Factors affecting the repeatability of gamma camera calibration for quantitative imaging applications using a sealed source.

Abstract:

:Several applications in nuclear medicine require absolute activity quantification of single photon emission computed tomography images. Obtaining a repeatable calibration factor that converts voxel values to activity units is essential for these applications. Because source preparation and measurement of the source activity using a radionuclide activity meter are potential sources of variability, this work investigated instrumentation and acquisition factors affecting repeatability using planar acquisition of sealed sources. The calibration factor was calculated for different acquisition and geometry conditions to evaluate the effect of the source size, lateral position of the source in the camera field-of-view (FOV), source-to-camera distance (SCD), and variability over time using sealed Ba-133 sources. A small region of interest (ROI) based on the source dimensions and collimator resolution was investigated to decrease the background effect. A statistical analysis with a mixed-effects model was used to evaluate quantitatively the effect of each variable on the global calibration factor variability. A variation of 1 cm in the measurement of the SCD from the assumed distance of 17 cm led to a variation of 1-2% in the calibration factor measurement using a small disc source (0.4 cm diameter) and less than 1% with a larger rod source (2.9 cm diameter). The lateral position of the source in the FOV and the variability over time had small impacts on calibration factor variability. The residual error component was well estimated by Poisson noise. Repeatability of better than 1% in a calibration factor measurement using a planar acquisition of a sealed source can be reasonably achieved. The best reproducibility was obtained with the largest source with a count rate much higher than the average background in the ROI, and when the SCD was positioned within 5 mm of the desired position. In this case, calibration source variability was limited by the quantum noise.

journal_name

Phys Med Biol

authors

Anizan N,Wang H,Zhou XC,Wahl RL,Frey EC

doi

10.1088/0031-9155/60/3/1325

subject

Has Abstract

pub_date

2015-02-07 00:00:00

pages

1325-37

issue

3

eissn

0031-9155

issn

1361-6560

journal_volume

60

pub_type

杂志文章
  • In vitro double-integrating-sphere optical properties of tissues between 630 and 1064 nm.

    abstract::The optical properties (absorption and scattering coefficients and the scattering anisotropy factor) were measured in vitro for cartilage, liver, lung, muscle, myocardium, skin, and tumour (colon adenocarcinoma CC 531) at 630, 632.8, 790, 850 and 1064 nm. Rabbits, rats, piglets, goats, and dogs were used to obtain the...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/42/11/017

    authors: Beek JF,Blokland P,Posthumus P,Aalders M,Pickering JW,Sterenborg HJ,van Gemert MJ

    更新日期:1997-11-01 00:00:00

  • Modelling infrared temperature measurements: implications for laser irradiation and cryogen cooling studies.

    abstract::The use of thermographic techniques has increased as infrared detector technology has evolved and improved. For laser-tissue interactions, thermal cameras have been used to monitor the thermal response of tissue to pulsed and continuous wave irradiation. It is important to note that the temperature indicated by the th...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/45/2/319

    authors: Choi B,Pearce JA,Welch AJ

    更新日期:2000-02-01 00:00:00

  • Electromagnetic redesign of the HYPERcollar applicator: toward improved deep local head-and-neck hyperthermia.

    abstract::Accumulating evidence shows that hyperthermia improves head-and-neck cancer treatment. Over the last decade, we introduced a radiofrequency applicator, named HYPERcollar, which enables local heating also of deep locations in this region. Based on clinical experience, we redesigned the HYPERcollar for improved comfort,...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/58/17/5997

    authors: Togni P,Rijnen Z,Numan WC,Verhaart RF,Bakker JF,van Rhoon GC,Paulides MM

    更新日期:2013-09-07 00:00:00

  • Locally linear constraint based optimization model for material decomposition.

    abstract::Dual spectral computed tomography (DSCT) has a superior material distinguishability than the conventional single spectral computed tomography (SSCT). However, the decomposition process is an illposed problem, which is sensitive to noise. Thus, the decomposed image quality is degraded, and the corresponding signal-to-n...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa8e13

    authors: Wang Q,Zhu Y,Yu H

    更新日期:2017-10-19 00:00:00

  • A comparison of direct reconstruction algorithms in proton computed tomography.

