SU-E-T-610: Impact of Variable Beam Spot Size on Treatment Time in Particle Therapy.

Abstract:

PURPOSE:The dosimetric advantage of particle therapy comes with a much higher infrastructure investment and operation costs. Increasing patient throughput is a key factor to manage operation costs. We investigate the impact of variable beam spot sizes on treatment time and discuss the tradeoffs involved. METHODS:The following realistic assumptions were used. (1) The beam traveling speed is independent of the beam spot size. (2) The beam spot is a 2D Gaussian. Changing the beam spot size implies varying the standard deviation. (3) The maximum beam intensity is a constant independent of the beam spot size. Increasing the beam spot reduces the fluence. (4) Varying the beam spot size incurs in a reset time penalty.A 2D tumor was used in the study. Dose calculations were based on pencil beam kernels from GEANT4.The total treatment time is divided into the beam travel time, the beam-on time, andthe time for changing the spot size. RESULTS:We found that: (1) Changing the beam spot size has no impact on the beam-on time, because the maximum beam intensity is independentof the beam spot and increasing the beam spot only reduces the fluence. (2) Larger beam spot size shortens the total travel time inversely proportional to the radius of the beam spot. (3) Plans with different beam spot sizes have similar dosimetric qualities. (4) If higher beam intensity could be used for larger beam spot size, savings in beam-on time would be inversely proportional to the intensity available. CONCLUSIONS:We have studied the interplay among beam intensity, travel time, and beam size reset time for a scanning beam with variable beam spot size. Our initial studies show necessary conditions for and limitations on savings in total treatment times. Further studies are being carried out to find additional time saving sources. Supported in part by NSF CBET-0853157.

journal_name

Med Phys

journal_title

Medical physics

authors

Riofrio D,Sellner S,Cabal G,Keyes R,Holzscheiter M,Jaekel O,Luan S

doi

10.1118/1.4735699

subject

Has Abstract

pub_date

2012-06-01 00:00:00

pages

3846

issue

6Part19

eissn

0094-2405

issn

2473-4209

journal_volume

39

pub_type

杂志文章
  • Suitability of microDiamond detectors for the determination of absorbed dose to water around high-dose-rate 192 Ir brachytherapy sources.

    abstract:PURPOSE:Experimental dosimetry of high-dose-rate (HDR) 192 Ir brachytherapy (BT) sources is complicated due to high dose and dose-rate gradients, and softening of photon energy spectrum with depth. A single crystal synthetic diamond detector microDiamond (PTW 60019, Freiburg, Germany) has a small active volume, high se...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12694

    authors: Kaveckyte V,Malusek A,Benmakhlouf H,Alm Carlsson G,Carlsson Tedgren Å

    更新日期:2018-01-01 00:00:00

  • Scatter-glare correction using a convolution algorithm with variable weighting.

    abstract::Several investigators have approximated the scatter component of digital radiographic images by applying a constant weighting factor to a convolved version of the detected image. The scatter approximation is then subtracted from the detected image. When a constant weighting factor is used, the scatter in areas of low ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596088

    authors: Naimuddin S,Hasegawa B,Mistretta CA

    更新日期:1987-05-01 00:00:00

  • Carpal tunnel syndrome diagnosis by a self-normalization process and ultrasound compound imaging.

    abstract:PURPOSE:Carpal tunnel syndrome (CTS) is the common entrapment neuropathy that occurs due to compression of the median nerve at the wrist. Ultrasound images have been used to highlight anatomical variants of the median nerve, and CTS is thought to be associated to enlargement of the cross-sectional area (CSA) of the med...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4767754

    authors: Liao YY,Wu CC,Kuo TT,Chen JP,Hsu YW,Yeh CK

    更新日期:2012-12-01 00:00:00

  • Study of dosimetry consistency for kilovoltage x-ray beams.

    abstract::In this paper, the consistency of kilovoltage (tube potentials between 40 and 300 kV) x-ray beam dosimetry using the "in-air" method and the in-phantom measurement has been studied. The procedures for the measurement of the central-axis depth-dose curve, which serve as a link between the dose at the reference depth to...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598448

    authors: Ma CM,Li XA,Seuntjens JP

    更新日期:1998-12-01 00:00:00

  • Thermoacoustic molecular tomography with magnetic nanoparticle contrast agents for targeted tumor detection.

