Accuracy of the SpaceLabs Medical 90217 ambulatory blood pressure monitor.

Abstract:

:OBJECTIVE: To test the SpaceLabs Medical 90217 ambulatory blood pressure monitor for compliance with The Association for the Advancement of Medical Instrumentation's standard and a modification of the British Hypertension Society (BHS) protocol. METHODS: The study concerned 85 subjects aged 17-88 years with various arm circumferences and ranges of blood pressure. Resting casual blood pressure was 135 +/- 24/76 +/- 14 mmHg (mean +/- SD). For each subject nine readings of the 90217 device were compared with auscultatory reference readings of two independent observers who used a dual stethoscope. The measurements were performed with subjects standing, sitting and supine. The manual reference readings were alternated with the 90217 monitor's readings. RESULTS: Average differences between manual (average of recordings by two clinicians) and automated readings were -5 +/- 4.3/0.6 +/- 4.6 mmHg (seated), 0.0 +/- 6.0/-1.8 +/- 4.7 mmHg (standing) and -1.6 +/- 5.1/-0.7 +/- 5.1 mmHg (supine).CONCLUSIONS: The device meets the Association for the Advancement of Medical Instrumentation's standard. Comparing the results with the modified BHS protocol, it was found that the device earned the highest BHS grade of 'A' both for systolic and for diastolic blood pressures.

journal_name

Blood Press Monit

authors

Baumgart P,Kamp J

subject

Has Abstract

pub_date

1998-10-01 00:00:00

pages

303-307

issue

5

eissn

1359-5237

issn

1473-5725

journal_volume

3

pub_type

杂志文章
  • The effect of crossing legs on blood pressure.

    abstract:OBJECTIVE:To determine whether crossing of the legs at the knee or at the ankles during blood pressure measurement in sitting position has an effect on blood pressure. METHODS:One hundred and eleven patients, 60 women, mean age 52+/-17 years (19-80): 49 chronically treated hypertensives, 28 treated diabetics and 34 no...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e3280b083a7

    authors: Adiyaman A,Tosun N,Elving LD,Deinum J,Lenders JW,Thien T

    更新日期:2007-06-01 00:00:00

  • Adequacy of validation of wide-range cuffs used with home blood pressure monitors: a systematic review.

    abstract:OBJECTIVES:Wide-range 'one-size-fits-all' blood pressure (BP) cuffs are commonly sold with home BP monitors. Assessment of the accuracy of these cuffs is important because they do not adhere to the basic principles of proper cuffing. The aim of this systematic review was to review the published validation data evaluati...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000344

    authors: Sprague E,Padwal RS

    更新日期:2018-10-01 00:00:00

  • Evaluation, management, and referral of elderly emergency department patients with elevated blood pressure.

    abstract:OBJECTIVES:To determine blood pressure (BP) reassessment rates and to describe the evaluation and outpatient referral rates of elderly emergency department (ED) patients with elevated BP. METHODS:This was a retrospective cohort of patients who were at least 60 years, presented with a systolic BP of at least 140 mmHg o...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,多中心研究

    doi:10.1097/MBP.0b013e328332fd40

    authors: Baumann BM,Cienki JJ,Cline DM,Egging D,Lehrmann JF,Tanabe P

    更新日期:2009-12-01 00:00:00

  • Unchanged 24 h ambulatory blood pressure during shjort-term salt restriction and salt repletion in normotensive volunteers.

    abstract::BACKGROUND: The lack of significant correlation between sodium consumption and blood pressure in a population has tentatively been explained by individual differences in salt sensitivity. Previously we have shown that one-third of essential hypertensived patients respond to a salt load of 9 g daily for 5 days with an ...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Raeder H,Aaberg M,Omvik P

    更新日期:1996-02-01 00:00:00

  • Beneficial response of blood pressure to short-term continuous positive airway pressure in Chinese patients with obstructive sleep apnea-hypopnea syndrome.

    abstract:BACKGROUND:Obstructive sleep apnea-hypopnea syndrome (OSAHS) is common in people with hypertension and cardiac rhythm disorder. The objectives of this study were to determine the effect of short-term continuous positive airway pressure (CPAP) on blood pressure (BP) and cardiac rhythm in Chinese patients with moderate t...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000324

    authors: Wang X,Qiu J,Wang Y,Cai Z,Lu X,Li T

    更新日期:2018-08-01 00:00:00

  • Canadian valsartan study in patients with mild-to-moderate hypertension.

