Intercomparison of methods for image quality characterization. I. Modulation transfer function.

Abstract:

:The modulation transfer function (MTF) and the noise power spectrum (NPS) are widely recognized as the most relevant metrics of resolution and noise performance in radiographic imaging. These quantities have commonly been measured using various techniques, the specifics of which can have a bearing on the accuracy of the results. As a part of a study aimed at comparing the relative performance of different techniques, in this paper we report on a comparison of two established MTF measurement techniques: one using a slit test device [Dobbins et al., Med. Phys. 22, 1581-1593 (1995)] and another using a translucent edge test device [Samei et al., Med. Phys. 25, 102-113 (1998)], with one another and with a third technique using an opaque edge test device recommended by a new international standard (IEC 62220-1, 2003). The study further aimed to substantiate the influence of various acquisition and processing parameters on the estimated MTF. The slit test device was made of 2 mm thick Pb slabs with a 12.5 microm opening. The translucent edge test device was made of a laminated and polished Pt(0.9)Ir(0.1). alloy foil of 0.1 mm thickness. The opaque edge test device was made of a 2 mm thick W slab. All test devices were imaged on a representative indirect flat-panel digital radiographic system using three published beam qualities: 70 kV with 0.5 mm Cu filtration, 70 kV with 19 mm Al filtration, and 74 kV with 21 mm Al filtration (IEC-RQA5). The latter technique was also evaluated in conjunction with two external beam-limiting apertures (per IEC 62220-1), and with the tube collimator limiting the beam to the same area achieved with the apertures. The presampled MTFs were deduced from the acquired images by Fourier analysis techniques, and the results analyzed for relative values and the influence of impacting parameters. The findings indicated that the measurement technique has a notable impact on the resulting MTF estimate, with estimates from the overall IEC method 4.0% +/- 0.2% lower than that of Dobbins et al. and 0.7% +/- 0.4% higher than that of Samei et al. averaged over the zero to cutoff frequency range. Over the same frequency range, keeping beam quality and limitation constant, the average MTF estimate obtained with the edge techniques differed by up to 5.2% +/- 0.2% from that of the slit, with the opaque edge providing lower MTF estimates at lower frequencies than those obtained with the translucent edge or slit. The beam quality impacted the average estimated MTF by as much as 3.7% +/- 0.9% while the use of beam limiting devices alone increased the average estimated MTF by as much as 7.0% +/- 0.9%. While the slit method is inherently very sensitive to misalignment, both edge techniques were found to tolerate misalignments by as much as 6 cm. The results suggest the use of the opaque edge test device and the tube internal collimator for beam limitation in order to achieve an MTF result most reflective of the overall performance of the imaging system and least susceptible to misalignment and scattered radiation. Careful attention to influencing factors is warranted to achieve accurate results.

journal_name

Med Phys

journal_title

Medical physics

authors

Samei E,Ranger NT,Dobbins JT 3rd,Chen Y

doi

10.1118/1.2188816

subject

Has Abstract

pub_date

2006-05-01 00:00:00

pages

1454-65

issue

5

eissn

0094-2405

issn

2473-4209

journal_volume

33

pub_type

杂志文章
  • Dosimetric characteristics of the Leipzig surface applicators used in the high dose rate brachy radiotherapy.

    abstract::The nucletron Leipzig applicator is designed for (HDR) 192Ir brachy radiotherapy of surface lesions. The dosimetric characteristics of this applicator were investigated using simulation method based on Monte Carlo N-particle (MCNP) code and phantom measurements. The simulation method was validated by comparing calcula...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1812609

    authors: Niu H,Hsi WC,Chu JC,Kirk MC,Kouwenhoven E

    更新日期:2004-12-01 00:00:00

  • SU-E-T-264: New Concrete Designed and Evaluation for Megavoltage X Radiotherapy Facilities (CONTEK-RFH2).

    abstract:PURPOSE:The most common material for shielding is concrete, which can be made using various materials of different densities as aggregates. New techniques in radiotherapy, as IMRT and VMAT, require more monitor units and it is important to develop specifically designed shielding materials. METHODS:Arraela S.L. has dev...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735331

    authors: Mera M,Pereira L,Mera M,Pereira L,Meilán E,Moral FD,Teijeiro A,Salgado M,Andrade B,Gomez F,Fuentes-Vázquez V,Caruncho J,Medina A

    更新日期:2012-06-01 00:00:00

  • Heating patterns of microwave applicators in inhomogeneous arm and thigh phantoms.

