Abstract:
:Low fresh gas flows (FGFs) decrease the use of anesthetic gases, but increase CO2 absorbent usage. CO2 absorbent usage remains poorly quantified. The goal of this study is to determine canister life of 8 commercially available CO2 absorbent prepacks with the Zeus®. Pre-packed CO2 canisters of 8 different brands were tested in vitro: Amsorb Plus, Spherasorb, LoFloSorb, LithoLyme, SpiraLith, SpheraSorb, Drägersorb 800+, Drägersorb Free, and CO2ntrol. CO2 (160 mL min- 1) flowed into the tip of a 2 L breathing bag that was ventilated with a tidal volume of 500 mL, a respiratory rate of 10/min, and an I:E ratio of 1:1 using the controlled mechanical ventilation mode of the Zeus® (Dräger, Lubeck, Germany). In part I, canister life of 5 canisters each of 2 different lots of each brand was determined with a 350 mL min- 1 FGF. Canister life is the time it takes for the inspired CO2 concentration (FICO2) to rise to 0.5%. In part II, canister life was measured accross a FGF range of 0.25 to 4 L min- 1 for Drägersorb 800+ (2 lots) and SpiraLith (1 lot). In part III, the calculated canister life per 100 g fresh granule content of the different brands was compared between the Zeus and (previously published data for) the Aisys. In vitro canister life of prefilled CO2 absorber canisters differed between brands, and depended on the amount of CO2 absorbent and chemical composition. Canister life expressed as FCU0.5 (the fraction of the canister used per hour) was proportional to FGF over 0.2-2 L min-1 range only, but was non-linear with higher FGF: FCU0.5 was larger than expected with FGF > 2 L min-1, and even with FGF > minute ventilation FCU0.5 did not become zero, indicating some CO2 was being absorbed. Canister life on a per weight basis of the same brand is higher with the Zeus than the Aisys. Canister life of prefilled CO2 absorber canisters differs between brands. The FCU0.5-FGF relationship is not linear across the entire FGF range. Canister life of prepacks of the same brand for the Zeus and Aisys differs, the exact etiology of which is probably multifactorial, and may include differences in the absolute amount of absorbent and different rebreathing characteristics of the machines.
journal_name
J Clin Monit Computjournal_title
Journal of clinical monitoring and computingauthors
Omer M,Hendrickx JFA,De Ridder S,De Houwer A,Carette R,De Cooman S,De Wolf AMdoi
10.1007/s10877-017-0088-xsubject
Has Abstractpub_date
2018-10-01 00:00:00pages
799-806issue
5eissn
1387-1307issn
1573-2614pii
10.1007/s10877-017-0088-xjournal_volume
32pub_type
杂志文章abstract::After introduction of the protoporphyrin IX-triplet state lifetime technique as a new method to measure mitochondrial oxygen tension in vivo, the development of a clinical monitor was started. This monitor is the "COMET", an acronym for Cellular Oxygen METabolism. The COMET is a non-invasive electrically powered optic...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-016-9966-x
更新日期:2017-12-01 00:00:00
abstract::The physiology underlying the intracranial pressure (ICP) curve morphology is not fully understood. Recent research has suggested that the morphology could be dependent on arterial cerebral inflow and the physiological and pathophysiological properties of the intracranial cavity. If understood, the ICP curve could pro...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-019-00342-8
更新日期:2020-06-01 00:00:00
abstract::The capnography is an important monitor to assess the patient's respiratory status. Importance of education for clinicians in interpretation of alarms and various capnography waveforms in different clinical scenarios and equipment malfunction is an understatement. Capnography waveforms due to sample line tube leak is ...
