Improved dose-calculation accuracy in proton treatment planning using a simplified Monte Carlo method verified with three-dimensional measurements in an anthropomorphic phantom.

Abstract:

:Treatment planning for proton tumor therapy requires a fast and accurate dose-calculation method. We have implemented a simplified Monte Carlo (SMC) method in the treatment planning system of the National Cancer Center Hospital East for the double-scattering beam delivery scheme. The SMC method takes into account the scattering effect in materials more accurately than the pencil beam algorithm by tracking individual proton paths. We confirmed that the SMC method reproduced measured dose distributions in a heterogeneous slab phantom better than the pencil beam method. When applied to a complex anthropomorphic phantom, the SMC method reproduced the measured dose distribution well, satisfying an accuracy tolerance of 3 mm and 3% in the gamma index analysis. The SMC method required approximately 30 min to complete the calculation over a target volume of 500 cc, much less than the time required for the full Monte Carlo calculation. The SMC method is a candidate for a practical calculation technique with sufficient accuracy for clinical application.

journal_name

Phys Med Biol

authors

Hotta K,Kohno R,Takada Y,Hara Y,Tansho R,Himukai T,Kameoka S,Matsuura T,Nishio T,Ogino T

doi

10.1088/0031-9155/55/12/018

subject

Has Abstract

pub_date

2010-06-21 00:00:00

pages

3545-56

issue

12

eissn

0031-9155

issn

1361-6560

pii

S0031-9155(10)46368-X

journal_volume

55

pub_type

杂志文章
  • A novel technique for measuring the low-dose envelope of pencil-beam scanning spot profiles.

    abstract::To investigate the profile measurement capabilities of an IBA-Dosimetry scintillation detector and to assess its feasibility for determining the low-intensity tails of pencil-beam scanning spots, the responses of the scintillation detector and Gafchromic EBT2 film to a 115 MeV proton spot were measured in-air at the i...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/58/12/N171

    authors: Lin L,Ainsley CG,Mertens T,De Wilde O,Talla PT,McDonough JE

    更新日期:2013-06-21 00:00:00

  • On the impact of longitudinal breathing motion randomness for tomotherapy delivery.

    abstract::The purpose of this study is to explain the unplanned longitudinal dose modulations that appear in helical tomotherapy (HT) dose distributions in the presence of irregular patient breathing. This explanation is developed by the use of longitudinal (1D) simulations of mock and surrogate data and tested with a fully 4D ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/53/18/001

    authors: Kissick MW,Flynn RT,Westerly DC,Hoban PW,Mo X,Soisson ET,McCall KC,Mackie TR,Jeraj R

    更新日期:2008-09-21 00:00:00

  • Dynamic (99m)Tc-MAG3 renography: images for quality control obtained by combining pharmacokinetic modelling, an anthropomorphic computer phantom and Monte Carlo simulated scintillation camera imaging.

    abstract::In dynamic renal scintigraphy, the main interest is the radiopharmaceutical redistribution as a function of time. Quality control (QC) of renal procedures often relies on phantom experiments to compare image-based results with the measurement setup. A phantom with a realistic anatomy and time-varying activity distribu...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/58/10/3145

    authors: Brolin G,Gleisner KS,Ljungberg M

    更新日期:2013-05-21 00:00:00

  • Three-dimensional cellular-level imaging using full-field optical coherence tomography.

    abstract::An ultrahigh-resolution full-field optical coherence tomography (OCT) system has been developed for cellular-level imaging of biological media. The system is based on a Linnik interference microscope illuminated with a tungsten halogen lamp, associated with a high-resolution CCD camera. En face tomographic images are ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/49/7/010

    authors: Dubois A,Moneron G,Grieve K,Boccara AC

    更新日期:2004-04-07 00:00:00

  • Assessment of the effects of pixel loss on image quality in direct digital radiography.

    abstract::Modern digital radiographic 'flat panel' detectors can exhibit a progressive form of image degradation arising from non-functioning pixels. The effect of these 'dead pixels' on the quantitive image quality measures of modulation transfer function (MTF), noise power spectrum (NPS) and detective quantum efficiency (DQE)...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/49/6/008

    authors: Padgett R,Kotre CJ

    更新日期:2004-03-21 00:00:00

  • Relative determination of W-values for alpha particles in tissue-equivalent and other gases.

