Abstract:
BACKGROUND:The aim of this study was to validate the Omron HEM-907 blood pressure (BP) measuring device according to the international validation protocol. DESIGN:The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 more subjects are selected making a total number of 33 subjects on which the final validation is performed. METHODS:For each subject, BP measurements were performed simultaneously by two trained observers using mercury sphygmomanometers alternately with the Omron HEM-907 device. In all, 99 measurements were obtained for comparison. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. RESULTS:The difference between the two observers was -1 +/- 2 mmHg for the systolic BP (SBP) and for the diastolic BP (DBP). The Omron HEM-907 passed the first phase of the validation process. For the second phase, the average differences between the device and mercury sphygmomanometer readings were -1 +/- 7 and -5 +/- 6 mmHg for SBP and DBP respectively. Readings for the HEM-907 device differed by less than 5 mmHg for 61 of the systolic readings and 52 of the diastolic readings; by less than 10 mmHg for 85 of the systolic readings and 85 of the diastolic readings; and by less than 15 mmHg for 94 of the systolic readings and 96 of the diastolic readings. CONCLUSIONS:The Omron HEM-907 device passes the two phases of the international validation protocol.
journal_name
Blood Press Monitjournal_title
Blood pressure monitoringauthors
El Assaad MA,Topouchian JA,Darné BM,Asmar RGdoi
10.1097/00126097-200208000-00006subject
Has Abstractpub_date
2002-08-01 00:00:00pages
237-41issue
4eissn
1359-5237issn
1473-5725journal_volume
7pub_type
杂志文章abstract:AIM:Blood pressure (BP), a key vital sign, monitors general health. Oscillometric devices are increasingly used for measurement, although their accuracy continues to be critically debated. A functional block diagram is used to review the components that affect accuracy. METHODS:A block diagram is presented covering th...
journal_title:Blood pressure monitoring
pub_type: 杂志文章,评审
doi:10.1097/MBP.0b013e32835026b0
更新日期:2012-04-01 00:00:00
abstract::In 2004-2005, the antihypertensive effects of telmisartan 80 mg versus valsartan 160 mg combined with hydrochlorothiazide 25 mg were assessed in a large placebo-controlled trial in patients with stages 1 and 2 hypertension and demonstrated that both agents were highly effective in lowering blood pressure (BP) compared...
journal_title:Blood pressure monitoring
pub_type: 杂志文章,随机对照试验
doi:10.1097/MBP.0b013e3282f3859d
更新日期:2008-02-01 00:00:00
abstract:OBJECTIVE:Vitamin D has beneficial effects on vascular endothelial function, blood pressure (BP) and arterial stiffness. Arterial stiffness increases in early-stage hypertensive patients and it is a strong predictor of cardiovascular morbidity and mortality. The purpose of this study was to assess the association betwe...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/MBP.0000000000000497
更新日期:2020-11-23 00:00:00
abstract:OBJECTIVES:The objective of this work was to test the hypothesis that the Finapres underestimates blood pressure during sympathetic peripheral vasoconstriction. METHODS:Measurements were made simultaneously with two Finapres devices and one radial artery tonometer during the onset of periodic lower body negative press...
journal_title:Blood pressure monitoring
pub_type: 临床试验,杂志文章
doi:10.1097/MBP.0b013e3282c9acc3
更新日期:2007-10-01 00:00:00
abstract::Masked hypertension (blood pressure that is normal in the physicians' office but elevated elsewhere) is a common phenomenon as prevalence among studies varies from 8 to 45% and is seen at all ages. large discrepancies, however, exist between studies that have dealt with masked hypertension. It is of high clinical impo...
journal_title:Blood pressure monitoring
pub_type: 杂志文章,评审
doi:10.1097/mbp.0b013e3280fb2792
更新日期:2007-08-01 00:00:00
abstract:BACKGROUND:Manufacturer-supplied blood pressure (BP) cuffs are part of the automatic oscillometric BP devices algorithm. MATERIALS AND METHODS:This study assessed the differences in BP values using the Omron HEM 907-XL (Omron) device with two types of cuffs: the Baum cuff (BC) and the supplied Omron cuff (OC). A sampl...
