Dose optimization in pediatric cardiac x-ray imaging.

Abstract:

PURPOSE:The aim of this research was to explore x-ray beam parameters with intent to optimize pediatric x-ray settings in the cardiac catheterization laboratory. This study examined the effects of peak x-ray tube voltage (kVp) and of copper (Cu) x-ray beam filtration independently on the image quality to dose balance for pediatric patient sizes. The impact of antiscatter grid removal on the image quality to dose balance was also investigated. METHODS:Image sequences of polymethyl methacrylate phantoms approximating chest sizes typical of pediatric patients were captured using a modern flat-panel receptor based x-ray imaging system. Tin was used to simulate iodine-based contrast medium used in clinical procedures. Measurements of tin detail contrast and flat field image noise provided the contrast to noise ratio. Entrance surface dose (ESD) and effective dose (E) measurements were obtained to calculate the figure of merit (FOM), CNR2/dose, which evaluated the dose efficiency of the x-ray parameters investigated. The kVp, tube current (mA), and pulse duration were set manually by overriding the system's automatic dose control mechanisms. Images were captured with 0, 0.1, 0.25, 0.4, and 0.9 mm added Cu filtration, for 50, 55, 60, 65, and 70 kVp with the antiscatter grid in place, and then with it removed. RESULTS:For a given phantom thickness, as the Cu filter thickness was increased, lower kVp was favored. Examining kVp alone, lower values were generally favored, more so for thinner phantoms. Considering ESD, the 8.5 cm phantom had the highest FOM at 50 kVp using 0.4 mm of Cu filtration. The 12 cm phantom had the highest FOM at 55 kVp using 0.9 mm Cu, and the 16 cm phantom had highest FOM at 55 kVp using 0.4 mm Cu. With regard to E, the 8.5 and 12 cm phantoms had the highest FOM at 50 kVp using 0.4 mm of Cu filtration, and the 16 cm phantom had the highest FOM at 50 kVp using 0.25 mm Cu. Antiscatter grid removal improved the FOM for a given set of x-ray conditions. Under aforesaid optimal settings, the 8.5 cm phantom FOM improved by 24% and 33% for ESD and E, respectively. Corresponding improvements were 26% and 24% for the 12 cm phantom and 6% and 15% for the 16 cm phantom. CONCLUSIONS:For pediatric patients, using 0.25-0.9 mm Cu filtration in the x-ray beam while maintaining 50-55 kVp, depending on patient size, provided optimal x-ray image quality to dose ratios. These settings, adjusted for x-ray tube loading limits and clinically acceptable image quality, should provide a useful strategy for optimizing iodine contrast agent based cardiac x-ray imaging. Removing the antiscatter grid improved the FOM for the 8.5 and 12 cm phantoms, therefore grid removal is recommended for younger children. Improvement for the 16 cm phantom declined into the estimated margin of error for the FOM; the need for grid removal for older children would depend on practical feasibility in the clinical environment.

journal_name

Med Phys

journal_title

Medical physics

authors

Gislason AJ,Davies AG,Cowen AR

doi

10.1118/1.3488911

subject

Has Abstract

pub_date

2010-10-01 00:00:00

pages

5258-69

issue

10

eissn

0094-2405

issn

2473-4209

journal_volume

37

pub_type

杂志文章
  • A restricted angular scattering model for electron penetration in dense media.

    abstract::A restricted angular scattering model for electron penetration in dense media is presented. In the model, the Fermi-Eyges transport equation is modified through the addition of an extra term which may be interpreted as representing an apparent force opposing the scattering of electrons into wider angles. The introduct...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596747

    authors: McLellan J,Sandison GA,Papiez L,Huda W

    更新日期:1991-01-01 00:00:00

  • Computerized detection of masses in digital mammograms: automated alignment of breast images and its effect on bilateral-subtraction technique.

    abstract::An automated technique for the alignment of right and left breast images has been developed for use in the computerized analysis of bilateral breast images. In this technique, the breast region is first identified in each digital mammogram by use of histogram analysis and morphological filtering operations. The anteri...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597307

    authors: Yin FF,Giger ML,Doi K,Vyborny CJ,Schmidt RA

    更新日期:1994-03-01 00:00:00

  • Correcting magnetic resonance k-space data for in-plane motion using an optical position tracking system.

    abstract:PURPOSE:Motion is a major confound of image quality in MRI. A method of retrospectively correcting the effects of rotations and translations on the acquired k-space data is presented. METHODS:In two phantom experiments of well-controlled translation and rotation, two MRI-compatible infrared cameras recorded motion dat...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3254189

    authors: Marxen M,Marmurek J,Baker N,Graham SJ

    更新日期:2009-12-01 00:00:00

  • Efficient generation of a magnetic field-free line.

