The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management.

Abstract:

:The increasing complexity of modern radiation therapy planning and delivery challenges traditional prescriptive quality management (QM) methods, such as many of those included in guidelines published by organizations such as the AAPM, ASTRO, ACR, ESTRO, and IAEA. These prescriptive guidelines have traditionally focused on monitoring all aspects of the functional performance of radiotherapy (RT) equipment by comparing parameters against tolerances set at strict but achievable values. Many errors that occur in radiation oncology are not due to failures in devices and software; rather they are failures in workflow and process. A systematic understanding of the likelihood and clinical impact of possible failures throughout a course of radiotherapy is needed to direct limit QM resources efficiently to produce maximum safety and quality of patient care. Task Group 100 of the AAPM has taken a broad view of these issues and has developed a framework for designing QM activities, based on estimates of the probability of identified failures and their clinical outcome through the RT planning and delivery process. The Task Group has chosen a specific radiotherapy process required for "intensity modulated radiation therapy (IMRT)" as a case study. The goal of this work is to apply modern risk-based analysis techniques to this complex RT process in order to demonstrate to the RT community that such techniques may help identify more effective and efficient ways to enhance the safety and quality of our treatment processes. The task group generated by consensus an example quality management program strategy for the IMRT process performed at the institution of one of the authors. This report describes the methodology and nomenclature developed, presents the process maps, FMEAs, fault trees, and QM programs developed, and makes suggestions on how this information could be used in the clinic. The development and implementation of risk-assessment techniques will make radiation therapy safer and more efficient.

journal_name

Med Phys

journal_title

Medical physics

authors

Huq MS,Fraass BA,Dunscombe PB,Gibbons JP Jr,Ibbott GS,Mundt AJ,Mutic S,Palta JR,Rath F,Thomadsen BR,Williamson JF,Yorke ED

doi

10.1118/1.4947547

subject

Has Abstract

pub_date

2016-07-01 00:00:00

pages

4209

issue

7

eissn

0094-2405

issn

2473-4209

journal_volume

43

pub_type

共识发展会议,杂志文章,实务指引
  • A boundary element approach for image-guided near-infrared absorption and scatter estimation.

    abstract::Multimodality NIR spectroscopy systems offer the possibility of region-based vascular and molecular characterization of tissue in vivo. However, computationally efficient 3D image reconstruction algorithms specific to these image-guided systems currently do not exist. Image reconstruction is often based on finite-elem...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2795832

    authors: Srinivasan S,Pogue BW,Carpenter C,Yalavarthy PK,Paulsen K

    更新日期:2007-11-01 00:00:00

  • Optimized knot placement for B-splines in deformable image registration.

    abstract:PURPOSE:To develop an automatic knot placement algorithm to enable the use of NonUniform Rational B-Splines (NURBS) in deformable image registration. METHODS:The authors developed a two-step approach to fit a known displacement vector field (DVF). An initial fit was made with uniform knot spacing. The error generated ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3609416

    authors: Jacobson TJ,Murphy MJ

    更新日期:2011-08-01 00:00:00

  • Radiation attenuation by lead and nonlead materials used in radiation shielding garments.

    abstract::The attenuating properties of several types of lead (Pb)-based and non-Pb radiation shielding materials were studied and a correlation was made of radiation attenuation, materials properties, calculated spectra and ambient dose equivalent. Utilizing the well-characterized x-ray and gamma ray beams at the National Rese...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2426404

    authors: McCaffrey JP,Shen H,Downton B,Mainegra-Hing E

    更新日期:2007-02-01 00:00:00

  • Absorbed dose distributions from ophthalmic 106 Ru/106 Rh plaques measured in water with radiochromic film.

    abstract:PURPOSE:Brachytherapy with 106 Ru/106 Rh plaques offers good outcomes for small-to-medium choroidal melanomas and retinoblastomas. The dose measurement of the plaques is challenging, due to the small range of the emitted beta particles and steep dose gradients involved. The scarce publications on film dosimetry of 106 ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12777

    authors: Hermida-López M,Brualla L

    更新日期:2018-04-01 00:00:00

  • Accelerated barrier optimization compressed sensing (ABOCS) reconstruction for cone-beam CT: phantom studies.

    abstract:PURPOSE:Recent advances in compressed sensing (CS) enable accurate CT image reconstruction from highly undersampled and noisy projection measurements, due to the sparsifiable feature of most CT images using total variation (TV). These novel reconstruction methods have demonstrated advantages in clinical applications wh...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4729837

    authors: Niu T,Zhu L

    更新日期:2012-07-01 00:00:00

  • Study of the secondary neutral radiation in proton therapy: toward an indirect in vivo dosimetry.

