SU-E-T-109: An Evaluation of a Newly Available Portal Dosimetry QA Tool for SRS Dynamic Arc and VMAT.

Abstract:

PURPOSE:In our institute, the MatriXX has been used for IMRT, VMAT and dynamic arc QA. However, the small fields (<3cm) shaped by HDMLC for SRS dynamic arc (DA) present QA measurement issues when using the MatriXX due to its limited resolution (7.6 mm). A recently available portal dosimetry (PD) software tool in Eclipse Version 10 is capable of performing QA for arc plans. This study is to evaluate this new tool for QA. METHODS:EPID calibration was performed using a Varian Novalis TX unit with HDMLCs for a 6X high dose rate SRS beam. The PDIP algorithm was configured using measured beam data. Brainlab iPlan was used to generate the DA patient treatment plans. The plans were then exported to Aria via DICOM RT. A verification plan was created in Eclipse and the predicted portal dose calculated. The comparison was done between predicted and measured portal dose distributions. RESULTS:Compared with chamber array analysis, the measured portal dose shows significant improvement in correlation with predicted dose. To eliminate the uncertainty of dose normalization, absolute CU is chosen for analysis. Gamma factor was studied. Thecriterion was set at 3%/1mm using a 10% dose threshold such that 95% of the pixels had a gamma <1. QA was performed on 9 plans and the average gamma passing rate is 98.5%, with passing rates ranging from 95% to 100%. CONCLUSIONS:With approximately 0.7mm resolution, it is shown that the newly available PD tool is capable of overcoming the limitations of chamber arrays for small field SRS QA and is also suitable for HDMLC QA to provide detailed dose comparison with highly modulated dose profiles within a small treatment field. In addition, the time needed to perform the QA procedure using the PD tool is significantly shortened compared to the MatriXX.

journal_name

Med Phys

journal_title

Medical physics

authors

Luo G,Crass J,Morales M,Ding G

doi

10.1118/1.4735167

subject

Has Abstract

pub_date

2012-06-01 00:00:00

pages

3728

issue

6Part11

eissn

0094-2405

issn

2473-4209

journal_volume

39

pub_type

杂志文章
  • Quantitative study of focused ultrasound enhanced doxorubicin delivery to prostate tumor in vivo with MRI guidance.

    abstract:PURPOSE:The purpose of this study was to investigate the potential of MR-guided pulsed focused ultrasound (pFUS) for the enhancement of drug uptake in prostate tumors in vivo using doxorubicin (Dox). METHODS:An antitumor drug Dox, an orthotopic animal prostate tumor model using human prostate cancer, LNCaP cell line, ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4705346

    authors: Chen X,Cvetkovic D,Ma CM,Chen L

    更新日期:2012-05-01 00:00:00

  • SU-E-T-142: Evaluation of the Energy Dependence of Gafchromic EBT3 Film for Electron Beams.

    abstract:PURPOSE:To evaluate the energy dependence of Gafchromic EBT3 film for establishing a quality assurance method of bolus electron conformal radiotherapy. METHODS:We irradiated electron beam to EBT3 films, which were set in the water tank. The linear accelerator used was Varian Clinac 21EX. The energy of electron beams w...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735200

    authors: Inoue T,Sugimoto S,Fukata K,Kurokawa C,Ozawa S,Sasai K

    更新日期:2012-06-01 00:00:00

  • Energy spectrum based calculation of the half and the tenth value layers for brachytherapy sources using a semiempirical parametrized mass attenuation coefficient formulism.

    abstract::As different types of radionuclides (e.g., 131Cs source) are introduced for clinical use in brachytherapy, the question is raised regarding whether a relatively simple method exists for the derivation of values of the half value layer (HVL) or the tenth value layer (TVL). For the radionuclide that has been clinically ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2905356

    authors: Yue NJ

    更新日期:2008-06-01 00:00:00

  • Variability of tumor area measurements for response assessment in malignant pleural mesothelioma.

    abstract:PURPOSE:The measurement of malignant pleural mesothelioma is critical to the assessment of tumor response to therapy. Current response assessment standards utilize summed linear measurements acquired on three computed tomography (CT) sections. The purpose of this study was to evaluate manual area measurements as an alt...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4810940

    authors: Labby ZE,Straus C,Caligiuri P,MacMahon H,Li P,Funaki A,Kindler HL,Armato SG 3rd

    更新日期:2013-08-01 00:00:00

  • A preclinical Talbot-Lau prototype for x-ray dark-field imaging of human-sized objects.

