Simultaneous validation of the SunTech 247 diagnostic station blood pressure measurement device according to the British Hypertension Society protocol, the International Protocol and the Association for the Advancement of Medical Instrumentation standards

Abstract:

OBJECTIVE:SunTech 247 is an oscillometric blood pressure (BP) monitor manufactured by SunTech Medical Inc., Morrisville, North Carolina, USA. The aims of this study were (i) to determine its accuracy using the 1993 modified British Hypertension Society (BHS) protocol, the 2002 International Protocol (IP) and the 2002/2003 Association for the Advancement of Medical Instrumentation (AAMI) standards, (ii) to compare the performance of BHS protocol and IP. METHODS:Systolic BP (SBP) and diastolic BP (DBP) were recorded to the nearest 2 mmHg and measured with the arm supported at heart level. For each participant, nine sequential same arm measurements were taken by two trained observers, comparing the device to standard mercury sphygmomanometers. Procedures and data analysis were carried out following protocols guidelines. RESULTS:First, we recruited 33 participants required by the IP. Then, data collection continued to obtain 85 participants for the BHS protocol/AAMI standard. Readings differing by less than 5, 10 and 15 mmHg for SBP and DBP fulfilled IP recommendations. According to the BHS protocol, for the better observer, the percentage of test device readings differing from the mercury standard by 5, 10 and 15 mmHg or less were: 92.9, 100 and 100% for SBP and 94.9, 98.8 and 99.6% for DBP. The mean differences +/- standard deviation between the device and mercury sphygmomanometer readings were 1.7 +/- 3.1 mmHg for SBP and 1.1 +/- 3.2 mmHg for DBP, achieving AAMI requirements. Standard deviation of mean differences were 2.4 for SBP and 2.1 for DBP, following the IP, and 3.1 and 3.2, respectively, for BHS protocol. The variance of all 99 differences (total deviance), with 98 degree of freedom, was 585.6 for SBP and 444.1 for DBP. The 33 differences obtained from the mean differences for each participant (between-subject variance), with 32 degree of freedom, were 275.6 and 152.1, respectively. F test was 1.83 with P = 0.0191 for SBP, and 1.07 (P = 0.3936) for DBP. CONCLUSION:(i) The device achieved the requirements stated by the 2002 IP, fulfilled the standards stated by the AAMI, and on the basis of the standards indicated by the 1993 modified BHS protocol, can be classified as 'A' grade both for SBP and DBP. Therefore, SunTech 247 may be recommended for clinical use, (ii) IP and BHS protocol provide complementary features for device validation; the comparison of total deviance and between-subject variance confirms the estimates of IP authors for SBP.

journal_name

Blood Press Monit

authors

Polo Friz H,Facchetti R,Primitz L,Beltrame L,Galbiati V,Ricioppo A,Bombelli M,Sega R

doi

10.1097/MBP.0b013e328330c873

subject

Has Abstract

pub_date

2009-10-01 00:00:00

pages

222-7

issue

5

eissn

1359-5237

issn

1473-5725

journal_volume

14

pub_type

杂志文章
  • Hypertension in psoriasis patients: based on office and ambulatory blood pressure measurement.

    abstract:OBJECTIVE:Psoriasis is a chronic inflammatory disorder, which is associated with high risk of metabolic comorbidities. Hypertension (HTN) is among the most common and serious associations. In this study, we aimed to evaluate HTN in psoriatic patients through office and ambulatory blood pressure monitoring (ABPM). MATE...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000489

    authors: Abedini R,Najafi MT,Azizpour A,Nasimi M,Darreshiri M,Teymourpour A

    更新日期:2020-12-01 00:00:00

  • Methods of data analysis of ambulatory blood pressure monitoring in clinical trials.

    abstract::The objectives of the analysis are to compare the blood pressure levels at different phases of the trial and to be sure that the effect will continue until the next dose. There are problems before the analysis for which consensus is necessary. If the overall 24 h study is the first generally recognized stage, it is di...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: de Gaudemaris R,Chau NP

