MRI reconstruction using deep Bayesian estimation.

Abstract:

PURPOSE:To develop a deep learning-based Bayesian estimation for MRI reconstruction. METHODS:We modeled the MRI reconstruction problem with Bayes's theorem, following the recently proposed PixelCNN++ method. The image reconstruction from incomplete k-space measurement was obtained by maximizing the posterior possibility. A generative network was utilized as the image prior, which was computationally tractable, and the k-space data fidelity was enforced by using an equality constraint. The stochastic backpropagation was utilized to calculate the descent gradient in the process of maximum a posterior, and a projected subgradient method was used to impose the equality constraint. In contrast to the other deep learning reconstruction methods, the proposed one used the likelihood of prior as the training loss and the objective function in reconstruction to improve the image quality. RESULTS:The proposed method showed an improved performance in preserving image details and reducing aliasing artifacts, compared with GRAPPA, ℓ 1 -ESPRiT, model-based deep learning architecture for inverse problems (MODL), and variational network (VN), last two were state-of-the-art deep learning reconstruction methods. The proposed method generally achieved more than 3 dB peak signal-to-noise ratio improvement for compressed sensing and parallel imaging reconstructions compared with the other methods. CONCLUSIONS:The Bayesian estimation significantly improved the reconstruction performance, compared with the conventional ℓ 1 -sparsity prior in compressed sensing reconstruction tasks. More importantly, the proposed reconstruction framework can be generalized for most MRI reconstruction scenarios.

journal_name

Magn Reson Med

authors

Luo G,Zhao N,Jiang W,Hui ES,Cao P

doi

10.1002/mrm.28274

subject

Has Abstract

pub_date

2020-10-01 00:00:00

pages

2246-2261

issue

4

eissn

0740-3194

issn

1522-2594

journal_volume

84

pub_type

杂志文章
  • Acceleration and motion-correction techniques for high-resolution intravascular MRI.

    abstract:PURPOSE:High-resolution intravascular (IV) MRI is susceptible to degradation from physiological motion and requires high frame-rates for true endoscopy. Traditional cardiac-gating techniques compromise efficiency by reducing the effective scan rate. Here we test whether compressed sensing (CS) reconstruction and ungate...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.25436

    authors: Hegde SS,Zhang Y,Bottomley PA

    更新日期:2015-08-01 00:00:00

  • Fast three-dimensional inner volume excitations using parallel transmission and optimized k-space trajectories.

    abstract:PURPOSE:To design short parallel transmission (pTx) pulses for excitation of arbitrary three-dimensional (3D) magnetization patterns. METHODS:We propose a joint optimization of the pTx radiofrequency (RF) and gradient waveforms for excitation of arbitrary 3D magnetization patterns. Our optimization of the gradient wav...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.26021

    authors: Davids M,Schad LR,Wald LL,Guérin B

    更新日期:2016-10-01 00:00:00

  • B factor dependence of the temporal characteristics of brain activation using dynamic apparent diffusion coefficient contrast.

    abstract::Functional MRI studies to date have been generally performed using the blood oxygenation level dependent (BOLD) contrast mechanism. Recently, it has been proposed that dynamic change in the apparent diffusion coefficient (ADC), measured using intravoxel incoherent motion (IVIM) weighting, can be used as a robust funct...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.20293

    authors: Harshbarger TB,Song AW

    更新日期:2004-12-01 00:00:00

  • Gradient rotating outer volume excitation (GROOVE): A novel method for single-shot two-dimensional outer volume suppression.

    abstract:PURPOSE:To introduce a new outer volume suppression (OVS) technique that uses a single pulse and rotating gradients to accomplish frequency-swept excitation. This new technique, which is called gradient rotating outer volume excitation (GROOVE), produces a circular or elliptical suppression band rather than suppressing...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.25101

    authors: Powell NJ,Jang A,Park JY,Valette J,Garwood M,Marjańska M

    更新日期:2015-01-01 00:00:00

  • Multiple-coil k-space interpolation enhances resolution in single-shot spatiotemporal MRI.

    abstract:PURPOSE:Spatio-temporal encoding (SPEN) experiments can deliver single-scan MR images without folding complications and with robustness to chemical shift and susceptibility artifacts. Further resolution improvements are shown to arise by relying on multiple receivers, to interpolate the sampled data along the low-bandw...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.26731

    authors: Liberman G,Solomon E,Lustig M,Frydman L

    更新日期:2018-02-01 00:00:00

  • On the use of Bayesian probability theory for analysis of exponential decay data: an example taken from intravoxel incoherent motion experiments.

