Clinical evaluation of a new type of centrifugal pump.

Abstract:

:The major problems with existing centrifugal pumps are leakage, mechanical trauma, and thrombus formation. In consideration of these problems, a new compact centrifugal pump system was developed. The purpose of this study was to evaluate the new centrifugal pump system clinically. Ten patients underwent open heart surgery with a centrifugal pump or a roller pump. During surgery, hemodynamic and hematological data were obtained. A pulsatile assist device in the pump circuit was used in patients with severe heart disease. There was neither operative death nor hospital mortality, and there was no difference with regard to hemodynamic data between the two groups. The centrifugal pump groups, however, had significantly lower hemolysis, especially during prolonged cardiopulmonary bypass. This centrifugal pump could also create sufficient pulsatile flow with a pulsatile assist device. Postoperative macroscopic and microscopic findings demonstrated the smooth surface of the pump without thrombus formation. This centrifugal pump system might be useful for prolonged cardiopulmonary bypass.

journal_name

Artif Organs

journal_title

Artificial organs

authors

Ninomiya J,Shoji T,Tanaka S,Ikeshita M,Ochi M,Yamauchi S,Yajima T,Yamauchi H,Sugimoto T,Aizawa T

doi

10.1111/j.1525-1594.1994.tb03403.x

subject

Has Abstract

pub_date

1994-09-01 00:00:00

pages

702-5

issue

9

eissn

0160-564X

issn

1525-1594

journal_volume

18

pub_type

杂志文章
  • Cerebral oxygen metabolism during total body flow and antegrade cerebral perfusion at deep and moderate hypothermia.

    abstract::The aim of this study is to evaluate the effect of temperature on cerebral oxygen metabolism at total body flow bypass and antegrade cerebral perfusion (ACP). Neonatal piglets were put on cardiopulmonary bypass (CPB) with the initial flow rate of 200mL/kg/min. After cooling to 18°C (n=6) or 25°C (n=7), flow was reduce...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2010.01131.x

    authors: Sasaki T,Boni L,Riemer RK,Yeung JT,Ramamoorthy C,Beckman R,Gisner C,Shuttleworth P,Hanley FL,Reddy VM

    更新日期:2010-11-01 00:00:00

  • Combination of plasma exchange and continuous hemofiltration as temporary metabolic support for patients with acute liver failure.

    abstract::The combination of plasma exchange and continuous hemofiltration was applied to patients with acute liver failure, and its clinically advantageous effects were evaluated. This procedure was found to be a powerful approach in decreasing patient morbidity; however, the effect was temporary and the patients finally died....

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1994.tb02217.x

    authors: Matsubara S

    更新日期:1994-05-01 00:00:00

  • Regenerative medicine for cardiovascular disorders-new milestones: adult stem cells.

    abstract::Cardiovascular disorders are the leading causes of mortality and morbidity in the developed world. Cell-based modalities have received considerable scientific attention over the last decade for their potential use in this clinical arena. This review was intended as a brief overview on the subject of therapeutic potent...

    journal_title:Artificial organs

    pub_type: 杂志文章,评审

    doi:10.1111/j.1525-1594.2006.00209.x

    authors: Akar AR,Durdu S,Corapcioglu T,Ozyurda U

    更新日期:2006-04-01 00:00:00

  • Partial artificial heart (ALVAD) use with subsequent cardiac and renal allografting in a patient with stone heart syndrome.

    abstract::The abdominal left ventricular assist device (ALVAD) is an order of magnitude more effective than conventional intra-aortic balloon pumping (IABP) in unloading and providing circulatory support to the failing left ventricle. This is a report of a unique case which demonstrates that in the absence of pulmonary vascular...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1978.tb01632.x

    authors: Norman JC,Dacso CC,Reul GJ,Massin EK,Klima T,Kahan BD,Keiser JT,Cooley DA

    更新日期:1978-11-01 00:00:00

  • Further development of hybrid functional electrical stimulation orthoses.

    abstract::In this study two aspects of hybrid functional electrical stimulation (FES) orthoses were investigated: joint motion constraints and FES control strategies. First, the effects of joint motion constraints on the gait of normal subjects were investigated using modern motion analysis systems, including electromyogram (EM...

    journal_title:Artificial organs

    pub_type: 临床试验,杂志文章

    doi:10.1111/j.1525-1594.1997.tb04648.x

    authors: Yang L,Granat MH,Paul JP,Condie DN,Rowley DI

    更新日期:1997-03-01 00:00:00

  • Binding of antithrombin to immobilized heparin under varying flow conditions.

