Calculation of lung mean dose and quantification of error for 90 Y-microsphere radioembolization using 99m Tc-MAA SPECT/CT and diagnostic chest CT.

Abstract:

PURPOSE:Current treatment planning for 90 Y radioembolization estimates lung mean dose (LMD) by measuring the lung shunt fraction (LSF) from 99m Tc-macroaggregated albumin (MAA) planar imaging and assuming a 1-kg lung mass. This methodology, however, overestimates LSF and LMD and could therefore unnecessarily limit the dose to target volume(s). We propose an improved LMD calculation that derives LSF from 99m Tc-MAA SPECT/CT and the patient-specific lung mass from diagnostic chest CT. Furthermore, we investigated the errors in lung mass, LSF, and LMD arising from contour variability in patient data in order to estimate the precision of our proposed methodology. METHODS:Our proposed LMD (LMDnew ) calculation consisted of the following steps: (a) estimate liver counts from the MAA SPECT/CT liver contour; (b) estimate total lung counts by multiplying density (counts/g) from the MAA SPECT/CT left-lung contour by the total lung mass (g) from the diagnostic CT lung contours; (c) compute LSFnew from liver and lung counts; (d) calculate LMDnew using LSFnew and the total lung mass from the diagnostic CT (Mnew ). LMDnew , LSFnew , and Mnew estimates were compared to standard model values (LMDclin , LSFclin , and 1 kg, respectively) in 52 consecutive patients with hepatocellular carcinoma who underwent radioembolization using 90 Y glass microspheres. The precision of our methodology was quantified by varying lung and liver contours in the same patient population and calculating the resulting relative errors in the liver count, lung count, and lung mass measurements. RESULTS:The median Mnew was 839 g (range, 550-1178 g) for men and 731 g (range, 548-869 g) for women. The median LSFnew was 0.02 (range, 0.01-0.11), while the median LMDnew was 4.9 Gy (range, 0.3-25.5 Gy). Mnew , LSFnew , and LMDnew were significantly lower than Mclin , LSFclin , and LMDclin , with respective relative mean (±SD) differences of -20% (±16%) for Mnew , -63% (±15%) for LSFnew , and -53% (±23%) for LMDnew . The estimated 1-sigma uncertainties in Mnew , LSFnew , and LMDnew were 9%, 10%, and 13%, respectively. CONCLUSIONS:We derived a method to calculate lung mass and LSF using routinely available diagnostic chest CT and 99m Tc-MAA SPECT/CT. More importantly, we systematically quantified the errors in our measurements to establish the precision of the estimated lung dose (13%). The proposed methodology provides a more accurate LMD and an estimate of its precision, which will improve treatment and retreatment planning for 90 Y radioembolizations.

journal_name

Med Phys

journal_title

Medical physics

authors

Lopez B,Mahvash A,Lam MGEH,Kappadath SC

doi

10.1002/mp.13575

subject

Has Abstract

pub_date

2019-09-01 00:00:00

pages

3929-3940

issue

9

eissn

0094-2405

issn

2473-4209

journal_volume

46

pub_type

杂志文章
  • Study of the secondary neutral radiation in proton therapy: toward an indirect in vivo dosimetry.

    abstract:PURPOSE:Secondary particles produced in the collision of protons with beam modifiers are of concern in proton therapy. Nevertheless, secondary radiation can provide information on the dosimetric parameters through its dependency on the modulating accessories (range shifter and range modulating wheel). Relatively little...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4765049

    authors: Carnicer A,Letellier V,Rucka G,Angellier G,Sauerwein W,Herault J

    更新日期:2012-12-01 00:00:00

  • Evaluating IMRT and VMAT dose accuracy: practical examples of failure to detect systematic errors when applying a commonly used metric and action levels.

    abstract:PURPOSE:This study (1) examines a variety of real-world cases where systematic errors were not detected by widely accepted methods for IMRT/VMAT dosimetric accuracy evaluation, and (2) drills-down to identify failure modes and their corresponding means for detection, diagnosis, and mitigation. The primary goal of detai...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4826166

    authors: Nelms BE,Chan MF,Jarry G,Lemire M,Lowden J,Hampton C,Feygelman V

    更新日期:2013-11-01 00:00:00

  • Mitigating differential baseline shifts in locally advanced lung cancer patients using an average anatomy model.

    abstract:BACKGROUND AND PURPOSE:Differential baseline shifts between primary tumor and involved lymph nodes in locally advanced lung cancer patients compromise the accuracy of radiotherapy. The purpose of this study was to evaluate the performance of an average anatomy model (AAM) derived from repeat imaging and deformable regi...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12271

    authors: Abdoli M,Van Kranen SR,Stankovic U,Rossi MMG,Belderbos JSA,Sonke JJ

    更新日期:2017-07-01 00:00:00

  • In vivo surface dosimetry with a scintillating fiber dosimeter in preclinical image-guided radiotherapy.

