Implementing RapidArc into clinical routine: a comprehensive program from machine QA to TPS validation and patient QA.

Abstract:

PURPOSE:With the increased commercial availability of intensity modulated arc therapy (IMAT) comes the need for comprehensive QA programs, covering the different aspects of this newly available technology. This manuscript proposes such a program for the RapidArc (RA) (Varian Medical Systems, Palo Alto) IMAT solution. METHODS:The program was developed and tested out for a Millennium120 MLC on iX Clinacs and a HighDefinition MLC on a Novalis TX, using a variety of measurement equipment including Gafchromic film, 2D ion chamber arrays (Seven29 and StarCheck, PTW, Freiburg, Germany) with inclinometer and Octavius phantom, the Delta4 systam (ScandiDos, Uppsala, Sweden) and the portal imager (EPID). First, a number of complementary machine QA tests were developed to monitor the correct interplay between the accelerating/decelerating gantry, the variable dose rate and the MLC position, straining the delivery to the maximum allowed limits. Second, a systematic approach to the validation of the dose calculation for RA was adopted, starting with static gantry and RA specific static MLC shapes and gradually moving to dynamic gantry, dynamic MLC shapes. RA plans were then optimized on a series of artificial structures created within the homogeneous Octavius phantom and within a heterogeneous lung phantom. These served the double purpose of testing the behavior of the optimization algorithm (PRO) as well as the precision of the forward dose calculation. Finally, patient QA on a series of clinical cases was performed with different methods. In addition to the well established in-phantom QA, we evaluated the portal dosimetry solution within the Varian approach. RESULTS:For routine machine QA, the "Snooker Cue" test on the EPID proved to be the most sensitive to overall problem detection. It is also the most practical one. The "Twinkle" and "Sunrise" tests were useful to obtain well differentiated information on the individual treatment delivery components. The AAA8.9 dose calculations showed excellent agreement with all corresponding measurements, except in areas where the 2.5 mm fixed fluence resolution was insufficient to accurately model the tongue and groove effect or the dose through nearly closed opposing leafs. Such cases benefited from the increased fluence resolution in AAA10.0. In the clinical RA fields, these effects were smeared out spatially and the impact of the fluence resolution was considerably less pronounced. The RA plans on the artificial structure sets demonstrated some interesting characteristics of the PRO8.9 optimizer, such as a sometimes unexpected dependence on the collimator rotation and a suboptimal coverage of targets within lung tissue. Although the portal dosimetry was successfully validated, we are reluctant to use it as a sole means of patient QA as long as no gantry angle information is embedded. CONCLUSIONS:The all-in validation program allows a systematic approach in monitoring the different levels of RA treatments. With the systematic approach comes a better understanding of both the capabilities and the limits of the used solution. The program can be useful for implementation, but also for the validation of major upgrades.

journal_name

Med Phys

journal_title

Medical physics

authors

Van Esch A,Huyskens DP,Behrens CF,Samsoe E,Sjolin M,Bjelkengren U,Sjostrom D,Clermont C,Hambach L,Sergent F

doi

10.1118/1.3622672

subject

Has Abstract

pub_date

2011-09-01 00:00:00

pages

5146-66

issue

9

eissn

0094-2405

issn

2473-4209

journal_volume

38

pub_type

杂志文章
  • Material-specific analysis using coherent-scatter imaging.

    abstract::Coherent-scatter computed tomography (CSCT) is a novel imaging method we are developing to produce cross-sectional images based on the low-angle (<10 degrees) scatter properties of tissue. At diagnostic energies, this scatter is primarily coherent with properties dependent upon the molecular structure of the scatterer...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1493216

    authors: Batchelar DL,Cunningham IA

    更新日期:2002-08-01 00:00:00

  • MO-F-BRA-02: Evaluation of 4D CT to 4D Cone-Beam CT Deformable Image Registration for Lung Cancer Adaptive Radiation Therapy.