    abstract:PURPOSE:Signal encoding in magnetic particle imaging (MPI) is achieved by moving a field-free point (FFP) through the region of interest. One way to increase the sensitivity of the method is to scan the region of interest with a field-free line (FFL) instead of the FFP. Recently, the first feasible FFL coil setup was i...

    journal_title:Medical physics

    pub_type: 信件

    doi:10.1118/1.3447726

    authors: Knopp T,Erbe M,Biederer S,Sattel TF,Buzug TM

    更新日期:2010-07-01 00:00:00

  • First study of on-treatment volumetric imaging during respiratory gated VMAT.

    abstract:PURPOSE:To obtain on-treatment volumetric patient anatomy during respiratory gated volumetric modulated arc therapy (VMAT). METHODS:On-board imaging device integrated with Linacs offers a viable tool for obtaining patient anatomy during radiation treatment delivery. In this study, the authors acquired beam-level kV im...

    journal_title:Medical physics

    pub_type: 信件

    doi:10.1118/1.4794925

    authors: Choi K,Xing L,Koong A,Li R

    更新日期:2013-04-01 00:00:00

  • Evaluation of dual energy quantitative CT for determining the spatial distributions of red marrow and bone for dosimetry in internal emitter radiation therapy.

    abstract:PURPOSE:To evaluate a three-equation three-unknown dual-energy quantitative CT (DEQCT) technique for determining region specific variations in bone spongiosa composition for improved red marrow dose estimation in radionuclide therapy. METHODS:The DEQCT method was applied to 80/140 kVp images of patient-simulating lumb...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4870378

    authors: Goodsitt MM,Shenoy A,Shen J,Howard D,Schipper MJ,Wilderman S,Christodoulou E,Chun SY,Dewaraja YK

    更新日期:2014-05-01 00:00:00

  • Geometric studies on variable radius spiral cone-beam scanning.

    abstract::The goal is to perform geometric studies on cone-beam CT scanning along a three-dimensional (3D) spiral of variable radius. First, the background for variable radius spiral cone-beam scanning is given in the context of electron-beam CT/micro-CT. Then, necessary and sufficient conditions are proved for existence and un...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1751251

    authors: Yea Y,Zhu J,Wang G

    更新日期:2004-06-01 00:00:00

  • Factors influencing the performance of patient specific quality assurance for pencil beam scanning IMPT fields.

    abstract:PURPOSE:A detailed analysis of 2728 intensity modulated proton therapy (IMPT) fields that were clinically delivered to patients between 2007 and 2013 at Paul Scherrer Institute (PSI) was performed. The aim of this study was to analyze the results of patient specific dosimetric verifications and to assess possible corre...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4964449

    authors: Trnková P,Bolsi A,Albertini F,Weber DC,Lomax AJ

    更新日期:2016-11-01 00:00:00

  • SU-E-T-144: Daily and Monthly Quality Assurance with TomoDose.

    abstract:PURPOSE:Examine the use of TomoDose for Tomotherapy quality assurance. METHODS:Baseline TomoDose profiles are acquired annually in accordance with TG-148. During monthly quality assurance, TomoDose is irradiated with a static beam and referenced to baseline. Differences between superimposed profiles are evaluated mont...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735202

    authors: Terrell J

    更新日期:2012-06-01 00:00:00

  • An analysis of the mechanical parameters used for finite element compression of a high-resolution 3D breast phantom.

    abstract:PURPOSE:The authors previously introduced a methodology to generate a realistic three-dimensional (3D), high-resolution, computer-simulated breast phantom based on empirical data. One of the key components of such a phantom is that it provides a means to produce a realistic simulation of clinical breast compression. In...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3637500

    authors: Hsu CM,Palmeri ML,Segars WP,Veress AI,Dobbins JT 3rd

    更新日期:2011-10-01 00:00:00

  • Low dose fraction behavior of high sensitivity radiochromic film.

    abstract::A high sensitivity (HS) model of radiochromic film is receiving increasing use. The film's linear sensitometric response in the range of 0.5-40 Gy would make this film an ideal candidate for complex dosimetry applications that require tissue equivalence. This study investigates the potential use for clinical dosimetry...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1883565

    authors: Hirata EY,Cunningham C,Micka JA,Keller H,Kissick MW,DeWerd LA

    更新日期:2005-04-01 00:00:00

  • Tomographic-based 3D scintillation dosimetry using a three-view plenoptic imaging system.

    abstract:PURPOSE:To demonstrate the feasibility of a three-plenoptic camera projection, scintillation-based dosimetry system for measuring three-dimensional (3D) dose distributions of static photon radiation fields. METHODS:Static x-ray photon beams were delivered to a cubic plastic scintillator volume embedded within acrylic ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14213

    authors: Rilling M,Allain G,Thibault S,Archambault L

    更新日期:2020-08-01 00:00:00

  • A method for reducing motion induced errors in T2-weighted magnetic resonance imaging.