    abstract::Rheologic factors are likely to influence the balance between thrombotic and antithrombotic forces at the level of the vascular wall. In this study, the effects of flow-velocity/wall shear stress on the interaction of antithrombin (AT) with surface-immobilized heparin were investigated. The binding of AT to low-affini...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:

    authors: Pasche B,Elgue G,Olsson P,Riesenfeld J,Rasmuson A

    更新日期:1991-12-01 00:00:00

  • Anemia in acute renal failure: role for erythropoiesis-stimulating proteins?

    abstract::The use of recombinant human erythropoietin (rHuEpo) has revolutionized anemia management of early and late stages of chronic kidney disease. Darbopoietin is also now available for the treatment of anemia of chronic kidney disease. In addition, rHuEpo has been used for the treatment of anemia observed in critical illn...

    journal_title:Artificial organs

    pub_type: 杂志文章,评审

    doi:10.1046/j.1525-1594.2003.07287.x

    authors: Liangos O,Pereira BJ,Jaber BL

    更新日期:2003-09-01 00:00:00

  • Oxidative Stress Increases Erythrocyte Sensitivity to Shear-Mediated Damage.

    abstract::Patients receiving mechanical circulatory support often present with heightened inflammation and free radical production associated with pre-existing conditions in addition to that which is due to blood interactions with nonbiological surfaces. The aim of this experimental laboratory study was to assess the deformabil...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12997

    authors: McNamee AP,Horobin JT,Tansley GD,Simmonds MJ

    更新日期:2018-02-01 00:00:00

  • Flow behavior within the 12-cc Penn State pulsatile pediatric ventricular assist device: an experimental study of the initial design.

    abstract::Planar particle image velocimetry was used to explore the flow behavior of the newly designed 12-cc Penn State pneumatic pediatric assist pump. Wall shear maps complemented the velocity data. Bjork-Shiley Monostrut 17-mm mechanical heart valves were used in the inlet and outlet ports. In comparison with larger Penn St...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2008.00565.x

    authors: Manning KB,Wivholm BD,Yang N,Fontaine AA,Deutsch S

    更新日期:2008-06-01 00:00:00

  • Development of a novel centrifugal pump: magnetic rotary pump.

    abstract::The rotational axis of the centrifugal pump has some associated problems such as blood destruction and sealing between the axis and pump housing. To improve upon these deficits we have developed a new type of blood pump, the magnetic rotary pump (MRP). The MRP has an original design with no rotational axis and no impe...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1997.tb03734.x

    authors: Naganuma S,Yambe T,Sonobe T,Kobayashi S,Nitta S

    更新日期:1997-07-01 00:00:00

  • Serum hyaluronic acid and interleukin-6 as possible markers of carpal tunnel syndrome in chronic hemodialysis patients.

    abstract::To elucidate the precise mechanism of carpal tunnel syndrome (CTS), serum hyaluronic acid (HA), interleukin-1 beta (IL-1 beta), interleukin-6 (IL-6), and tumor necrosis factor-alpha (TNF-alpha) were measured in 71 chronic hemodialysis patients with or without CTS and/or shoulder pain. Patients were divided into two gr...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1994.tb02227.x

    authors: Takasu S,Takatsu S,Kunitomo K,Kokumai Y

    更新日期:1994-06-01 00:00:00

  • Pulse conductance and flow-induced hemolysis during pulsatile cardiopulmonary bypass.

    abstract::In this study, the hypothesis was tested that a low-resistant, high-compliant oxygenator provides better pulse conductance and less hemolysis than a high-resistant, low-compliant oxygenator during pulsatile cardiopulmonary bypass. Forty adults undergoing coronary artery bypass surgery were randomly divided into two gr...

    journal_title:Artificial organs

    pub_type: 杂志文章,随机对照试验

    doi:10.1111/j.1525-1594.2009.00853.x

    authors: Simons AP,Wortel P,van Kan RA,van der Veen FH,Weerwind PW,Maessen JG

    更新日期:2010-04-01 00:00:00

  • Circulatory loop design and components introduce artifacts impacting in vitro evaluation of ventricular assist device thrombogenicity: A call for caution.

    abstract::Mechanical circulatory support (MCS) devices continue to be hampered by thrombotic adverse events (AEs), a consequence of device-imparted supraphysiologic shear stresses, leading to shear-mediated platelet activation (SMPA). In advancing MCS devices from design to clinical use, in vitro circulatory loops containing th...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.13626

    authors: Li M,Walk R,Roka-Moiia Y,Sheriff J,Bluestein D,Barth EJ,Slepian MJ

    更新日期:2020-06-01 00:00:00

  • Influence of antihypertensive drugs and steroid hormones on protein adsorption/desorption on polycarbonate.