    abstract::Heating rate (HR) patterns in cylindrical structures were studied with inhomogeneous limb phantoms. These phantoms, arm and thigh models consisting of fat, bone, and muscle material, were heated with Clini-Therm L, M, and MS applicators at 915 MHz. The thigh model is 18 cm in diameter with 2.65-cm-thick fat on the out...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596627

    authors: Chou CK,McDougall JA,Chan KW,Luk KH

    更新日期:1991-11-01 00:00:00

  • Classifier performance prediction for computer-aided diagnosis using a limited dataset.

    abstract::In a practical classifier design problem, the true population is generally unknown and the available sample is finite-sized. A common approach is to use a resampling technique to estimate the performance of the classifier that will be trained with the available sample. We conducted a Monte Carlo simulation study to co...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2868757

    authors: Sahiner B,Chan HP,Hadjiiski L

    更新日期:2008-04-01 00:00:00

  • TU-E-213AB-01: JMPSLC Has Morphed into Medical Physics Licensure and Regulatory Recognition Subcommittee.

    abstract::Based on direction from AAM Board of Directors, JMPLSC has redefined its scope and name. On November 4, 2011 the subcommittee renamed itself the Medical Physics Licensure and Regulatory Recognition Subcommittee. Its newly defined charge is to promote the protection of the public through the recognition of the professi...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735947

    authors: Pizzutiello R

    更新日期:2012-06-01 00:00:00

  • An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

    abstract:PURPOSE:Radiochromic film provides dose measurement at high spatial resolution, but often is not preferred for routine evaluation of patient-specific intensity modulated radiation therapy (IMRT) plans owing to ease-of-use factors. The authors have established an efficient protocol that combines calibration and measurem...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4754797

    authors: Lewis D,Micke A,Yu X,Chan MF

    更新日期:2012-10-01 00:00:00

  • A novel total variation based ring artifact suppression method for CBCT imaging with two-dimensional antiscatter grids.

    abstract:PURPOSE:Two-dimensional antiscatter grids (2DASG) for cone beam computed tomography (CBCT) is a new area of research to reduce scatter intensity, and consequently improve CBCT image quality. One of the challenges in implementation of 2DASGs is their septa shadows that are impinged on the projections. If these artifacts...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13456

    authors: Alexeev T,Kavanagh B,Miften M,Altunbas C

    更新日期:2019-05-01 00:00:00

  • Modeling acute and chronic hypoxia using serial images of 18F-FMISO PET.

    abstract:PURPOSE:Two types of tumor hypoxia most likely exist in human cancers: Chronic hypoxia due to the paucity of blood capillaries and acute hypoxia due to temporary shutdoWn of microvasculatures or fluctuation in the red cell flux. In a recent hypoxia imaging study using 18F-FMISO PET, the authors observed variation in tr...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3213092

    authors: Wang K,Yorke E,Nehmeh SA,Humm JL,Ling CC

    更新日期:2009-10-01 00:00:00

  • SU-E-T-267: Construction and Evaluation of a Neutron Wall to Shield a 15 MV Linac in a Low-Energy Vault.

    abstract:PURPOSE:To design and quantify the shielding efficacy of an inner Borated Polyethylene (BPE)wall for a 15 MV linac in a low energy vault. METHODS:A Varian TrueBeam linac with a maximum photon energy of 15 MV was installed in asmaller, preexisting vault. This vault originally housed a low-energy machine and did not hav...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735334

    authors: Speiser M,Hager F,Foster R,Solberg T

    更新日期:2012-06-01 00:00:00

  • Are we doing any good by doing really well? (Where's the Bacon?).

    abstract::Francis Bacon, who with Rene Decartes laid the intellectual foundations for Western science in the seventeenth century, asserted that the purpose of all knowledge is "action in the production of works for ... the relief of man's estate." We assess briefly several aspects of a few of the current efforts directed to the...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1555493

    authors: Herbert D

    更新日期:2003-04-01 00:00:00

  • Effective and equivalent organ doses in patients undergoing coronary angiography and percutaneous coronary intervention.

    abstract:PURPOSE:Recent recommendations of the International Commission on Radiological Protection state that the use of effective dose (E) for assessing patient exposure has severe limitations, though it can be kept for dose comparisons. In cardiology procedures, the equivalent dose (H(T)) is one of the most appropriate dose q...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3561508

    authors: Compagnone G,Ortolani P,Domenichelli S,Ovi V,Califano G,Dall'Ara G,Marzocchi A

    更新日期:2011-04-01 00:00:00

  • Neutron and gamma spectroscopy for clinical dosimetry.

    abstract::Neutron and photon energy spectra were measured in different depths and lateral positions in tissue-equivalent phantoms exposed to a therapeutical 14.1 MeV neutron beam. Intensity weighted mass energy absorption coefficients were derived for application in ionization chamber dosimetry in mixed neutron-gamma fields. Th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594744

    authors: Schmidt R,Magiera E,Scobel W

    更新日期:1980-09-01 00:00:00

  • SU-E-I-103: Detecting Anomalous Pixels and Correlated Artifacts in Digital Detectors from Flat-Field Images.

