Performance of a 41 x 41 cm2 amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications.

Abstract:

:We measured the physical imaging performance of a 41 x 41 cm2 amorphous silicon flat panel detector designed for angiographic and R&F imaging applications using methods from the emerging IEC standard for the measurement of detective quantum efficiency (DQE) in digital radiographic detectors. Measurements on 12 production detectors demonstrate consistent performance. The mean DQE at the detector center is about 0.77 at zero frequency and 0.27 at the Nyquist frequency (2.5 cycles/mm) when measured with a 7 mm of Al HVL spectrum at about 3.6 microGy. The mean MTF at the center of the detector for this spectrum is 0.24 at the Nyquist frequency. For radiographic operation all 2048 x 2048 detector elements are read out individually. For fluoroscopy, the detector operates in two 30 frame per second modes: either the center 1024 x 1024 detector elements are read out or the entire detector is read out with 2 x 2 pixel binning. A model was developed to predict differences in performance between the modes, and measurements demonstrate agreement with the model. Lag was measured using a quasi-equilibrium exposure method and was found to be 0.044 in the first frame and less than 0.007 after 1 s. We demonstrated that it is possible to use the lag data to correct for temporal correlation in images when measuring DQE with a fluoroscopic imaging technique. Measurements as a function of position on the detector demonstrate a high degree of uniformity. We also characterized dependences on spectrum, exposure level, and direction. Finally, we measured the DQE of a current state of the art image intensifier/CCD system using the same method as for the flat panel. We found the image intensifier system to have lower DQE than the flat panel at high exposure levels and approximately equivalent DQE at fluoroscopic levels.

journal_name

Med Phys

journal_title

Medical physics

authors

Granfors PR,Aufrichtig R,Possin GE,Giambattista BW,Huang ZS,Liu J,Ma B

doi

10.1118/1.1609151

subject

Has Abstract

pub_date

2003-10-01 00:00:00

pages

2715-26

issue

10

eissn

0094-2405

issn

2473-4209

journal_volume

30

pub_type

杂志文章
  • Modeling acute and chronic hypoxia using serial images of 18F-FMISO PET.

    abstract:PURPOSE:Two types of tumor hypoxia most likely exist in human cancers: Chronic hypoxia due to the paucity of blood capillaries and acute hypoxia due to temporary shutdoWn of microvasculatures or fluctuation in the red cell flux. In a recent hypoxia imaging study using 18F-FMISO PET, the authors observed variation in tr...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3213092

    authors: Wang K,Yorke E,Nehmeh SA,Humm JL,Ling CC

    更新日期:2009-10-01 00:00:00

  • Plan averaging for multicriteria navigation of sliding window IMRT and VMAT.

    abstract:PURPOSE:To describe a method for combining sliding window plans [intensity modulated radiation therapy (IMRT) or volumetric modulated arc therapy (VMAT)] for use in treatment plan averaging, which is needed for Pareto surface navigation based multicriteria treatment planning. METHODS:The authors show that by taking an...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4859295

    authors: Craft D,Papp D,Unkelbach J

    更新日期:2014-02-01 00:00:00

  • Correction factors for water-proofing sleeves in kilovoltage x-ray beams.

    abstract::This paper investigates the effect of the waterproofing sleeve on the calibration of kilovoltage photon beams (50-300 kV). The sleeve effect correction factor, ps has been calculated using the Monte Carlo method as the ratios of the air kerma in an air cavity of a cylindrical chamber without the waterproofing sleeve t...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598039

    authors: Ma CM,Seuntjens JP

    更新日期:1997-09-01 00:00:00

  • A voxel-based multiscale model to simulate the radiation response of hypoxic tumors.

    abstract:PURPOSE:In radiotherapy, it is important to predict the response of tumors to irradiation prior to the treatment. This is especially important for hypoxic tumors, which are known to be highly radioresistant. Mathematical modeling based on the dose distribution, biological parameters, and medical images may help to impr...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4903298

    authors: Espinoza I,Peschke P,Karger CP

    更新日期:2015-01-01 00:00:00

  • Systematic feasibility analysis of performing elastography using reduced dose CT lung image pairs.

    abstract:PURPOSE:Elastography using computer tomography (CT) is a promising methodology that can provide patient-specific regional distributions of lung biomechanical properties. The purpose of this paper is to investigate the feasibility of performing elastography using simulated lower dose CT scans. METHODS:A cohort of eight...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14112

    authors: Hasse K,Hsieh SS,O'Connell D,Stiehl B,Min Y,Neylon J,Low DA,Santhanam AP