journal_title:Journal of clinical monitoring and computing
pub_type: 信件
doi:10.1007/s10877-019-00388-8
更新日期:2020-08-01 00:00:00
abstract::An intraoperative automated closed-loop system for goal-directed fluid therapy has been successfully tested in silico, in vivo and in a clinical case-control matching. This trial compared intraoperative cardiac output (CO) in patients managed with this closed-loop system versus usual practice in an academic medical ce...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章,随机对照试验
doi:10.1007/s10877-018-0106-7
更新日期:2018-12-01 00:00:00
abstract::Surgical pleth index (SPI) has been widely investigated in assessing the nociceptive level. The aim of this study was to investigate the relationship between SPI level and patient responses to trachea intubation and skin incision. A total of 40 patients undergoing open abdominal general surgery were recruited for anal...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-019-00425-6
更新日期:2020-12-01 00:00:00
abstract::One of the main goals of hemodynamic support is to preserve tissue perfusion. However issue perfusion is related more to microvascular perfusion than aortic blood flow. Monitoring the microcirculation has long been difficult. Recent technologic advances have made feasible monitoring of the microcirculation at bedside ...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章,评审
doi:10.1007/s10877-012-9383-8
更新日期:2012-10-01 00:00:00
abstract::In recent years students have increasingly objected to laboratory exercises involving animal subjects. We have replaced the valuable animal experiments with demonstrations using a full-scale human patient simulator. In small groups first-year medical students observe realistic clinical situations such as opioid-induce...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1023/a:1011401011887
更新日期:2000-01-01 00:00:00
abstract::Severe traumatic brain injury (TBI) patients are monitored with continuous arterial blood pressure (ABP) and intracranial pressure (ICP). The pressure reactivity index (PRx) is a frequently used correlation coefficient between ABP and ICP to inform clinicians at the bedside about trends in global cerebrovascular press...
journal_title:Journal of clinical monitoring and computing
pub_type: 信件
doi:10.1007/s10877-020-00471-5
更新日期:2020-02-08 00:00:00
abstract:PURPOSE:Introducing advanced hemodynamic monitoring might be beneficial during Helicopter Emergency Medical Service (HEMS) care. However, it should not increase the on-scene-time, it should be easy to use and should be non-invasive. The goal of this study was to investigate the feasibility of non-invasive cardiac outpu...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-021-00657-5
更新日期:2021-01-23 00:00:00
abstract::Oesophageal balloon calibration improves the oesophageal pressure (Pes) assessment during invasive controlled mechanical ventilation. The primary aim of the present investigation was to ascertain the feasibility of oesophageal balloon calibration during pressure support ventilation (PSV). Secondarily, the calibrated P...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-019-00436-3
更新日期:2020-12-01 00:00:00
abstract::This report presents a method for tracking Motor Evoked Potential (MEP) amplitudes over the course of a case using a moving least squares linear regression (LSMAs). During a case, newly obtained MEP amplitudes are compared to those predicted by a just previous linear regression (least squares moving average or LSMA). ...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-018-0181-9
更新日期:2019-06-01 00:00:00
abstract::Total shoulder arthroplasty (TSA) is typically performed in the beach-chair position. Maintenance of adequate mean arterial pressure is required to provide appropriate cerebral perfusion pressure and prevent cerebral ischemia. Placement of an arterial line to facilitate invasive monitoring is discretionary, based on c...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-016-9939-0
更新日期:2017-10-01 00:00:00
abstract::Ultrasound scanning of bladder volume is used for prevention of postoperative urinary retention (POUR). Accurate assessment of bladder volume is needed to allow clinical decision-making regarding the need for postoperative catheterization. Two commonly used ultrasound devices, the BladderScan® BVI 9400 and the newly r...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-018-0123-6
更新日期:2018-12-01 00:00:00
abstract::To develop and validate a mathematical model for predicting intracranial pressure (ICP) noninvasively using phase-contrast cine MRI (PC-MRI). We performed a retrospective analysis of PC-MRI from patients with communicating hydrocephalus (n = 138). The patients were recruited from Shenzhen Second People's Hospital betw...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-020-00598-5
更新日期:2020-10-01 00:00:00
abstract::Changes have been made to the AnaConDa device (Sedana Medical, Stockholm, Sweden), decreasing its size to reduce dead space and carbon dioxide (CO2) retention. However, this also involves a decrease in the surface area of the activated carbon filter. The CO2 elimination and sevoflurane (SEV) reflection of the old devi...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-019-00285-0
更新日期:2020-02-01 00:00:00
abstract:OBJECTIVE:To define the longest sampling interval which will faithfully record the time course of episodes of severe hypotension and hypoxia in anesthetised patients. METHODS:Electronic anesthetic records of 1501 patients were analyzed retrospectively for hypotension where measured systolic blood pressure fell to 60 m...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1023/a:1009978414365
更新日期:1998-07-01 00:00:00