    更新日期:1996-06-01 00:00:00

  • A psychophysiologic reaction to application of ambulatory blood pressure monitor: fact or artifact?

    abstract:BACKGROUND:The study investigated the clinical significance of relatively high blood pressure (BP) elevations during the calibration of the ambulatory blood pressure monitor (ABPM). METHODS:Inferences were drawn from the analysis of differences between 'M', the average of five manually triggered responses, and 'W', th...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e328340ff51

    authors: Yanovski A,Townsend RR,Virginia F,Michael C

    更新日期:2011-02-01 00:00:00

  • The effects of various definitions of dipping and daytime and night-time on the characterization of 24 h profiles of blood pressure.

    abstract:BACKGROUND:Ambulatory blood pressure monitoring allows characterization of the patient's average blood pressure load as well as various profiles of blood pressure. Results from clinical studies suggest that dippers (patients whose blood pressures decrease during night-time) have a lower risk of cardiovascular events th...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200002000-00004

    authors: Butkevich A,Phillips RA,Sheinart KF,Tuhrim S

    更新日期:2000-02-01 00:00:00

  • Propaq Neonatal monitor used with its own single- or with Critikon twin-hose cuffs: does it matter?.

    abstract::Some oscillometric non-invasive blood pressure (NIBP) monitors use twin-hose cuffs, whereas others use single-hose cuffs. The incompatibility hinders the transfer of patients from monitors using a single-hose cuff to monitors using twin-hose cuffs and vice versa. In particular, the Propaq Encore uses a single-hose cuf...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Amoore JN

    更新日期:1997-12-01 00:00:00

  • Accuracy of the SpaceLabs Medical 90217 ambulatory blood pressure monitor.

    abstract::OBJECTIVE: To test the SpaceLabs Medical 90217 ambulatory blood pressure monitor for compliance with The Association for the Advancement of Medical Instrumentation's standard and a modification of the British Hypertension Society (BHS) protocol. METHODS: The study concerned 85 subjects aged 17-88 years with various ar...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Baumgart P,Kamp J

    更新日期:1998-10-01 00:00:00

  • Effects of race and sex on blood pressure and hemodynamic stress response as a function of the menstrual cycle.

    abstract::OBJECTIVE: To examine the influence of race and sex on responses to cardiovascular stress during two menstrual cycle phases. METHODS: Subjects were exposed to two stressors, a mirror star-tracing task and a speaking task. Blood pressure, heart rate, and impedance cardiography measures of reactivity were recorded. Wome...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Ahwal WN,Mills PJ,Kalshan DA,Nelesen RA

    更新日期:1997-08-01 00:00:00

  • Masked hypertension, a review of the literature.

    abstract::Masked hypertension (blood pressure that is normal in the physicians' office but elevated elsewhere) is a common phenomenon as prevalence among studies varies from 8 to 45% and is seen at all ages. large discrepancies, however, exist between studies that have dealt with masked hypertension. It is of high clinical impo...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/mbp.0b013e3280fb2792

    authors: Verberk WJ,Thien T,de Leeuw PW

    更新日期:2007-08-01 00:00:00

  • Potential hypotensive effects of Umezu polyphenols: a 14-week community-based, double-masked and placebo-controlled trial.

    abstract:OBJECTIVE:To evaluate blood pressure (BP)-lowering effects of Umezu polyphenols, polyphenols contained in Japanese plums, in a community-based sample by double-masked and placebo-controlled design. METHODS:Seventy-two Japanese community-dwellers who were interested in prevention or control of their BP (preferably high...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0000000000000476

    authors: Takemura S,Yoshimasu K,Tsuno K,Kuroda M,Kishida K,Mitani T,Miyashita K

    更新日期:2020-12-01 00:00:00

  • Evaluation, management, and referral of elderly emergency department patients with elevated blood pressure.

