Dose measurements near a non-radioactive gold seed using radiographic film.

Abstract:

:The dose distribution near a non-radioactive gold seed under a 6 MV photon beam was measured using radiographic film, water equivalent bolus and solid water slabs. This type of small seed is typically used as a marker in target positional verification using a portal imager for conformal prostate treatment such as intensity modulated radiation therapy. A stack of three films was placed on top of the seed located on a soft bolus. Solid water slabs were then placed on top of the film. The films were exposed using a small 1x1 cm2 field. Then, using a similar experimental set-up and exposure, another stack of three films was placed under the seed, which was then covered by the soft bolus and solid water slabs. The cross-plane axial beam profiles at different depths, depending on the thickness of the film package, were measured. From the group of beam profiles above and below the seed, the dose distribution along a selected vertical line within the profiles was easily plotted. Compared to the dose with no seed at the isocentre and 5 cm of solid water, there was about a 21% increase in dose at 0.35 mm above the seed. On the other hand, there was about a 22% decrease in dose at the same distance below the seed. The dosimetry of the calibrated film was verified with a MOSFET detector. The change in dose due to the seed by varying the incident beam angles was also measured for this note.

journal_name

Phys Med Biol

authors

Chow JC,Grigorov GN

doi

10.1088/0031-9155/50/18/N02

subject

Has Abstract

pub_date

2005-09-21 00:00:00

pages

N227-34

issue

18

eissn

0031-9155

issn

1361-6560

pii

S0031-9155(05)98201-8

journal_volume

50

pub_type

杂志文章
  • Effect of dominant features on neural network performance in the classification of mammographic lesions.

    abstract::Two different classifiers, an artificial neural network (ANN) and a hybrid system (one step rule-based method followed by an artificial neural network) have been investigated to merge computer-extracted features in the task of differentiating between malignant and benign masses. A database consisting of 65 cases (38 m...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/44/10/315

    authors: Huo Z,Giger ML,Metz CE

    更新日期:1999-10-01 00:00:00

  • Volumetric characterization of the Aurora magnetic tracker system for image-guided transorbital endoscopic procedures.

    abstract::In some medical procedures, it is difficult or impossible to maintain a line of sight for a guidance system. For such applications, people have begun to use electromagnetic trackers. Before a localizer can be effectively used for an image-guided procedure, a characterization of the localizer is required. The purpose o...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/53/16/009

    authors: Atuegwu NC,Galloway RL

    更新日期:2008-08-21 00:00:00

  • Tetrazolium salts-Pluronic F-127 gels for 3D radiotherapy dosimetry.

    abstract::This work is a follow-up study for a recently-proposed 3D radiochromic gel dosimeter that contains a tetrazolium salt and a physical gel matrix made of poly(ethylene oxide)-block-poly(propylene oxide)-block-poly(ethylene oxide) (Pluronic F-127). Several tetrazolium salts were examined in this work, including tetrazoli...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aabbb6

    authors: Kwiatos K,Maras P,Kadlubowski S,Stempień Z,Dudek M,Kozicki M

    更新日期:2018-05-04 00:00:00

  • Respiratory motion prediction by using the adaptive neuro fuzzy inference system (ANFIS).

    abstract::The quality of radiation therapy delivered for treating cancer patients is related to set-up errors and organ motion. Due to the margins needed to ensure adequate target coverage, many breast cancer patients have been shown to develop late side effects such as pneumonitis and cardiac damage. Breathing-adapted radiatio...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/50/19/020

    authors: Kakar M,Nyström H,Aarup LR,Nøttrup TJ,Olsen DR

    更新日期:2005-10-07 00:00:00

  • A further assessment of an electromagnetic method to measure body composition.

    abstract::Two experimental investigations of the electromagnetic method to measure body composition are presented. In the first experiment, the impedance of a single-loop transmitting antenna, located at the centre of a 150 m helical receiving coil, was observed to have a sharp minimum at 2.40 MHz. The impedance of the antenna ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/40/4/011

    authors: Sutcliffe JF,Smye SW,Smith MA

    更新日期:1995-04-01 00:00:00

  • Prior image constrained scatter correction in cone-beam computed tomography image-guided radiation therapy.