    abstract:PURPOSE:Talbot-Lau x-ray interferometry provides information about the scattering and refractive properties of an object - in addition to the object's attenuation features. Until recently, this method was ineligible for imaging human-sized objects as it is challenging to adapt Talbot-Lau interferometers (TLIs) to the r...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12889

    authors: Hauke C,Bartl P,Leghissa M,Ritschl L,Sutter SM,Weber T,Zeidler J,Freudenberger J,Mertelmeier T,Radicke M,Michel T,Anton G,Meinel FG,Baehr A,Auweter S,Bondesson D,Gaass T,Dinkel J,Reiser M,Hellbach K

    更新日期:2018-06-01 00:00:00

  • First clinical implementation of real-time, real anatomy tracking and radiation beam control.

    abstract:PURPOSE:We describe the acceptance testing, commissioning, periodic quality assurance, and workflow procedures developed for the first clinically implemented magnetic resonance imaging-guided radiation therapy (MR-IGRT) system for real-time tracking and beam control. METHODS:The system utilizes real-time cine imaging ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13002

    authors: Green OL,Rankine LJ,Cai B,Curcuru A,Kashani R,Rodriguez V,Li HH,Parikh PJ,Robinson CG,Olsen JR,Mutic S,Goddu SM,Santanam L

    更新日期:2018-05-28 00:00:00

  • Performance characteristics of A 150 plastic-equivalent gases in A 150 plastic proportional counters for 14.8-MeV neutrons.

    abstract::Two recently developed A 150 plastic-equivalent gas mixtures have been tested for suitability in proportional counter applications. Methane- and propane-based "tissue-equivalent" gases were also included for comparison purposes. Event-size weighted dose distributions were measured in a 14.8-MeV neutron beam. Resolutio...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595536

    authors: DeLuca PM Jr,Schell MC,Pearson DW,Higgins PD,Attix FH

    更新日期:1984-07-01 00:00:00

  • Measurements of the neutron yields from 7Li(p,n)7Be reaction (thick target) with incident energies from 1.885 to 2.0 MeV.

    abstract::Accelerator-based neutron source have been considered to be practical for boron neutron capture therapy (BNCT). Based on experience with a parameters of the Brookhaven National Laboratory BMRR reactor neutron source, which has been used in treatment experiments, the future accelerator-based neutron source for BNCT sho...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598299

    authors: Yu W,Yue G,Han X,Chen J,Tian B

    更新日期:1998-07-01 00:00:00

  • MO-F-213AB-01: Improving Dose Uniformity in Patch-Field Proton Therapy Using Beam Current Modulation.

    abstract:PURPOSE:To improve the dose homogeneity in patch-field proton therapy using beam current modulation. METHODS:MCNPX simulations were performed for various configurations of the Mevion S250 beamline. Spread-out Bragg peak (SOBP) dose distributions were calculated, and lateral and distal dose profiles were extracted for ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735806

    authors: Hill P,Klein E,Bloch C

    更新日期:2012-06-01 00:00:00

  • Breast tumor segmentation in 3D automatic breast ultrasound using Mask scoring R-CNN.

    abstract:PURPOSE:Automatic breast ultrasound (ABUS) imaging has become an essential tool in breast cancer diagnosis since it provides complementary information to other imaging modalities. Lesion segmentation on ABUS is a prerequisite step of breast cancer computer-aided diagnosis (CAD). This work aims to develop a deep learnin...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14569

    authors: Lei Y,He X,Yao J,Wang T,Wang L,Li W,Curran WJ,Liu T,Xu D,Yang X

    更新日期:2020-10-30 00:00:00

  • Clinical implementation of a Monte Carlo treatment planning system.

    abstract::The purpose of this study was to implement the Monte Carlo method for clinical radiotherapy dose calculations. We used the EGS4/BEAM code to obtain the phase-space data for 6-20 MeV electron beams and 4, 6, and 15 MV photon beams for Varian Clinac 1800, 2100C, and 2300CD accelerators. A multiple-source model was used ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598729

    authors: Ma CM,Mok E,Kapur A,Pawlicki T,Findley D,Brain S,Forster K,Boyer AL

    更新日期:1999-10-01 00:00:00

  • Gradient corrections for reference dosimetry using Farmer-type ionization chambers in single-layer scanned proton fields.

    abstract:PURPOSE:The local depth dose gradient and the displacement correction factor for Farmer-type ionization chambers are quantified for reference dosimetry at shallow depth in single-layer scanned proton fields. METHOD:Integrated radial profiles as a function of depth (IRPDs) measured at three proton therapy centers were ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14554

    authors: Palmans H,Medin J,Trnková P,Vatnitsky S

    更新日期:2020-12-01 00:00:00

  • The line spread function and modulation transfer function of a computed tomographic scanner.

