Multileaf collimator performance monitoring and improvement using semiautomated quality control testing and statistical process control.

Abstract:

PURPOSE:High-quality radiation therapy using highly conformal dose distributions and image-guided techniques requires optimum machine delivery performance. In this work, a monitoring system for multileaf collimator (MLC) performance, integrating semiautomated MLC quality control (QC) tests and statistical process control tools, was developed. The MLC performance monitoring system was used for almost a year on two commercially available MLC models. Control charts were used to establish MLC performance and assess test frequency required to achieve a given level of performance. MLC-related interlocks and servicing events were recorded during the monitoring period and were investigated as indicators of MLC performance variations. METHODS:The QC test developed as part of the MLC performance monitoring system uses 2D megavoltage images (acquired using an electronic portal imaging device) of 23 fields to determine the location of the leaves with respect to the radiation isocenter. The precision of the MLC performance monitoring QC test and the MLC itself was assessed by detecting the MLC leaf positions on 127 megavoltage images of a static field. After initial calibration, the MLC performance monitoring QC test was performed 3-4 times/week over a period of 10-11 months to monitor positional accuracy of individual leaves for two different MLC models. Analysis of test results was performed using individuals control charts per leaf with control limits computed based on the measurements as well as two sets of specifications of ± 0.5 and ± 1 mm. Out-of-specification and out-of-control leaves were automatically flagged by the monitoring system and reviewed monthly by physicists. MLC-related interlocks reported by the linear accelerator and servicing events were recorded to help identify potential causes of nonrandom MLC leaf positioning variations. RESULTS:The precision of the MLC performance monitoring QC test and the MLC itself was within ± 0.22 mm for most MLC leaves and the majority of the apparent leaf motion was attributed to beam spot displacements between irradiations. The MLC QC test was performed 193 and 162 times over the monitoring period for the studied units and recalibration had to be repeated up to three times on one of these units. For both units, rate of MLC interlocks was moderately associated with MLC servicing events. The strongest association with the MLC performance was observed between the MLC servicing events and the total number of out-of-control leaves. The average elapsed time for which the number of out-of-specification or out-of-control leaves was within a given performance threshold was computed and used to assess adequacy of MLC test frequency. CONCLUSIONS:A MLC performance monitoring system has been developed and implemented to acquire high-quality QC data at high frequency. This is enabled by the relatively short acquisition time for the images and automatic image analysis. The monitoring system was also used to record and track the rate of MLC-related interlocks and servicing events. MLC performances for two commercially available MLC models have been assessed and the results support monthly test frequency for widely accepted ± 1 mm specifications. Higher QC test frequency is however required to maintain tighter specification and in-control behavior.

journal_name

Med Phys

journal_title

Medical physics

authors

Létourneau D,Wang A,Amin MN,Pearce J,McNiven A,Keller H,Norrlinger B,Jaffray DA

doi

10.1118/1.4901520

subject

Has Abstract

pub_date

2014-12-01 00:00:00

pages

121713

issue

12

eissn

0094-2405

issn

2473-4209

journal_volume

41

pub_type

杂志文章
  • Comparative performance of multiview stereoscopic and mammographic display modalities for breast lesion detection.

    abstract:PURPOSE:Mammography is known to be one of the most difficult radiographic exams to interpret. Mammography has important limitations, including the superposition of normal tissue that can obscure a mass, chance alignment of normal tissue to mimic a true lesion and the inability to derive volumetric information. It has b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3562901

    authors: Webb LJ,Samei E,Lo JY,Baker JA,Ghate SV,Kim C,Soo MS,Walsh R

    更新日期:2011-04-01 00:00:00

  • Thermoluminescent dosimetry of the SourceTech Medical model STM1251 125I seed.

    abstract::Many new models of 125I seeds are being introduced, mainly due to the increase in prostate seed implants. We have evaluated the SourceTech Medical (STM), model STM1251, 125I seed using thermoluminescent dosimeters (TLDs) in a solid water phantom. TLD cubes, LiF TLD-100, with dimension 1 mm on each edge, were irradiate...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1577251

    authors: Chiu-Tsao ST,Duckworth TL,Hsiung CY,Li Z,Williamson J,Patel NS,Harrison LB

    更新日期:2003-07-01 00:00:00

  • Selective enhancement filters for nodules, vessels, and airway walls in two- and three-dimensional CT scans.

    abstract::Computer-aided diagnostic (CAD) schemes have been developed to assist radiologists in the early detection of lung cancer in radiographs and computed tomography (CT) images. In order to improve sensitivity for nodule detection, many researchers have employed a filter as a preprocessing step for enhancement of nodules. ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1581411

    authors: Li Q,Sone S,Doi K

    更新日期:2003-08-01 00:00:00

  • MR acoustic radiation force imaging: in vivo comparison to ultrasound motion tracking.