    abstract::Traditionally, the method of nonlinear least squares (NLLS) analysis has been used to estimate the parameters obtained from exponential decay data. In this study, we evaluated the use of Bayesian probability theory to analyze such data; specifically, that resulting from intravoxel incoherent motion NMR experiments. An...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910290510

    authors: Neil JJ,Bretthorst GL

    更新日期:1993-05-01 00:00:00

  • Electrocardiogram acquisition during MR examinations for patient monitoring and sequence triggering.

    abstract::We have developed a method for measuring the electrocardiogram (ECG) continuously during MR examinations. In contrast to ECG acquisition by wires, our new method is to amplify and convert the ECG into an optical signal directly above the patient's heart. The optical signal is transmitted out of the magnet bore by opti...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910320416

    authors: Felblinger J,Lehmann C,Boesch C

    更新日期:1994-10-01 00:00:00

  • Sodium imaging of human brain at 7 T with 15-channel array coil.

    abstract::Signal-to-noise ratio (SNR) is a major challenge to sodium magnetic resonance imaging. Phased array coils have been shown significantly improving SNR in proton imaging over volume coils. This study investigates SNR advantage of a 15-channel array head coil (birdcage volume coil for transmit/receive and 15-channel arra...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.24192

    authors: Qian Y,Zhao T,Wiggins GC,Wald LL,Zheng H,Weimer J,Boada FE

    更新日期:2012-12-01 00:00:00

  • Reduction of B1 sensitivity in selective single-slab 3D turbo spin echo imaging with very long echo trains.

    abstract::Single-slab 3D turbo/fast spin echo (SE) imaging with very long echo trains was recently introduced with slab selection using a highly selective excitation pulse and short, nonselective refocusing pulses with variable flip angles for high imaging efficiency. This technique, however, is vulnerable to image degradation ...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.22070

    authors: Park J,Mugler JP 3rd,Hughes T

    更新日期:2009-10-01 00:00:00

  • Biexponential diffusion attenuation in various states of brain tissue: implications for diffusion-weighted imaging.

    abstract::Diffusion-weighted single voxel experiments conducted at b-values up to 1 x 10(4) smm-2 yielded biexponential signal attenuation curves for both normal and ischemic brain. The relative fractions of the rapidly and slowly decaying components (f1, f2) are f1 = 0.80 +/- 0.02, f2 = 0.17 +/- 0.02 in healthy adult rat brain...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910360607

    authors: Niendorf T,Dijkhuizen RM,Norris DG,van Lookeren Campagne M,Nicolay K

    更新日期:1996-12-01 00:00:00

  • New method for contrast manipulation in DNP-enhanced MRI.

    abstract::The magnetization subtraction technique (MS), which is equivalent to the inversion recovery technique in strong magnetic fields, has been implemented in dynamic nuclear polarization-enhanced magnetic resonance imaging (DNPI). The general theoretical basis of the MS method, which can be applied to DNPI or to prepolariz...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910350316

    authors: Planinsic G,Grucker D,Stepisnik J

    更新日期:1996-03-01 00:00:00

  • Automatic selection of mask and arterial phase images for temporally resolved MR digital subtraction angiography.

    abstract::For time-resolved background-subtracted contrast-enhanced magnetic resonance angiography, the bright and sparse arterial signal allows unique identification of contrast bolus arrival in the arteries. This article presents an automatic filtering algorithm using such arterial characterization for selecting arterial phas...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.10358

    authors: Kim J,Prince MR,Zabih R,Bezanson J,Watts R,Erel HE,Wang Y

    更新日期:2002-12-01 00:00:00

  • MR elastography for evaluating regeneration of tissue-engineered cartilage in an ectopic mouse model.

    abstract:PURPOSE:The purpose of the present study was to apply noninvasive methods for monitoring regeneration and mechanical properties of tissue-engineered cartilage in vivo at different growth stages using MR elastography (MRE). METHODS:Three types of scaffolds, including silk, collagen, and gelatin seeded by human mesenchy...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.25745

    authors: Khalilzad-Sharghi V,Han Z,Xu H,Othman SF

    更新日期:2016-03-01 00:00:00

  • Continuous assessment of relative cerebral blood volume in transient ischemia using steady state susceptibility-contrast MRI.

    abstract::The utility of a noninvasive steady state susceptibility-contrast MRI technique for continuous measurement of relative cerebral blood volume (rCBV) during global transient ischemia and subsequent hyperemia in a feline ischemia model is demonstrated. The measurements were obtained during a 10-min period of occlusion an...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910350207

    authors: Hamberg LM,Boccalini P,Stranjalis G,Hunter GJ,Huang Z,Halpern E,Weisskoff RM,Moskowitz MA,Rosen BR