    abstract::Several direct algorithms have been proposed to take into account the non-linear path of protons in the reconstruction of a proton CT (pCT) image. This paper presents a comparison between five of them, in terms of spatial resolution and relative stopping power (RSP) accuracy. Our comparison includes (1) a distance-dri...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab7d53

    authors: Khellaf F,Krah N,Létang JM,Collins-Fekete CA,Rit S

    更新日期:2020-06-01 00:00:00

  • Influence of trace elements in human tissue in low-energy photon brachytherapy dosimetry.

    abstract::The aim of this paper is to determine the dosimetric impact of trace elements in human tissues for low-energy photon sources used in brachytherapy. Monte Carlo dose calculations were used to investigate the dosimetric effect of trace elements present in normal or cancerous human tissues. The effect of individual trace...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/57/11/3585

    authors: White SA,Landry G,van Gils F,Verhaegen F,Reniers B

    更新日期:2012-06-07 00:00:00

  • Improved scatter correction using adaptive scatter kernel superposition.

    abstract::Accurate scatter correction is required to produce high-quality reconstructions of x-ray cone-beam computed tomography (CBCT) scans. This paper describes new scatter kernel superposition (SKS) algorithms for deconvolving scatter from projection data. The algorithms are designed to improve upon the conventional approac...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/55/22/007

    authors: Sun M,Star-Lack JM

    更新日期:2010-11-21 00:00:00

  • Relative Patlak plot for dynamic PET parametric imaging without the need for early-time input function.

    abstract::The Patlak graphical method is widely used in parametric imaging for modeling irreversible radiotracer kinetics in dynamic PET. The net influx rate of radiotracer can be determined from the slope of the Patlak plot. The implementation of the standard Patlak method requires the knowledge of full-time input function fro...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aad444

    authors: Zuo Y,Qi J,Wang G

    更新日期:2018-08-10 00:00:00

  • Elastostatics of a spherical inclusion in homogeneous biological media.

    abstract::A three-dimensional spherical inclusion model that approximates a lesion bonded to a tissue matrix is proposed for biomedical elastography. Analytical formulae describing spatial strain and stress distributions generated in infinite media by uniform loading are given under a linear, homogeneous, isotropic elasticity a...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/43/1/001

    authors: Bilgen M,Insana MF

    更新日期:1998-01-01 00:00:00

  • Percentile-specific computational phantoms constructed from ICRP mesh-type reference computational phantoms (MRCPs).

    abstract::Recently, the Task Group 103 of the International Commission on Radiological Protection (ICRP) has developed new mesh-type reference computational phantoms (MRCPs) for adult male and female. When compared to the current voxel-type reference computational phantoms in ICRP Publication 110, the MRCPs have several advanta...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aafcdb

    authors: Lee H,Yeom YS,Nguyen TT,Choi C,Han H,Shin B,Zhang X,Kim CH,Chung BS,Zankl M

    更新日期:2019-02-05 00:00:00

  • PDT dose dosimetry for Photofrin-mediated pleural photodynamic therapy (pPDT).

    abstract::Photosensitizer fluorescence excited by photodynamic therapy (PDT) treatment light can be used to monitor the in vivo concentration of the photosensitizer and its photobleaching. The temporal integral of the product of in vivo photosensitizer concentration and light fluence is called PDT dose, which is an important do...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa9874

    authors: Ong YH,Kim MM,Finlay JC,Dimofte A,Singhal S,Glatstein E,Cengel KA,Zhu TC

    更新日期:2017-12-29 00:00:00

  • Low density contrast agents for x-ray phase contrast imaging: the use of ambient air for x-ray angiography of excised murine liver tissue.

    abstract::We report a new preparative method for providing contrast through reduction in electron density that is uniquely suited for propagation-based differential x-ray phase contrast imaging. The method, which results in an air or fluid filled vasculature, makes possible visualization of the smallest microvessels, roughly do...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/53/23/017

    authors: Laperle CM,Hamilton TJ,Wintermeyer P,Walker EJ,Shi D,Anastasio MA,Derdak Z,Wands JR,Diebold G,Rose-Petruck C

    更新日期:2008-12-07 00:00:00

  • Numerical calculation of energy deposition by broad high-energy electron beams.

    abstract::The feasibility is demonstrated of a numerical method to calculate dose deposition by broad high-energy electron beams in homogeneous matter or in heterogeneous matter in which the heterogeneities are arranged in slabs perpendicular to the beam axis. The method is based only on the basic physical interaction processes...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/34/10/003

    authors: Huizenga H,Storchi PR

    更新日期:1989-10-01 00:00:00

  • Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration.