    abstract:PURPOSE:New preclinical image-guided irradiators and treatment planning systems represent a huge progress in radiobiology. Nevertheless, quality control of preclinical treatments is not as advanced as in clinical radiotherapy and in vivo dosimetry is less developed. In this study, we evaluate the use of a scintillating...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13903

    authors: Le Deroff C,Pérès EA,Ledoux X,Toutain J,Frelin-Labalme AM

    更新日期:2020-01-01 00:00:00

  • Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging.

    abstract:PURPOSE:The spatial resolution of iterative reconstruction (IR) in computed tomography (CT) is contrast- and noise-dependent because of the nonlinear regularization. Due to the severe noise contamination, it is challenging to perform precise spatial-resolution measurements at very low-contrast levels. The purpose of th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4916802

    authors: Yu L,Vrieze TJ,Leng S,Fletcher JG,McCollough CH

    更新日期:2015-05-01 00:00:00

  • Beam modeling and verification of a photon beam multisource model.

    abstract::Dose calculations for treatment planning of photon beam radiotherapy require a model of the beam to drive the dose calculation models. The beam shaping process involves scattering and filtering that yield radiation components which vary with collimator settings. The necessity to model these components has motivated th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1898485

    authors: Ahnesjö A,Weber L,Murman A,Saxner M,Thorslund I,Traneus E

    更新日期:2005-06-01 00:00:00

  • Investigating the feasibility of photon-counting K-edge imaging at high x-ray fluxes using nonlinearity corrections.

    abstract:PURPOSE:Pulse pileup occurring at high x-ray fluxes can severely degrade the energy resolution provided by a photon counting detector, which can represent a problem in spectroscopic CT when performing quantitative material discrimination tasks. As the effects of pileup can be most easily seen as a degradation of a dete...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4820535

    authors: Rink K,Oelfke U,Fiederle M,Zuber M,Koenig T

    更新日期:2013-10-01 00:00:00

  • Neutron and gamma spectroscopy for clinical dosimetry.

    abstract::Neutron and photon energy spectra were measured in different depths and lateral positions in tissue-equivalent phantoms exposed to a therapeutical 14.1 MeV neutron beam. Intensity weighted mass energy absorption coefficients were derived for application in ionization chamber dosimetry in mixed neutron-gamma fields. Th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594744

    authors: Schmidt R,Magiera E,Scobel W

    更新日期:1980-09-01 00:00:00

  • Technical and dosimetric realization of in vivo x-ray microbeam irradiations at the Munich Compact Light Source.

    abstract:PURPOSE:X-ray microbeam radiation therapy is a preclinical concept for tumor treatment promising tissue sparing and enhanced tumor control. With its spatially separated, periodic micrometer-sized pattern, this method requires a high dose rate and a collimated beam typically available at large synchrotron radiation faci...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14433

    authors: Burger K,Urban T,Dombrowsky AC,Dierolf M,Günther B,Bartzsch S,Achterhold K,Combs SE,Schmid TE,Wilkens JJ,Pfeiffer F

    更新日期:2020-10-01 00:00:00

  • Quantitative ionization chamber alignment to a water surface: Theory and simulation.

    abstract:PURPOSE:To examine the response properties of cylindrical cavity ionization chambers (ICs) in the depth-ionization buildup region so as to obtain a robust chamber-signal - based method for definitive water surface identification, hence absolute ionization chamber depth localization. METHOD & MATERIALS:An analytical mo...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12319

    authors: Siebers JV,Ververs JD,Tessier F

    更新日期:2017-07-01 00:00:00

  • A new open-source GPU-based microscopic Monte Carlo simulation tool for the calculations of DNA damages caused by ionizing radiation - Part II: sensitivity and uncertainty analysis.

    abstract:PURPOSE:Calculations of deoxyribonucleic acid (DNA) damages involve many parameters in the computation process. As these parameters are often subject to uncertainties, it is of central importance to comprehensively quantify their impacts on DNA single-strand break (SSB) and double-strand break (DSB) yields. This has be...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14036

    authors: Lai Y,Tsai MY,Tian Z,Qin N,Yan C,Hung SH,Chi Y,Jia X

    更新日期:2020-04-01 00:00:00

  • A filtering method for signal equalization in region-of-interest fluoroscopy.

    abstract::A method to significantly reduce the exposure area product in fluoroscopy using a pre-patient region-of-interest (ROI) attenuator is presented. The attenuator has a thin central region and a gradually increasing thickness away from the center. It is shown that the unwanted brightening artifact caused by the attenuator...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1470205

    authors: Robert N,Komljenovic PT,Rowlands JA

    更新日期:2002-05-01 00:00:00

  • Predicting high-intensity focused ultrasound thalamotomy lesions using 2D magnetic resonance thermometry and 3D Gaussian modeling.