    abstract:PURPOSE:The primary feasibility steps of demonstrating the ability of microwave-induced thermoacoustic (TA) in phantoms have been previously reported. However, none were shown to target a diseased site in living subjects in thermoacoustic tomography (TAT) field so far. To determine the expressions of oncogenic surface ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3466696

    authors: Nie L,Ou Z,Yang S,Xing D

    更新日期:2010-08-01 00:00:00

  • TU-E-217BCD-10: Dose Reduction in Digital Breast Tomosynthesis with the Dose Reduction Prior Image Constrained Compressed Sensing (DR-PICCS) Algorithm.

    abstract:PURPOSE:To reduce image noise and radiation dose in Digital Breast Tomosynthesis (DBT) reconstructions. METHODS:A retrospective study was performed on clinical data sets acquired at a normal dose with Hologic Selenia Dimensions DBT systems. The Prior Image Constrained Compressed Sensing (PICCS) algorithm was used to r...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735982

    authors: Garrett J,Tang J,Zhang Y,Ruth C,Jing Z,Chen GH

    更新日期:2012-06-01 00:00:00

  • Characterization of distensibility, plaque burden, and composition of the atherosclerotic carotid artery using magnetic resonance imaging.

    abstract:PURPOSE:Arterial distensibility is a marker that can measure vessel wall functional and structural changes resulting from atherosclerosis with applications including estimation of mechanical properties of the wall. We sought to assess the feasibility of using magnetic resonance imaging (MRI) to include wall distensibil...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4754302

    authors: Canton G,Hippe DS,Sun J,Underhill HR,Kerwin WS,Tang D,Yuan C

    更新日期:2012-10-01 00:00:00

  • ALTIS: A fast and automatic lung and trachea CT-image segmentation method.

    abstract:PURPOSE:The automated segmentation of each lung and trachea in CT scans is commonly taken as a solved problem. Indeed, existing approaches may easily fail in the presence of some abnormalities caused by a disease, trauma, or previous surgery. For robustness, we present ALTIS (implementation is available at http://lids....

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13773

    authors: Sousa AM,Martins SB,Falcão AX,Reis F,Bagatin E,Irion K

    更新日期:2019-11-01 00:00:00

  • The generalized geometry of eye plaque therapy.

    abstract::A calculation is described that enables the rapid assessment of dose rate at various points of interest within the eye (lens, optic nerve, etc.) for the treatment of choroidal melanoma by plaque therapy. 125I seeds are used as the radiation source. The location of the plaque and its associated seeds relative to the ey...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596234

    authors: Kepka AG,Johnson PM,Kline RW

    更新日期:1988-05-01 00:00:00

  • Accuracy of deformable image registration on magnetic resonance images in digital and physical phantoms.

    abstract:PURPOSE:Accurate deformable image registration is necessary for longitudinal studies. The error associated with commercial systems has been evaluated using computed tomography (CT). Several in-house algorithms have been evaluated for use with magnetic resonance imaging (MRI), but there is still relatively little inform...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12406

    authors: Ger RB,Yang J,Ding Y,Jacobsen MC,Fuller CD,Howell RM,Li H,Jason Stafford R,Zhou S,Court LE

    更新日期:2017-10-01 00:00:00

  • Optimized low-kV spectrum of dual-energy CT equipped with high-kV tin filtration for electron density measurements.

    abstract:PURPOSE:This paper describes the low-kV spectral optimization of dual-energy CT (DECT) equipped with high-kV tin filtration for the quantitative acquisition of electron density information, which is essential for treatment planning in radiotherapy. In addition, an analytical DECT image simulation was preliminarily perf...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3584200

    authors: Saito M

    更新日期:2011-06-01 00:00:00

  • Dose homogeneity specification for reference dosimetry of nonstandard fields.

    abstract:PURPOSE:To investigate the sensitivity of the plan-class specific correction factor to dose distributions in composite nonstandard field dosimetry. METHODS:A cylindrical water-filled PMMA phantom was constructed at the center of which reference absorbed dose could be measured. Ten different TomoTherapy(®)-based IMRT f...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3669487

    authors: Chung E,Soisson E,Seuntjens J

    更新日期:2012-01-01 00:00:00

  • A technical solution to avoid partial scan artifacts in cardiac MDCT.

    abstract::Quantitative evaluation of cardiac image data obtained using multidetector row computed tomography (CT) is compromised by partial scan reconstructions, which improve the temporal resolution but significantly increase image-to-image CT number variations for a fixed region of interest compared to full reconstruction ima...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2805476

    authors: Primak AN,Dong Y,Dzyubak OP,Jorgensen SM,McCollough CH,Ritman EL

    更新日期:2007-12-01 00:00:00

  • CeBr3 scintillators for 4 He prompt gamma spectroscopy: Results from a Monte Carlo optimization study.