    更新日期:2020-08-01 00:00:00

  • Detection nonuniformity measurements and corrections for cone beam transmission CT on a gamma camera.

    abstract::Cone beam transmission CT (CB-CT) improves SPECT imaging by providing high-quality attenuation maps for attenuation compensation and for correlated SPECT and CT imaging. The present work measures the detection nonuniformity for CB-CT implemented with a gamma camera, and applies nonuniformity corrections to make CB-CT ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596838

    authors: Manglos SH,Capone RB,Bassano DA

    更新日期:1992-03-01 00:00:00

  • A sparse orthogonal collimator for small animal intensity-modulated radiation therapy part I: Planning system development and commissioning.

    abstract:PURPOSE:To achieve more translatable preclinical research results, small animal irradiation needs to more closely simulate human radiotherapy. Although the clinical gold standard is intensity-modulated radiation therapy (IMRT), the direct translation of this method for small animals is impractical. In this study we des...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13872

    authors: Woods K,Nguyen D,Neph R,Ruan D,O'Connor D,Sheng K

    更新日期:2019-12-01 00:00:00

  • Fast time-of-flight camera based surface registration for radiotherapy patient positioning.

    abstract:PURPOSE:This work introduces a rigid registration framework for patient positioning in radiotherapy, based on real-time surface acquisition by a time-of-flight (ToF) camera. Dynamic properties of the system are also investigated for future gating/tracking strategies. METHODS:A novel preregistration algorithm, based on...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3664006

    authors: Placht S,Stancanello J,Schaller C,Balda M,Angelopoulou E

    更新日期:2012-01-01 00:00:00

  • Determination of electron energy, spectral width, and beam divergence at the exit window for clinical megavoltage x-ray beams.

    abstract::Monte Carlo simulations of x-ray beams typically take parameters of the electron beam in the accelerating waveguide to be free parameters. In this paper, a methodology is proposed and implemented to determine the energy, spectral width, and beam divergence of the electron source. All treatment head components were rem...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3070547

    authors: Sawkey DL,Faddegon BA

    更新日期:2009-03-01 00:00:00

  • SU-E-T-619: A Network-Flow Solution Approach to VMAT Treatment Plan Optimization.

    abstract:PURPOSE:To add mathematical rigor to the merging phase of the recently published two-stage VMAT optimization method called VMERGE. Using an exact merging method, we are able to better characterize the tradeoff between delivery efficiency and dose quality. METHODS:VMERGE begins with an IMRT plan that uses 180 equi-spac...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735709

    authors: Salari E,Craft D

    更新日期:2012-06-01 00:00:00

  • Comparison of breast density measured on MR images acquired using fat-suppressed versus nonfat-suppressed sequences.

    abstract:PURPOSE:To investigate the difference of MR percent breast density measured from fat-suppressed versus nonfat-suppressed imaging sequences. METHODS:Breast magnetic resonance imaging (MRI) with and without fat suppression was acquired from 38 subjects. Breasts were divided into subgroups of different morphological patt...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3646756

    authors: Chang DH,Chen JH,Lin M,Bahri S,Yu HJ,Mehta RS,Nie K,Hsiang DJ,Nalcioglu O,Su MY

    更新日期:2011-11-01 00:00:00

  • Skeletal dosimetry in cone beam computed tomography.

    abstract::Cone beam computed tomography (CBCT) is a relatively new patient imaging technique that has proved invaluable for treatment target verification and patient positioning during image-guided radiotherapy (IGRT). It has been shown that CBCT results in additional dose to bone that may amount to 10% of the prescribed dose. ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3147143

    authors: Walters BR,Ding GX,Kramer R,Kawrakow I

    更新日期:2009-07-01 00:00:00

  • The accuracy of SPECT brain activation images: propagation of registration errors.

    abstract::Functional single photon emission computed tomography (SPECT) images of brain activation are based on a comparison of base line and activation images. The correctness of the functional images depends, among other factors, on the accurate spatial registration (alignment) of the base line and activation image data. The ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597228

    authors: Sychra JJ,Pavel DG,Chen Y,Jani A

    更新日期:1994-12-01 00:00:00

  • A formal study of lateral magnification and its influence on mammographic imaging sharpness.