    abstract::W (the average energy to form an ion pair) for 5.4 MeV 241Am alpha particles in a Rossi-type tissue-equivalent (TE) gas, argon and methane was determined to an accuracy better than 0.2% using a new automated data handling system. A vibrating reed electrometer and current digitiser were used to measure the current prod...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/24/2/005

    authors: Krieger GL,Baum JW,Varma MN,Beach JL

    更新日期:1979-03-01 00:00:00

  • Reducing scan angle using adaptive prior knowledge for a limited-angle intrafraction verification (LIVE) system for conformal arc radiotherapy.

    abstract::The purpose of this study is to develop an adaptive prior knowledge guided image estimation technique to reduce the scan angle needed in the limited-angle intrafraction verification (LIVE) system for 4D-CBCT reconstruction. The LIVE system has been previously developed to reconstruct 4D volumetric images on-the-fly du...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa6913

    authors: Zhang Y,Yin FF,Zhang Y,Ren L

    更新日期:2017-05-07 00:00:00

  • Feasibility of polymer gel-based measurements of radiation isocenter accuracy in magnetic fields.

    abstract::For conventional irradiation devices, the radiation isocenter accuracy is determined by star shot measurements on films. In magnetic resonance (MR)-guided radiotherapy devices, the results of this test may be altered by the magnetic field and the need to align the radiation and imaging isocenter may require a modifica...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aac228

    authors: Dorsch S,Mann P,Lang C,Haering P,Runz A,Karger CP

    更新日期:2018-05-29 00:00:00

  • A neural network regression model for relative dose computation.

    abstract::This paper describes a neural network (NN) regression model for relative dose computation. The input signal of the NN model includes depths and field sizes and the output is the relative dose, for example, percentage depth dose (%DD) or tissue-air ratio (TAR) in this paper. After a functional link has been created, th...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/45/4/307

    authors: Wu X,Zhu Y

    更新日期:2000-04-01 00:00:00

  • Automatic bi-objective parameter tuning for inverse planning of high-dose-rate prostate brachytherapy.

    abstract::We present an automatic bi-objective parameter-tuning approach for inverse planning methods for high-dose-rate prostate brachytherapy, which aims to overcome the difficult and time-consuming manual parameter tuning that is currently required to obtain patient-specific high-quality treatment plans. We modelled treatmen...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab7362

    authors: Maree SC,Bosman PAN,van Wieringen N,Niatsetski Y,Pieters BR,Bel A,Alderliesten T

    更新日期:2020-04-02 00:00:00

  • A fast and sensitive TLD method for measurement of energy and homogeneity of electron beams using transmitted radiation through lead.

    abstract::A simple and fast, but sensitive TLD method for the measurement of energy and homogeneity of therapeutically used electron beams has been developed and tested. This method is based on the fact that when small thicknesses of high-Z absorbers such as lead are interposed in the high-energy electron beams, the transmitted...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/39/9/005

    authors: Pradhan AS,Quast U,Sharma PK

    更新日期:1994-09-01 00:00:00

  • Dead time correction and counting statistics for positron tomography.

    abstract::A correction for loss of events due to dead time in dynamic positron emission tomography (PET) is presented. The model employs a paralysing dead time to describe the behaviour of a tomograph over the range of event rates normally encountered in patient studies (up to 200 000 events/s per detector layer). The Donner 28...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/30/5/002

    authors: Mazoyer BM,Roos MS,Huesman RH

    更新日期:1985-05-01 00:00:00

  • Calibration and analysis of a multimodal micro-CT and structured light imaging system for the evaluation of excised breast tissue.

    abstract::A multimodal micro-computed tomography (CT) and multi-spectral structured light imaging (SLI) system is introduced and systematically analyzed to test its feasibility to aid in margin delineation during breast conserving surgery (BCS). Phantom analysis of the micro-CT yielded a signal-to-noise ratio of 34, a contrast ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa94b6

    authors: McClatchy DM,Rizzo EJ,Meganck J,Kempner J,Vicory J,Wells WA,Paulsen KD,Pogue BW

    更新日期:2017-11-10 00:00:00

  • Collapsed cone convolution and analytical anisotropic algorithm dose calculations compared to VMC++ Monte Carlo simulations in clinical cases.