    abstract::X-ray scatter is a significant problem in cone-beam computed tomography when thicker objects and larger cone angles are used, as scattered radiation can lead to reduced contrast and CT number inaccuracy. Advances have been made in x-ray computed tomography (CT) by incorporating a high quality prior image into the imag...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/56/4/009

    authors: Brunner S,Nett BE,Tolakanahalli R,Chen GH

    更新日期:2011-02-21 00:00:00

  • Improved iterative tomographic reconstruction for x-ray imaging with edge-illumination.

    abstract::Iterative tomographic reconstruction has been established as a viable alternative for data analysis in phase-sensitive x-ray imaging based on the edge-illumination principle. However, previously published approaches did not account for drifts of optical elements during a scan, which can lead to artefacts. Up to now, t...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab439d

    authors: Modregger P,Meganck J,Hagen CK,Massimi L,Olivo A,Endrizzi M

    更新日期:2019-10-10 00:00:00

  • Optical-thermal characterization of cutaneous transilluminators.

    abstract::In recent years, there has been an increase in the popularity of light-emitting diode (LED)-based, battery-powered transilluminators (BPTs) for facilitating transdermal vascular access in adults and neonates. BPTs are believed to have lower potential for inducing skin burns than prior devices based on high-power broad...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/54/22/008

    authors: Pfefer TJ,Mehrabi A,James R,Landry R,Weininger S,Chang I,Kaufman D,Miller S

    更新日期:2009-11-21 00:00:00

  • Monte Carlo-based evaluation of S-values in mouse models for positron-emitting radionuclides.

    abstract::In addition to being a powerful clinical tool, Positron emission tomography (PET) is also used in small laboratory animal research to visualize and track certain molecular processes associated with diseases such as cancer, heart disease and neurological disorders in living small animal models of disease. However, dosi...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/58/1/169

    authors: Xie T,Zaidi H

    更新日期:2013-01-07 00:00:00

  • PDE-constrained LDDMM via geodesic shooting and inexact Gauss-Newton-Krylov optimization using the incremental adjoint Jacobi equations.

    abstract::The class of non-rigid registration methods proposed in the framework of PDE-constrained large deformation diffeomorphic metric mapping is a particularly interesting family of physically meaningful diffeomorphic registration methods. Inexact Gauss-Newton-Krylov optimization has shown an excellent numerical accuracy an...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aaf598

    authors: Hernandez M

    更新日期:2019-01-07 00:00:00

  • Improved precision of syndesmophyte measurement for the evaluation of ankylosing spondylitis using CT: a phantom and patient study.

    abstract::Ankylosing spondylitis is a disease characterized by abnormal bone formation (syndesmophyte) at the margins of inter-vertebral disc spaces. Syndesmophyte growth is currently typically monitored by the visual inspection of radiographs. The limitations inherent to the modality (2D projection of a 3D object) and rater (q...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/57/14/4683

    authors: Tan S,Yao J,Yao L,Ward MM

    更新日期:2012-07-21 00:00:00

  • 4D digital tomosynthesis image reconstruction using brute force-based adaptive total variation (BF-ATV) in a prototype LINAC system.

    abstract::Respiratory-correlated cone-beam CT (CBCT) not only inhibits rapid scanning due to the slow speed of the LINAC head gantry rotation, but its implementation for routine patient imaging is impractical because of the high radiation dose delivered during the process. Digital tomosynthesis (DTS) is a potentially faster tec...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab0d50

    authors: Choi S,Lee S,Kang YN,Hsieh SS,Kim HJ

    更新日期:2019-05-02 00:00:00

  • Modeling the radiation-induced cell death in a therapeutic proton beam using thermoluminescent detectors and radiation transport simulations.

    abstract::Changes in the relative biological effectiveness (RBE) of the radiation-induced cell killing of human salivary glands (HSG) were assessed along the Bragg peak of a 60 MeV clinical proton beam by means of coupling biophysical models with the results of Monte Carlo radiation transport simulations and experimental measur...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab491f

    authors: Parisi A,Olko P,Swakoń J,Horwacik T,Jabłoński H,Malinowski L,Nowak T,Struelens L,Vanhavere F

    更新日期:2020-01-10 00:00:00

  • Implementation of pencil kernel and depth penetration algorithms for treatment planning of proton beams.

    abstract::The implementation of two algorithms for calculating dose distributions for radiation therapy treatment planning of intermediate energy proton beams is described. A pencil kernel algorithm and a depth penetration algorithm have been incorporated into a commercial three dimensional treatment planning system (Helax-TMS,...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/45/1/302

    authors: Russell KR,Isacsson U,Saxner M,Ahnesjö A,Montelius A,Grusell E,Dahlgren CV,Lorin S,Glimelius B

