Design and evaluation of quantitative MRI phantoms to mimic the simultaneous presence of fat, iron, and fibrosis in the liver.

Abstract:

PURPOSE:To design, construct, and evaluate quantitative MR phantoms that mimic MRI signals from the liver with simultaneous control of three parameters: proton-density fat fraction (PDFF), R 2 ∗ , and T1 . These parameters are established biomarkers of hepatic steatosis, iron overload, and fibrosis/inflammation, respectively, which can occur simultaneously in the liver. METHODS:Phantoms including multiple vials were constructed. Peanut oil was used to modulate PDFF, MnCl2 and iron microspheres were used to modulate R 2 ∗ , and NiCl2 was used to modulate the T1 of water (T1,water ). Phantoms were evaluated at both 1.5 T and 3 T using stimulated-echo acquisition-mode MRS and chemical shift-encoded MRI. Stimulated-echo acquisition-mode MRS data were processed to estimate T1,water , T1,fat , R 2 , water ∗ , and R 2 , fat ∗ for each vial. Chemical shift-encoded MRI data were processed to generate PDFF and R 2 ∗ maps, and measurements were obtained in each vial. Measurements were evaluated using linear regression and Bland-Altman analysis. RESULTS:High-quality PDFF and R 2 ∗ maps were obtained with homogeneous values throughout each vial. High correlation was observed between imaging PDFF with target PDFF (slope = 0.94-0.97, R2 = 0.994-0.997) and imaging R 2 ∗ with target R 2 ∗ (slope = 0.84-0.88, R2 = 0.935-0.943) at both 1.5 T and 3 T. The values of R 2 , fat ∗ and R 2 , water ∗ were highly correlated with slope close to 1.0 at both 1.5 T (slope = 0.90, R2 = 0.988) and 3 T (slope = 0.99, R2 = 0.959), similar to the behavior observed in vivo. The value of T1,water (500-1200 ms) was controlled with varying NiCl2 concentration, while T1,fat (300 ms) was independent of NiCl2 concentration. CONCLUSION:Novel quantitative MRI phantoms that mimic the simultaneous presence of fat, iron, and fibrosis in the liver were successfully developed and validated.

journal_name

Magn Reson Med

authors

Zhao R,Hamilton G,Brittain JH,Reeder SB,Hernando D

doi

10.1002/mrm.28452

subject

Has Abstract

pub_date

2021-02-01 00:00:00

pages

734-747

issue

2

eissn

0740-3194

issn

1522-2594

journal_volume

85

pub_type

杂志文章
  • Field camera measurements of gradient and shim impulse responses using frequency sweeps.

    abstract:PURPOSE:Applications of dynamic shimming require high field fidelity, and characterizing the shim field dynamics is therefore necessary. Modeling the system as linear and time-invariant, the purpose of this work was to measure the impulse response function with optimal sensitivity. THEORY AND METHODS:Frequency-swept p...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.24934

    authors: Vannesjo SJ,Dietrich BE,Pavan M,Brunner DO,Wilm BJ,Barmet C,Pruessmann KP

    更新日期:2014-08-01 00:00:00

  • Continuously moving table MRI with golden angle radial sampling.

    abstract:PURPOSE:Continuously moving table (CMT) MRI is a high throughput technique that has multiple applications in whole-body imaging. In this work, CMT MRI based on golden angle (GA, 111.246° azimuthal step) radial sampling is developed at 3 Tesla (T), with the goal of increased flexibility in image reconstruction using arb...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.25531

    authors: Sengupta S,Smith DS,Welch EB

    更新日期:2015-12-01 00:00:00

  • MR microimaging of articular cartilage and contrast enhancement by manganese ions.

    abstract::Saline solutions of manganese ions (Mn2+) were used as articular contrast agents in magnetic resonance microscopy (9.4 T) of cartilage of chicken femoral condyles and pig temporomandibular joints. The diffusion of Mn2+ from the articular surface into the cartilage matrix led to a strong contrast enhancement in the car...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910240114

    authors: Kusaka Y,Gründer W,Rumpel H,Dannhauer KH,Gersonde K

    更新日期:1992-03-01 00:00:00

  • Spatially resolved assessment of serotonin-induced bronchoconstrictive responses in the rat lung using 3He ventilation MRI under spontaneous breathing conditions.

    abstract::The effect on lung ventilation of bronchoconstriction induced by serotonin (intravenous injection of 50 microg/kg of serotonin) was imaged using a hyperpolarized (3)He MR ventilation protocol in spontaneously breathing rats. Lung function maps assessing airflow obstruction, a key feature in clinical pneumology, were d...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.22391

    authors: Mosbah K,Stupar V,Berthezène Y,Beckmann N,Crémillieux Y

    更新日期:2010-06-01 00:00:00

  • Cerebral T1rho relaxation time increases immediately upon global ischemia in the rat independently of blood glucose and anoxic depolarization.