    abstract:OBJECTIVE:This study was designed mainly to establish the rates of response to valsartan 80mg once daily (qd) and to valsartan 160mg qd given to non-responders to 80mg qd, as well as to determine the safety of valsartan and the blood pressure control achieved using valsartan over a period of 24 h or more, using ambulat...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章,多中心研究

    doi:10.1097/00126097-200104000-00005

    authors: Lasko BH,Laplante A,Hébert D,Bonnefis-Boyer S

    更新日期:2001-04-01 00:00:00

  • Dialysis-induced hypotension is associated with impaired aortic elasticity in patients undergoing chronic hemodialysis.

    abstract:BACKGROUND:The pathogenesis of dialysis-induced hypotension (DIH) is multifactorial and may include increased aortic stiffness. This study was undertaken to determine aortic elastic properties in patients undergoing hemodialysis with and without DIH, and to assess the effects of DIH on mortality. METHODS:Aortic stiffn...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,多中心研究

    doi:10.1097/MBP.0b013e3282f13f75

    authors: Duman D,Demirtunc R,Erdogan B,Sahin GM,Karadag B

    更新日期:2008-04-01 00:00:00

  • 24 h non-invasive ambulatory blood pressure monitoring: do the number of recordings per hour and/or ways of analyzing day and night matter?

    abstract:OBJECTIVE:To analyze whether the number of blood pressure registrations per hour and/or different ways of defining day and night have any influence on the result of ambulatory blood pressure monitoring. DESIGN AND METHODS:The study used blood pressure data from 40 normotensive subjects (20 men) and 40 (20 men) treated...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200110000-00005

    authors: Enström IE,Pennert KM

    更新日期:2001-10-01 00:00:00

  • Frequency of spontaneous fist clenching during routine blood pressure measurements and its effect on measurement accuracy.

    abstract:BACKGROUND:Guidelines for office blood pressure reading techniques do not warn us about the possibility that patients may clench their fist during cuff-inflation. It is unknown how often patients do this, and what effect it has on measured blood pressure readings. DESIGN AND METHODS:We registered double blood pressure...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200206000-00001

    authors: Smulders YM,Godfried MH,van Montfrans GA

    更新日期:2002-06-01 00:00:00

  • Determination of accuracies of 10 models of home blood pressure monitors using an oscillometric simulator.

    abstract:BACKGROUND:Patients and doctors often use home blood pressure monitoring (HBPM) to assess the control of hypertension. Despite its popularity there has always been some uncertainty with regard to its accuracy, reliability, reproducibility, and comparability. Although there are pre-market HBPM standards of accuracy, the...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-199904010-00007

    authors: Yarows SA,Amerena JV

    更新日期:1999-02-01 00:00:00

  • Conventional and ambulatory measurements of blood pressure in old patients with isolated systolic hypertension: baseline observations in the Syst-Eur trial.

    abstract::OBJECTIVES: To compare clinic and am measurements of blood pressure in old patients with isolated systolic hypertension and their reproducibilities. PATIENTS: In total 610 patients aged >/= 60 years with isolated systolic hypertension detected by clinic measurement were monitored during the placebo run-in phase of the...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Emelianov D,Thijs L,Staessen JA,Celis H,Clement D,Davidson C,Gasowski J,Gil-Extremera B,Fogari R,Jääskivi M,Lehtonen A,Nedogoda S,O'Brien E,Palatini P,Parati G,Salvetti A,Vanhanen H,Webster J,Fagard R

    更新日期:1998-06-01 00:00:00

  • Which is the correct term for blood pressure measurements taken at home?

    abstract::The terms 'self-blood pressure' and 'home blood pressure' are being used to describe measurements of blood pressure taken by the patients at home. However, home measurements are not always self-measurements, because these are often taken by the patients' relatives. There is little evidence on the effect of self-measur...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/00126097-200308000-00007

    authors: Stergiou GS

    更新日期:2003-08-01 00:00:00

  • Percentiles of waist-to-sitting-height ratio and its relationship with obesity and elevated blood pressure among children and adolescents in Shandong, China.