    abstract::It is known that motion of an object imaged by magnetic resonance imaging leads to errors and artifacts in the image. This effect is especially important when a pulse sequence with long echo time (TE) and long TR is used to provide T2-weighted images. We have modified the usual spin warp imaging pulse sequence in a wa...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596026

    authors: Joseph PM,Shetty A,Bonaroti EA

    更新日期:1987-07-01 00:00:00

  • Low-dose cone-beam CT via raw counts domain low-signal correction schemes: Performance assessment and task-based parameter optimization (Part II. Task-based parameter optimization).

    abstract:PURPOSE:Different low-signal correction (LSC) methods have been shown to efficiently reduce noise streaks and noise level in CT to provide acceptable images at low-radiation dose levels. These methods usually result in CT images with highly shift-variant and anisotropic spatial resolution and noise, which makes the par...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12855

    authors: Gomez-Cardona D,Hayes JW,Zhang R,Li K,Cruz-Bastida JP,Chen GH

    更新日期:2018-05-01 00:00:00

  • Sci-Sat AM(1): Imaging-04: Respiratory errors in cardiac PET/CT with manual alignment of the CT image.

    abstract::Respiratory motion can produce misregistration errors between CT and PET images in cardiac PET/CT imaging. The objective of this study was to determine if manual registration of a single-phase end-expiration CT scan to the PET image would eliminate respiratory-induced artifacts. Listmode data from 71 cardiac PET patie...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2965988

    authors: Wells RG

    更新日期:2008-07-01 00:00:00

  • Vision 20/20: Mammographic breast density and its clinical applications.

    abstract::Breast density is a strong predictor of the failure of mammography screening to detect breast cancer and is a strong predictor of the risk of developing breast cancer. The many imaging options that are now available for imaging dense breasts show great promise, but there is still the question of determining which wome...

    journal_title:Medical physics

    pub_type: 杂志文章,评审

    doi:10.1118/1.4935141

    authors: Ng KH,Lau S

    更新日期:2015-12-01 00:00:00

  • Kymogram detection and kymogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart.

    abstract::Subsecond single-slice, multi-slice or cone-beam spiral computed tomography (SSCT, MSCT, CBCT) offer great potential for improving heart imaging. Together with the newly developed phase-correlated cardiac reconstruction algorithms 180 degrees MCD and 180 degrees MCI [Med. Phys. 27, 1881-1902 (2000)] or related algorit...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1487861

    authors: Kachelriess M,Sennst DA,Maxlmoser W,Kalender WA

    更新日期:2002-07-01 00:00:00

  • A Monte Carlo investigation of low-Z target image quality generated in a linear accelerator using Varian's VirtuaLinac.

    abstract:PURPOSE:The focus of this work was the demonstration and validation of VirtuaLinac with clinical photon beams and to investigate the implementation of low-Z targets in a TrueBeam linear accelerator (Linac) using Monte Carlo modeling. METHODS:VirtuaLinac, a cloud based web application utilizing Geant4 Monte Carlo code,...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4861818

    authors: Parsons D,Robar JL,Sawkey D

    更新日期:2014-02-01 00:00:00

  • Conceptual design of beryllium targets for the generation of neutron beams for radiation therapy by the (p,n) reaction.

    abstract::A conceptual design is presented showing that, by judiciously choosing the beryllium target thickness and the backstop material, improvements can be achieved in dose-rate, skin sparing, and penetration of neutron beams generated by the same proton accelerator. ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594748

    authors: Awschalom M,Rosenberg I

    更新日期:1980-09-01 00:00:00

  • Computer-aided measurement of liver volumes in CT by means of geodesic active contour segmentation coupled with level-set algorithms.

    abstract:PURPOSE:Computerized liver extraction from hepatic CT images is challenging because the liver often abuts other organs of a similar density. The purpose of this study was to develop a computer-aided measurement of liver volumes in hepatic CT. METHODS:The authors developed a computerized liver extraction scheme based o...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3395579

    authors: Suzuki K,Kohlbrenner R,Epstein ML,Obajuluwa AM,Xu J,Hori M

    更新日期:2010-05-01 00:00:00

  • Some possible effects of a catheter on the arterial wall.

    abstract::When a catheter is inserted through an artery, the wall shear stress increases and the secondary flow exhibits an increased number of instantaneous vortices. These effects become more evident when the annular gap between the catheter and the artery becomes more narrow. Accordingly it is suggested that these factors ma...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596448

    authors: Karahalios GT

    更新日期:1990-09-01 00:00:00

  • Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy.