    更新日期:1996-02-01 00:00:00

  • Indirect 17(O)-magnetic resonance imaging of cerebral blood flow in the rat.

    abstract::Proton T(1rho)-dispersion MRI is demonstrated for indirect, in vivo detection of (17)O in the brain. This technique, which may be readily implemented on any clinical MRI scanner, is applied towards high-resolution, quantitative mapping of cerebral blood flow (CBF) in the rat by monitoring the clearance of (17)O-enrich...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.10403

    authors: Tailor DR,Roy A,Regatte RR,Charagundla SR,McLaughlin AC,Leigh JS,Reddy R

    更新日期:2003-03-01 00:00:00

  • x-f Choice: reconstruction of undersampled dynamic MRI by data-driven alias rejection applied to contrast-enhanced angiography.

    abstract::A technique for reconstructing dynamic undersampled MRI data, termed "x-f choice," was developed and applied to dynamic contrast-enhanced MR angiography (DCE-MRA). Regular undersampling in k-t space (a hybrid of k-space and time) creates aliasing in the conjugate x-f space that must be resolved. When regions in the ob...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.21008

    authors: Malik SJ,Schmitz S,O'Regan D,Larkman DJ,Hajnal JV

    更新日期:2006-10-01 00:00:00

  • Relaxation effects of ferucarbotran-labeled mesenchymal stem cells at 1.5T and 3T: discrimination of viable from lysed cells.

    abstract::Human mesenchymal stem cells (hMSCs) were labeled with Ferucarbotran by simple incubation and cultured for up to 14 d. Iron content was determined by spectrometry and the intracellular localization of the contrast agent uptake was studied by electron and confocal microscopy. At various time points after labeling, rang...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.22011

    authors: Henning TD,Wendland MF,Golovko D,Sutton EJ,Sennino B,Malek F,Bauer JS,McDonald DM,Daldrup-Link H

    更新日期:2009-08-01 00:00:00

  • Off-resonance-dependent slice profile effects in balanced steady-state free precession imaging.

    abstract::The purpose of this study was the simulation and measurement of balanced steady-state free precession (bSSFP) slice profiles for a detailed analysis of the influence of off-resonance effects on slice profile shape and bSSFP signal intensity. Due to the frequency response function of the bSSFP sequence, measurements th...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.21557

    authors: Staehle F,Leupold J,Hennig J,Markl M

    更新日期:2008-05-01 00:00:00

  • Graphical programming interface: A development environment for MRI methods.

    abstract:PURPOSE:To introduce a multiplatform, Python language-based, development environment called graphical programming interface for prototyping MRI techniques. METHODS:The interface allows developers to interact with their scientific algorithm prototypes visually in an event-driven environment making tasks such as paramet...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.25528

    authors: Zwart NR,Pipe JG

    更新日期:2015-11-01 00:00:00

  • Porous medium 3D flow simulation of contrast media washout in cardiac MRI reflects myocardial injury.

    abstract:PURPOSE:Myocardial blood-flow simulation based on laws of fluid mechanics is a valuable tool for understanding tissue behavior. Our aim is to evaluate the ability of a porous-media flow model approach to reflect disturbed washout of contrast media (CM) from the myocardium as observed by cardiovascular MR. METHODS:A co...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.27756

    authors: Riazy L,Schaeffter T,Olbrich M,Schueler J,von Knobelsdorff-Brenkenhoff F,Niendorf T,Schulz-Menger J

    更新日期:2019-08-01 00:00:00

  • Eight-channel phased array coil and detunable TEM volume coil for 7 T brain imaging.

    abstract::An eight-channel receive-only brain coil and table-top detunable volume transmit coil were developed and tested at 7 T for human imaging. Optimization of this device required attention to sources of interaction between the array elements, between the transmit and receive coils and minimization of common mode currents ...

    journal_title:Magnetic resonance in medicine

    pub_type: 临床试验,杂志文章

    doi:10.1002/mrm.20547

    authors: Wiggins GC,Potthast A,Triantafyllou C,Wiggins CJ,Wald LL

    更新日期:2005-07-01 00:00:00

  • Magnetic resonance imaging with T1 dispersion contrast.