    abstract::Multileaf collimator (MLC) calibration and quality control is a time-consuming procedure typically involving the processing, scanning and analysis of films to measure leaf and collimator positions. Faster and more reliable calibration procedures are required for these tasks, especially with the introduction of intensi...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/50/7/003

    authors: Baker SJ,Budgell GJ,MacKay RI

    更新日期:2005-04-07 00:00:00

  • Effects of f-number on the histotripsy intrinsic threshold and cavitation bubble cloud behavior.

    abstract::Histotripsy is an ultrasound ablation method that depends on the initiation of a cavitation bubble cloud to fractionate soft tissue. Although previous work has provided significant insight into the process of intrinsic threshold histotripsy, the majority of these studies have used highly focused (i.e. f-number  <  0.6...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa54c7

    authors: Vlaisavljevich E,Gerhardson T,Hall T,Xu Z

    更新日期:2017-02-21 00:00:00

  • Multi-frequency characterization of the speed of sound and attenuation coefficient for longitudinal transmission of freshly excised human skulls.

    abstract::For medical applications of ultrasound inside the brain, it is necessary to understand the relationship between the apparent density of skull bone and its corresponding speed of sound and attenuation coefficient. Although there have been previous studies exploring this phenomenon, there is still a need to extend the m...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/56/1/014

    authors: Pichardo S,Sin VW,Hynynen K

    更新日期:2011-01-07 00:00:00

  • Changes in lung tumor shape during respiration.

    abstract::Evidence that some lung tumors change shape during respiration is derived from respiratory gated CT data by statistical shape modeling and image manipulation. Some tumors behave as rigid objects while others show systematic shape changes. Two views of lung motion are presented to allow analysis of the results. In the ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/57/4/919

    authors: Kyriakou E,McKenzie DR

    更新日期:2012-02-21 00:00:00

  • An extended diffraction-enhanced imaging method for implementing multiple-image radiography.

    abstract::Diffraction-enhanced imaging (DEI) is an analyser-based x-ray imaging method that produces separate images depicting the projected x-ray absorption and refractive properties of an object. Because the imaging model of DEI does not account for ultra-small-angle x-ray scattering (USAXS), the images produced in DEI can co...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/52/7/011

    authors: Chou CY,Anastasio MA,Brankov JG,Wernick MN,Brey EM,Connor DM,Zhong Z

    更新日期:2007-04-07 00:00:00

  • Volume estimation of low-contrast lesions with CT: a comparison of performances from a phantom study, simulations and theoretical analysis.

    abstract::Measurements of lung nodule volume with multi-detector computed tomography (MDCT) have been shown to be more accurate and precise compared to conventional lower dimensional measurements. Quantifying the size of lesions is potentially more difficult when the object-to-background contrast is low as with lesions in the l...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/60/2/671

    authors: Li Q,Gavrielides MA,Zeng R,Myers KJ,Sahiner B,Petrick N

    更新日期:2015-01-21 00:00:00

  • A study of the impact of x-ray tube performance on angiography system imaging efficiency.

    abstract::This work compared the impact of x-ray tube performance and automatic dose rate control (ADRC) parameter selection on system imaging efficiency of two Siemens angiography systems: a Siemens Megalix x-ray tube installed on an Artis Zee system (denoted 'MEGALIX') and a newer generation Gigalix x-ray tube installed on an...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/abbb7a

    authors: Dehairs M,Bosmans H,Marshall NW

    更新日期:2020-11-24 00:00:00

  • Dielectric characteristics of packed human erythrocytes with haemoglobins F, AA, AS and SS.

    abstract::The radiofrequency dielectric behaviour of packed human erythrocytes has been investigated in normal haemoglobins F and AA and also abnormal haemoglobins homozygous SS and heterozygous AS. Dielectric measurements were made in tightly packed blood cells at an average room temperature of 26.5 +/- 0.5 degrees C in the fr...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/28/4/002

    authors: Laogun AA,Ajayi NO,Okafor LO,Osamo NO

    更新日期:1983-04-01 00:00:00

  • A novel linear programming approach to fluence map optimization for intensity modulated radiation therapy treatment planning.