    abstract:PURPOSE:Flattening filter free (FFF) beams generated by medical linear accelerators have recently started to be used in radiotherapy clinical practice. Such beams present fundamental differences with respect to the standard filter flattened (FF) beams, making the generally used dosimetric parameters and definitions not...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4754799

    authors: Fogliata A,Garcia R,Knoos T,Nicolini G,Clivio A,Vanetti E,Khamphan C,Cozzi L

    更新日期:2012-10-01 00:00:00

  • Parametrization of head-scatter factors for rectangular photon fields using an equivalent square formalism.

    abstract::Head-scatter factors of symmetric square and rectangular fields (field center on the central beam axis) defined by the upper (Y) and lower (X) jaws for 6 and 15 MV photon beams from 2300CD and 600C accelerators (Varian Associates, Inc., Palo Alto, CA) were measured, as well as those for fields shaped by the Y jaws and...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597617

    authors: Yu MK,Murray B,Sloboda R

    更新日期:1995-08-01 00:00:00

  • Repeatability of SUV measurements in serial PET.

    abstract:PURPOSE:The standardized uptake value (SUV) is a quantitative measure of FDG tumor uptake frequently used as a tool to monitor therapeutic response. This study aims to (i) assess the reproducibility and uncertainty of SUV max and SUV mean, due to purely statistical, i.e., nonbiological, effects and (ii) to establish th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3578604

    authors: Schwartz J,Humm JL,Gonen M,Kalaigian H,Schoder H,Larson SM,Nehmeh SA

    更新日期:2011-05-01 00:00:00

  • Trapping volumetric measurement by multidetector CT in chronic obstructive pulmonary disease: effect of CT threshold.

    abstract:PURPOSE:The purpose of this study was to evaluate the effect of various computed tomography (CT) thresholds on trapping volumetric measurements by multidetector CT in chronic obstructive pulmonary disease (COPD). METHODS:Twenty-three COPD patients were scanned with a 64-slice CT scanner in both the inspiratory and exp...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4813906

    authors: Wang X,Duan J,Du Y,Shen N,He B,Yuan H

    更新日期:2013-08-01 00:00:00

  • Oxygen distribution in tumors: a qualitative analysis and modeling study providing a novel Monte Carlo approach.

    abstract:PURPOSE:To construct a Monte Carlo (MC)-based simulation model for analyzing the dependence of tumor oxygen distribution on different variables related to tumor vasculature [blood velocity, vessel-to-vessel proximity (vessel proximity), and inflowing oxygen partial pressure (pO2)]. METHODS:A voxel-based tissue model c...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4892386

    authors: Lagerlöf JH,Kindblom J,Bernhardt P

    更新日期:2014-09-01 00:00:00

  • MO-F-213AB-01: Improving Dose Uniformity in Patch-Field Proton Therapy Using Beam Current Modulation.

    abstract:PURPOSE:To improve the dose homogeneity in patch-field proton therapy using beam current modulation. METHODS:MCNPX simulations were performed for various configurations of the Mevion S250 beamline. Spread-out Bragg peak (SOBP) dose distributions were calculated, and lateral and distal dose profiles were extracted for ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735806

    authors: Hill P,Klein E,Bloch C

    更新日期:2012-06-01 00:00:00

  • Normalized dose data for upper gastrointestinal tract contrast studies performed to infants.

    abstract::The aim of the current study was to (a) provide normalized dose data for the estimation of the radiation dose from upper gastrointestinal tract contrast (UGIC) studies carried out to infants and (b) estimate the average patient dose and risks associated with radiation from UGIC examinations performed in our institutio...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2181297

    authors: Damilakis J,Stratakis J,Raissaki M,Perisinakis K,Kourbetis N,Gourtsoyiannis N

    更新日期:2006-04-01 00:00:00

  • Sparse and optimal acquisition design for diffusion MRI and beyond.

    abstract:PURPOSE:Diffusion magnetic resonance imaging (MRI) in combination with functional MRI promises a whole new vista for scientists to investigate noninvasively the structural and functional connectivity of the human brain-the human connectome, which had heretofore been out of reach. As with other imaging modalities, diffu...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3700166

    authors: Koay CG,Ozarslan E,Johnson KM,Meyerand ME

    更新日期:2012-05-01 00:00:00

  • Heating patterns of microwave applicators in inhomogeneous arm and thigh phantoms.

    abstract::Heating rate (HR) patterns in cylindrical structures were studied with inhomogeneous limb phantoms. These phantoms, arm and thigh models consisting of fat, bone, and muscle material, were heated with Clini-Therm L, M, and MS applicators at 915 MHz. The thigh model is 18 cm in diameter with 2.65-cm-thick fat on the out...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596627

    authors: Chou CK,McDougall JA,Chan KW,Luk KH

    更新日期:1991-11-01 00:00:00