    abstract:OBJECTIVES:To determine blood pressure (BP) reassessment rates and to describe the evaluation and outpatient referral rates of elderly emergency department (ED) patients with elevated BP. METHODS:This was a retrospective cohort of patients who were at least 60 years, presented with a systolic BP of at least 140 mmHg o...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,多中心研究

    doi:10.1097/MBP.0b013e328332fd40

    authors: Baumann BM,Cienki JJ,Cline DM,Egging D,Lehrmann JF,Tanabe P

    更新日期:2009-12-01 00:00:00

  • Relationship between noninvasive and invasive blood pressure values in end-stage renal disease patients on dialysis.

    abstract:BACKGROUND:End-stage renal disease (ESRD) is complicated by high blood pressure (BP) and severe atherosclerosis, which may lead to different values of BP measured by invasive and noninvasive methods. However, the differences in BP values between these two methods and their possible mechanisms in ESRD have not been inve...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000026

    authors: Chen Y,Zhu Y,Chen C,Chen X,Pan H,Chen S,Tan X

    更新日期:2014-04-01 00:00:00

  • Repeatability of blood pressure measurements during treadmill exercise.

    abstract:OBJECTIVE:Blood pressure measurements during exercise are difficult to obtain but are used widely in clinical practice. The objective of this study was to investigate the repeatability of blood pressure measurements and gas exchange parameters during treadmill exercise. METHODS:Participants were 10 healthy and nonsmok...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章

    doi:10.1097/MBP.0b013e328351de5c

    authors: Instebø A,Helgheim V,Greve G

    更新日期:2012-04-01 00:00:00

  • The effect of crossing legs on blood pressure.

    abstract:OBJECTIVE:To determine whether crossing of the legs at the knee or at the ankles during blood pressure measurement in sitting position has an effect on blood pressure. METHODS:One hundred and eleven patients, 60 women, mean age 52+/-17 years (19-80): 49 chronically treated hypertensives, 28 treated diabetics and 34 no...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e3280b083a7

    authors: Adiyaman A,Tosun N,Elving LD,Deinum J,Lenders JW,Thien T

    更新日期:2007-06-01 00:00:00

  • Effect of fasting on 24-h blood pressure values of individuals with no previous history of hypertension.

    abstract:OBJECTIVE:We aimed to analyze the difference in 24-h blood pressure values during Ramadan of fasting and nonfasting individuals with no previous history of hypertension. PATIENTS AND METHODS:This study was planned as a multicenter research study in the cities of Izmit, Zonguldak, Sivas, and Adana. The percentage chang...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,多中心研究

    doi:10.1097/MBP.0000000000000244

    authors: Seker A,Demirci H,Ocakoglu G,Aydin U,Ucar H,Yildiz G,Yaman O

    更新日期:2017-10-01 00:00:00

  • The impact of arm circumference on noninvasive oscillometric blood pressure referenced with intra-aortic blood pressure.

    abstract:OBJECTIVE:To investigate the influence of arm circumference (AC) on the brachial blood pressure (BP) measured with an adult cuff. PARTICIPANTS AND METHODS:This study included 208 patients (150 men, 61.0±8.0 years) for coronary angiography. Intra-aortic BP and noninvasive right brachial oscillometric BP (noninvasive BP...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000141

    authors: Shangguan Q,Wu Y,Xu J,Su H,Li J,Hong K,Cheng X

    更新日期:2015-12-01 00:00:00

  • Comparison of two automated sphygmomanometers for use in the office setting.

    abstract:OBJECTIVE:To compare readings obtained using two automated blood pressure (BP) recording devices, BpTRU and Omron 907, in a clinical setting. METHODS:Two series of 50 patients attending a hypertension unit for 24-h ambulatory BP monitoring had BP recorded either every 1 or every 2 min using BpTRU and Omron 907 devices...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e32831e314f

    authors: Myers MG,Valdivieso M,Kiss A,Tobe SW

    更新日期:2009-02-01 00:00:00

  • Validation of the A&D UA-774 (UA-767Plus) device for self-measurement of blood pressure.