    更新日期:2000-01-01 00:00:00

  • A new method of body habitus correction for total body potassium measurements.

    abstract::This paper describes an accurate and time-efficient method for the determination of total body potassium via a combination of measurements in the Birmingham whole body counter and the use of the Monte Carlo n-particle (MCNP) simulation code. In developing this method, MCNP has also been used to derive values for some ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/51/17/009

    authors: O'Hehir S,Green S,Beddoe AH

    更新日期:2006-09-07 00:00:00

  • Measurements of angular distributions of Rayleigh and Mie scattering events in biological models.

    abstract::Remission spectroscopy in living tissues can only be performed because scattering processes induce a pronounced amount of backscattered light. New types of scattering chambers were constructed in order to investigate the angular distribution of light intensities caused by scattering events typical for living tissues. ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/34/12/012

    authors: Frank KH,Kessler M,Appelbaum K,Albrecht HP,Mauch ED

    更新日期:1989-12-01 00:00:00

  • Comparison of two-compartment exchange and continuum models of dMRI in skeletal muscle.

    abstract::Clinical diffusion MRI (dMRI) is sensitive to micrometer scale spin displacements, but the image resolution is  ∼mm, so the biophysical interpretation of the signal relies on establishing appropriate subvoxel tissue models. A class of two-compartment exchange models originally proposed by Kärger have been used success...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab2aa6

    authors: Naughton NM,Georgiadis JG

    更新日期:2019-08-01 00:00:00

  • Analytic treatment of resolution precision in electronically collimated SPECT imaging involving multiple-interaction gamma rays.

    abstract::The widely applied single-interaction analytic expression characterizing the energy resolution component of the angular resolution precision for an electronically collimated point source is extended to include multiple-interaction Compton scatter sequences as well as sequences terminated by photoelectric absorption. T...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/43/10/021

    authors: Durkee JW Jr,Antich PP,Tsyganov EN,Constantinescu A,Kulkarni PV,Smith BJ,Arbique GM,Lewis MA,Nguyen T,Raheja A,Thambi G,Parkey RW

    更新日期:1998-10-01 00:00:00

  • Laser speckle imaging of blood flow in microcirculation.

    abstract::Monitoring the spatio-temporal characteristics of microcirculation is crucial for studying the functional activities of biotissue and the mechanism of disease. However, conventional methods used to measure blood flow suffer from limited spatial resolution or the injection of exogenous substances or the need of scannin...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/49/7/020

    authors: Cheng H,Luo Q,Liu Q,Lu Q,Gong H,Zeng S

    更新日期:2004-04-07 00:00:00

  • A multi-scale framework with unsupervised joint training of convolutional neural networks for pulmonary deformable image registration.

    abstract::To achieve accurate and fast deformable image registration (DIR) for pulmonary CT, we proposed a Multi-scale DIR framework with unsupervised Joint training of Convolutional Neural Network (MJ-CNN). MJ-CNN contains three models at multi-scale levels for a coarse-to-fine DIR to avoid being trapped in a local minimum. It...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab5da0

    authors: Jiang Z,Yin FF,Ge Y,Ren L

    更新日期:2020-01-13 00:00:00

  • Microwave thermal ablation using CT-scanner for predicting the variation of ablated region over time: advantages and limitations.

    abstract::This study aims at investigating in real-time the structural and dynamical changes occurring in an ex vivo tissue during a microwave thermal ablation (MTA) procedure. The experimental set-up was based on ex vivo liver tissue inserted in a dedicated box, in which 3 fibre-optic (FO) temperature probes were introduced to...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab1a67

    authors: Strigari L,Minosse S,D'Alessio D,Farina L,Cavagnaro M,Cassano B,Pinto R,Vallati G,Lopresto V

    更新日期:2019-05-31 00:00:00

  • Computational errors of the induced electric field in voxelized and tetrahedral anatomical head models exposed to spatially uniform and localized magnetic fields.

    abstract::At low and intermediate frequencies, the strength of the induced electric field is used as dosimetric quantity for human protection in the International Commission on Non-Ionizing Radiation Protection (ICNIRP) guidelines. To compute the induced electric field, numerical methods based on anatomically realistic voxel mo...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab5dfb

    authors: Soldati M,Laakso I

    更新日期:2020-01-10 00:00:00

  • Evaluation of the basic properties of the BANGkit gel dosimeter.