    abstract:BACKGROUND:Waist circumference (WC) and related anthropometric indices have been recognized as useful measures of abdominal obesity. However, little is known about the association of waist-to-sitting-height ratio (WSHtR) with obesity and elevated blood pressure (BP). PARTICIPANTS AND METHODS:Data for this study were o...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000154

    authors: Zhang YX,Zhao JS,Chu ZH

    更新日期:2016-02-01 00:00:00

  • Validation of the Omron HEM-907 device for blood pressure measurement.

    abstract:BACKGROUND:The aim of this study was to validate the Omron HEM-907 blood pressure (BP) measuring device according to the international validation protocol. DESIGN:The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, ...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200208000-00006

    authors: El Assaad MA,Topouchian JA,Darné BM,Asmar RG

    更新日期:2002-08-01 00:00:00

  • Validation protocols for blood pressure-measuring devices: status quo and development needs.

    abstract::Hypertension is a major risk factor for cardiovascular morbidity and mortality. Therefore, blood pressure self-measuring devices have emerged as valuable tools in patient care and the accuracy of these instruments is of fundamental importance. For this reason, several validation procedures for assessing the efficacy o...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/MBP.0000000000000150

    authors: Beime B,Deutsch C,Gomez T,Zwingers T,Mengden T,Bramlage P

    更新日期:2016-02-01 00:00:00

  • Masked hypertension, a review of the literature.

    abstract::Masked hypertension (blood pressure that is normal in the physicians' office but elevated elsewhere) is a common phenomenon as prevalence among studies varies from 8 to 45% and is seen at all ages. large discrepancies, however, exist between studies that have dealt with masked hypertension. It is of high clinical impo...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/mbp.0b013e3280fb2792

    authors: Verberk WJ,Thien T,de Leeuw PW

    更新日期:2007-08-01 00:00:00

  • Impact of choosing the closest rather than the mean auscultatory blood pressure value on the European Society of Hypertension International Protocol device validation results.

    abstract:OBJECTIVE:When calculating the difference between alternating auscultatory reference standard (ARS) and the automated device-under-test (DUT) blood pressure (BP) measurements, the European Society of Hypertension International Protocol (ESH-IP) allows investigators to choose the closest ARS value to the DUT value, rath...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000454

    authors: Padwal R,Wood PW,McLean D,Ringrose JS

    更新日期:2020-08-01 00:00:00

  • Automated office blood pressure - being alone and not location is what matters most.

    abstract:OBJECTIVE:Measurement of office blood pressure using a fully automated sphygmomanometer that takes multiple readings with the patient resting quietly alone has been called automated office blood pressure (AOBP). Almost all AOBP research has involved the patient resting alone in an examining room, which is often impract...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章

    doi:10.1097/MBP.0000000000000133

    authors: Armstrong D,Matangi M,Brouillard D,Myers MG

    更新日期:2015-08-01 00:00:00

  • Blood pressure monitoring in sleep: time to wake up.

    abstract::Hypertension is a highly common condition with well-established adverse consequences. Ambulatory blood pressure monitoring has repeatedly been shown to better predict cardiovascular outcomes and mortality, compared to single office visit blood pressure. Non-dipping of sleep-time blood pressure is an independent marker...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/MBP.0000000000000426

    authors: Kwon Y,Stafford PL,Lim DC,Park S,Kim SH,Berry RB,Calhoun DA

    更新日期:2020-04-01 00:00:00

  • Oscillometric sphygmomanometers: a critical appraisal of current technology.

    abstract:AIM:Blood pressure (BP), a key vital sign, monitors general health. Oscillometric devices are increasingly used for measurement, although their accuracy continues to be critically debated. A functional block diagram is used to review the components that affect accuracy. METHODS:A block diagram is presented covering th...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/MBP.0b013e32835026b0

    authors: Amoore JN

    更新日期:2012-04-01 00:00:00

  • Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT.

    abstract:BACKGROUND:Two devices for self-measurement of blood pressure at the brachial artery-the Omron M5-I and the Omron 705IT-were evaluated according to the international protocol of the European Society of Hypertension. DESIGN:The international validation protocol is divided into two phases: the first phase is performed o...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200306000-00006

    authors: El Assaad MA,Topouchian JA,Asmar RG

    更新日期:2003-06-01 00:00:00

  • The impact of the day of the week on home blood pressure: the Finn-Home study.