    abstract::The purpose of this work was to study and quantify the differences in dose distributions computed with some of the newest dose calculation algorithms available in commercial planning systems. The study was done for clinical cases originally calculated with pencil beam convolution (PBC) where large density inhomogeneit...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/52/13/002

    authors: Hasenbalg F,Neuenschwander H,Mini R,Born EJ

    更新日期:2007-07-07 00:00:00

  • Dose rate calculations around 192Ir brachytherapy sources using a Sievert integration model.

    abstract::The classical Sievert integral method is a valuable tool for dose rate calculations around brachytherapy sources, combining simplicity with reasonable computational times. However, its accuracy in predicting dose rate anisotropy around 192Ir brachytherapy sources has been repeatedly put into question. In this work, we...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/45/2/309

    authors: Karaiskos P,Angelopoulos A,Baras P,Rozaki-Mavrouli H,Sandilos P,Vlachos L,Sakelliou L

    更新日期:2000-02-01 00:00:00

  • Application of the adaptive wavelet transform for analysis of blood flow oscillations in the human skin.

    abstract::An original method for the analysis of oscillations of cutaneous blood flow has been developed, which makes use of laser Doppler flowmetry (LDF) data and is based on the continuous wavelet transform and adaptive wavelet theory. The potential of the method has been demonstrated in experiments with the response of micro...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/53/21/005

    authors: Tankanag A,Chemeris N

    更新日期:2008-11-07 00:00:00

  • Characterization of cirrhosis and hepatocellular carcinoma using low-angle x-ray scattering signatures of serum.

    abstract::The diagnosis of hepatocellular carcinoma (HCC) usually occurs at late stages in the disease when there are few effective treatment options. The measurement of the concentration of tumour markers in the serum of patients is a complementary tool frequently used for the interpretation of diagnostic imaging results. It i...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/48/17/401

    authors: Elshemey WM,Desouky OS,Mohammed MS,Elsayed AA,el-Houseini ME

    更新日期:2003-09-07 00:00:00

  • Correction factors for parallel-plate chambers used in plastic phantoms in electron dosimetry.

    abstract::In electron beam dosimetry using parallel-plate chambers solid phantoms are sometimes necessary. To obtain the dose to water from the ionization obtained in the solid phantom, fluence correction factors and perturbation factors have to be applied. In this study fluence factors in a perturbation free geometry have been...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/42/11/007

    authors: Nilsson B,Montelius A,Andreo P,Johansson J

    更新日期:1997-11-01 00:00:00

  • Numerical study of oxygen transport in a carotid bifurcation.

    abstract::This study investigates the oxygen mass transport in the region around the human carotid bifurcation, particularly addressing the effects of bifurcation geometry and pulsatile blood flow on the oxygen transport between the blood flow and artery wall tissue, coupled with the metabolic oxygen consumption and oxygen diff...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/55/14/004

    authors: Tada S

    更新日期:2010-07-21 00:00:00

  • Automated stroke lesion segmentation in non-contrast CT scans using dense multi-path contextual generative adversarial network.

    abstract::Stroke lesion volume is a key radiologic measurement in assessing prognosis of acute ischemic stroke (AIS) patients. The aim of this paper is to develop an automated segmentation method for accurately segmenting follow-up ischemic and hemorrhagic lesion from multislice non-contrast CT (NCCT) volumes of AIS patients. T...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aba166

    authors: Kuang H,Menon BK,Qiu W

    更新日期:2020-11-05 00:00:00

  • On the surviving fraction in irradiated multicellular tumour spheroids: calculation of overall radiosensitivity parameters, influence of hypoxia and volume effects.

    abstract::We model the heterogeneous response to radiation of multicellular tumour spheroids assuming position- and volume-dependent radiosensitivity. We propose a method to calculate the overall radiosensitivity parameters to obtain the surviving fraction of tumours. A mathematical model of a spherical tumour with a hypoxic co...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/50/8/005

    authors: Horas JA,Olguin OR,Rizzotto MG

    更新日期:2005-04-21 00:00:00

  • Photothermal effects of laser tissue soldering.