    abstract:PURPOSE:To develop a method of using two-dimensional (2D) magnetic resonance thermometry, and three-dimensional (3D) Gaussian modeling to predict the volume, shape, and location of 1 day postoperative T1w high-intensity focused ultrasound lesions in medication refractory tremor patients; thereby facilitating a better c...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13868

    authors: Seasons GM,Mazerolle EL,Sankar T,Martino D,Kiss ZHT,Pichardo S,Pike GB

    更新日期:2019-12-01 00:00:00

  • Fluid-filled dynamic bowtie filter: Description and comparison with other modulators.

    abstract:PURPOSE:A dynamic bowtie filter can modulate flux along both fan and view angles for reduced patient dose, scatter, and required photon flux, which is especially important for photon counting detectors (PCDs). Among the proposed dynamic bowtie designs, the piecewise-linear attenuator (Hsieh and Pelc, Med Phys. 2013;40:...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13272

    authors: Shunhavanich P,Hsieh SS,Pelc NJ

    更新日期:2019-01-01 00:00:00

  • Anisotropy functions for 103Pd, 125I, and 192Ir interstitial brachytherapy sources.

    abstract::Anisotropy of dose distributions around 103Pd, 125I, and 192Ir sources for interstitial brachytherapy was examined. Dose rates around 125I models 6702 and 6711 and 192Ir sources were measured using lithium fluoride thermoluminescence dosimeters (LiF TLDs) in a water-equivalent, solid phantom. From these measured data ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597110

    authors: Nath R,Meigooni AS,Muench P,Melillo A

    更新日期:1993-09-01 00:00:00

  • Sci-Fri PM: Planning-04: Dose escalation study using anatomy-based aperture IMRT and SPECT perfusion images for lung cancer.

    abstract::In the case of non-small cell lung cancer, doses typically prescribed (60-66 Gy) are not sufficient to ensure a satisfactory tumor control probability. Dose escalation needs to be realized, but dose to organs at risk (OARs) must be kept under widely accepted clinical thresholds. Also, lung functionality is not homogen...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2965976

    authors: St-Hilaire J,Lavoie C,Beaulieu F,Dagnault A,Morin F,Gingras L,Tremblay D,Beaulieu L

    更新日期:2008-07-01 00:00:00

  • Smoothing intensity-modulated treatment delivery under hardware constraints.

    abstract::A method is developed to smooth intensity-modulated beam delivery while considering hardware constraints. The method uses matrix algebra to model MLC leaf motion and hardware constraints such as interleaf digitization, synchronization requirements, and intersegment field abutment to deliver intensity modulated beam se...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1521121

    authors: Ma L

    更新日期:2002-12-01 00:00:00

  • Technical Note: Comparison of peripheral patient dose from MR-guided 60 Co therapy and 6 MV linear accelerator IGRT.

    abstract:PURPOSE:The use of X-ray imaging in radiation therapy can give a substantial dose to the patient. A Cobalt machine combined with an magnetic resonance imaging (MRI) was recently introduced to clinical work. One positive aspect of using non-ionizing imaging devices is the reduction of the patient exposure. The purpose o...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12293

    authors: Hauri P,Hälg RA,Schneider U

    更新日期:2017-07-01 00:00:00

  • Evaluations of magnetic resonance imaging parameters with simple phantoms.

    abstract::Imaging parameters associated with a clinical magnetic resonance imaging (MRI) system, evaluated with the use of simple and inexpensive phantoms, are described in evaluation of a clinical MRI system. Images, obtained with a Diasonics 0.35-T MRI system using an elliptical whole-body radio frequency coil, are presented ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595894

    authors: Brateman L,Jennings LW,Nunnally RL,Vaughan JT

    更新日期:1986-07-01 00:00:00

  • Modeling the dosimetry of organ-at-risk in head and neck IMRT planning: an intertechnique and interinstitutional study.

    abstract:PURPOSE:To build a statistical model to quantitatively correlate the anatomic features of structures and the corresponding dose-volume histogram (DVH) of head and neck (HN) Tomotherapy (Tomo) plans. To study if the model built upon one intensity modulated radiation therapy (IMRT) technique (such as conventional Linac) ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4828788

    authors: Lian J,Yuan L,Ge Y,Chera BS,Yoo DP,Chang S,Yin F,Wu QJ

    更新日期:2013-12-01 00:00:00

  • Automated matching and segmentation of lymphoma on serial CT examinations.

    abstract::In patients with lymphoma, identification and quantification of the tumor extent on serial CT examinations is critical for assessing tumor response to therapy. In this paper, we present a computer method to automatically match and segment lymphomas in follow-up CT images. The method requires that target lymph nodes in...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2404617

    authors: Yan J,Zhao B,Curran S,Zelenetz A,Schwartz LH

    更新日期:2007-01-01 00:00:00

  • Dosimetric problems at low monitor unit settings for scanned and scattering foil electron beams.