    abstract:PURPOSE:Range uncertainties limit the potential of charged particle therapy. In vivo and online range verification techniques could increase the confidence in the dose delivery distribution and lead to more conformal treatments. Prompt gamma imaging and prompt gamma spectroscopy (PGS) have been demonstrated for such a ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12795

    authors: Dal Bello R,Magalhaes Martins P,Seco J

    更新日期:2018-04-01 00:00:00

  • SU-E-J-17: Evaluation of Metal Artifact Reduction in MVCTs Using a Model Based Image Reconstruction Method.

    abstract:PURPOSE:To evaluate the performance of a model based image reconstruction in reducing metal artifacts in MVCT systems, and to compare with filtered-back projection (FBP) technique. METHODS:Iterative maximum likelihood polychromatic algorithm for CT (IMPACT) is used with pair/triplet production process and the energy d...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734850

    authors: Paudel M,MacKenzie M,Fallone B,Rathee S

    更新日期:2012-06-01 00:00:00

  • Micro-angiography for neuro-vascular imaging. II. Cascade model analysis.

    abstract::A micro-angiographic detector was designed and its performance was previously tested to evaluate its feasibility as an improvement over current x-ray detectors for neuro-interventional imaging. The detector was shown to have a modulation transfer function value of about 2% at the Nyquist frequency of 10 cycles/mm and ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1617550

    authors: Ganguly A,Rudin S,Bednarek DR,Hoffmann KR

    更新日期:2003-11-01 00:00:00

  • A multipinhole small animal SPECT system with submillimeter spatial resolution.

    abstract::Single photon emission computed tomography (SPECT) is an important technology for molecular imaging studies of small animals. In this arena, there is an increasing demand for high performance imaging systems that offer improved spatial resolution and detection efficiency. We have designed a multipinhole small animal i...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2190332

    authors: Funk T,Després P,Barber WC,Shah KS,Hasegawa BH

    更新日期:2006-05-01 00:00:00

  • SAPHIRE (scintillator avalanche photoconductor with high resolution emitter readout) for low dose x-ray imaging: spatial resolution.

    abstract::An indirect flat panel imager (FPI) with programmable avalanche gain and field emitter array (FEA) readout is being investigated for low-dose and high resolution x-ray imaging. It is made by optically coupling a structured x-ray scintillator, e.g., thallium (Tl) doped cesium iodide (CsI), to an amorphous selenium (a-S...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2937652

    authors: Li D,Zhao W

    更新日期:2008-07-01 00:00:00

  • Technical Note: A proposal of air ventilation system design criteria for a clinical room in a heavy-ion medical facility.

    abstract:PURPOSE:An optimized air ventilation system design for a treatment room in Heavy-ion Medical Facility is an important issue in the aspects of nuclear safety because the activated air produced in a treatment room can directly affect the medical staff and the general public in the radiation-free area. METHODS:Optimized ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12914

    authors: Kum O

    更新日期:2018-06-01 00:00:00

  • In vivo functional chronic imaging of a small animal model using optical-resolution photoacoustic microscopy.

    abstract::Optical-resolution photoacoustic microscopy (OR-PAM) has been validated as a valuable tool for label-free volumetric microvascular imaging. More importantly, the advantages of noninvasiveness and measurement consistency suggest the use of OR-PAM for chronic imaging of intact microcirculation. Here, such chronic imagin...

    journal_title:Medical physics

    pub_type: 信件

    doi:10.1118/1.3137572

    authors: Hu S,Maslov K,Wang LV

    更新日期:2009-06-01 00:00:00

  • Sci-Thur AM: Planning - 10: Improved dosimetric accuracy for patient specific quality assurance using a dual-detector measurement method for cyberknife output factors.

    abstract::The measurement of output factors for small fields is challenging and can lead to large dose errors in patient treatments if corrections for detector size and scatter from high-Z material are not applied. Due to its high spatial resolution and near tissue equivalence, GAFCHROMIC® film potentially provides a correction...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4740095

    authors: Vandervoort E,La Russa D,Ploquin N,Kadir IAE,Szanto J

    更新日期:2012-07-01 00:00:00

  • Sci-Thurs PM: Delivery-03: Optical imaging of microscopic radiation dose gradients using a digital microscope.