    abstract:OBJECTIVES:To determine the accuracy of the UA-774 (UA-767Plus) blood pressure monitor developed by the A&D Company. METHODS:Device evaluation was performed according to the modified British Hypertension Society (BHS) protocol released in 1993. Eighty-five subjects with characteristics outlined in the BHS protocol wer...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200408000-00008

    authors: Verdecchia P,Angeli F,Poeta F,Reboldi GP,Borgioni C,Pittavini L,Porcellati C

    更新日期:2004-08-01 00:00:00

  • Clinical test of a blood pressure measurement device with a wrist cuff.

    abstract::OBJECTIVE: To test a semi-automatic oscillometric blood pressure measurement device with a wrist cuff. METHODS: Two different test methods, partly based on European and American standards for clinical investigation of non-invasive sphygmomanometers, were worked out and compared. Mostly German guidelines were considere...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Wonka F,Thümmler M,Schöppe A

    更新日期:1996-08-01 00:00:00

  • Development of a simulator for the validation of noninvasive blood pressure-monitoring devices.

    abstract::This research aimed to develop a simulator capable of oscillometric pressure pulses recorded from the participants for the validation of oscillometric noninvasive blood pressure (NIBP) devices. The simulator generates the pressure pulses to the cuff connected to NIBP devices depending on the oscillometric waveforms ob...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章

    doi:10.1097/MBP.0000000000000174

    authors: Doh I,Lim HK,Ahn B

    更新日期:2016-06-01 00:00:00

  • The German Hypertension League (Deutsche Hochdruckliga) Quality Seal Protocol for blood pressure-measuring devices: 15-year experience and results from 105 devices for home blood pressure control.

    abstract:OBJECTIVE:The German Hypertension League (Deutsche Hochdruckliga) established a program to assess the accuracy and reliability of blood pressure (BP)-measuring devices in 1999 (Quality Seal Protocol). Here, we report on the results of a testing series of 105 devices designed for BP self-measurement. METHODS:The test p...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000186

    authors: Tholl U,Lüders S,Bramlage P,Dechend R,Eckert S,Mengden T,Nürnberger J,Sanner B,Anlauf M

    更新日期:2016-08-01 00:00:00

  • Comparison of central blood pressure devices on the basis of a modified protocol of the European Society of Hypertension: application to the Centron cBP301.

    abstract:OBJECTIVES:Central blood pressure (cBP) and pulse pressure amplification (PPA) are receiving renewed interest with the increase in the availability of noninvasive techniques that enable its measurement. However, to date, there is no standardized protocol to validate their accuracy. Although invasive comparison seems in...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000028

    authors: Agnoletti D,Millasseau S,Topouchian J,Safar ME,Blacher J

    更新日期:2014-04-01 00:00:00

  • Circadian variation of blood pressure and the assessment of antihypertensive therapy.

    abstract::The inherent variability of activity and the awake-sleep cycle creates changes in hemodynamics that may influence the outcome of a variety of cardiovascular disorders. Recently, data suggest that increased variability of blood pressure may promote excessive hypertensive target organ disease. Several epidemiologic stud...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:

    authors: White WB

    更新日期:1999-01-01 00:00:00

  • Cardiovascular triggers and morning events.

    abstract::Sudden cardiac death, myocardial infarction, unstable angina, ventricular tachyarrhythmias, fatal pulmonary thromboembolism, rupture of thoracic aorta, and ischemic and hemorrhagic cerebrovascular accidents exhibit a prominent circadian pattern with events more frequently during the morning (0600-1200 h)after awakenin...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Chasen C,Muller JE

    更新日期:1998-02-01 00:00:00

  • Serum copeptin levels in predicting nondippers in newly diagnosed hypertension.