    abstract::Radiographic magnification is often employed in mammography to improve detection of small anatomical details of clinical interest, such as microcalcifications. The image size depends on the conventional magnification factor, on the focal spot size, and on the x-ray angulation. In this paper magnification effects are i...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597288

    authors: Schiabel H,Ventura A,Frère AF

    更新日期:1994-02-01 00:00:00

  • Flattening filter removal for improved image quality of megavoltage fluoroscopy.

    abstract:PURPOSE:Removal of the linear accelerator (linac) flattening filter enables a high rate of dose deposition with reduced treatment time. When used for megavoltage imaging, an unflat beam has reduced primary beam scatter resulting in sharper images. In fluoroscopic imaging mode, the unflat beam has higher photon count pe...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4812678

    authors: Christensen JD,Kirichenko A,Gayou O

    更新日期:2013-08-01 00:00:00

  • High resolution digital autoradiographic and dosimetric analysis of heterogeneous radioactivity distribution in xenografted prostate tumors.

    abstract:PURPOSE:The first main aim of this study was to illustrate the absorbed dose rate distribution from 177Lu in sections of xenografted prostate cancer (PCa) tumors using high resolution digital autoradiography (DAR) and compare it with hypothetical identical radioactivity distributions of 90Y or 7 MeV alpha-particles. Th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4967877

    authors: Timmermand OV,Nilsson J,Strand SE,Elgqvist J

    更新日期:2016-12-01 00:00:00

  • Unequal weighting of given doses in opposed fields in treatment of cancer of the tonsillar region using 60Co, 4-, 8-, 15-, 24-MVp photons.

    abstract::Weighting of given doses from parallel opposed fields is a widespread practice in radiation therapy of cancer in the tonsillar region. In order to determine when weighting is useful, dose distributions on the central axis were calculated for various energies (60Co, 4-, 8-, 15-, and MVp photons), weighting factors, and...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594276

    authors: Wollin M,Gilbert HA,Kagan AR

    更新日期:1976-03-01 00:00:00

  • A model to simulate the oxygen distribution in hypoxic tumors for different vascular architectures.

    abstract:PURPOSE:As hypoxic cells are more resistant to photon radiation, it is desirable to obtain information about the oxygen distribution in tumors prior to the radiation treatment. Noninvasive techniques are currently not able to provide reliable oxygenation maps with sufficient spatial resolution; therefore mathematical m...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4812431

    authors: Espinoza I,Peschke P,Karger CP

    更新日期:2013-08-01 00:00:00

  • A method to compare supra-pubic ultrasound and CT images of the prostate: technique and early clinical results.

    abstract::We describe a unique method that allows the comparison of spatially registered ultrasound (SRUS) images and computed tomography-derived contours (CTDCs) that were acquired with a minimal time lapse. As such, we have a tool that will provide validation of the spatial accuracy of the US system and that will allow compar...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1644515

    authors: Molloy JA,Srivastava S,Schneider BF

    更新日期:2004-03-01 00:00:00

  • Three-dimensional dose reconstruction of breast cancer treatment using portal imaging.

    abstract::In this study, we present an algorithm for three-dimensional (3-D) dose reconstruction using portal images obtained with an electronic portal imaging device (EPID). For this purpose an algorithm for 2-D dose reconstruction, which was previously developed in our institution, was adapted. The external contour of the pat...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1589496

    authors: Louwe RJ,Damen EM,van Herk M,Minken AW,Törzsök O,Mijnheer BJ

    更新日期:2003-09-01 00:00:00

  • Semiautomated four-dimensional computed tomography segmentation using deformable models.

    abstract::The purpose of this work is to demonstrate a proof of feasibility of the application of a commercial prototype deformable model algorithm to the problem of delineation of anatomic structures on four-dimensional (4D) computed tomography (CT) image data sets. We acquired a 4D CT image data set of a patient's thorax that...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1929207

    authors: Ragan D,Starkschall G,McNutt T,Kaus M,Guerrero T,Stevens CW

    更新日期:2005-07-01 00:00:00

  • Effective point of measurement of thimble ion chambers in megavoltage photon beams.

    abstract:PURPOSE:Determine the effective point of measurement (EPOM) of 12 thimble ion chambers, including miniature chambers and three models widely used for clinical reference dosimetry. The EPOM is the point at which the measured dose would arise in the measurement medium in the absence of the probe: For cylindrical chambers...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3266750

    authors: Tessier F,Kawrakow I

    更新日期:2010-01-01 00:00:00

  • A generic TG-186 shielded applicator for commissioning model-based dose calculation algorithms for high-dose-rate 192 Ir brachytherapy.