    abstract:PURPOSE:Secondary particles produced in the collision of protons with beam modifiers are of concern in proton therapy. Nevertheless, secondary radiation can provide information on the dosimetric parameters through its dependency on the modulating accessories (range shifter and range modulating wheel). Relatively little...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4765049

    authors: Carnicer A,Letellier V,Rucka G,Angellier G,Sauerwein W,Herault J

    更新日期:2012-12-01 00:00:00

  • Task Group 284 Report: Magnetic Resonance Imaging Simulation in Radiotherapy: Considerations for Clinical Implementation, Optimization, and Quality Assurance.

    abstract::The use of dedicated magnetic resonance simulation (MR-SIM) platforms in Radiation Oncology has expanded rapidly, introducing new equipment and functionality with the overall goal of improving the accuracy of radiation treatment planning. However, this emerging technology presents a new set of challenges that need to ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14695

    authors: Glide-Hurst CK,Paulson ES,McGee K,Tyagi N,Hu Y,Balter J,Bayouth J

    更新日期:2021-01-01 00:00:00

  • A scintillator-based approach to monitor secondary neutron production during proton therapy.

    abstract:PURPOSE:The primary objective of this work is to measure the secondary neutron field produced by an uncollimated proton pencil beam impinging on different tissue-equivalent phantom materials using organic scintillation detectors. Additionally, the Monte Carlo code mcnpx-PoliMi was used to simulate the detector response...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4963813

    authors: Clarke SD,Pryser E,Wieger BM,Pozzi SA,Haelg RA,Bashkirov VA,Schulte RW

    更新日期:2016-11-01 00:00:00

  • Characteristics of sensitometric curves of radiographic films.

    abstract::A new type of radiographic film, EDR (extended dose range) film, has been recently become available for film dosimetry. It is particularly attractive for composite isodose verification of intensity modulated radiation therapy because of its low sensitivity relative to the more common Kodak XV film. For XV film, the re...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1568979

    authors: Zhu XR,Yoo S,Jursinic PA,Grimm DF,Lopez F,Rownd JJ,Gillin MT

    更新日期:2003-05-01 00:00:00

  • Analytical considerations of beam hardening in medical accelerator photon spectra.

    abstract::Beam hardening is a well-known phenomenon for therapeutic accelerator beams passing through matter in narrow beam geometry. This study assesses quantitatively the magnitude of beam hardening of therapeutic beams in water. A formal concept of beam hardening is proposed which is based on the decrease of the mean attenua...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598704

    authors: Kleinschmidt C

    更新日期:1999-09-01 00:00:00

  • Monte Carlo modeling of the response of NRC's 90Sr/90Y primary beta standard.

    abstract::The BEAMnrc/EGSnrc Monte Carlo code system is employed to develop a model of the National Research Council of Canada primary standard of absorbed dose to tissue in a beta radiation field, comprising an extrapolation chamber and 90Sr/90Y beta source. We benchmark the model against the measured response of the chamber i...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1997347

    authors: Selvam TP,Saull PR,Rogers DW,National Research Council of Canada.

    更新日期:2005-10-01 00:00:00

  • Matching photon and electron fields with dynamic intensity modulation.

    abstract::A technique was developed to reduce the size and magnitude of the hot and cold spots in the abutting regions of photon and electron fields. The photon and electron fields were set up such that the photon field extended approximately 2 cm into the electron field in the abutting region. The region of the photon beam tha...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598753

    authors: Li JG,Xing L,Boyer AL,Hamilton RJ,Spelbring DR,Turian JV

    更新日期:1999-11-01 00:00:00

  • Sci-Fri PM: Planning-04: Dose escalation study using anatomy-based aperture IMRT and SPECT perfusion images for lung cancer.

    abstract::In the case of non-small cell lung cancer, doses typically prescribed (60-66 Gy) are not sufficient to ensure a satisfactory tumor control probability. Dose escalation needs to be realized, but dose to organs at risk (OARs) must be kept under widely accepted clinical thresholds. Also, lung functionality is not homogen...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2965976

    authors: St-Hilaire J,Lavoie C,Beaulieu F,Dagnault A,Morin F,Gingras L,Tremblay D,Beaulieu L

    更新日期:2008-07-01 00:00:00

  • SU-E-T-398: Verification of Gamma Knife EXtend System Based Fractionated Treatment Planning Using EBT2 Film.