    abstract:PURPOSE:Anomalous pixels may be defined as those pixels whose exposure response relationship is deviant from the typical, expected or calibrated response. A group of anomalous pixels may Result in visible correlated artifacts. Here we demonstrate an approach to identify anomalous pixels and correlated artifacts using f...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734820

    authors: Dave J,Gingold E,Yorkston J,Bercha I,Goldman L,Walz-Flannigan A,Willis C

    更新日期:2012-06-01 00:00:00

  • Technical Note: Influence of Compton currents on profile measurements in small-volume ion chambers.

    abstract:PURPOSE:This work is to evaluate the effects of Compton current generation in three small-volume ionization chambers on measured beam characteristics for electron fields. METHODS:Beam scans were performed using Exradin A16, A26, and PTW 31014 microchambers. Scans with varying chamber components shielded were performed...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4929555

    authors: Tanny S,Holmes S,Sperling N,Parsai EI

    更新日期:2015-10-01 00:00:00

  • SU-E-T-302: A Simulation Study with Geant4 Investigating the Secondary Prompt Gamma Emissions from Incident 40 MeV Protons Onto Various Materials.

    abstract:PURPOSE:This study aims to investigate the prompt secondary gamma spectrum and thedepth-dose distributions of incident 40 MeV protons onto various phantom materials.The goal is to find a relationship between the secondary prompt gamma emissions andthe range of the incident particles. METHODS:An application was constru...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735388

    authors: Lau A,Chen Y,Ahmad S

    更新日期:2012-06-01 00:00:00

  • Correcting and reweighting false label masks in brain tumor segmentation.

    abstract:PURPOSE:Recently, brain tumor segmentation has made important progress. However, the quality of manual labels plays an important role in the performance, while in practice, it could vary greatly and in turn could substantially mislead the learning process and decrease the accuracy. We need to design a mechanism to comb...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14480

    authors: Cheng G,Ji H,He L

    更新日期:2020-09-24 00:00:00

  • SU-E-J-157: Simulation and Design of a Real-Time 6D Head Motion Compensation Platform Based on a Stewart Platform Approach.

    abstract:PURPOSE:Real-time sub-millimeter head motion compensation during frameless SRS delivery has the potential to achieve the accuracy of frame-based SRS while being significantly less invasive. Previously, we demonstrated real-time 6D head motion monitoring using an optical camera, however, at the time we were limited to o...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734995

    authors: Belcher AH,Wiersma RD

    更新日期:2012-06-01 00:00:00

  • Spectra and air-kerma strength for encapsulated 192Ir sources.

    abstract::The photon spectra in vacuum around four types of 192Ir HDR brachytherapy sources are calculated using the Monte Carlo code EGS4 and the most recent spectral information on 192Ir decay. The air-kerma strengths per unit activity are calculated based on the photon fluence around a bare 192Ir source and around each of fo...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598763

    authors: Borg J,Rogers DW

    更新日期:1999-11-01 00:00:00

  • WE-G-213CD-03: A Dual Complementary Verification Method for Dynamic Tumor Tracking on Vero SBRT.

    abstract:PURPOSE:to use complementary cine EPID and gimbals log file analysis for in-vivo tracking accuracy monitoring. METHODS:A clinical prototype of dynamic tracking (DT) was installed on the Vero SBRT system. This prototype version allowed tumor tracking by gimballed linac rotations using an internal-external correspondenc...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4736204

    authors: Poels K,Depuydt T,Verellen D,De Ridder M

    更新日期:2012-06-01 00:00:00

  • Performance of a 41 x 41 cm2 amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications.

    abstract::We measured the physical imaging performance of a 41 x 41 cm2 amorphous silicon flat panel detector designed for angiographic and R&F imaging applications using methods from the emerging IEC standard for the measurement of detective quantum efficiency (DQE) in digital radiographic detectors. Measurements on 12 product...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1609151

    authors: Granfors PR,Aufrichtig R,Possin GE,Giambattista BW,Huang ZS,Liu J,Ma B

    更新日期:2003-10-01 00:00:00

  • Local patient dose diagnostic reference levels in pediatric interventional cardiology in Chile using age bands and patient weight values.

    abstract:PURPOSE:To present the results of a patient dose evaluation program in pediatric cardiology and propose local diagnostic reference levels (DRLs) for different types of procedure and age range, in addition to suggesting approaches to correlate patient dose values with patient weight. This study was the first conducted i...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4905116

    authors: Ubeda C,Miranda P,Vano E

    更新日期:2015-02-01 00:00:00

  • Broad line quantitative chemical shift spectroscopy.