    abstract:PURPOSE:To evaluate two deformable image registration (DIR) algorithms for the purpose of contour mapping to support image guided adaptive radiotherapy (IGART) with 4D cone beam CT (4DCBCT). METHODS:Eleven locally advanced non-small cell lung cancer (NSCLC) patients underwent one planning 4D fan- beam CT (4DFBCT) and ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735821

    authors: Balik S,Hugo G,Weiss E,Jan N,Roman N,Sleeman W,Fatyga M,Christensen G,Murphy M,Lu J,Keall P,Williamson J

    更新日期:2012-06-01 00:00:00

  • Resolution of pinhole collimators used with gamma cameras.

    abstract::The line spread function (LSF) of an imaging system incorporating a pinhole of 0.275 cm radius and a scintillation camera has been measured and its full width at half maximum (FWHM) compared to the values predicted by an expression based on the resolution index of Anger. In most cases, the actual resolution of the sys...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594755

    authors: Goulet P,Joy ML

    更新日期:1980-09-01 00:00:00

  • SU-E-J-84: Investigating the Use of Deformable Algorithms to Register and Accumulate Dose to the Spinal Cord for Head-And-Neck Radiotherapy Patients.

    abstract:PURPOSE:Mega-voltage cone-beam CT (MVCBCT) imaging can be used to monitor changes in neck flexion and calculate the true spinal cord dose during radiotherapy. To sum the dose over multiple fractions and compare with the planned dose requires deformable image registration. This study investigates the accuracy of deforma...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734919

    authors: Ueda U,Chen J,Kirby N,Quivey J,Yom S,Pouliot J

    更新日期:2012-06-01 00:00:00

  • Technical Note: 4D cone-beam CT reconstruction from sparse-view CBCT data for daily motion assessment in pencil beam scanned proton therapy (PBS-PT).

    abstract:PURPOSE:The number of pencil beam scanned proton therapy (PBS-PT) facilities equipped with cone-beam computed tomography (CBCT) imaging treating thoracic indications is constantly rising. To enable daily internal motion monitoring during PBS-PT treatments of thoracic tumors, we assess the performance of Motion-Aware Re...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14521

    authors: den Otter LA,Chen K,Janssens G,Meijers A,Both S,Langendijk JA,Rosen LR,Wu HT,Knopf AC

    更新日期:2020-12-01 00:00:00

  • Monte Carlo simulation of diagnostic x-ray scatter.

    abstract::A Monte Carlo method was developed and implemented to simulate x-ray photon transport. Simulations consisted of a pencil beam of monoenergetic photons with energies from 50 to 110 keV incident on water and aluminum slabs. The dependence of scatter fraction and multiple scattering on x-ray energy, scatterer thickness, ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596268

    authors: Papin PJ,Rielly PS

    更新日期:1988-11-01 00:00:00

  • Direct-response ultraviolet thermoluminescent dosimeter.

    abstract::An integrating solid-state dosimeter has been developed to detect ultraviolet radiation using the process of "direct" ultraviolet stimulation. This process yields a thermoluminescent signal after an exposure to ultraviolet radiation, without prior exposure to ionizing radiation. Magnesium oxide crystals were found to ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594259

    authors: Dhar A,DeWerd LA,Stoebe TG

    更新日期:1976-11-01 00:00:00

  • A four-alternative forced choice (4AFC) methodology for evaluating microcalcification detection in clinical full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) systems using an inkjet-printed anthropomorphic phantom.

    abstract:PURPOSE:The advent of three-dimensional breast imaging systems such as digital breast tomosynthesis (DBT) has great promise for improving the detection and diagnosis of breast cancer. With these new technologies comes an essential need for testing methods to assess the resultant image quality. Although randomized clini...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13629

    authors: Ikejimba LC,Salad J,Graff CG,Ghammraoui B,Cheng WC,Lo JY,Glick SJ

    更新日期:2019-09-01 00:00:00

  • An optimized ultrasound detector for photoacoustic breast tomography.