    abstract::We evaluated the basic properties of a commercially available BANGkit gel dosimeter, which is a normoxic type of BANG gel. This gel-kit has the same composition as the BANG 3 gel, but is fully oxygenated. To exclude oxygen, oxygen scavenging ascorbic acid and copper sulfate as a catalyst are used. The properties that ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/52/8/017

    authors: Murakami Y,Nakashima T,Watanabe Y,Akimitsu T,Matsuura K,Kenjo M,Kaneyasu Y,Wadasaki K,Hirokawa Y,Ito K

    更新日期:2007-04-21 00:00:00

  • Experimental evaluation of fiber-interspaced antiscatter grids for large patient imaging with digital x-ray systems.

    abstract::Radiographic imaging of large patients is compromised by x-ray scatter. Optimization of digital x-ray imaging systems used for projection radiography requires the use of the best possible antiscatter grid. The performance of antiscatter grids used in conjunction with digital x-ray imaging systems can be characterized ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/52/16/010

    authors: Fetterly KA,Schueler BA

    更新日期:2007-08-21 00:00:00

  • Estimation of torque on mechanical heart valves due to magnetic resonance imaging including an estimation of the significance of the Lenz effect using a computational model.

    abstract::Mitral and aortic valve replacement is a procedure which is common in cardiac surgery. Some of these replacement valves are mechanical and contain moving metal parts. Should the patient in whom such a valve has been implanted be involved in magnetic resonance imaging, there is a possible dangerous interaction between ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/45/12/320

    authors: Robertson NM,Diaz-Gomez M,Condon B

    更新日期:2000-12-01 00:00:00

  • Optimization of tumour control probability for heterogeneous tumours in fractionated radiotherapy treatment protocols.

    abstract::We find the dose distribution that maximizes the tumour control probability (TCP) for a fixed mean tumour dose per fraction. We consider a heterogeneous tumour volume having a radiation response characterized by the linear quadratic model with heterogeneous radiosensitivity and repopulation rate that may vary in time....

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/49/3/005

    authors: Levin-Plotnik D,Hamilton RJ

    更新日期:2004-02-07 00:00:00

  • The noise navigator: a surrogate for respiratory-correlated 4D-MRI for motion characterization in radiotherapy.

    abstract::Respiratory-correlated 4D-MRI can characterize respiratory-induced motion of tumors and organs-at-risk for radiotherapy treatment planning and is a necessity for image guidance of moving tumors treated on an MRI-linac. Essential for 4D-MRI generation is a robust respiratory surrogate signal. We investigated the feasib...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab5c62

    authors: Navest RJM,Mandija S,Bruijnen T,Stemkens B,Tijssen RHN,Andreychenko A,Lagendijk JJW,van den Berg CAT

    更新日期:2020-01-13 00:00:00

  • Factors affecting in vivo measurement precision and accuracy of 109Cd K x-ray fluorescence measurements.

    abstract::109Cd K x-ray fluorescence (XRF) measurement systems from two research centres were used to measure tibia lead content in a population (n = 530) of young adults. The group mean bone lead contents (+/-SEM) determined by McMaster University (n = 214) and the University of Maryland (n = 316) were 2.80 +/- 0.51 and 2.33 +...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/44/9/313

    authors: McNeill FE,Stokes L,Chettle DR,Kaye WE

    更新日期:1999-09-01 00:00:00

  • Applying a variable relative biological effectiveness (RBE) might affect the analysis of clinical trials comparing photon and proton therapy for prostate cancer.

    abstract::The purpose of this study was the evaluation of the impact of a variable relative biological effectiveness (RBE) compared to a constant RBE value of 1.1 in proton therapy prostate trials due to uncertainties in α/β ratio. Twenty patients receiving passive scattered proton therapy (PSPT) and fifteen patients receiving ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab2144

    authors: Marteinsdottir M,Paganetti H

    更新日期:2019-06-05 00:00:00

  • Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution.

    abstract::Developments in positron emission tomography (PET) technology have resulted in systems with finer detector elements designed to further improve spatial resolution. However, there is a limit to what extent reducing detector element size will improve spatial resolution in PET. The spatial resolution of PET imaging is li...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/44/3/019

    authors: Levin CS,Hoffman EJ

    更新日期:1999-03-01 00:00:00