    abstract::Time-dependent changes of T1 in the rotating frame (T1rho), diffusion, T2, and magnetization transfer contrast on cardiac arrest-induced global ischemia in rat were investigated. T1rho, as acquired with spin lock amplitudes >0.6 G, started to increase 10-20 sec after cardiac arrest followed by an increase within 3-4 m...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1228

    authors: Kettunen MI,Gröhn OH,Penttonen M,Kauppinen RA

    更新日期:2001-09-01 00:00:00

  • Rapid radiofrequency field mapping in vivo using single-shot STEAM MRI.

    abstract::Higher field strengths entail less homogeneous RF fields. This may influence quantitative MRI and MRS. A method for rapidly mapping the RF field in the human head with minimal distortion was developed on the basis of a single-shot stimulated echo acquisition mode (STEAM) sequence. The flip angle of the second RF pulse...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.21676

    authors: Helms G,Finsterbusch J,Weiskopf N,Dechent P

    更新日期:2008-09-01 00:00:00

  • Robust motion correction in CEST imaging exploiting low-rank approximation of the z-spectrum.

    abstract:PURPOSE:To introduce and evaluate an image registration technique for robust quantification of CEST acquisitions corrupted by motion. METHODS:The proposed iterative algorithm exploits a low-rank approximation of the z-spectrum (LRAZ), to gradually separate the contrast variation due to saturation at different off-reso...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.27206

    authors: Wech T,Köstler H

    更新日期:2018-11-01 00:00:00

  • MRI visualization of rodent liver structure and peritoneal adhesion with dialyzate enhancement.

    abstract::This study investigated the use of peritoneal dialysis fluid (dialyzate) as a MR contrast agent to visualize the liver structure and peritoneal adhesion in rats at 7 T. Intraperitoneal injection of dialyzate (approximately 0.1 ml/g) yielded excellent and consistent intraperitoneal enhancement that delineated the liver...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.21506

    authors: Cheung JS,Guo H,Leung JC,Man K,Lai KN,Wu EX

    更新日期:2008-05-01 00:00:00

  • Proton NMR spectroscopy of solvent-saturable resonances: a new approach to study pH effects in situ.

    abstract::It is shown that the effect of pH changes can be measured in proton NMR spectra through the pH sensitivity of the signal intensities of metabolite protons exchanging with water. To observe this phenomenon, pulse sequences must be used that can sensitively observe these exchangeable protons under physiological conditio...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910400105

    authors: Mori S,Eleff SM,Pilatus U,Mori N,van Zijl PC

    更新日期:1998-07-01 00:00:00

  • Accurate quantification of in vivo 31P NMR signals using the variable projection method and prior knowledge.

    abstract::Free induction decay signals are analyzed by fitting a model function directly in the time domain. No starting values are needed for linear model parameters, and omission of corrupted data points poses no problems. A significant gain of accuracy is achieved by imposing prior knowledge about the model parameters. ...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910060111

    authors: van der Veen JW,de Beer R,Luyten PR,van Ormondt D

    更新日期:1988-01-01 00:00:00

  • Fast 1H spectroscopic imaging using a multi-element head-coil array.

    abstract::Fast proton magnetic resonance spectroscopic imaging (MRSI) using a multi-element head-coil array is examined with respect to three aspects: the coil design, the use of an appropriate signal combination method, and the design of the MRSI pulse sequence itself. An eight-element head-coil array has been developed to inc...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910400204

    authors: Schäffter T,Börnert P,Leussler C,Carlsen IC,Leibfritz D

    更新日期:1998-08-01 00:00:00

  • Investigation of human cervical and upper thoracic spinal cord motion: implications for imaging spinal cord structure and function.

    abstract::Spinal cord (SC) motion is thought to be the dominant source of error in current diffusion and spinal functional MRI (fMRI) methods. However, until now, such motion has not been well characterized in three dimensions. While previous studies have predominantly examined motion in the superior/inferior (S/I) direction, t...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.21260

    authors: Figley CR,Stroman PW

    更新日期:2007-07-01 00:00:00

  • Fast acquisition-weighted three-dimensional proton MR spectroscopic imaging of the human prostate.

    abstract::The clinical application of 3D proton spectroscopic imaging (3D SI) of the human prostate requires a robust suppression of periprostatic lipid signal contamination, minimal intervoxel signal contamination, and the shortest possible measurement time. In this work, a weighted elliptical sampling of k-space, combined wit...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.20103

    authors: Scheenen TW,Klomp DW,Röll SA,Fütterer JJ,Barentsz JO,Heerschap A

    更新日期:2004-07-01 00:00:00

  • Myocardial effective transverse relaxation time T2* Correlates with left ventricular wall thickness: A 7.0 T MRI study.