    abstract::We present a novel linear programming (LP) based approach for efficiently solving the intensity modulated radiation therapy (IMRT) fluence-map optimization (FMO) problem to global optimality. Our model overcomes the apparent limitations of a linear-programming approach by approximating any convex objective function by...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/48/21/005

    authors: Romeijn HE,Ahuja RK,Dempsey JF,Kumar A,Li JG

    更新日期:2003-11-07 00:00:00

  • Magnetic separation of iron oxide nanoparticles to improve their application for magnetic particle imaging.

    abstract::Magnetic Particle Imaging (MPI) is a promising medical imaging technique for visualizing the three-dimensional distribution of tracer materials, specifically iron oxide nanoparticles (IONP). The optimization of magnetic nanoparticles (MNP) plays an essential role to improve the image resolution and sensitivity of imag...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/abcd19

    authors: Arsalani S,Loewa N,Kosch O,Radon P,Baffa O,Wiekhorst F

    更新日期:2020-11-23 00:00:00

  • The production of positron emitters with millisecond half-life during helium beam radiotherapy.

    abstract::Therapy with helium ions is currently receiving significantly increasing interest because helium ions have a sharper penumbra than protons and undergo less fragmentation than carbon ions and thus require less complicated dose calculations. For any ion of interest in hadron therapy, the accuracy of dose delivery is lim...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab51c3

    authors: Ozoemelam I,van der Graaf E,Brandenburg S,Dendooven P

    更新日期:2019-12-05 00:00:00

  • Dosimetric verification of open asymmetric photon fields calculated with a treatment planning system based on dose-to-energy-fluence concepts.

    abstract::Output normalized dose profiles for asymmetric open photon fields has been calculated using a commercial treatment planning system (TPS) based on a dose-to-energy-fluence concept. The model does not require any additional measurements for off-axis fields. Calculations are compared with measurements for quadratic field...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/41/8/003

    authors: Hurkmans C,Knöös T,Nilsson P

    更新日期:1996-08-01 00:00:00

  • A novel microbubble construct for intracardiac or intravascular MR manometry: a theoretical study.

    abstract::It has been demonstrated that gas-filled microbubble contrast agents, based on their volume changes, can serve as pressure probes in an MR field. It was recently reported that such an MR-based pressure measurement with microbubbles at 1.5 T must make use of microbubbles that have a volumetric magnetic susceptibility d...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/50/20/001

    authors: Dharmakumar R,Plewes DB,Wright GA

    更新日期:2005-10-21 00:00:00

  • Performance of a multi-axis ionization chamber array in a 1.5 T magnetic field.

    abstract::At the UMC Utrecht a prototype MR-linac has been installed. The system consists of an 8 MV Elekta linear accelerator and a 1.5 T Philips MRI system. This paper investigates the performance of the IC PROFILER™, a multi-axis ionization chamber array, in a 1.5 T magnetic field. The influence of the magnetic field on the ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/59/7/1845

    authors: Smit K,Kok JG,Lagendijk JJ,Raaymakers BW

    更新日期:2014-04-07 00:00:00

  • Exact analytic reconstruction in x-ray fluorescence CT and approximated versions.

    abstract::X-ray fluorescence computed tomography (XFCT) is a relatively new synchrotron-based imaging modality aiming at reconstructing the distribution of nonradiative elements within a sample irradiated with high-intensity monochromatic x-rays. In a recent paper La Rivière (2004 Phys. Med. Biol. 49 2391-405) presented an appr...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/55/4/007

    authors: Miqueles EX,De Pierro AR

    更新日期:2010-02-21 00:00:00

  • Dose conversion coefficients for paediatric CT examinations with automatic tube current modulation.

    abstract::A common dose-saving technique used in modern CT devices is automatic tube current modulation (TCM), which was originally designed to also reduce the dose in paediatric CT patients. In order to be able to deduce detailed organ doses of paediatric models, dose conversion coefficients normalized to CTDI(vol) for an eigh...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/57/20/6309

    authors: Schlattl H,Zankl M,Becker J,Hoeschen C

    更新日期:2012-10-21 00:00:00

  • Electric fields induced in humans and rodents by 60 Hz magnetic fields.

    abstract::Numerical computations are used to evaluate electric field dosimetry for high-resolution anatomically based inhomogeneous models of a human male child, and male and female rats and mice, under exposure to 60 Hz uniform magnetic field sources of three perpendicular orientations. The goal is to compare the child data to...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/47/14/313

    authors: Dawson TW,Caputa K,Stuchly MA

    更新日期:2002-07-21 00:00:00