    abstract::Proton NMR spectroscopy was applied to quantitate the measurement of total body water/fat distributions in vivo. A special MR protocol was developed to excite a thick slab of tissue and display the magnitude NMR spectrum of the collected response signal. Very short echo time TE (8 ms) and long repetition time TR (4000...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596935

    authors: Sobol WT,Elster AD,Hinson WH,Chwals WJ

    更新日期:1992-01-01 00:00:00

  • The use of a priori information in the detection of mammographic microcalcifications to improve their classification.

    abstract::In this work, we present a calcification-detection scheme that automatically localizes calcifications in a previously detected cluster in order to generate the input for a cluster-classification scheme developed in the past. The calcification-detection scheme makes use of three pieces of a priori information: the loca...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1559884

    authors: Salfity MF,Nishikawa RM,Jiang Y,Papaioannou J

    更新日期:2003-05-01 00:00:00

  • Computational modeling of 915 MHz microwave ablation: Comparative assessment of temperature-dependent tissue dielectric models.

    abstract:PURPOSE:The objective of this study is to develop a computational model for simulating 915 MHz microwave ablation (MWA), and verify the simulation predictions of transient temperature profiles against experimental measurements. Due to the limited experimental data characterizing temperature-dependent changes of tissue ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12359

    authors: Deshazer G,Hagmann M,Merck D,Sebek J,Moore KB,Prakash P

    更新日期:2017-09-01 00:00:00

  • On calculating kerma, collision kerma and radiative yields.

    abstract:PURPOSE:To study the relationships between dose (D), kerma (K), and collision kerma ( K col ) in photon beams and to investigate total radiative yields for electrons and positrons as a functio...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13744

    authors: Rogers DWO,Townson RW

    更新日期:2019-11-01 00:00:00

  • TU-G-BRB-01: Continuous Path Optimization for Non-Coplanar Variant SAD IMRT Delivery Using C-Arm Machines.

    abstract:PURPOSE:To develop and investigate a continuous path optimization methodology to traverse prescribed non-coplanar IMRT beams with variant SADs, by orchestrating the couch and gantry movement with zero-collision, minimal patient motion consequence and machine travel time. METHODS:We convert the given collision zone def...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735996

    authors: Ruan D,Dong P,Low D,Sheng K

    更新日期:2012-06-01 00:00:00

  • Liver segmentation in contrast enhanced CT data using graph cuts and interactive 3D segmentation refinement methods.

    abstract:PURPOSE:Liver segmentation is an important prerequisite for the assessment of liver cancer treatment options like tumor resection, image-guided radiation therapy (IGRT), radiofrequency ablation, etc. The purpose of this work was to evaluate a new approach for liver segmentation. METHODS:A graph cuts segmentation metho...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3682171

    authors: Beichel R,Bornik A,Bauer C,Sorantin E

    更新日期:2012-03-01 00:00:00

  • Potential use of P-32 ophthalmic applicator: Monte Carlo simulations for design and dosimetry.

    abstract::Postoperative beta-irradiation after pterygium excision has been considered a valuable therapeutic procedure to reduce the recurrence rate. Recently, it was reported that beta-irradiation also substantially reduced the risk of surgical failure after glaucoma surgery. Pure beta-irradiation using a 90Sr/Y applicator has...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2896076

    authors: Park YK,Ye SJ,Kim IH,Wee WR,Kim MK,Han HS,Son KJ,Park UJ

    更新日期:2008-05-01 00:00:00

  • Mitigating differential baseline shifts in locally advanced lung cancer patients using an average anatomy model.

    abstract:BACKGROUND AND PURPOSE:Differential baseline shifts between primary tumor and involved lymph nodes in locally advanced lung cancer patients compromise the accuracy of radiotherapy. The purpose of this study was to evaluate the performance of an average anatomy model (AAM) derived from repeat imaging and deformable regi...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12271

    authors: Abdoli M,Van Kranen SR,Stankovic U,Rossi MMG,Belderbos JSA,Sonke JJ

    更新日期:2017-07-01 00:00:00

  • Pharmacokinetic modeling.

    abstract::For radiation dosimetry calculations of radiolabeled monoclonal antibodies, (MAB), pharmacokinetics are critical. Specifically, pharmacokinetic modeling is a useful component of estimation of cumulated activity in various source organs in the body. It is thus important to formulate general methods of pharmacokinetic m...

    journal_title:Medical physics

    pub_type: 杂志文章,评审

    doi:10.1118/1.597047

    authors: Strand SE,Zanzonico P,Johnson TK

    更新日期:1993-03-01 00:00:00