    abstract::To develop artificial materials for prolonged use in the vascular system, the complicated process of surface-induced thrombosis needs to be better understood. It is documented that certain antihypertensive drugs can inhibit platelet adhesion to an artificial surface; on the other hand, steroid drugs increase platelet ...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:

    authors: Chandy T,Sharma CP

    更新日期:1991-06-01 00:00:00

  • Influence of reuse and of reuse sterilants on the first-use syndrome.

    abstract::Over a 4 year period, five of 98 patients at our dialysis unit developed signs and symptoms consistent with first-use syndrome (FUS). Marked improvement was noted after subjecting new dialyzers to automated processing using either formaldehyde or peracetic acid. No episodes of FUS occurred in patients being treated wi...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1987.tb02645.x

    authors: Vanholder R,Ringoir S

    更新日期:1987-04-01 00:00:00

  • A new smartphone-based method for wound area measurement.

    abstract::Proper wound healing can be assessed by monitoring the wound surface area. Its reduction by 10 or 50% should be achieved after 1 or 4 weeks, respectively, from the start of the applied therapy. There are various methods of wound area measurement, which differ in terms of the cost of the devices and their accuracy. Thi...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12169

    authors: Foltynski P,Ladyzynski P,Wojcicki JM

    更新日期:2014-04-01 00:00:00

  • Influence of different heparin concentrations on the results of in vitro investigations in plasmaseparation technology using capillary membrane filters.

    abstract::Capillary membrane filters are devices commonly used for plasmapheresis and recirculating detoxification systems, including plasmaseparation units. Besides clinical trials and case reports, research on plasmafiltration techniques is carried out using in vitro systems. Notably, such hemoperfused in vitro systems requir...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.2003.07132.x

    authors: Unger JK,Haltern C,Dohmen B,Rossaint R

    更新日期:2003-07-01 00:00:00

  • Circulatory support for patients experiencing postcardiotomy cardiogenic shock: analysis of risk factors for death and long-term follow-up.

    abstract::Over the last 5 years, 12 patients received ventricular assist devices (VADs) while in postcardiotomy cardiogenic shock. There were 7 male and 5 female patients ranging in age from 22 to 73 years (average age, 54). Eight patients underwent surgery for valve replacements, 3 for coronary artery bypass graftings, and 1 f...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1993.tb00608.x

    authors: Murakami T,Nakayama H,Irie H,Kino K,Hisamochi K,Koomoto T,Takagaki M,Sugawara E,Senoo Y,Teramoto S

    更新日期:1993-07-01 00:00:00

  • A bedside micromethod for the determination of total carbon dioxide in biological fluids.

    abstract::The quality control of bicarbonate hemodialysis requires the determination of TCO2, either in blood or in dialysate. The use of the M + D TCO2 kit, a bedside micromethod, gave the authors an economical and reliable solution to this problem. ...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1985.tb04404.x

    authors: Canaud B,Araujo A,Nguyen QV,Flavier JL,Mion C

    更新日期:1985-11-01 00:00:00

  • Ripple reduction control of the undulation pump total artificial heart.

    abstract::An undulation pump total artificial heart (UPTAH) in which the revolutions of the motor are converted to undulation motion of a disk has been developed. In an experiment, a goat using the UPTAH survived for 54 days. However, a large ripple was observed in the device's output pressure and flow waveform. In calculating ...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.2002.06798.x

    authors: Saito I,Chinzei T,Mochizuki S,Abe Y,Isoyama T,Iwasaki K,Suzuki T,Karita T,Ono T,Kouno A,Ishimaru M,Baba A,Ozeki T,Tohyama T,Kobayashi Si S,Imachi K

    更新日期:2002-01-01 00:00:00

  • Quantification of Shear-Induced Platelet Activation: High Shear Stresses for Short Exposure Time.

    abstract::Thrombosis and thromboembolism are the life-threatening clinical complications for patients supported or treated with prosthetic cardiovascular devices. The high mechanical shear stress within these devices is believed to be the major contributing factor to cause platelet activation (PA) and function alteration, leadi...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12438

    authors: Ding J,Chen Z,Niu S,Zhang J,Mondal NK,Griffith BP,Wu ZJ

    更新日期:2015-07-01 00:00:00

  • Atrial versus ventricular cannulation for a rotary ventricular assist device.

    abstract::The ventricular assist device inflow cannulation site is the primary interface between the device and the patient. Connecting these cannulae to either atria or ventricles induces major changes in flow dynamics; however, there are little data available on precise quantification of these changes. The objective of this i...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2010.01093.x

    authors: Timms D,Gregory S,Hsu PL,Thomson B,Pearcy M,McNeil K,Fraser J,Steinseifer U