    abstract::A simple method to measure the line spread function of computed tomographic scanners has been developed. The line spread function of the EMI-Scanner was measured and the MTF of the scanner was calculated. The reproducibility of the method indicated this technique could be utilized to compare the spatial resolution of ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594283

    authors: Judy PF

    更新日期:1976-07-01 00:00:00

  • Analyzing the impact of intrafraction motion: correlation of different dose metrics with changes in target D95%.

    abstract:PURPOSE:A number of techniques are available to determine the dosimetric impact of intrafraction motion during intensity modulated radiation therapy (IMRT). Motion-induced dose perturbations can be determined both computationally and experimentally using a number of different dosimetric metrics. However, these measures...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3605633

    authors: Waghorn BJ,Meeks SL,Langen KM

    更新日期:2011-08-01 00:00:00

  • Incorporation of treatment plan spatial and temporal dose patterns into a prostate intrafractional motion management strategy.

    abstract:PURPOSE:Periodic MV∕KV radiographs taken during volumetric modulated arc therapy (VMAT) for hypofractionated treatment provide guidance in intrafractional motion management. The choice of imaging frequency and timing are key components in delivering the desired dose while reducing associated overhead such as imaging do...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4742846

    authors: Zhang P,Hunt M,Happersett L,Cox B,Mageras G

    更新日期:2012-09-01 00:00:00

  • Wedge-shaped dose distributions by computer-controlled collimator motion.

    abstract::We have recently installed a linear accelerator, modified to allow computer control of several machine parameters during irradiation of the patient. As an initial feasibility study of computer-controlled radiation therapy, its application to produce wedge-shaped dose distributions by moving the collimator jaws has bee...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594440

    authors: Kijewski PK,Chin LM,Bjärngard BE

    更新日期:1978-09-01 00:00:00

  • A nonvoxel-based dose convolution/superposition algorithm optimized for scalable GPU architectures.

    abstract:PURPOSE:Real-time adaptive planning and treatment has been infeasible due in part to its high computational complexity. There have been many recent efforts to utilize graphics processing units (GPUs) to accelerate the computational performance and dose accuracy in radiation therapy. Data structure and memory access pat...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4895822

    authors: Neylon J,Sheng K,Yu V,Chen Q,Low DA,Kupelian P,Santhanam A

    更新日期:2014-10-01 00:00:00

  • Dose and detectability for a cone-beam C-arm CT system revisited.

    abstract:PURPOSE:The authors had previously published measurements of the detectability of disk-shaped contrast objects in images obtained from a C-arm CT system. A simple approach based on Rose's criterion was used to scale the date, assuming the threshold for the smallest diameter detected should be inversely proportional to ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3397465

    authors: Ganguly A,Yoon S,Fahrig R

    更新日期:2010-05-01 00:00:00

  • Evaluation of CT-based lean-body SUV.

    abstract:PURPOSE:The authors introduce a novel method for defining standardized uptake values (SUVs) in PET∕CT based on routinely collected CT data. The goal of the study is to reduce, if possible, the variability of SUV in a heterogeneous population. Two well established methods for defining SUV are based on body weight (BW) a...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4816656

    authors: Hamill JJ,Sunderland JJ,LeBlanc AK,Kojima CJ,Wall J,Martin EB

    更新日期:2013-09-01 00:00:00

  • The role of water desorption on optical clearing of biotissue: studied with near infrared reflectance spectroscopy.

    abstract::In order to understand the role of water desorption, i.e., the opposite of water adsorption, in optical clearing of biological tissue created by the application of hyperosmotic agents, dynamics of water loss in fresh porcine muscle tissue administrated with glycerol or ethylene glycol was investigated with the near-in...

    journal_title:Medical physics

    pub_type:

    doi:10.1118/1.1576228

    authors: Xu X,Wang RK

    更新日期:2003-06-01 00:00:00

  • Reconstruction of electron spectra using singular component decomposition.

    abstract::Reconstruction of electron spectra of medical accelerators from measured depth dose distributions is an attractive tool for commissioning of a Monte Carlo treatment planning system. However, the reconstruction method is an inverse radiation transport problem which is poorly conditioned, in the sense it may become unst...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1461840

    authors: Chvetsov AV,Sandison GA

    更新日期:2002-04-01 00:00:00

  • Monte Carlo dose calculations in homogeneous media and at interfaces: a comparison between GEPTS, EGSnrc, MCNP, and measurements.