    abstract:PURPOSE:Myocardial effective relaxation time T2* is commonly regarded as a surrogate for myocardial tissue oxygenation. However, it is legitimate to assume that there are multiple factors that influence T2*. To this end, this study investigates the relationship between T2* and cardiac macromorphology given by left vent...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.26312

    authors: Huelnhagen T,Hezel F,Serradas Duarte T,Pohlmann A,Oezerdem C,Flemming B,Seeliger E,Prothmann M,Schulz-Menger J,Niendorf T

    更新日期:2017-06-01 00:00:00

  • Free breathing three-dimensional late gadolinium enhancement cardiovascular magnetic resonance using outer volume suppressed projection navigators.

    abstract:PURPOSE:To develop a three-dimensional, free-breathing, late gadolinium enhancement (3D FB-LGE) cardiovascular magnetic resonance (CMR) technique, and to compare it with clinically used two-dimensional breath-hold LGE (2D BH-LGE). METHODS:The proposed 3D FB-LGE method consisted of inversion preparation, inversion dela...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.26234

    authors: Menon RG,Miller GW,Jeudy J,Rajagopalan S,Shin T

    更新日期:2017-04-01 00:00:00

  • Dynamic 13C NMR analysis of pyruvate and lactate oxidation in the in vivo canine myocardium: evidence of reduced utilization with increased work.

    abstract::In this work, substrate selection was monitored in the left ventricle of the canine myocardium by following pyruvate and lactate oxidation under in vivo conditions at basal and elevated workloads. These studies were conducted in the open chest model using dynamic 13C NMR techniques in the presence and absence of dichl...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910380608

    authors: Rath DP,Zhu H,Tong X,Jiang Z,Hamlin RL,Robitaille PM

    更新日期:1997-12-01 00:00:00

  • More accurate estimation of diffusion tensor parameters using diffusion Kurtosis imaging.

    abstract::With diffusion tensor imaging, the diffusion of water molecules through brain structures is quantified by parameters, which are estimated assuming monoexponential diffusion-weighted signal attenuation. The estimated diffusion parameters, however, depend on the diffusion weighting strength, the b-value, which hampers t...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.22603

    authors: Veraart J,Poot DH,Van Hecke W,Blockx I,Van der Linden A,Verhoye M,Sijbers J

    更新日期:2011-01-01 00:00:00

  • CINE turbo spin echo imaging.

    abstract::High-resolution turbo spin echo (TSE) images have demonstrated important details of carotid artery morphology; however, it is evident that pulsatile blood and wall motion related to the cardiac cycle are still significant sources of image degradation. Although ECG gating can reduce artifacts due to cardiac-induced pul...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.22909

    authors: Mendes J,Parker DL,Hulet J,Treiman GS,Kim SE

    更新日期:2011-11-01 00:00:00

  • Prospective multiaxial motion correction for fMRI.

    abstract::Corruption of the image time series due to interimage head motion limits the clinical utility of functional MRI. This paper presents a method for real-time prospective correction of rotation and translation in all six degrees of rigid body motion. By incorporating an orbital navigator (ONAV) echo for each of the sagit...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/(sici)1522-2594(200003)43:3<459::aid-mrm19

    authors: Ward HA,Riederer SJ,Grimm RC,Ehman RL,Felmlee JP,Jack CR Jr

    更新日期:2000-03-01 00:00:00

  • A 1H-NMR method for determining temperature in cell culture perfusion systems.

    abstract::This report describes a noninvasive 1H-NMR method for measuring absolute temperatures (+/- 0.2 degrees C) in biological samples and, in particular, in cell culture perfusion systems, utilizing the linear temperature dependence of the water chemical shift relative to the temperature-independent shift of one of the comp...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910290120

    authors: Lutz NW,Kuesel AC,Hull WE

    更新日期:1993-01-01 00:00:00

  • Quantification of atherosclerotic plaque components using in vivo MRI and supervised classifiers.

    abstract::In this work we aimed to study the possibility of using supervised classifiers to quantify the main components of carotid atherosclerotic plaque in vivo on the basis of multisequence MRI data. MRI data consisting of five MR weightings were obtained from 25 symptomatic subjects. Histological micrographs of endarterecto...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.20828

    authors: Hofman JM,Branderhorst WJ,ten Eikelder HM,Cappendijk VC,Heeneman S,Kooi ME,Hilbers PA,ter Haar Romeny BM

    更新日期:2006-04-01 00:00:00

  • Phase-updated regularized SENSE for navigator-free multishot diffusion imaging.