    abstract:OBJECTIVE:The impact of the day of the week on home blood pressure (BP) level and day-to-day BP profile is unknown. Our objectives were to examine (i) how the initial measurement day of the week affects 3-day and 7-day mean home BP and (ii) the BP variation between different days of the week. PARTICIPANTS AND METHODS:...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000156

    authors: Juhanoja EP,Puukka PJ,Johansson JK,Niiranen TJ,Jula AM

    更新日期:2016-04-01 00:00:00

  • Prognostic value of average real variability of systolic blood pressure in elderly treated hypertensive patients.

    abstract:OBJECTIVE:The independent prognostic significance of ambulatory blood pressure variability in the elderly is incompletely clear. We investigated the prognostic value of average real variability of 24-hour blood pressure in elderly treated hypertensive patients. METHODS:The occurrence of a combined end-point including ...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000381

    authors: Coccina F,Pierdomenico AM,Cuccurullo C,Pierdomenico SD

    更新日期:2019-08-01 00:00:00

  • Blood pressure cuff comparability study.

    abstract:BACKGROUND:Manufacturer-supplied blood pressure (BP) cuffs are part of the automatic oscillometric BP devices algorithm. MATERIALS AND METHODS:This study assessed the differences in BP values using the Omron HEM 907-XL (Omron) device with two types of cuffs: the Baum cuff (BC) and the supplied Omron cuff (OC). A sampl...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0000000000000208

    authors: Ostchega Y,Nwankwo T,Zhang G,Chiappa M

    更新日期:2016-12-01 00:00:00

  • Calibration of blood pressure data after replacement of the standard mercury sphygmomanometer by an oscillometric device and concurrent change of cuffs.

    abstract::Oscillometric blood pressure (BP) measurement devices are increasingly replacing standard mercury sphygmomanometers and generalizability of validation studies to other environments, for example, national survey environments, is assumed. We compared BP measurements according to two highly standardized German national s...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000081

    authors: Neuhauser HK,Ellert U,Thamm M,Adler C

    更新日期:2015-02-01 00:00:00

  • Circadian variation of blood pressure and the assessment of antihypertensive therapy.

    abstract::The inherent variability of activity and the awake-sleep cycle creates changes in hemodynamics that may influence the outcome of a variety of cardiovascular disorders. Recently, data suggest that increased variability of blood pressure may promote excessive hypertensive target organ disease. Several epidemiologic stud...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:

    authors: White WB

    更新日期:1999-01-01 00:00:00

  • Ambulatory blood pressure monitoring of chronically dialyzed pediatric patients.

    abstract:BACKGROUND:Background Ambulatory blood pressure monitoring (ABPM) has been shown to be more representative of blood pressure levels in adult patients than are casual measurements of blood pressure. OBJECTIVE:To evaluate, by means of ABPM, the behavior of blood pressure in children with chronic renal failure submitted ...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Koch VH,Furusawa EA,Ignes E,Okay Y,Mion Junior D

    更新日期:1999-10-01 00:00:00

  • Effect of bedtime melatonin ingestion on blood pressure of normotensive subjects.

    abstract::OBJECTIVE: To determine whether the evening intake of 5 mg melatonin affects 24 h arterial blood pressure and heart-rate profiles. METHODS: Twenty-one young normotensive subjects were administered placebo or melatonin for 4 weeks, according to a cross-over double-blind design, and were subjected to ambulatory blood pr...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Lusardi P,Preti P,Savino S,Piazza E,Zoppi A,Fogari R

    更新日期:1997-04-01 00:00:00

  • Reproducibility of ambulatory blood pressure monitoring.

    abstract::OBJECTIVE: To investigate whether blood pressure monitoring is reproducible. DESIGN: Reproducibility of ambulatory blood pressure monitoring data was assessed by means of traditional and relatively new statistical methods, namely correlation coefficients, regression analysis and agreement analysis. METHODS: Ninety-one...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Musso NR,Lotti G

    更新日期:1996-04-01 00:00:00

  • Frequency domain of heart rate and blood pressure variability in essential hypertensive patients during sleep: differences between dippers and non-dippers.

    abstract::METHODS: Autonomic nervous function was evaluated by means of power spectral analysis of heart rate and blood pressure variability in dipper (n = 10) and non-dipper (n = 9) essential hypertensive subjects during sleep. The non-dipper subjects were defined as those in whom the nocturnal decrease in blood pressure was <...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Frisina N,Pedullà M,Mento G,Lanuzza B

    更新日期:1996-10-01 00:00:00