    abstract::Three Monte Carlo photon/electron transport codes (GEPTS, EGSnrc, and MCNP) are bench-marked against dose measurements in homogeneous (both low- and high-Z) media as well as at interfaces. A brief overview on physical models used by each code for photon and electron (positron) transport is given. Absolute calorimetric...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1473134

    authors: Chibani O,Li XA

    更新日期:2002-05-01 00:00:00

  • Kymogram detection and kymogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart.

    abstract::Subsecond single-slice, multi-slice or cone-beam spiral computed tomography (SSCT, MSCT, CBCT) offer great potential for improving heart imaging. Together with the newly developed phase-correlated cardiac reconstruction algorithms 180 degrees MCD and 180 degrees MCI [Med. Phys. 27, 1881-1902 (2000)] or related algorit...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1487861

    authors: Kachelriess M,Sennst DA,Maxlmoser W,Kalender WA

    更新日期:2002-07-01 00:00:00

  • Absorbed-dose beam quality conversion factors for cylindrical chambers in high energy photon beams.

    abstract::Recent working groups of the AAPM [Almond et al., Med. Phys. 26, 1847 (1999)] and the IAEA (Andreo et al., Draft V.7 of "An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water," IAEA, 2000) have described guidelines to base reference dosimetry of high energy photon beams on absorb...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1328081

    authors: Seuntjens JP,Ross CK,Shortt KR,Rogers DW

    更新日期:2000-12-01 00:00:00

  • Ideal-observer detectability in photon-counting differential phase-contrast imaging using a linear-systems approach.

    abstract:PURPOSE:To provide a cascaded-systems framework based on the noise-power spectrum (NPS), modulation transfer function (MTF), and noise-equivalent number of quanta (NEQ) for quantitative evaluation of differential phase-contrast imaging (Talbot interferometry) in relation to conventional absorption contrast under equal-...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4739195

    authors: Fredenberg E,Danielsson M,Stayman JW,Siewerdsen JH,Aslund M

    更新日期:2012-09-01 00:00:00

  • A methodology to determine margins by EPID measurements of patient setup variation and motion as applied to immobilization devices.

    abstract::Assessment of clinic and site specific margins are essential for the effective use of three-dimensional and intensity modulated radiation therapy. An electronic portal imaging device (EPID) based methodology is introduced which allows individual and population based CTV-to-PTV margins to be determined and compared wit...

    journal_title:Medical physics

    pub_type: 临床试验,杂志文章

    doi:10.1118/1.1800712

    authors: Prisciandaro JI,Frechette CM,Herman MG,Brown PD,Garces YI,Foote RL

    更新日期:2004-11-01 00:00:00

  • A contrast-detail analysis of diagnostic ultrasound imaging.

    abstract::For diagnostic ultrasound imaging, as in computed tomography, a feature of prime importance is the detection of focal lesions of varying size and contrast (echo amplitude) from surrounding tissue. This study describes a new tissue-stimulating phantom which has been used to measure the threshold detection of varying co...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595218

    authors: Smith SW,Lopez H

    更新日期:1982-01-01 00:00:00

  • Determination of the reference air kerma rate for 192Ir brachytherapy sources and the related uncertainty.

    abstract::Different methods exist to determine the air kerma calibration factor of an ionization chamber for the spectrum of a 192Ir high-dose-rate (HDR) or pulsed-dose-rate (PDR) source. An analysis of two methods to obtain such a calibration factor was performed: (i) the method recommended by [Goetsch et al., Med. Phys. 18, 4...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1791352

    authors: van Dijk E,Kolkman-Deurloo IK,Damen PM

    更新日期:2004-10-01 00:00:00

  • Monte Carlo assessment of computed tomography dose to tissue adjacent to the scanned volume.

    abstract::The assessment of the radiation dose to internal organs or to an embryo or fetus is required on occasion for risk assessment or for comparing imaging studies. Limited resources hinder the ability to accurately assess the radiation dose received to locations outside the tissue volume actually scanned during computed to...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1312809

    authors: Boone JM,Cooper VN 3rd,Nemzek WR,McGahan JP,Seibert JA

    更新日期:2000-10-01 00:00:00

  • Determination of electron energy, spectral width, and beam divergence at the exit window for clinical megavoltage x-ray beams.

    abstract::Monte Carlo simulations of x-ray beams typically take parameters of the electron beam in the accelerating waveguide to be free parameters. In this paper, a methodology is proposed and implemented to determine the energy, spectral width, and beam divergence of the electron source. All treatment head components were rem...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3070547

    authors: Sawkey DL,Faddegon BA

    更新日期:2009-03-01 00:00:00