    更新日期:2010-09-01 00:00:00

  • In Vitro Evaluation of an Alternative Neonatal Extracorporeal Life Support Circuit on Hemodynamic Performance and Bubble Trap.

    abstract::The objective of this study was to evaluate an alternative neonatal extracorporeal life support (ECLS) circuit with a RotaFlow centrifugal pump and Better-Bladder (BB) for hemodynamic performance and gaseous microemboli (GME) capture in a simulated neonatal ECLS system. The circuit consisted of a Maquet RotaFlow centr...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12748

    authors: Spencer SB,Wang S,Woitas K,Glass K,Kunselman AR,Ündar A

    更新日期:2017-01-01 00:00:00

  • Optimal continuous retrograde coronary perfusion flow rate with beating heart in normal temperature.

    abstract:PURPOSE:We were led to the premise that with retrograde oxygenated blood perfusion, it would be possible to perform operations on the beating heart. METHODS:The experiment consisted of two stages: procedure 1 and 2. In procedure 1, six pigs were used. Retrograde perfusion flow was controlled to be in the range of 5.0-...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2004.47279.x

    authors: Tsukioka T,Tomita S,Watanabe G,Takemura H

    更新日期:2004-03-01 00:00:00

  • Microencapsulation of crystalline insulin or islets of Langerhans: an insulin diffusion study.

    abstract::Microcapsules containing insulin crystals or islets of Langerhans were made by extruding a mixture of insulin crystals or islets and sodium alginate into a calcium chloride solution, and then coating it with poly-l-lysine. When these microcapsules were incubated at 37 degrees C, insulin could be detected readily in th...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1983.tb04187.x

    authors: Leung YF,O'Shea GM,Goosen MF,Sun AM

    更新日期:1983-05-01 00:00:00

  • Automatic Control of Veno-Venous Extracorporeal Lung Assist.

    abstract::Veno-venous extracorporeal lung assist (ECLA) can provide sufficient gas exchange even in most severe cases of acute respiratory distress syndrome. Commercially available systems are manually controlled, although an automatically controlled ECLA could allow individualized and continuous adaption to clinical requiremen...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12664

    authors: Kopp R,Bensberg R,Stollenwerk A,Arens J,Grottke O,Walter M,Rossaint R

    更新日期:2016-10-01 00:00:00

  • An In Vitro Feasibility Study of the Influence of Configurations and Leaflet Thickness on the Hydrodynamics of Deformed Transcatheter Aortic Valve.

    abstract::Clinically, the percutaneous transcatheter aortic valve (TAV) has been reported to be deformed in a noncircular configuration after its implant. The deformation is universal and various, and it leads to serious leakage and durability problems. Even in the same deformation, the leaflets made in different tissue thickne...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12833

    authors: Feng W,Yang X,Liu Y,Fan Y

    更新日期:2017-08-01 00:00:00

  • Initial experience with Impella RP in a quaternary transplant center.

    abstract::Right ventricular failure is one of the most common complications encountered after left ventricular assist device implantation and heart transplantation. It has been reported to have an incidence up to 30%. It increases morbidity and short-term mortality. Impella RP is a small pump that can provide up to 4L/min of fl...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.13610

    authors: Monteagudo-Vela M,Simon A,Panoulas V

    更新日期:2020-05-01 00:00:00

  • A Pseudo-One Compartment Model of Phosphorus Kinetics During Hemodialysis: Further Supporting Evidence.

    abstract::A pseudo-one compartment model has been proposed to describe phosphorus kinetics during hemodialysis and the immediate post-dialysis period. This model assumes that phosphorus mobilization from tissues is proportional to the difference between the pre-dialysis serum concentration (a constant) and the instantaneous ser...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12897

    authors: Leypoldt JK,Agar BU,Cheung AK,Bernardo AA

    更新日期:2017-11-01 00:00:00

  • Biomechanical properties of decellularized porcine pulmonary valve conduits.

    abstract::Tissue-engineered heart valves constructed from a xenogeneic or allogeneic decellularized matrix might overcome the disadvantages of current heart valve substitutes. One major necessity besides effective decellularization is to preserve the biomechanical properties of the valve. Native and decellularized porcine pulmo...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2007.00452.x

    authors: Seebacher G,Grasl C,Stoiber M,Rieder E,Kasimir MT,Dunkler D,Simon P,Weigel G,Schima H

    更新日期:2008-01-01 00:00:00