    abstract:PURPOSE:Photoacoustic imaging has proven to be able to detect vascularization-driven optical absorption contrast associated with tumors. In order to detect breast tumors located a few centimeter deep in tissue, a sensitive ultrasound detector is of crucial importance for photoacoustic mammography. Further, because the ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4792462

    authors: Xia W,Piras D,van Hespen JC,van Veldhoven S,Prins C,van Leeuwen TG,Steenbergen W,Manohar S

    更新日期:2013-03-01 00:00:00

  • Correlation between physical measurements and observer evaluations of image quality in digital chest radiography.

    abstract:PURPOSE:The aim of this paper was to investigate the relationship between the physical and subjective (observer) image quality metrics in digital chest radiography. METHODS:Five digital radiography systems, four with indirect flat panel detector and one with storage phosphor-based computed radiography system, were use...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14244

    authors: Yalcin A,Olgar T,Sancak T,Atac GK,Akyar S

    更新日期:2020-09-01 00:00:00

  • A Monte Carlo investigation of low-Z target image quality generated in a linear accelerator using Varian's VirtuaLinac.

    abstract:PURPOSE:The focus of this work was the demonstration and validation of VirtuaLinac with clinical photon beams and to investigate the implementation of low-Z targets in a TrueBeam linear accelerator (Linac) using Monte Carlo modeling. METHODS:VirtuaLinac, a cloud based web application utilizing Geant4 Monte Carlo code,...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4861818

    authors: Parsons D,Robar JL,Sawkey D

    更新日期:2014-02-01 00:00:00

  • Technical Note: exploring the limit for the conversion of energy-subtracted CT number to electron density for high-atomic-number materials.

    abstract:PURPOSE:For accurate tissue inhomogeneity correction in radiotherapy treatment planning, the authors had previously proposed a novel conversion of the energy-subtracted CT number to an electron density (ΔHU-ρe conversion), which provides a single linear relationship between ΔHU and ρe over a wide ρe range. The purpose ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4881327

    authors: Saito M,Tsukihara M

    更新日期:2014-07-01 00:00:00

  • On the accuracy of techniques for obtaining the calibration coefficient N(K) of 192Ir HDR brachytherapy sources.

    abstract::The accuracy of interpolation or averaging procedures for obtaining the calibration coefficient N(K) for 192Ir high-dose-rate brachytherapy sources has been investigated using the EGSnrc Monte Carlo simulation system. It is shown that the widely used two-point averaging procedure of Goetsch et al. [Med. Phys. 18, 462 ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2239198

    authors: Mainegra-Hing E,Rogers DW

    更新日期:2006-09-01 00:00:00

  • Technical Note: Virtual phantom analyses for preprocessing evaluation and detection of a robust feature set for MRI-radiomics of the brain.

    abstract:PURPOSE:The purpose of the paper was to use a virtual phantom to identify a set of radiomic features from T1-weighted and T2-weighted magnetic resonance imaging (MRI) of the brain which is stable to variations in image acquisition parameters and to evaluate the effect of image preprocessing on radiomic features stabili...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13834

    authors: Bologna M,Corino V,Mainardi L

    更新日期:2019-11-01 00:00:00

  • Four-dimensional cone-beam computed tomography and digital tomosynthesis reconstructions using respiratory signals extracted from transcutaneously inserted metal markers for liver SBRT.

    abstract:PURPOSE:Respiration-induced intrafraction target motion is a concern in liver cancer radiotherapy, especially in stereotactic body radiotherapy (SBRT), and therefore, verification of its motion is necessary. An effective means to localize the liver cancer is to insert metal fiducial markers to or near the tumor with si...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3544369

    authors: Park JC,Park SH,Kim JH,Yoon SM,Kim SS,Kim JS,Liu Z,Watkins T,Song WY

    更新日期:2011-02-01 00:00:00

  • Evaluating IMRT and VMAT dose accuracy: practical examples of failure to detect systematic errors when applying a commonly used metric and action levels.