    abstract:PURPOSE:A joint working group was created by the American Association of Physicists in Medicine (AAPM), the European Society for Radiotherapy and Oncology (ESTRO), and the Australasian Brachytherapy Group (ABG) with the charge, among others, to develop a set of well-defined test case plans and perform calculations and ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12459

    authors: Ma Y,Vijande J,Ballester F,Tedgren ÅC,Granero D,Haworth A,Mourtada F,Fonseca GP,Zourari K,Papagiannis P,Rivard MJ,Siebert FA,Sloboda RS,Smith R,Chamberland MJP,Thomson RM,Verhaegen F,Beaulieu L

    更新日期:2017-11-01 00:00:00

  • Tissue equivalence in clinical neutron dosimetry: comparison of dose distributions in nine tissue substitutes for a d(14)Be neutron beam.

    abstract::Depth dose distributions for a d(14)Be neutron beam, measured separately for total and gamma absorbed dose, are presented for nine tissue substitutes and for two field sizes. These data are used to examine methods to transform depth dose curves from one material to another. Best results are found when the local depths...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595695

    authors: Hensley FW,Rassow J,Temme A

    更新日期:1985-05-01 00:00:00

  • Sinusoidal modulation analysis for optical system MTF measurements.

    abstract::The modulation transfer function (MTF) is a commonly used metric for defining the spatial resolution characteristics of imaging systems. While the MTF is defined in terms of how an imaging system demodulates the amplitude of a sinusoidal input, this approach has not been in general use to measure MTFs in the medical i...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597840

    authors: Boone JM,Yu T,Seibert JA

    更新日期:1996-12-01 00:00:00

  • A time-domain wavelet-based approach for fluorescence diffuse optical tomography.

    abstract:PURPOSE:In the context of fluorescence diffuse optical tomography, determining the optimal way to exploit the time-resolved information has been receiving much attention and different features of the time-resolved signals have been introduced. In this article, the authors revisit and generalize the notion of feature, c...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3431571

    authors: Ducros N,Da Silva A,Dinten JM,Seelamantula CS,Unser M,Peyrin F

    更新日期:2010-06-01 00:00:00

  • Anniversary paper: Role of medical physicists and the AAPM in improving geometric aspects of treatment accuracy and precision.

    abstract::The last 50 years have seen great advances in the accuracy of external beam radiation therapy. Geometrical uncertainties have been reduced from a centimeter or more in presimulation, skin-mark guided days to 1-2 mm in today's image-guided radiation therapy treatments. Medical physicists, with the support and guidance ...

    journal_title:Medical physics

    pub_type: 杂志文章,评审

    doi:10.1118/1.2836420

    authors: Yorke ED,Keall P,Verhaegen F

    更新日期:2008-03-01 00:00:00

  • Radiologic validation of a fast neutron multileaf collimator.

    abstract::Teletherapy with high linear energy transfer radiations (LET), perhaps more than with low LET types, requires careful beam collimation to limit effects to normal structures. Intensity modulated techniques may also hold promise in this regard. Accordingly, a remote computer-controlled, high-resolution multileaf collima...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2760026

    authors: Farr JB,Maughan RL,Yudelev M,Blosser E,Brandon J,Horste T,Forman JD

    更新日期:2007-09-01 00:00:00

  • Loss of field flatness in AECL teletherapy units form misadjustment of the collimator.

    abstract::Atomic Energy of Canada Limited 60Co teletherapy units have adjustable collimator hinges to adapt to source diameters from 0.75 to 2.50 cm. Adjustment of the collimator hinge to the manfacturer's settings can cause a loss of radiation field flatness as high as 20% for large fields. The recommended settings appear to c...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594170

    authors: Heintz PH

    更新日期:1975-03-01 00:00:00

  • FRoG: An independent dose and LETd prediction tool for proton therapy at ProBeam® facilities.

    abstract:PURPOSE:Particle therapy is becoming increasingly available world-wide for precise tumor targeting, its favorable depth dose deposition, and increased biological damage to tumor tissue compared to conventional photon therapy. As demand increases for improved robustness and conformality, next-generation secondary dose c...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14417

    authors: Kopp B,Fuglsang Jensen M,Mein S,Hoffmann L,Nyström H,Falk M,Haberer T,Abdollahi A,Debus J,Mairani A

    更新日期:2020-10-01 00:00:00