    abstract:OBJECTIVES:The nondipper pattern in hypertension is associated with worse cardiovascular outcomes. In this study, we hypothesized that high copeptin levels could predict nondippers among hypertensive patients and investigated the associations between copeptin levels and nondipper pattern in newly diagnosed hypertensive...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章

    doi:10.1097/MBP.0000000000000123

    authors: Uzun F,Biyik I,Akturk IF,Yalcin AA,Erturk M,Oner E,Kalkan AK,Yalcin B,Atmaca H

    更新日期:2015-08-01 00:00:00

  • Frequency of spontaneous fist clenching during routine blood pressure measurements and its effect on measurement accuracy.

    abstract:BACKGROUND:Guidelines for office blood pressure reading techniques do not warn us about the possibility that patients may clench their fist during cuff-inflation. It is unknown how often patients do this, and what effect it has on measured blood pressure readings. DESIGN AND METHODS:We registered double blood pressure...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200206000-00001

    authors: Smulders YM,Godfried MH,van Montfrans GA

    更新日期:2002-06-01 00:00:00

  • Reference values for ambulatory blood pressure and self-measured blood pressure based on prospective outcome data.

    abstract::Although self-measured blood pressure and ambulatory blood pressure are widely used for research and clinical purposes, reference values are still scarcely supported by prospective outcome data. For self-measured blood pressure, values of 135 mmHg systolic and 85 mmHg diastolic (an average of two measurements in the m...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/00126097-200112000-00011

    authors: Verdecchia P

    更新日期:2001-12-01 00:00:00

  • Angiotensin II receptor blockers following intravenous nicardipine administration to lower blood pressure in patients with hypertensive intracerebral hemorrhage: a prospective randomized study.

    abstract:BACKGROUND AND OBJECTIVE:In patients with hypertensive intracerebral hemorrhage (HICH), intravenous nicardipine is primarily used to lower blood pressure (BP). However, there are few studies investigating the role of oral antihypertensives administered after intravenous nicardipine to prevent BP from rising. Angiotensi...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0000000000000225

    authors: Inamasu J,Nakae S,Adachi K,Hirose Y

    更新日期:2017-02-01 00:00:00

  • Effect of bedtime melatonin ingestion on blood pressure of normotensive subjects.

    abstract::OBJECTIVE: To determine whether the evening intake of 5 mg melatonin affects 24 h arterial blood pressure and heart-rate profiles. METHODS: Twenty-one young normotensive subjects were administered placebo or melatonin for 4 weeks, according to a cross-over double-blind design, and were subjected to ambulatory blood pr...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Lusardi P,Preti P,Savino S,Piazza E,Zoppi A,Fogari R

    更新日期:1997-04-01 00:00:00

  • Quality of blood pressure phenotype in the Nigerian Population Research on Environment Gene and Health.

    abstract:OBJECTIVE:In the ongoing Nigerian Population Research on Environment Gene and Health (NIPREGH), we are applying standardized epidemiologic methods to determine cardiovascular phenotypes including blood pressure (BP) among adult Black Africans of Nigerian origin. We present the quality control of the conventionally meas...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章

    doi:10.1097/MBP.0000000000000053

    authors: Odili AN,Ameh VO,Ogedengbe JO,Staessen JA

    更新日期:2014-08-01 00:00:00

  • Do we undertreat hypertensive smokers? A comparison between smoking and non-smoking hypertensives.

    abstract::The acute effect of smoking is a rise in blood pressure and heart rate. Nevertheless, in several epidemiologic studies, smokers tend to have a slightly lower office blood pressure than non-smokers. We studied the 24 h ambulatory blood pressure consecutively in hypertensive smokers (> or = 10 cigarettes per day, n = 26...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200010000-00004

    authors: Bang LE,Buttenschøn L,Kristensen KS,Svendsen TL

    更新日期:2000-10-01 00:00:00