journal_title:Blood pressure monitoring
pub_type: 杂志文章,随机对照试验
doi:10.1097/MBP.0000000000000208
更新日期:2016-12-01 00:00:00
abstract:OBJECTIVE:Previous studies have indicated moderate-to-poor reproducibility of heart rate variability (HRV) but the reproducibility of blood pressure variability (BPV) and spectral measures of baroreceptor sensitivity (BRS) are not well established. METHODS:We measured normal-to-normal heart beat (RR) interval and fing...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/00126097-200502000-00005
更新日期:2005-02-01 00:00:00
abstract::BACKGROUND: Blood pressure measurements taken in the clinic or laboratory are assumed to generalize to the world outside. This is true both of casual blood pressure measurements and of changes in blood pressure responses to stress. Such generalizability is crucial to the usefulness of blood pressure measurements as pr...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:
更新日期:1996-12-01 00:00:00
abstract::The sympathetic nervous system is intimately involved in both the origin and the perpetuation of the hypertensive state, although there is some debate as to its precise contribution to the long-term hypertensive condition. Catecholamines are leading candidates to be both the pressor mechanism that initiates the rise i...
journal_title:Blood pressure monitoring
pub_type: 杂志文章,评审
doi:10.1097/00126097-200010002-00005
更新日期:2000-01-01 00:00:00
abstract:BACKGROUND:Data on the reproducibility of serial measurements of ambulatory blood pressure in hypertensive patients are lacking. The purpose of this study was to examine (1) the reproducibility of four consecutive ambulatory blood pressure measurements, and (2) the reproducibility of nocturnal falls in blood pressure i...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/00126097-200102000-00007
更新日期:2001-02-01 00:00:00
abstract:OBJECTIVES:Vitamin D deficiency and high parathyroid hormone (PTH) levels have been linked with hypertension. Nondipper hypertension is associated with increased morbidity and mortality. We aimed to investigate the relationship of vitamin D and PTH levels with nondipper hypertension and nocturnal decline in untreated h...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/MBP.0000000000000286
更新日期:2017-12-01 00:00:00
abstract::Thomas G. Pickering spent most of his scientific career in carrying out research on clinical hypertension and blood pressure (BP) measurement. In our review of Pickering's seminal work, we first focused on white-coat hypertension and masked hypertension, two terms that he had introduced. Next, we highlighted the early...
journal_title:Blood pressure monitoring
pub_type: 传,历史文章,杂志文章
doi:10.1097/MBP.0b013e3283380e1e
更新日期:2010-04-01 00:00:00
abstract::BACKGROUND: Hypertension is responsible for substantial morbidity in end-stage renal disease. Ambulatory blood pressure monitoring has been used to understand the behavior of blood pressure in renal failure, especially during the interdialytic period for patients on hemodialysis. OBJECTIVE: To review the published dat...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:
更新日期:1997-12-01 00:00:00
abstract:BACKGROUND:We have developed a new blood pressure (BP)-measuring device with a double-cuff, and evaluated its accuracy and applicability for practical use. DESIGN:The double-cuff method features a small cuff placed in the centre of the compression cuff to detect the oscillation of arterial pulsation, by which algorith...
journal_title:Blood pressure monitoring
pub_type: 临床试验,杂志文章
doi:10.1097/00126097-200310000-00005
更新日期:2003-10-01 00:00:00
abstract:BACKGROUND:In acute ischemic stroke, a transient elevation of blood pressure is common; its significance and its relationship with the neurological outcome are still unclear. METHODS:In 71 consecutive patients with acute ischemic stroke, aged 25-94 years, admitted to our unit, we investigated the relationships between...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/01.mbp.0000209077.23084.93
更新日期:2006-08-01 00:00:00
abstract:OBJECTIVE:The aim of the present study was to validate the blood pressure (BP) measurement device Erkameter 125 PRO according to the International Protocol revision 2010 of the European Society of Hypertension (ESH-IP2). PATIENTS AND METHODS:In 33 patients aged 32-79 years, BP measurements were performed alternately u...
journal_title:Blood pressure monitoring
pub_type: 临床试验,杂志文章
doi:10.1097/MBP.0000000000000216
更新日期:2016-12-01 00:00:00
abstract:OBJECTIVE:The ankle-brachial index (ABI) is the most useful and efficient tool for assessing the presence of peripheral artery disease (PAD). The aim of this study was to evaluate the effect of multiple sequential blood pressure (BP) measurements on the estimation of the ABI and the consequent diagnosis of PAD in a hig...
