A dosimetry study of the Oncoseed 6711 using glass rod dosimeters and EGS5 Monte Carlo code in a geometry lacking radiation equilibrium scatter conditions.

Abstract:

PURPOSE:The aim of this study was to develop a dose calculation method which is applicable to the interseed attenuation and the geometry lacking the equilibrium radiation scatter conditions in brachytherapy. METHODS:The dose obtained from measurement with a radiophotoluminescent glass rod dosimeter (GRD) was compared to the dose calculated with the Monte Carlo (MC) code "EGS5," using the 125I source structure detailed in by Kennedy et al. The GRDs were irradiated with 125I Oncoseed 6711 in a human head phantom. The phantom was a cylinder made of 2 mm thick PMMA with a diameter of 18 cm and length of 16 cm. Some of the GRD positions were so close to the phantom surface that the backscatter margin was less than 5 cm, insufficient for photons. RESULTS:The EGS5 simulations were found to reproduce the relative dose distributions as measured with the GRDs to within 25% uncertainty in the geometry lacking the equilibrium radiation scatter conditions. The absolute value of the GRD measurement agreed with the American Association of Physicist in Medicine Task Group No 43 Updated Protocol (AAPM-TG43U1) formalism to within 3% of the reference point (r = 1 cm, theta = 90 degrees), where the TG43U1 is especially reliable because of the abundant data accumulation in composing the formalism. The factor to normalize the measured or calculated dose to the TG43U1 estimate at the reference point was evaluated to be 0.97 for the GRD measurement and 1.8 for the MC calculation, which uses the integration of the apparent activity with the time as the amount of disintegration during the irradiation. Also, F(r,theta) and g(r) estimated by this calculation method were consistent with those proposed in the TG43U1. CONCLUSIONS:The results of this investigation support the validity of both the MC calculation method and GRD measurement in this study as well as the TG-43U1 formalism. Also, this calculation is applicable to interseed attenuation and the geometry lacking the equilibrium radiation scatter.

journal_name

Med Phys

journal_title

Medical physics

authors

Tanaka K,Tateoka K,Asanuma O,Kamo K,Bengua G,Sato K,Ueda T,Takeda H,Takagi M,Hareyama M,Takada J

doi

10.1118/1.3590370

subject

Has Abstract

pub_date

2011-06-01 00:00:00

pages

3069-76

issue

6

eissn

0094-2405

issn

2473-4209

journal_volume

38

pub_type

杂志文章
  • Patient motion effects on the quantification of regional myocardial blood flow with dynamic PET imaging.

    abstract:PURPOSE:Patient motion is a common problem during dynamic positron emission tomography (PET) scans for quantification of myocardial blood flow (MBF). The purpose of this study was to quantify the prevalence of body motion in a clinical setting and evaluate with realistic phantoms the effects of motion on blood flow qua...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4943565

    authors: Hunter CR,Klein R,Beanlands RS,deKemp RA

    更新日期:2016-04-01 00:00:00

  • RF tissue-heating near metallic implants during magnetic resonance examinations: an approach in the ac limit.

    abstract:PURPOSE:State of the art to access radiofrequency (RF) heating near implants is computer modeling of the devices and solving Maxwell's equations for the specific setup. For a set of input parameters, a fixed result is obtained. This work presents a theoretical approach in the alternating current (ac) limit, which can p...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3637495

    authors: Ballweg V,Eibofner F,Graf H

    更新日期:2011-10-01 00:00:00

  • Noise, sampling, and the number of projections in cone-beam CT with a flat-panel detector.

    abstract:PURPOSE:To investigate the effect of the number of projection views on image noise in cone-beam CT (CBCT) with a flat-panel detector. METHODS:This fairly fundamental consideration in CBCT system design and operation was addressed experimentally (using a phantom presenting a uniform medium as well as statistically moti...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4875688

    authors: Zhao Z,Gang GJ,Siewerdsen JH

    更新日期:2014-06-01 00:00:00

  • Rigid 2D/3D slice-to-volume registration and its application on fluoroscopic CT images.

    abstract::Registration of single slices from FluoroCT, CineMR, or interventional magnetic resonance imaging to three dimensional (3D) volumes is a special aspect of the two-dimensional (2D)/3D registration problem. Rather than digitally rendered radiographs (DRR), single 2D slice images obtained during interventional procedures...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2401661

    authors: Birkfellner W,Figl M,Kettenbach J,Hummel J,Homolka P,Schernthaner R,Nau T,Bergmann H