    abstract:PURPOSE:This study (1) examines a variety of real-world cases where systematic errors were not detected by widely accepted methods for IMRT/VMAT dosimetric accuracy evaluation, and (2) drills-down to identify failure modes and their corresponding means for detection, diagnosis, and mitigation. The primary goal of detai...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4826166

    authors: Nelms BE,Chan MF,Jarry G,Lemire M,Lowden J,Hampton C,Feygelman V

    更新日期:2013-11-01 00:00:00

  • Noise, resolution, and sensitivity considerations in the design of a single-slice emission-transmission computed tomographic system.

    abstract::A prototype Emission-Transmission Computed Tomography (ETCT) system is being developed that will acquire single-slice x-ray transmission CT images simultaneously with single photon emission computed tomography (SPECT) images. This system will permit the correlation of anatomical information from x-ray CT with function...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596643

    authors: Liew SC,Hasegawa BH

    更新日期:1991-09-01 00:00:00

  • Effects on image quality of a 2D antiscatter grid in x-ray digital breast tomosynthesis: Initial experience using the dual modality (x-ray and molecular) breast tomosynthesis scanner.

    abstract:PURPOSE:Radiation scattered from the breast in digital breast tomosynthesis (DBT) causes image degradation, including loss of contrast between cancerous and background tissue. Unlike in 2-dimensional (2D) mammography, an antiscatter grid cannot readily be used in DBT because changing alignment between the tube and dete...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4943632

    authors: Patel T,Peppard H,Williams MB

    更新日期:2016-04-01 00:00:00

  • Low-dose CT reconstruction using spatially encoded nonlocal penalty.

    abstract:PURPOSE:Computed tomography (CT) is one of the most used imaging modalities for imaging both symptomatic and asymptomatic patients. However, because of the high demand for lower radiation dose during CT scans, the reconstructed image can suffer from noise and artifacts due to the trade-off between the image quality and...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12523

    authors: Kim K,El Fakhri G,Li Q

    更新日期:2017-10-01 00:00:00

  • Primary dose in photon beams with lateral electron disequilibrium.

    abstract::It is shown that in narrow monoenergetic photon beams under conditions of lateral electron disequilibrium, primary absorbed dose P(r) is a simple function of beam radius r: P(r) = P lambda.(1 - e(-gamma.r)), where P lambda is the primary dose in broad beams for which complete lateral electron equilibrium exists, and g...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596741

    authors: Nizin PS,Chang XS

    更新日期:1991-07-01 00:00:00

  • Trajectory analysis for field free line magnetic particle imaging.

    abstract:PURPOSE:Magnetic particle imaging (MPI) is a relatively new method to image the spatial distribution of magnetic nanoparticle (MNP) tracers administered to the body with high spatial and temporal resolution using an inhomogeneous magnetic field. The spatial information of the MNP's is encoded using a field free point (...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13411

    authors: Top CB,Güngör A,Ilbey S,Güven HE

    更新日期:2019-04-01 00:00:00

  • A scintillator-based approach to monitor secondary neutron production during proton therapy.

    abstract:PURPOSE:The primary objective of this work is to measure the secondary neutron field produced by an uncollimated proton pencil beam impinging on different tissue-equivalent phantom materials using organic scintillation detectors. Additionally, the Monte Carlo code mcnpx-PoliMi was used to simulate the detector response...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4963813

    authors: Clarke SD,Pryser E,Wieger BM,Pozzi SA,Haelg RA,Bashkirov VA,Schulte RW

    更新日期:2016-11-01 00:00:00

  • Robust calculation of effective atomic numbers: the Auto-Z(eff) software.