    abstract::MR acoustic radiation force (ARF) imaging was developed for measuring tiSsue elastic properties using focused ultrasound to deliver a localized tissue motion. In this study, an imaging ultrasound transducer was mounted on the focused ultrasound transducer and ultrasound motion tracking was performed simultaneously to ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3120289

    authors: Huang Y,Curiel L,Kukic A,Plewes DB,Chopra R,Hynynen K

    更新日期:2009-06-01 00:00:00

  • Calculation of photon energy deposition kernels and electron dose point kernels in water.

    abstract::Effects of changes in the physics of EGSnrc compared to EGS4/PRESTA on energy deposition kernels for monoenergetic photons and on dose point kernels for beta sources in water are investigated. In the diagnostic energy range, Compton binding corrections were found to increase the primary energy fraction up to 4.5% at 3...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1861412

    authors: Mainegra-Hing E,Rogers DW,Kawrakow I

    更新日期:2005-03-01 00:00:00

  • Fast time-of-flight camera based surface registration for radiotherapy patient positioning.

    abstract:PURPOSE:This work introduces a rigid registration framework for patient positioning in radiotherapy, based on real-time surface acquisition by a time-of-flight (ToF) camera. Dynamic properties of the system are also investigated for future gating/tracking strategies. METHODS:A novel preregistration algorithm, based on...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3664006

    authors: Placht S,Stancanello J,Schaller C,Balda M,Angelopoulou E

    更新日期:2012-01-01 00:00:00

  • A novel approach in electron beam radiation therapy of lips carcinoma: a Monte Carlo study.

    abstract:PURPOSE:Squamous cell carcinoma (SCC) is commonly treated by electron beam radiotherapy (EBRT) followed by a boost via brachytherapy. Considering the limitations associated with brachytherapy, in this study, a novel boosting technique in EBRT of lip carcinoma using an internal shield as an internal dose enhancer tool (...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4795756

    authors: Shokrani P,Baradaran-Ghahfarokhi M,Zadeh MK

    更新日期:2013-04-01 00:00:00

  • "Battered data"--some clinical effects of the abuse of multiple regression methods: the NSD.

    abstract::The NSD equation, D = 1850 x T0.11 x N0.24, is a celebrated transmogrification of Cohen's two well-known collations of data on response to clinical irradiations (3 degrees erythema and 0.90 ablation of skin cancer) in which the relation of D and T is fixed by the data selected by Cohen and the additional constraint th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595034

    authors: Herbert DE Jr

    更新日期:1981-11-01 00:00:00

  • Patch-based generation of a pseudo CT from conventional MRI sequences for MRI-only radiotherapy of the brain.

    abstract:PURPOSE:In radiotherapy (RT) based on magnetic resonance imaging (MRI) as the only modality, the information on electron density must be derived from the MRI scan by creating a so-called pseudo computed tomography (pCT). This is a nontrivial task, since the voxel-intensities in an MRI scan are not uniquely related to e...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4914158

    authors: Andreasen D,Van Leemput K,Hansen RH,Andersen JA,Edmund JM

    更新日期:2015-04-01 00:00:00

  • A new framework of image reconstruction from fan beam projections.

    abstract::In this paper, we present a unified framework to reconstruct images for both fan beam and cone beam projections. The important feature of our theoretical framework is that it does not depend on the classical concept of the Radon transform at all. This property allows us to directly generalize the ideas and techniques ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1577252

    authors: Chen GH

    更新日期:2003-06-01 00:00:00

  • Carpal tunnel syndrome diagnosis by a self-normalization process and ultrasound compound imaging.

    abstract:PURPOSE:Carpal tunnel syndrome (CTS) is the common entrapment neuropathy that occurs due to compression of the median nerve at the wrist. Ultrasound images have been used to highlight anatomical variants of the median nerve, and CTS is thought to be associated to enlargement of the cross-sectional area (CSA) of the med...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4767754

    authors: Liao YY,Wu CC,Kuo TT,Chen JP,Hsu YW,Yeh CK

    更新日期:2012-12-01 00:00:00

  • Neutron energies in medical electron accelerator rooms.

    abstract::Two experimental methods are described for obtaining average energies of neutron spectra produced by an electron accelerator in a medical treatment room. In the first method, the dose equivalent is measured by means of indium activation foils in an Andersson-Braun moderator. Fluences are determined by the usual activa...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595662

    authors: LaRiviere PD

    更新日期:1985-11-01 00:00:00

  • Effect of finite sample size on feature selection and classification: a simulation study.