journal_title:Blood pressure monitoring
pub_type: 杂志文章,随机对照试验
doi:10.1097/MBP.0b013e328351de79
更新日期:2012-04-01 00:00:00
abstract:OBJECTIVE:To investigate the influence of arm circumference (AC) on the brachial blood pressure (BP) measured with an adult cuff. PARTICIPANTS AND METHODS:This study included 208 patients (150 men, 61.0±8.0 years) for coronary angiography. Intra-aortic BP and noninvasive right brachial oscillometric BP (noninvasive BP...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/MBP.0000000000000141
更新日期:2015-12-01 00:00:00
abstract:BACKGROUND:Patients and doctors often use home blood pressure monitoring (HBPM) to assess the control of hypertension. Despite its popularity there has always been some uncertainty with regard to its accuracy, reliability, reproducibility, and comparability. Although there are pre-market HBPM standards of accuracy, the...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/00126097-199904010-00007
更新日期:1999-02-01 00:00:00
abstract:BACKGROUND:Measurement of blood pressure remains the most commonly performed screening test in medical practice. With the likely removal of mercury sphygmomanometers from the workplace alternative devices are required. Of these the aneroid sphygmomanometer is popular both in the community and hospital setting. We inves...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/00126097-200212000-00003
更新日期:2002-12-01 00:00:00
abstract:OBJECTIVES:Nondipping pattern of circadian blood pressure (BP) is associated with increased cardiovascular morbidity and mortality; however, limited data are available among obese African-Americans. We, therefore, aimed to evaluate the pattern of circadian BP variation and to identify clinical conditions associated wit...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/MBP.0b013e328346a871
更新日期:2011-06-01 00:00:00
abstract:BACKGROUND:Ambulatory blood pressure monitoring allows characterization of the patient's average blood pressure load as well as various profiles of blood pressure. Results from clinical studies suggest that dippers (patients whose blood pressures decrease during night-time) have a lower risk of cardiovascular events th...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/00126097-200002000-00004
更新日期:2000-02-01 00:00:00
abstract:OBJECTIVES:A number of operator-independent oscillometric devices to measure hemodynamics and arterial stiffness became available recently, but some and in particular VaSera VS-1500 do not provide estimates of aortic pressures and aortic pulse wave velocity (aPWV). The aim of this work was the retrospective application...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/MBP.0000000000000137
更新日期:2015-10-01 00:00:00
abstract:BACKGROUND:The accuracy of home blood pressure measurement (HBPM) depends on adherence to the measurement schedule. We investigated the number of deviations from the requested schedule using an HBPM device equipped with a diagnostic mode that only allows patients to take a fixed number of BP readings at preset times. ...
journal_title:Blood pressure monitoring
pub_type: 杂志文章,随机对照试验
doi:10.1097/MBP.0b013e328357352a
更新日期:2012-10-01 00:00:00
abstract:OBJECTIVES:Wide-range 'one-size-fits-all' blood pressure (BP) cuffs are commonly sold with home BP monitors. Assessment of the accuracy of these cuffs is important because they do not adhere to the basic principles of proper cuffing. The aim of this systematic review was to review the published validation data evaluati...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/MBP.0000000000000344
更新日期:2018-10-01 00:00:00
abstract:BACKGROUND:To test the accuracy of a new algorithm for the BPM-100, an automated oscillometric blood pressure (BP) monitor, using stored data from an independently conducted validation trial comparing the BPM-100(Beta) with a mercury sphygmomanometer. DESIGN:Raw pulse wave and cuff pressure data were stored electronic...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/00126097-200106000-00008
更新日期:2001-06-01 00:00:00
abstract::Hypertension is highly prevalent in pediatric patients with end-stage renal disease (ESRD). Ambulatory blood pressure monitoring (ABPM) has been used to characterize the altered circadian blood pressure patterns in ESRD in order to improve our understanding of the blood pressure dysregulation that occurs in these chil...
journal_title:Blood pressure monitoring
pub_type: 杂志文章,评审
doi:
更新日期:1999-06-01 00:00:00
abstract:OBJECTIVE:The validation of sphygmomanometer is important in accurate blood pressure measurement. This study presents the validation results by the Medipro MediCare 100f upper arm blood pressure monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010. METHODS:The ESH-IP ...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/MBP.0b013e328346d715
更新日期:2011-08-01 00:00:00
abstract:OBJECTIVES:The purpose of this study was to validate the ScottCare 320 ambulatory blood pressure monitor (ABPM) using both group-level and individual-level validation procedures. The group-level validation followed a modified protocol of the European Hypertension Society's validation protocol. The individual-level vali...
journal_title:Blood pressure monitoring
pub_type: 临床试验,杂志文章
doi:10.1097/MBP.0b013e32835f4fc7
更新日期:2013-06-01 00:00:00
abstract:OBJECTIVES:To determine the accuracy of the UA-774 (UA-767Plus) blood pressure monitor developed by the A&D Company. METHODS:Device evaluation was performed according to the modified British Hypertension Society (BHS) protocol released in 1993. Eighty-five subjects with characteristics outlined in the BHS protocol wer...
journal_title:Blood pressure monitoring
pub_type: 杂志文章
doi:10.1097/00126097-200408000-00008
更新日期:2004-08-01 00:00:00