    abstract:PURPOSE:The most appropriate method of evaluating the effective atomic number necessitates consideration of energy-dependent behavior. Previously, this required quite laborious calculation, which is why many scientists revert to over-simplistic power-law methods. The purpose of this work is to develop user-friendly sof...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3689810

    authors: Taylor ML,Smith RL,Dossing F,Franich RD

    更新日期:2012-04-01 00:00:00

  • Tomotherapy dose distribution verification using MAGIC-f polymer gel dosimetry.

    abstract:PURPOSE:This paper presents the application of MAGIC-f gel in a three-dimensional dose distribution measurement and its ability to accurately measure the dose distribution from a tomotherapy unit. METHODS:A prostate intensity-modulated radiation therapy (IMRT) irradiation was simulated in the gel phantom and the treat...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4704496

    authors: Pavoni JF,Pike TL,Snow J,DeWerd L,Baffa O

    更新日期:2012-05-01 00:00:00

  • Electron backscatter corrections for parallel-plate chambers.

    abstract::The wall of an ionization chamber is commonly assumed to have a negligible effect on chamber response in electron beams. For cylindrical chambers with thin walls this assumption is valid. However, parallel-plate chambers commonly possess large mechanical supports which may affect chamber response in a manner not accou...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596165

    authors: Hunt MA,Kutcher GJ,Buffa A

    更新日期:1988-01-01 00:00:00

  • Predicting radiotherapy-induced cardiac perfusion defects.

    abstract::The purpose of this work is to compare the efficacy of mathematical models in predicting the occurrence of radiotherapy-induced left ventricular perfusion defects assessed using single-photon emission computed tomography (SPECT). The basis of this study is data from 73 left-sided breast/ chestwall patients treated wit...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1823571

    authors: Das SK,Baydush AH,Zhou S,Miften M,Yu X,Craciunescu O,Oldham M,Light K,Wong T,Blazing M,Borges-Neto S,Dewhirst MW,Marks LB

    更新日期:2005-01-01 00:00:00

  • Performance evaluation of a diode array for enhanced dynamic wedge dosimetry.

    abstract::The performance of a diode array (Profiler) was evaluated by comparing its enhanced dynamic wedge (EDW) profiles measured at various depths with point measurements using a 0.03 cm3 ionization chamber on a commercial linear accelerator. The Profiler, which covers a 22.5 cm width, was used to measure larger field widths...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598019

    authors: Zhu TC,Ding L,Liu CR,Palta JR,Simon WE,Shi J

    更新日期:1997-07-01 00:00:00

  • A novel computer-aided lung nodule detection system for CT images.

    abstract:PURPOSE:The paper presents a complete computer-aided detection (CAD) system for the detection of lung nodules in computed tomography images. A new mixed feature selection and classification methodology is applied for the first time on a difficult medical image analysis problem. METHODS:The CAD system was trained and t...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3633941

    authors: Tan M,Deklerck R,Jansen B,Bister M,Cornelis J

    更新日期:2011-10-01 00:00:00

  • Potential use of P-32 ophthalmic applicator: Monte Carlo simulations for design and dosimetry.

    abstract::Postoperative beta-irradiation after pterygium excision has been considered a valuable therapeutic procedure to reduce the recurrence rate. Recently, it was reported that beta-irradiation also substantially reduced the risk of surgical failure after glaucoma surgery. Pure beta-irradiation using a 90Sr/Y applicator has...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2896076

    authors: Park YK,Ye SJ,Kim IH,Wee WR,Kim MK,Han HS,Son KJ,Park UJ

    更新日期:2008-05-01 00:00:00

  • The role of humidity and other correction factors in the AAPM TG-21 dosimetry protocol.

    abstract::A detailed derivation is presented of the formulas required to determine Ngas and Dmed in the AAPM TG-21 dosimetry protocol. This protocol specifies how to determine the absorbed dose in an electron or photon beam when using exposure or absorbed dose calibrated ion chambers. It is shown that the expression given in TG...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596292

    authors: Rogers DW,Ross CK

    更新日期:1988-01-01 00:00:00