    abstract:PURPOSE:The small number of samples available for training and testing is often the limiting factor in finding the most effective features and designing an optimal computer-aided diagnosis (CAD) system. Training on a limited set of samples introduces bias and variance in the performance of a CAD system relative to that...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3284974

    authors: Way TW,Sahiner B,Hadjiiski LM,Chan HP

    更新日期:2010-02-01 00:00:00

  • Development and evaluation of a case-based reasoning classifier for prediction of breast biopsy outcome with BI-RADS lexicon.

    abstract::Approximately 70-85% of breast biopsies are performed on benign lesions. To reduce this high number of biopsies performed on benign lesions, a case-based reasoning (CBR) classifier was developed to predict biopsy results from BI-RADS findings. We used 1433 (931 benign) biopsy-proven mammographic cases. CBR similarity ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1501140

    authors: Bilska-Wolak AO,Floyd CE Jr

    更新日期:2002-09-01 00:00:00

  • Determining Pion, the correction factor for recombination losses in an ionization chamber.

    abstract::The 1983 AAPM protocol for the determination of absorbed dose from high-energy photon and electron beams recommends using Pion (the reciprocal of collection efficiency), as determined by the two-voltage technique, to correct for recombination losses in ionization chambers. Methods and data for the determination of ion...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595574

    authors: Weinhous MS,Meli JA

    更新日期:1984-11-01 00:00:00

  • Technical Note: Gray tracking in medical color displays-A report of Task Group 196.

    abstract:PURPOSE:The authors discuss measurement methods and instrumentation useful for the characterization of the gray tracking performance of medical color monitors for diagnostic applications. The authors define gray tracking as the variability in the chromaticity of the gray levels in a color monitor. METHODS:The authors ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4953186

    authors: Badano A,Wang J,Boynton P,Le Callet P,Cheng WC,Deroo D,Flynn MJ,Matsui T,Penczek J,Revie C,Samei E,Steven PM,Swiderski S,Van Hoey G,Yamaguchi M,Hasegawa M,Nagy BV

    更新日期:2016-07-01 00:00:00

  • An approximate short scan helical FDK cone beam algorithm based on nutating curved surfaces satisfying the Tuy's condition.

    abstract::Traditionally, short scan helical FDK algorithms have been implemented based on horizontal transaxial slices. However, not every point on the horizontal transaxial slice satisfies Tuy's condition for the corresponding (pi+fan angle) segment of helix, which means that some points on the horizontal slices are incomplete...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1916077

    authors: Hu J,Tam K,Qi J

    更新日期:2005-06-01 00:00:00

  • Calculation of monitor units for a linear accelerator with asymmetric jaws.

    abstract::A simple approach was developed that calculates the output factors and tissue maximum ratio of an asymmetric field at any point within the open field, and specifically both at the central axis (when the jaws do not shadow it) and at the effective center of the open field, using the existing tables for symmetric fields...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597524

    authors: Rosenberg I,Chu JC,Saxena V

    更新日期:1995-01-01 00:00:00

  • DMLC tracking and gating can improve dose coverage for prostate VMAT.

    abstract:PURPOSE:To assess and compare the dosimetric impact of dynamic multileaf collimator (DMLC) tracking and gating as motion correction strategies to account for intrafraction motion during conventionally fractionated prostate radiotherapy. METHODS:A dose reconstruction method was used to retrospectively assess the dose d...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4892605

    authors: Colvill E,Poulsen PR,Booth JT,O'Brien RT,Ng JA,Keall PJ

    更新日期:2014-09-01 00:00:00

  • X-ray imaging technique for in vitro tissue composition measurements using saline/iodine displacement: technique optimization.

    abstract::An in vitro radiographic technique which uses saline/iodine displacement has been developed to study the thickness of bone-equivalent and soft-tissue-equivalent materials within atherosclerotic plaques in arterial specimens which have been cut open longitudinally and laid flat. Results concerning the optimization of t...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597916

    authors: Moreau M,Holdsworth DW,Fenster A

    更新日期:1997-01-01 00:00:00

  • Influence of gold nanoparticles embedded in water on nanodosimetry for keV photon irradiation.

    abstract:PURPOSE:For the past two decades, high-Z nanoparticles have been of high interest to improve the therapeutic outcomes of radiation therapy, especially for low-energy X-rays. Monte Carlo (MC) simulations have been used to evaluate the boost of dose deposition induced by Auger electrons near the nanoparticle surface, by ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14576

    authors: Poignant F,Monini C,Testa É,Beuve M

    更新日期:2020-11-04 00:00:00

  • Image resolution of a microchannel plate x-ray image intensifier.