    abstract:PURPOSE:Either SENSE+CG or POCS-ICE methods can be used to correct for motion-induced phase errors in navigator-free multishot diffusion imaging. SENSE+CG has the advantage of a fast convergence, however, occasionally the convergence can be unstable, thus degrading the image quality. POCS-ICE has a stable convergence a...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.26361

    authors: Hu Z,Ma X,Truong TK,Song AW,Guo H

    更新日期:2017-07-01 00:00:00

  • A subspace-based coil combination method for phased-array magnetic resonance imaging.

    abstract:PURPOSE:Coil-by-coil reconstruction methods are followed by coil combination to obtain a single image representing a spin density map. Typical coil combination methods, such as square-root sum-of-squares and adaptive coil combining, yield images that exhibit spatially varying modulation of image intensity. Existing pra...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.25664

    authors: Gol Gungor D,Potter LC

    更新日期:2016-02-01 00:00:00

  • Global and peak local specific absorption rate control on parallel transmit systems using k-means SAR compression model.

    abstract:PURPOSE:To improve the specific absorption rate (SAR) compression model capability in parallel transmission (pTx) MRI systems. METHODS:A k-means clustering method is proposed to group voxels with similar SAR behaviors in the scanned object, providing a controlled upper-bounded estimation of peak local SARs. This k-mea...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.28456

    authors: Mao X,Bari S,Love DJ,Rispoli JV

    更新日期:2021-02-01 00:00:00

  • Analysis of the evolution of focal cerebral ischemia in the rat using the eigenimage filter.

    abstract::The eigenimage filter was used to evaluate the results of a MRI study of cerebral ischemia in a rat model. This linear filter segments a desired feature in an image sequence from other features which may interfere with its observation. The animals were imaged temporally, after occlusion of the middle cerebral artery, ...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910260207

    authors: Peck DJ,Spickler EM,Knight RA,Hearshen DO,Windham JP

    更新日期:1992-08-01 00:00:00

  • Iterative reweighted linear least squares for accurate, fast, and robust estimation of diffusion magnetic resonance parameters.

    abstract:PURPOSE:Diffusion-weighted magnetic resonance imaging suffers from physiological noise, such as artifacts caused by motion or system instabilities. Therefore, there is a need for robust diffusion parameter estimation techniques. In the past, several techniques have been proposed, including RESTORE and iRESTORE (Chang e...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.25351

    authors: Collier Q,Veraart J,Jeurissen B,den Dekker AJ,Sijbers J

    更新日期:2015-06-01 00:00:00

  • Arterial spin labeling perfusion fMRI with very low task frequency.

    abstract::Functional magnetic resonance imaging (fMRI) has become the most widely used modality for visualizing regional brain activation in response to sensorimotor or cognitive tasks. While the majority of fMRI studies have used blood oxygenation level-dependent (BOLD) contrast as a marker for neural activation, baseline drif...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.10437

    authors: Wang J,Aguirre GK,Kimberg DY,Roc AC,Li L,Detre JA

    更新日期:2003-05-01 00:00:00

  • Transverse relaxivity of iron oxide nanocrystals clustered in nanoemulsions: Experiment and theory.

    abstract:PURPOSE:To compare experimental transverse relaxivities of iron oxide nanocrystals (IONC) as a function of clustering and magnetic field strength with different theoretical model predictions. THEORY AND METHODS:Well-defined IONC clusters in nanoemulsions (NEs) of which both size and IONC loading could be judiciously t...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.25465

    authors: Hak S,Goa PE,Stenmark S,Bjerkholt FF,Haraldseth O

    更新日期:2015-09-01 00:00:00

  • MR angiography by selective inversion recovery.

    abstract::A modified inversion-recovery sequence is introduced which performs subtraction angiography by varying time-of-flight effects of blood flowing into an imaged slab. The selective 180 degrees excitation inverts different regions between measurements to isolate arterial and/or venous blood. On normal human subjects, high...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910040214

    authors: Nishimura DG,Macovski A,Pauly JM,Conolly SM

    更新日期:1987-02-01 00:00:00

  • In vivo validation of MR pulse pressure measurement in an aortic flow model: preliminary results.

    abstract::MR imaging experiments were conducted to investigate the feasibility of estimating vascular pulse pressure waveforms from measurements of blood flow rates and vessel cross-sectional area. Blood flow waveforms were measured in the aorta's of three 25-30-kg pigs at multiple imaging sections using phase-contrast velocity...

    journal_title:Magnetic resonance in medicine

    pub_type: 杂志文章

    doi:10.1002/mrm.1910380209

    authors: Urchuk SN,Fremes SE,Plewes DB

    更新日期:1997-08-01 00:00:00