Radiofrequency (RF) sequences are therefore often designed conservatively with large s...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/57/24/8153

    authors: Jin J,Liu F,Weber E,Crozier S

    更新日期:2012-12-21 00:00:00

  • Bulk optical properties of healthy female breast tissue.

    abstract::We have measured the bulk optical properties of healthy female breast tissues in vivo in the parallel plate, transmission geometry. Fifty-two volunteers were measured. Blood volume and blood oxygen saturation were derived from the optical property data using a novel method that employed a priori spectral information t...

    journal_title:Physics in medicine and biology

    pub_type: 临床试验,杂志文章

    doi:10.1088/0031-9155/47/16/302

    authors: Durduran T,Choe R,Culver JP,Zubkov L,Holboke MJ,Giammarco J,Chance B,Yodh AG

    更新日期:2002-08-21 00:00:00

  • Effect of blood perfusion on the ablation of liver parenchyma with high-intensity focused ultrasound.

    abstract::This paper discusses the effect of blood perfusion on the ablation of rat liver tissue with high-intensity focused ultrasound (HIFU). For this study a practical method has been developed, in which the liver blood flow can be reduced by ligation of the hepatic artery and portal vein. During the treatment the rat liver ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/38/11/011

    authors: Chen L,ter Haar G,Hill CR,Dworkin M,Carnochan P,Young H,Bensted JP

    更新日期:1993-11-01 00:00:00

  • Acquiring a four-dimensional computed tomography dataset using an external respiratory signal.

    abstract::Four-dimensional (4D) methods strive to achieve highly conformal radiotherapy, particularly for lung and breast tumours, in the presence of respiratory-induced motion of tumours and normal tissues. Four-dimensional radiotherapy accounts for respiratory motion during imaging, planning and radiation delivery, and requir...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/48/1/304

    authors: Vedam SS,Keall PJ,Kini VR,Mostafavi H,Shukla HP,Mohan R

    更新日期:2003-01-07 00:00:00

  • Influence of the physical state and straggling on the computation of the radiation dose due to radon daughters deposited in the lung.

    abstract::The effect of the physical state (phase) of the absorbing medium and the energy straggling of the alpha particles on the calculation of the radiation dose due to the daughter products of radon deposited in the lung have been studied in detail. The stopping power data for alpha particles in water and water vapour have ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/32/8/004

    authors: Haque AK,Al-affan IA

    更新日期:1987-08-01 00:00:00

  • Measurement of patient entrance surface dose rates for fluoroscopic x-ray units.

    abstract::Measurements of patient entrance surface dose rate provide valuable data for interpreting results from dose-area product studies on fluoroscopic x-ray equipment. Methods for measurement of entrance surface dose rate with backscatter and incident dose rate without backscatter have been investigated. Entrance surface do...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/40/5/008

    authors: Martin CJ

    更新日期:1995-05-01 00:00:00