    abstract:PURPOSE:To present EBT2 film verification of treatment planning with the eXtend System, a relocatable frame system for multiple-fraction or serial multiple-session radiosurgery. METHODS:A human head shaped phantom simulated the verification process for fractionated Gamma Knife (GK) treatment. Phantom preparation for e...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735487

    authors: Gopishankar N,Bisht RK

    更新日期:2012-06-01 00:00:00

  • SU-E-T-462: Calculation of Prescribed Dose for Permanent Implant with Cs-131 Using LQ Equation including Resensitization.

    abstract:PURPOSE:To develop a new biological equivalent dose (BED) formulism that includes repair, repopulation and resensitization (redistribution and reoxygenation), and can be used to calculate prescribed dose for permanent implant with Cs-131. METHODS:Because of lack of sufficient clinical data, the prescribed dose with ne...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735551

    authors: Luo W,Aryal P,Randall M

    更新日期:2012-06-01 00:00:00

  • A phantom for visualization of three-dimensional drug release by ultrasound-induced mild hyperthermia.

    abstract:PURPOSE:Ultrasound-induced mild hyperthermia has advantages for noninvasive, localized and controlled drug delivery. In this study, a tissue-mimicking agarose-based phantom with a thermally sensitive indicator was developed for studying the spatial drug delivery profile using ultrasound-induced mild hyperthermia. METH...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4813299

    authors: Lai CY,Kruse D,Seo JW,Kheirolomoom A,Ferrara KW

    更新日期:2013-08-01 00:00:00

  • A framework for defining FLASH dose rate for pencil beam scanning.

    abstract:PURPOSE:To develop a method of (a) calculating the dose rate of voxels within a proton field delivered using pencil beam scanning (PBS), and (b) reporting a representative dose rate for the PBS treatment field that enables correspondence between multiple treatment modalities. This method takes into account the unique s...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14456

    authors: Folkerts MM,Abel E,Busold S,Perez JR,Krishnamurthi V,Ling CC

    更新日期:2020-12-01 00:00:00

  • 3T MRI evaluation of the accuracy of atlas-based subthalamic nucleus identification.

    abstract::Modulation of the activity of the subthalamic nucleus (STN) using deep brain stimulation (DBS) in patients with advanced Parkinson's disease is the most common procedure performed today by functional neurosurgeons. The STN contours cannot be entirely identified on common 1.5 T images; in particular, the ventromedial b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2936229

    authors: Stancanello J,Muacevic A,Sebastiano F,Modugno N,Cerveri P,Ferrigno G,Uggeri F,Romanelli P

    更新日期:2008-07-01 00:00:00

  • Optimization of radiography applications using x-ray beams emitted by compact accelerators. Part I. Monte Carlo study of the hard x-ray spectrum.

    abstract:PURPOSE:A 3-year project called LABSYNC has been recently funded by the European Commission, with the aim of designing a radiation facility based on a compact light source, i.e., a laboratory-sized commercial synchrotron, capable of accelerating electrons up to 6 or 20 MeV. An accurate spectral description of hard x ra...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3223357

    authors: Marziani M,Taibi A,Di Domenico G,Gambaccini M

    更新日期:2009-10-01 00:00:00

  • Surface dose rate calibration of Sr-90 plane ophthalmic applicators.

    abstract::Calibration of an imported strontium-90 ophthalmic applicator at the U.S. National Bureau of Standards (now the National Institute of Standards and Technology) has disclose a significant discrepancy in dose rate calibration (32%-35%) with that quoted by the manufacturer. The University of Wisconsin has investigated th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596702

    authors: Goetsch SJ,Sunderland KS

    更新日期:1991-03-01 00:00:00

  • A Novel method to generate on-board 4D MRI using prior 4D MRI and on-board kV projections from a conventional LINAC for target localization in liver SBRT.

    abstract:PURPOSE:On-board MRI can provide superb soft tissue contrast for improving liver SBRT localization. However, the availability of on-board MRI in clinics is extremely limited. On the contrary, on-board kV imaging systems are widely available on radiotherapy machines, but its capability to localize tumors in soft tissue ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12998

    authors: Harris W,Wang C,Yin FF,Cai J,Ren L

    更新日期:2018-07-01 00:00:00

  • Incorporation of treatment plan spatial and temporal dose patterns into a prostate intrafractional motion management strategy.

    abstract:PURPOSE:Periodic MV∕KV radiographs taken during volumetric modulated arc therapy (VMAT) for hypofractionated treatment provide guidance in intrafractional motion management. The choice of imaging frequency and timing are key components in delivering the desired dose while reducing associated overhead such as imaging do...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4742846

    authors: Zhang P,Hunt M,Happersett L,Cox B,Mageras G

    更新日期:2012-09-01 00:00:00

  • Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams.