    abstract::Low-strength anastomoses and thermal damage of tissue are major concerns in laser tissue welding techniques where laser energy is used to induce thermal changes in the molecular structure of the tissues being joined, hence allowing them to bond together. Laser tissue soldering, on the other hand, is a bonding techniqu...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/44/4/013

    authors: McNally KM,Sorg BS,Welch AJ,Dawes JM,Owen ER

    更新日期:1999-04-01 00:00:00

  • Determination of some parameters for pion radiobiology studies.

    abstract::An experimental investigation of the central axis depth-dose and stopping rate distribution of the SIN biomedical pion beam is reported. The pion stopping rate in a thin disc of tissue-equivalent plastic was determined using a counter telescope. The dose rate at the position of this dic 'target' was measured using a s...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/22/3/007

    authors: Nordell B,Baarli J,Sullivan AH,Zielczynski M

    更新日期:1977-05-01 00:00:00

  • Imputing radiobiological parameters of the linear-quadratic dose-response model from a radiotherapy fractionation plan.

    abstract::The objective in cancer radiotherapy is to maximize tumor-kill while limiting toxic effects of radiation dose on nearby organs-at-risk (OAR). Given a fixed number of treatment sessions, planners thus face the problem of finding a dosing sequence that achieves this goal. This is called the fractionation problem, and ha...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/abb935

    authors: Ghate A

    更新日期:2020-09-16 00:00:00

  • Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution.

    abstract::Developments in positron emission tomography (PET) technology have resulted in systems with finer detector elements designed to further improve spatial resolution. However, there is a limit to what extent reducing detector element size will improve spatial resolution in PET. The spatial resolution of PET imaging is li...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/44/3/019

    authors: Levin CS,Hoffman EJ

    更新日期:1999-03-01 00:00:00

  • Dose measurements near a non-radioactive gold seed using radiographic film.

    abstract::The dose distribution near a non-radioactive gold seed under a 6 MV photon beam was measured using radiographic film, water equivalent bolus and solid water slabs. This type of small seed is typically used as a marker in target positional verification using a portal imager for conformal prostate treatment such as inte...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/50/18/N02

    authors: Chow JC,Grigorov GN

    更新日期:2005-09-21 00:00:00

  • 4D cone beam CT-based dose assessment for SBRT lung cancer treatment.

    abstract::The purpose of this research is to develop a 4DCBCT-based dose assessment method for calculating actual delivered dose for patients with significant respiratory motion or anatomical changes during the course of SBRT. To address the limitation of 4DCT-based dose assessment, we propose to calculate the delivered dose us...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/61/2/554

    authors: Cai W,Dhou S,Cifter F,Myronakis M,Hurwitz MH,Williams CL,Berbeco RI,Seco J,Lewis JH

    更新日期:2016-01-21 00:00:00

  • Comparative study of a low-Z cone-beam computed tomography system.

    abstract::Computed tomography images have been acquired using an experimental (low atomic number (Z) insert) megavoltage cone-beam imaging system. These images have been compared with standard megavoltage and kilovoltage imaging systems. The experimental system requires a simple modification to the 4 MeV electron beam from an E...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/56/14/014

    authors: Roberts DA,Hansen VN,Thompson MG,Poludniowski G,Niven A,Seco J,Evans PM

    更新日期:2011-07-21 00:00:00

  • An analysis of breast motion using high-frequency, dense surface points captured by an optical sensor during radiotherapy treatment delivery.

    abstract::Patient motion is an important factor affecting the quality of external beam radiotherapy in breast patients. We analyse the motion of a dense set of surface points on breast patients throughout their treatment schedule to assess the magnitude and stability of motion, in particular, with respect to breast volume. We u...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/54/21/005

    authors: Price GJ,Sharrock PJ,Marchant TE,Parkhurst JM,Burton D,Jain P,Price P,Moore CJ

    更新日期:2009-11-07 00:00:00

  • Rinsing paired-agent model (RPAM) to quantify cell-surface receptor concentrations in topical staining applications of thick tissues.

    abstract::Conventional molecular assessment of tissue through histology, if adapted to fresh thicker samples, has the potential to enhance cancer detection in surgical margins and monitoring of 3D cell culture molecular environments. However, in thicker samples, substantial background staining is common despite repeated rinsing...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa6cf1

    authors: Xu X,Wang Y,Xiang J,Liu JTC,Tichauer KM

    更新日期:2017-06-21 00:00:00