    abstract::Using superior localization and immobilization methods, stereotactic radiosurgery is capable of delivering spheres of dose as small as a few millimetres in diameter to intracranial targets. For targets abutting critical structures, the most conformal treatments minimize adverse radiation side effects and it is importa...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2965910

    authors: Keller BM,Peressotti C,Pignol JP

    更新日期:2008-07-01 00:00:00

  • Lag measurement in an a-Se active matrix flat-panel imager.

    abstract::Lag and residual contrast have been quantified in an amorphous selenium (a-Se) active-matrix flat-panel imager (AMFPI) as a function of frame time, kilovoltage (kV) and megavoltage (MV) x-ray photon energies and amount of radiation incident on the detector. The AMFPI contains a 200 microm thick a-Se layer deposited on...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1713298

    authors: Schroeder C,Stanescu T,Rathee S,Fallone BG

    更新日期:2004-05-01 00:00:00

  • A comparative contrast-detail study of five medical displays.

    abstract::The objective of this study was to compare the contrast-detail performance of five different commercial liquid crystal displays (LCDs) to other LCD and cathode-ray tube (CRT) displays for medical applications. A contrast-detail phantom, supplemented with 5 in. of acrylic, was imaged on a commercial digital radiographi...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2868780

    authors: Samei E,Ranger NT,Delong DM

    更新日期:2008-04-01 00:00:00

  • A comparison of gantry-mounted x-ray-based real-time target tracking methods.

    abstract:PURPOSE:Most modern radiotherapy machines are built with a 2D kV imaging system. Combining this imaging system with a 2D-3D inference method would allow for a ready-made option for real-time 3D tumor tracking. This work investigates and compares the accuracy of four existing 2D-3D inference methods using both motion tr...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12765

    authors: Montanaro T,Nguyen DT,Keall PJ,Booth J,Caillet V,Eade T,Haddad C,Shieh CC

    更新日期:2018-03-01 00:00:00

  • Validating FMEA output against incident learning data: A study in stereotactic body radiation therapy.

    abstract:PURPOSE:Though failure mode and effects analysis (FMEA) is becoming more widely adopted for risk assessment in radiation therapy, to our knowledge, its output has never been validated against data on errors that actually occur. The objective of this study was to perform FMEA of a stereotactic body radiation therapy (SB...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4919440

    authors: Yang F,Cao N,Young L,Howard J,Logan W,Arbuckle T,Sponseller P,Korssjoen T,Meyer J,Ford E

    更新日期:2015-06-01 00:00:00

  • Technical Note: Influence of Compton currents on profile measurements in small-volume ion chambers.

    abstract:PURPOSE:This work is to evaluate the effects of Compton current generation in three small-volume ionization chambers on measured beam characteristics for electron fields. METHODS:Beam scans were performed using Exradin A16, A26, and PTW 31014 microchambers. Scans with varying chamber components shielded were performed...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4929555

    authors: Tanny S,Holmes S,Sperling N,Parsai EI

    更新日期:2015-10-01 00:00:00

  • SU-D-211-06: The Effect of Motion in RapidArc Lung SBRT Delivery.

    abstract:PURPOSE:RapidArc is routinely used for stereotactic radiotherapy for lung cancer. While treatment dose is optimized and calculated on a static CT image, the motion of the target in conjunction with the motion of the MLC may Result in a delivered dose deviating from the planed dose. In this study, we investigate the dos...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734660

    authors: Zou J,Yin L,Shen J,Corradetti M,Kirk M,Rengan R,Teo BK

    更新日期:2012-06-01 00:00:00

  • Poster - Thur Eve - 18: Characterization of a camera and LED lightbox imaging system for radiochromic film dosimetry.

    abstract::Radiation therapy treatment modalities continue to develop and have become increasingly complex. With this, dose verification and quality assurance (QA) is of great importance to ensure that a prescribed dose is accurately and precisely delivered to a patient. Radiochromic film dosimetry has been adopted as a convenie...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4740126

    authors: Alexander K,Percy E,Olding T,Schreiner LJ,Salomons G

    更新日期:2012-07-01 00:00:00

  • Experimental investigation on the accuracy of plastic scintillators and of the spectrum discrimination method in small photon fields.

    abstract:PURPOSE:To investigate the accuracy of output factor measurements using a commercial (Exradin W1, SI) and a prototype, "in-house" developed, plastic scintillation dosimeter (PSD) in small photon fields. METHODS:Repetitive detector-specific output factor OFdet measurements were performed in water (parallel to the CAX) ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12064

    authors: Papaconstadopoulos P,Archambault L,Seuntjens J

    更新日期:2017-02-01 00:00:00