    更新日期:2007-01-01 00:00:00

  • Calibrated breast density methods for full field digital mammography: a system for serial quality control and inter-system generalization.

    abstract:PURPOSE:The authors are developing a system for calibrated breast density measurements using full field digital mammography (FFDM). Breast tissue equivalent (BTE) phantom images are used to establish baseline (BL) calibration curves at time zero. For a given FFDM unit, the full BL dataset is comprised of approximately ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4903299

    authors: Lu B,Smallwood AM,Sellers TA,Drukteinis JS,Heine JJ,Fowler EE

    更新日期:2015-02-01 00:00:00

  • The DosiMap, a new 2D scintillating dosimeter for IMRT quality assurance: characterization of two Cerenkov discrimination methods.

    abstract::New radiation therapy techniques such as IMRT present significant efficiency due to their highly conformal dose distributions. A consequence of the complexity of their dose distributions (high gradients, small irradiation fields, low dose distribution, ...) is the requirement for better precision quality assurance tha...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2897966

    authors: Frelin AM,Fontbonne JM,Ban G,Colin J,Labalme M,Batalla A,Vela A,Boher P,Braud M,Leroux T

    更新日期:2008-05-01 00:00:00

  • Regularized iterative weighted filtered backprojection for helical cone-beam CT.

    abstract::Contemporary reconstruction methods employed for clinical helical cone-beam computed tomography (CT) are analytical (noniterative) but mathematically nonexact, i.e., the reconstructed image contains so called cone-beam artifacts, especially for higher cone angles. Besides cone artifacts, these methods also suffer from...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2966353

    authors: Sunnegårdh J,Danielsson PE

    更新日期:2008-09-01 00:00:00

  • The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management.

    abstract::The increasing complexity of modern radiation therapy planning and delivery challenges traditional prescriptive quality management (QM) methods, such as many of those included in guidelines published by organizations such as the AAPM, ASTRO, ACR, ESTRO, and IAEA. These prescriptive guidelines have traditionally focuse...

    journal_title:Medical physics

    pub_type: 共识发展会议,杂志文章,实务指引

    doi:10.1118/1.4947547

    authors: Huq MS,Fraass BA,Dunscombe PB,Gibbons JP Jr,Ibbott GS,Mundt AJ,Mutic S,Palta JR,Rath F,Thomadsen BR,Williamson JF,Yorke ED

    更新日期:2016-07-01 00:00:00

  • Selective enhancement filters for nodules, vessels, and airway walls in two- and three-dimensional CT scans.

    abstract::Computer-aided diagnostic (CAD) schemes have been developed to assist radiologists in the early detection of lung cancer in radiographs and computed tomography (CT) images. In order to improve sensitivity for nodule detection, many researchers have employed a filter as a preprocessing step for enhancement of nodules. ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1581411

    authors: Li Q,Sone S,Doi K

    更新日期:2003-08-01 00:00:00

  • Absolute diameter measurements of coronary arteries based on the first zero crossing of the Fourier spectrum.

    abstract::Current edge detection methods used to determine coronary artery dimensions from digital (DSA) images suffer a strong dependence on the system's modulation transfer function (MTF). The videodensitometric algorithms are less sensitive to MTF blurring, yet still result in an overestimation of the vessel size of 10% to 2...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596370

    authors: Weber DM

    更新日期:1989-03-01 00:00:00

  • Nonlocal atlas-guided multi-channel forest learning for human brain labeling.

    abstract:PURPOSE:It is important for many quantitative brain studies to label meaningful anatomical regions in MR brain images. However, due to high complexity of brain structures and ambiguous boundaries between different anatomical regions, the anatomical labeling of MR brain images is still quite a challenging task. In many ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4940399

    authors: Ma G,Gao Y,Wu G,Wu L,Shen D

    更新日期:2016-02-01 00:00:00

  • Determination of the half-life of I-125 encapsulated clinical seeds using a Si(Li) detector.

    abstract::The half-life of two I-125 encapsulated clinical seeds was measured with a Si(Li) semiconductor detector. Two measurements of each seed yielded an average half-life of 59.56 +/- 0.17 days. The results are discussed in terms of the influence of possible I-126 contamination and leakage of I-125 activity through encapsul...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595423

    authors: Kubo H

    更新日期:1983-11-01 00:00:00

  • Higher energy: is it necessary, is it worth the cost for radiation oncology?