    abstract::The resolution has been evaluated for an experimental x-ray image intensifier employing three microchannel plates (MCPs) as the photon absorber and electron multipliers. The line spread function (LSF) was measured and used for determination of the modulation transfer function (MTF). The MTF was found to be independent...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594463

    authors: Gould RG

    更新日期:1978-01-01 00:00:00

  • Determination of the half-life of I-125 encapsulated clinical seeds using a Si(Li) detector.

    abstract::The half-life of two I-125 encapsulated clinical seeds was measured with a Si(Li) semiconductor detector. Two measurements of each seed yielded an average half-life of 59.56 +/- 0.17 days. The results are discussed in terms of the influence of possible I-126 contamination and leakage of I-125 activity through encapsul...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595423

    authors: Kubo H

    更新日期:1983-11-01 00:00:00

  • Linear motion correction in three dimensions applied to dynamic gadolinium enhanced breast imaging.

    abstract::Quantitative analysis of dynamic gadolinium-DTPA (diethylenetriamine pentaacetic acid) enhanced magnetic resonance imaging (MRI) is emerging as a highly sensitive tool for detecting malignant breast tissue. Three-dimensional rapid imaging techniques, such as keyhole MRI, yield high temporal sampling rates to accuratel...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598576

    authors: Krishnan S,Chenevert TL,Helvie MA,Londy FL

    更新日期:1999-05-01 00:00:00

  • Confirmation of a realistic reactor model for BNCT dosimetry at the TRIGA Mainz.

    abstract:PURPOSE:In order to build up a reliable dose monitoring system for boron neutron capture therapy (BNCT) applications at the TRIGA reactor in Mainz, a computer model for the entire reactor was established, simulating the radiation field by means of the Monte Carlo method. The impact of different source definition techni...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4897242

    authors: Ziegner M,Schmitz T,Khan R,Blaickner M,Palmans H,Sharpe P,Hampel G,Böck H

    更新日期:2014-11-01 00:00:00

  • Are we doing any good by doing really well? (Where's the Bacon?).

    abstract::Francis Bacon, who with Rene Decartes laid the intellectual foundations for Western science in the seventeenth century, asserted that the purpose of all knowledge is "action in the production of works for ... the relief of man's estate." We assess briefly several aspects of a few of the current efforts directed to the...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1555493

    authors: Herbert D

    更新日期:2003-04-01 00:00:00

  • Microdosimetric study for secondary neutrons in phantom produced by a 290 MeV/nucleon carbon beam.

    abstract::Absorbed doses from main charged-particle beams and charged-particle fragments have been measured with high accuracy for particle therapy, but there are few reports for doses from neutron components produced as fragments. This study describes the measurements on neutron doses produced by carbon beams; microdosimetric ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2767933

    authors: Endo S,Tanaka K,Takada M,Onizuka Y,Miyahara N,Sato T,Ishikawa M,Maeda N,Hayabuchi N,Shizuma K,Hoshi M

    更新日期:2007-09-01 00:00:00

  • Technical Note: Spot characteristic stability for proton pencil beam scanning.

    abstract:PURPOSE:The spot characteristics for proton pencil beam scanning (PBS) were measured and analyzed over a 16 month period, which included one major site configuration update and six cyclotron interventions. The results provide a reference to establish the quality assurance (QA) frequency and tolerance for proton pencil ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4939663

    authors: Chen CC,Chang C,Moyers MF,Gao M,Mah D

    更新日期:2016-02-01 00:00:00

  • Dose buildup for obliquely incident photon beams.

    abstract::For obliquely incident photon beams, the buildup of dose with depth is markedly different from normally incident beams. However, relatively little data on this topic exists for high-energy photon beams of energy greater than 6 MV. Measurements of dose in the buildup region were made using a plane-parallel ionization c...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596055

    authors: Gerbi BJ,Meigooni AS,Khan FM

    更新日期:1987-05-01 00:00:00

  • A topographic leaf-sequencing algorithm for delivering intensity modulated radiation therapy.

    abstract::Topographic treatment is a radiation therapy delivery technique for fixed-gantry (nonrotational) treatments on a helical tomotherapy system. The intensity-modulated fields are created by moving the treatment couch relative to a fan-beam positioned at fixed gantry angles. The delivered dose distribution is controlled b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2216876

    authors: Desai D,Ramsey CR,Breinig M,Mahan SL

    更新日期:2006-08-01 00:00:00