    abstract::Prepolarized MRI uses pulsed magnetic fields to produce MR images by polarizing the sample at one field strength (approximately 0.5 T) before imaging at a much lower field (approximately 50 mT). Contrast reflecting the T(1) of the sample at an intermediate field strength is achieved by polarizing the sample and then a...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.20910

    authors: Ungersma SE,Matter NI,Hardy JW,Venook RD,Macovski A,Conolly SM,Scott GC

    更新日期:2006-06-01 00:00:00

  • SMS-HSL: Simultaneous multislice aliasing separation exploiting hankel subspace learning.

    abstract:PURPOSE:To develop a novel, simultaneous multislice reconstruction method that exploits Hankel subspace learning (SMS-HSL) for aliasing separation in the slice direction. METHODS:An SMS signal model with the Hankel-structured matrix was proposed. To efficiently suppress interslice leakage artifacts from a signal subsp...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.26527

    authors: Park S,Park J

    更新日期:2017-10-01 00:00:00

  • Input parameter sensitivity analysis and comparison of quantification models for continuous arterial spin labeling.

    abstract::The regional cerebral blood flow (rCBF) values determined using continuous arterial spin labeling (CASL) are subject to several sources of variability, including natural physiologic variations, sensitivity to the input parameters, and the use of different quantification models. To date, a thorough analysis of the impa...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.20440

    authors: Steger TR,White RA,Jackson EF

    更新日期:2005-04-01 00:00:00

  • Diffusion tensor imaging of in vivo and excised rat spinal cord at 7 T with an icosahedral encoding scheme.

    abstract::Regional values of fractional anisotropy (FA) and mean diffusivity (D(av)) of in vivo and excised rat spinal cords were measured using an iscosahedral encoding scheme that is based on 21 uniformly distributed and alternating gradient directions with an echo planar imaging (EPI) readout. Based on the water phantom stud...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.20304

    authors: Madi S,Hasan KM,Narayana PA

    更新日期:2005-01-01 00:00:00

  • Volume-localized measurement of oxygen extraction fraction in the brain using MRI.

    abstract:PURPOSE:T2 -relaxation-under-spin-tagging (TRUST) is an MR technique for the non-invasive assessment of whole-brain cerebral oxygen extraction fraction (OEF), through measurement of the venous blood T2 relaxation time in the sagittal sinus. A key limitation of TRUST, however, is the lack of spatial specificity of the m...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.27823

    authors: O'Brien C,Okell TW,Chiew M,Jezzard P

    更新日期:2019-10-01 00:00:00

  • Polymetallic macromolecules are potential contrast agents of improved efficiency.

    abstract::The commercial synthetic homopolypeptide polyaspartate (average MW = 30,000, approximately 220 monomers) in solution quantitatively binds up to 40 mol Gd3+ ions per mole of polyaspartate. The water proton relaxivity of the solutions is far higher than that of clinically used or commonly investigated gadolinium(III) co...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910310109

    authors: Cavagna F,Luchinat C,Scozzafava A,Xia Z

    更新日期:1994-01-01 00:00:00

  • Retrospective artifact elimination in MEGA-PRESS using a correlation approach.

    abstract:PURPOSE:To develop a method for retrospective artifact elimination of MRS data. This retrospective method was based on an approach that combines jackknife analyses with the correlation of spectral windows, and therefore termed "JKC." METHODS:Twelve healthy volunteers performed 3 separate measurement protocols using a ...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.27590

    authors: Tapper S,Tisell A,Helms G,Lundberg P

    更新日期:2019-04-01 00:00:00

  • T2 of endotoxin lung injury with and without methylprednisolone treatment.

    abstract::NMR relaxation times (T1 and T2) and the water content (WC) of in vitro rat lungs were measured during the course of endotoxin lung injury in rats. Measurements of normal lungs, untreated endotoxin-injured lungs, and endotoxin-injured lungs treated with methylprednisolone (MPSL) were compared. The untreated endotoxin ...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910150204

    authors: Shioya S,Haida M,Tsuji C,Ono Y,Miyairi A,Fukuzaki M,Ohta Y,Yamabayashi H

    更新日期:1990-08-01 00:00:00

  • Utilization of an intra-oral diamagnetic passive shim in functional MRI of the inferior frontal cortex.

    abstract::Due to the presence of gross magnetic susceptibility artifacts, functional MRI (fMRI) has proved problematic in studies of the human inferior frontal cortex (IFC). There is a strong desire, therefore, to employ techniques that mitigate susceptibility artifacts in the IFC while preserving the imaging parameters of an f...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.10626

    authors: Wilson JL,Jezzard P

    更新日期:2003-11-01 00:00:00