    abstract::The physical characteristics of the interactions of megavoltage photons and electrons with matter provide distinct advantages, relative to low-energy (orthovoltage) x rays, that lead to better radiation dose distributions in patients. Use of these high-energy radiations has resulted in better patient care, which has b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596779

    authors: Das IJ,Kase KR

    更新日期:1992-07-01 00:00:00

  • Characterization of mesothelioma and tissues present in contrast-enhanced thoracic CT scans.

    abstract:PURPOSE:The purpose of this study was to characterize the Hounsfield unit (HU) distributions of mesothelioma and other tissues present in contrast-enhanced thoracic CT scans, to compare the HU distributions of mesothelioma, muscle, and liver by scanner and reconstruction filter/kernel combination, and to assess interpa...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3537610

    authors: Corson N,Sensakovic WF,Straus C,Starkey A,Armato SG 3rd

    更新日期:2011-02-01 00:00:00

  • Modification of the University of Washington Neutron Radiotherapy Facility for optimization of neutron capture enhanced fast-neutron therapy.

    abstract::A modified neutron production target assembly has been developed to provide improved performance of the proton-cyclotron-based neutron radiotherapy facility at the University of Washington for applications involving neutron capture enhanced fast-neutron therapy. The new target produces a neutron beam that yields essen...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598839

    authors: Nigg DW,Wemple CA,Risler R,Hartwell JK,Harker YD,Laramore GE

    更新日期:2000-02-01 00:00:00

  • Model-based calculations of off-axis ratio of conic beams for a dedicated 6 MV radiosurgery unit.

    abstract:PURPOSE:Because the small-radius photon beams shaped by cones in stereotactic radiosurgery (SRS) lack lateral electronic equilibrium and a detector's finite cross section, direct experimental measurement of dosimetric data for these beams can be subject to large uncertainties. As the dose calculation accuracy of a trea...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3488900

    authors: Yang JN,Ding X,Du W,Pino R

    更新日期:2010-10-01 00:00:00

  • Computation of beam quality parameters for Mo/Mo, Mo/Rh, Rh/Rh, and W/Al target/filter combinations in mammography.

    abstract::A computer program was implemented to predict mammography x-ray beam parameters in the range 20-40 kV for Mo/Mo, Mo/Rh, Rh/Rh, and W/Al target/filter combinations. The computation method used to simulate mammography x-ray spectra is based on the Boone et al. model. The beam quality parameters such as the half-value la...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1608498

    authors: Kharrati H,Zarrad B

    更新日期:2003-10-01 00:00:00

  • An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetry.

    abstract::The relationship between the pixel value and exit dose was investigated for a new commercially available amorphous silicon electronic portal imaging device. The pixel to dose mapping function was established to be linear for detector distances between 116.5 cm to 150 cm from the source, radiation field sizes from 5 x ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1508108

    authors: Grein EE,Lee R,Luchka K

    更新日期:2002-10-01 00:00:00

  • Calculations of two-dimensional electric field and potential distributions around electrodes in conducting media.

    abstract::Tumor cells can be killed by heating to 42.5 degrees C. It is convenient to localize heating by passing electric current at 500 kHz through tissue between a pair of electrodes. It is desirable to determine, for a given electrode-tissue configuration, what the resulting field and, therefore, current distribution will b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594235

    authors: Hutson RL

    更新日期:1976-07-01 00:00:00

  • Oxygen distribution in tumors: a qualitative analysis and modeling study providing a novel Monte Carlo approach.

    abstract:PURPOSE:To construct a Monte Carlo (MC)-based simulation model for analyzing the dependence of tumor oxygen distribution on different variables related to tumor vasculature [blood velocity, vessel-to-vessel proximity (vessel proximity), and inflowing oxygen partial pressure (pO2)]. METHODS:A voxel-based tissue model c...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4892386

    authors: Lagerlöf JH,Kindblom J,Bernhardt P

    更新日期:2014-09-01 00:00:00

  • The delta envelope: a technique for dose distribution comparison.

    abstract::The gamma index is a tool that compares a dose distribution with a reference distribution by combining dose-difference and distance-to-agreement criteria. It has been widely used for ten years despite its high computational cost. This cost is due to both a search process for each reference point and the necessity to r...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3070546

    authors: Blanpain B,Mercier D

    更新日期:2009-03-01 00:00:00

  • Material elemental decomposition in dual and multi-energy CT via a sparsity-dictionary approach for proton stopping power ratio calculation.