    abstract:PURPOSE:Dose verification in highly conformal radiation therapy such as IMRT or proton therapy can benefit from the high spatial resolution offered by radio-chromic films such as Gafchromic EBT or EBT2. Recently, a new generation of these films, EBT3, has become available. The composition and thickness of the sensitive...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4737890

    authors: Reinhardt S,Hillbrand M,Wilkens JJ,Assmann W

    更新日期:2012-08-01 00:00:00

  • Two-dimensional antiscatter grid: A novel scatter rejection device for Cone-beam computed tomography.

    abstract:PURPOSE:Scattered radiation remains to be a major cause of image quality degradation in Flat Panel Detector (FPD)-based Cone-beam computed tomography (CBCT). We have been investigating a novel two-dimensional antiscatter grid (2D-ASG) concept to reduce scatter intensity, and hence improve CBCT image quality. We present...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12724

    authors: Alexeev T,Kavanagh B,Miften M,Altunbas C

    更新日期:2018-02-01 00:00:00

  • Determination of the half-life of I-125 encapsulated clinical seeds using a Si(Li) detector.

    abstract::The half-life of two I-125 encapsulated clinical seeds was measured with a Si(Li) semiconductor detector. Two measurements of each seed yielded an average half-life of 59.56 +/- 0.17 days. The results are discussed in terms of the influence of possible I-126 contamination and leakage of I-125 activity through encapsul...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595423

    authors: Kubo H

    更新日期:1983-11-01 00:00:00

  • Differences among Monte Carlo codes in the calculations of voxel S values for radionuclide targeted therapy and analysis of their impact on absorbed dose evaluations.

    abstract::Several updated Monte Carlo (MC) codes are available to perform calculations of voxel S values for radionuclide targeted therapy. The aim of this work is to analyze the differences in the calculations obtained by different MC codes and their impact on absorbed dose evaluations performed by voxel dosimetry. Voxel S val...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3103401

    authors: Pacilio M,Lanconelli N,Lo MS,Betti M,Montani L,Torres AL,Coca PM

    更新日期:2009-05-01 00:00:00

  • Feasibility of estimating volumetric breast density from mammographic x-ray spectra using a cadmium telluride photon-counting detector.

    abstract:PURPOSE:Mammographic density of glandular breast tissue has a masking effect that can reduce lesion detection accuracy and is also a strong risk factor for breast cancer. Therefore, accurate quantitative estimation of breast density is clinically important. In this study, we investigate experimentally the feasibility o...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13031

    authors: Ghammraoui B,Badal A,Glick SJ

    更新日期:2018-06-03 00:00:00

  • Characterization of difference of Gaussian filters in the detection of mammographic regions.

    abstract::In this article, we present a characterization of the effect of difference of Gaussians (DoG) filters in the detection of mammographic regions. DoG filters have been used previously in mammographic mass computer-aided detection (CAD) systems. As DoG filters are constructed from the subtraction of two bivariate Gaussia...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2358326

    authors: Catarious DM Jr,Baydush AH,Floyd CE Jr

    更新日期:2006-11-01 00:00:00

  • Validation and application of a fast Monte Carlo algorithm for assessing the clinical impact of approximations in analytical dose calculations for pencil beam scanning proton therapy.

    abstract:PURPOSE:Monte Carlo (MC) dose calculation is generally superior to analytical dose calculation (ADC) used in commercial TPS to model the dose distribution especially for heterogeneous sites, such as lung and head/neck patients. The purpose of this study was to provide a validated, fast, and open-source MC code, MCsquar...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13231

    authors: Huang S,Souris K,Li S,Kang M,Barragan Montero AM,Janssens G,Lin A,Garver E,Ainsley C,Taylor P,Xiao Y,Lin L

    更新日期:2018-12-01 00:00:00

  • Comparison of two methods for accurate measurement of modulation transfer functions of screen-film systems.

    abstract::The modulation transfer function (MTF) of a screen-film system can be measured by two methods, i.e., a slit method with Fourier transform on the line spread function and a square-wave response function (SWRF) method. However, it is still uncertain whether MTFs obtained by the two methods are identical. In this study, ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597456

    authors: Morishita J,Doi K,Bollen R,Bunch PC,Hoeschen D,Sirand-rey G,Sukenobu Y

    更新日期:1995-02-01 00:00:00