abstract::Changes in expired nitric oxide (NO) occur in airway inflammation and have proved to be important in the monitoring of inflammatory disease processes such as asthma. We set out to develop a novel hand-held NO-analyzer with a performance comparable to the present more costly and complex chemiluminescence instruments. T...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-005-1158-z
更新日期:2004-12-01 00:00:00
abstract::Audible pulse tones, based on a variable-pitch frequency scale, allow the anesthesiologist to determine the patient's oxygen saturation without constant visual examination of the monitor display. The ability to reliably detect oxygen saturation levels based on audible pulse tones may be compromised when multiple pulse...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1023/b:jocm.0000032698.47717.06
更新日期:2004-04-01 00:00:00
abstract::With the AnaConDa™ and the MIRUS™ system, volatile anesthetics can be administered for inhalation sedation in intensive care units. Instead of a circle system, both devices use anesthetic reflectors to save on the anesthetic agent. We studied the efficiency of desflurane reflection with both devices using different ti...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-016-9902-0
更新日期:2018-08-01 00:00:00
abstract:OBJECTIVE:This study tests the hypothesis that a graphical object display (a data display consisting of meaningful shapes) will affect the ability of anesthesiologists to perform a diagnostic task rapidly and correctly. The diagnostic tasks studied were recognition and differentiation of five etiologies of shock--anaph...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1023/a:1009914019889
更新日期:1999-01-01 00:00:00
abstract::Bioreactance is a novel non-invasive method for cardiac output measurement that involves the analysis of blood flow-dependent changes in the phase shifts of electrical currents applied across the chest. The present study (1) compared resting and exercise cardiac outputs determined by bioreactance and bioimpedance meth...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-012-9334-4
更新日期:2012-04-01 00:00:00
abstract::The mesenteric traction syndrome (MTS) is associated with prostacyclin (PGI2) facilitated systemic vasodilatation during surgery and is identified by facial flushing. We hypothesized that severe facial flushing would be related to the highest concentrations of plasma PGI2 and accordingly to the highest levels of skin ...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-018-0226-0
更新日期:2019-10-01 00:00:00
abstract::To evaluate a compact and easily interpretable 4-parameter model describing the shape of the volumetric capnogram, and the resulting estimates of anatomical dead space (VDAW) and Phase III (alveolar plateau) slope (SIII). Data from of 8 mildly-endotoxemic pre-acute respiratory distress syndrome sheep were fitted to th...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-019-00451-4
更新日期:2020-12-01 00:00:00
abstract::The goal of this study was to evaluate the risk factors associated with positioning-related SSEP changes (PRSC). The study investigated the association between 18 plausible risk factors and the occurrence of intraoperative PRSC. Risk factors investigated included demographic variables, comorbidities, and procedure rel...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-018-0148-x
更新日期:2019-04-01 00:00:00
abstract::Recent advances in technology have led to the introduction of a variety of innovative devices, each with their own platform for data display, into the operating room (OR). While these innovative applications are expanding the traditional boundaries of the surgical space and enhancing treatment capabilities, the introd...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-006-9017-0
更新日期:2006-06-01 00:00:00
abstract::Cerebrovascular autoregulation (CA) is often impaired following traumatic brain injury. Established technologies and metrics used to assess CA are invasive and conducive for measurement, but not for continuous monitoring. We developed a trans-ocular brain impedance (TOBI) method that may provide non-invasive and conti...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-020-00556-1
更新日期:2020-07-14 00:00:00
abstract::Perioperative goal-directed therapy is considered to improve patient outcomes after high-risk surgery. The association of compliance with perioperative goal-directed therapy protocols and postoperative outcomes is unclear. The purpose of this study is to determine the effect of protocol compliance on postoperative out...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-020-00585-w
更新日期:2020-09-12 00:00:00
abstract::Excessive pressures exerted on the tongue during intubation may be associated with serious complications and may make intubation more difficult. The aim of this study was to evaluate five different intubating devices in the hands of experienced anaesthetists during simulated conditions of reduced cervical spine mobili...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-017-0095-y
更新日期:2018-10-01 00:00:00
abstract::Loss of endotracheal cuff pressure can lead to airway compromise and loss of volume delivery in mechanical ventilation requiring immediate intervention. A 40 years old male posted for bilateral interpositional arthroplasty was intubated nasally with 7.5 mm flexometallic endotracheal tube. In the intraoperative period ...
journal_title:Journal of clinical monitoring and computing
pub_type: 信件
doi:10.1007/s10877-020-00607-7
更新日期:2020-10-23 00:00:00
abstract::The aim of this study was to compare the cardiac function index (CFI) and global ejection fraction (GEF) obtained by VolumeView/EV1000™, with the left ventricular ejection fraction (LVEF) by echocardiography in septic shock patients. A prospective observational study was conducted in a medical intensive care unit of a...
journal_title:Journal of clinical monitoring and computing
pub_type: 杂志文章
doi:10.1007/s10877-018-0152-1
更新日期:2019-04-01 00:00:00