    abstract:PURPOSE:Accurate calculation of proton stopping power ratio (SPR) relative to water is crucial to proton therapy treatment planning, since SPR affects prediction of beam range. Current standard practice derives SPR using a single CT scan. Recent studies showed that dual-energy CT (DECT) offers advantages to accurately ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12796

    authors: Shen C,Li B,Chen L,Yang M,Lou Y,Jia X

    更新日期:2018-04-01 00:00:00

  • Calculation of photon energy deposition kernels and electron dose point kernels in water.

    abstract::Effects of changes in the physics of EGSnrc compared to EGS4/PRESTA on energy deposition kernels for monoenergetic photons and on dose point kernels for beta sources in water are investigated. In the diagnostic energy range, Compton binding corrections were found to increase the primary energy fraction up to 4.5% at 3...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1861412

    authors: Mainegra-Hing E,Rogers DW,Kawrakow I

    更新日期:2005-03-01 00:00:00

  • Impact of calibration methods for color medical displays on interpreting brain SPECT images.

    abstract:PURPOSE:Color images are visualized on medical monitors that are adjusted by a grayscale standard display function (GSDF) or γ2.2. Although the GSDF is visually displayed as a linear graded grayscale, it does not specify how color medical images should be presented. On the other hand, the usual gamma setting for color ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13532

    authors: Shibutani T,Onoguchi M,Yoneyama H,Nakajima K,Matsuo S,Bamba Y,Saito S

    更新日期:2019-06-01 00:00:00

  • Task Group 284 Report: Magnetic Resonance Imaging Simulation in Radiotherapy: Considerations for Clinical Implementation, Optimization, and Quality Assurance.

    abstract::The use of dedicated magnetic resonance simulation (MR-SIM) platforms in Radiation Oncology has expanded rapidly, introducing new equipment and functionality with the overall goal of improving the accuracy of radiation treatment planning. However, this emerging technology presents a new set of challenges that need to ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14695

    authors: Glide-Hurst CK,Paulson ES,McGee K,Tyagi N,Hu Y,Balter J,Bayouth J

    更新日期:2021-01-01 00:00:00

  • The properties of the ultramicrocylindrical ionization chamber for small field used in stereotactic radiosurgery.

    abstract::Accurate dosimetry of small-field photon beams tends to be difficult to perform due to the presence of lateral electronic disequilibrium and steep dose gradients. In stereotactic radiosurgery (SRS), small fields of 6-30 mm in diameter are used. Generally thermoluminescence dosimetry chips, Farmer, Thimble ion chamber,...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1350437

    authors: Vahc YW,Chung WK,Park KR,Lee JY,Lee YH,Kwon O,Kim S

    更新日期:2001-03-01 00:00:00

  • Performance characteristics of transmission imaging using a uniform sheet source with parallel-hole collimation.

    abstract::Transmission imaging is receiving increasing attention in SPECT due to the need to compensate for nonuniform attenuation in cardiac-chest SPECT. The quality of a transmission image has an important effect on the measured attenuation distribution. To improve image quality, knowledge of the performance characteristics o...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596752

    authors: Cao Z,Tsui BM

    更新日期:1992-09-01 00:00:00

  • Gradient corrections for reference dosimetry using Farmer-type ionization chambers in single-layer scanned proton fields.

    abstract:PURPOSE:The local depth dose gradient and the displacement correction factor for Farmer-type ionization chambers are quantified for reference dosimetry at shallow depth in single-layer scanned proton fields. METHOD:Integrated radial profiles as a function of depth (IRPDs) measured at three proton therapy centers were ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14554

    authors: Palmans H,Medin J,Trnková P,Vatnitsky S

    更新日期:2020-12-01 00:00:00

  • Use of proximal operator graph solver for radiation therapy inverse treatment planning.

    abstract:PURPOSE:Most radiation therapy optimization problems can be formulated as an unconstrained problem and solved efficiently by quasi-Newton methods such as the Limited-memory Broyden-Fletcher-Goldfarb-Shanno (L-BFGS) algorithm. However, several next generation planning techniques such as total variation regularization- b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12165

    authors: Liu X,Pelizzari C,Belcher AH,Grelewicz Z,Wiersma RD

    更新日期:2017-04-01 00:00:00

  • Correcting TG 119 confidence limits.

    abstract:PURPOSE:Task Group 119 (TG-119) has been adopted for evaluating the adequacy of intensity-modulated radiation therapy (IMRT) commissioning and for establishing patient-specific IMRT quality assurance (QA) passing criteria in clinical practice. TG-119 establishes 95% confidence limits (CLs), which help clinics identify ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12759

    authors: Kearney V,Solberg T,Jensen S,Cheung J,Chuang C,Valdes G

    更新日期:2018-03-01 00:00:00