    abstract::Electron beam dosimetry at low monitor unit (MU) settings is important for dosimetric applications. Dose linearity, beam flatness, and beam energies were studied at low MU settings with various dose rates for different types of linear accelerators. It is observed that for the scattering foil units, the dose/MU is a sm...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597328

    authors: Das IJ,Harrington JC,Akber SF,Tomer AF,Murray JC,Cheng CW

    更新日期:1994-06-01 00:00:00

  • A novel approach to neutron dosimetry.

    abstract:PURPOSE:Having been overlooked for many years, research is now starting to take into account the directional distribution of neutron workplace fields. Existing neutron dosimetry instrumentation does not account for this directional distribution, resulting in conservative estimates of dose in neutron workplace fields (b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4964456

    authors: Balmer MJ,Gamage KA,Taylor GC

    更新日期:2016-11-01 00:00:00

  • Material-specific analysis using coherent-scatter imaging.

    abstract::Coherent-scatter computed tomography (CSCT) is a novel imaging method we are developing to produce cross-sectional images based on the low-angle (<10 degrees) scatter properties of tissue. At diagnostic energies, this scatter is primarily coherent with properties dependent upon the molecular structure of the scatterer...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1493216

    authors: Batchelar DL,Cunningham IA

    更新日期:2002-08-01 00:00:00

  • Feature selection using factor analysis for Alzheimer's diagnosis using 18F-FDG PET images.

    abstract:PURPOSE:This article presents a computer-aided diagnosis technique for improving the accuracy of the early diagnosis of Alzheimer's disease (AD). Two hundred and ten 18F-FDG PET images from the ADNI initiative [52 normal controls (NC), 114 mild cognitive impairment (MCI), and 53 AD subjects] are studied. METHODS:The p...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3488894

    authors: Salas-Gonzalez D,Górriz JM,Ramírez J,Illán IA,López M,Segovia F,Chaves R,Padilla P,Puntonet CG

    更新日期:2010-11-01 00:00:00

  • An optimized ultrasound detector for photoacoustic breast tomography.

    abstract:PURPOSE:Photoacoustic imaging has proven to be able to detect vascularization-driven optical absorption contrast associated with tumors. In order to detect breast tumors located a few centimeter deep in tissue, a sensitive ultrasound detector is of crucial importance for photoacoustic mammography. Further, because the ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4792462

    authors: Xia W,Piras D,van Hespen JC,van Veldhoven S,Prins C,van Leeuwen TG,Steenbergen W,Manohar S

    更新日期:2013-03-01 00:00:00

  • Sci-Thurs PM: Delivery-03: Optical imaging of microscopic radiation dose gradients using a digital microscope.

    abstract::Using superior localization and immobilization methods, stereotactic radiosurgery is capable of delivering spheres of dose as small as a few millimetres in diameter to intracranial targets. For targets abutting critical structures, the most conformal treatments minimize adverse radiation side effects and it is importa...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2965910

    authors: Keller BM,Peressotti C,Pignol JP

    更新日期:2008-07-01 00:00:00

  • Low dose fraction behavior of high sensitivity radiochromic film.

    abstract::A high sensitivity (HS) model of radiochromic film is receiving increasing use. The film's linear sensitometric response in the range of 0.5-40 Gy would make this film an ideal candidate for complex dosimetry applications that require tissue equivalence. This study investigates the potential use for clinical dosimetry...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1883565

    authors: Hirata EY,Cunningham C,Micka JA,Keller H,Kissick MW,DeWerd LA

    更新日期:2005-04-01 00:00:00

  • Technical Note: Spot characteristic stability for proton pencil beam scanning.

    abstract:PURPOSE:The spot characteristics for proton pencil beam scanning (PBS) were measured and analyzed over a 16 month period, which included one major site configuration update and six cyclotron interventions. The results provide a reference to establish the quality assurance (QA) frequency and tolerance for proton pencil ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4939663

    authors: Chen CC,Chang C,Moyers MF,Gao M,Mah D

    更新日期:2016-02-01 00:00:00

  • Eye tracking and gating system for proton therapy of orbital tumors.

    abstract:PURPOSE:A new motion-based gated proton therapy for the treatment of orbital tumors using real-time eye-tracking system was designed and evaluated. METHODS:We developed our system by image-pattern matching, using a normalized cross-correlation technique with LabVIEW 8.6 and Vision Assistant 8.6 (National Instruments, ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4729708

    authors: Shin D,Yoo SH,Moon SH,Yoon M,Lee SB,Park SY

    更新日期:2012-07-01 00:00:00