On the accuracy of techniques for obtaining the calibration coefficient N(K) of 192Ir HDR brachytherapy sources.

Abstract:

:The accuracy of interpolation or averaging procedures for obtaining the calibration coefficient N(K) for 192Ir high-dose-rate brachytherapy sources has been investigated using the EGSnrc Monte Carlo simulation system. It is shown that the widely used two-point averaging procedure of Goetsch et al. [Med. Phys. 18, 462 (1991)] has some conceptual problems. Most importantly, they recommended, as did the IAEA, averaging A(wall)N(K) values whereas one should average 1/N(K) values. In practice this and other issues are shown to have little effect except for Goetsch et al.'s methods for determining A(wall) values. Their method of generalizing the A(wall) values measured in one geometry to other geometries is incorrect by up to 2%. However, these errors in A(wall) values cause systematic errors of only 0.3% in 192Ir calibration coefficients. It is shown that A(wall) values need not be included in the averaging technique at all, thereby simplifying the technique considerably. It is demonstrated that as long as ion chambers with a flat response are used and/or very heavily filtered 250 kV (or higher) beams of x rays are used in the averaging, then almost all techniques can provide adequate accuracy.

journal_name

Med Phys

journal_title

Medical physics

authors

Mainegra-Hing E,Rogers DW

doi

10.1118/1.2239198

subject

Has Abstract

pub_date

2006-09-01 00:00:00

pages

3340-7

issue

9

eissn

0094-2405

issn

2473-4209

journal_volume

33

pub_type

杂志文章
  • WE-F-211-01: The Evolving Landscape of Scientific Publishing.

    abstract::The dissemination of scientific advances has changed little since the first peer-reviewed journal was published in 1665 - that is, until this past decade. The print journal, delivered by mail and stored on office shelves and in library reading rooms around the world, has been transformed by immediate, on-demand access...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4736173

    authors: Armato S,Hendee W,Marshall C,Curran B

    更新日期:2012-06-01 00:00:00

  • Dose planning with comparison to in vivo dosimetry for epithermal neutron irradiation of the dog brain.

    abstract::Boron neutron capture therapy (BNCT) is an experimental type of radiotherapy, presently being used to treat glioblastoma and melanoma. To improve patient safety and to determine the radiobiological characteristics of the epithermal neutron beam of Finnish BNCT facility (FiR 1) dose-response studies were carried on the...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1517048

    authors: Seppälä T,Auterinen I,Aschan C,Serén T,Benczik J,Snellman M,Huiskamp R,Ramadan UA,Kankaanranta L,Joensuu H,Savolainen S

    更新日期:2002-11-01 00:00:00

  • A stochastic approach to estimate the uncertainty of dose mapping caused by uncertainties in b-spline registration.

    abstract:PURPOSE:In fractionated radiation therapy, image guidance with daily tomographic imaging becomes more and more clinical routine. In principle, this allows for daily computation of the delivered dose and for accumulation of these daily dose distributions to determine the actually delivered total dose to the patient. How...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3697524

    authors: Hub M,Thieke C,Kessler ML,Karger CP

    更新日期:2012-04-01 00:00:00

  • Upgrade and benchmarking of a 4D treatment planning system for scanned ion beam therapy.

    abstract:PURPOSE:Upgrade and benchmarking of a research 4D treatment planning system (4DTPS) suitable for realistic patient treatment planning and treatment simulations taking into account specific requirements for scanned ion beam therapy, i.e., modeling of dose heterogeneities due to interplay effects and range changes caused...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4800802

    authors: Richter D,Schwarzkopf A,Trautmann J,Krämer M,Durante M,Jäkel O,Bert C

    更新日期:2013-05-01 00:00:00

  • SU-E-I-94: External Beam Radiation Cherenkov Emission in Tissue Used for Tissue Oxygen Sensing.

    abstract:PURPOSE:To show that Cherenkov emission is generated by external radiotherapy beam in tissue, and could serve as optical source to excite an oxygen sensitive phosphor, Oxyphor G4, within tissue. The intensity and lifetime of the phosphorescence was measured with a time-gated system and reveals the oxygenation levels in...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734811

    authors: Zhang R,Kanick S,Vinogradov S,Esipova T,Pogue B

    更新日期:2012-06-01 00:00:00

  • An empirical model for independent dose verification of the Gamma Knife treatment planning.

    abstract::A formalism for an independent dose verification of the Gamma Knife treatment planning is developed. It is based on the approximation that isodose distribution for a single shot is in the shape of an ellipsoid in three-dimensional space. The dose profiles for a phantom along each of the three major axes are fitted to ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1501139

    authors: Phaisangittisakul N,Ma L

    更新日期:2002-09-01 00:00:00

  • A comparative contrast-detail study of five medical displays.

    abstract::The objective of this study was to compare the contrast-detail performance of five different commercial liquid crystal displays (LCDs) to other LCD and cathode-ray tube (CRT) displays for medical applications. A contrast-detail phantom, supplemented with 5 in. of acrylic, was imaged on a commercial digital radiographi...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2868780

    authors: Samei E,Ranger NT,Delong DM

    更新日期:2008-04-01 00:00:00

  • Development of PET/MRI with insertable PET for simultaneous PET and MR imaging of human brain.

    abstract:PURPOSE:The purpose of this study was to develop a dual-modality positron emission tomography (PET)/magnetic resonance imaging (MRI) with insertable PET for simultaneous PET and MR imaging of the human brain. METHODS:The PET detector block was composed of a 4 × 4 matrix of detector modules, each consisting of a 4 × 4 ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4918321

    authors: Jung JH,Choi Y,Jung J,Kim S,Lim HK,Im KC,Oh CH,Park HW,Kim KM,Kim JG

    更新日期:2015-05-01 00:00:00

  • MO-F-BRCD-01: Stereotactic Body Radiation Therapy: Updates on Clinical, Biological, and Physics/QA:SBRT (Part 1): Biological and Clinical Updates.

    abstract::Stereotactic Body Radiation Therapy (SBRT) is an important form of cancer therapy with increasingly broad application across a spectrum of tumor types in primary and metastatic settings. In this presentation the radiation biology, clinical experience from various trials, and cautionary updates on normal tissue toleran...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735812

    authors: Read P

    更新日期:2012-06-01 00:00:00

  • Quantitative cone-beam CT imaging in radiation therapy using planning CT as a prior: first patient studies.

    abstract:PURPOSE:Quantitative cone-beam CT (CBCT) imaging is on increasing demand for high-performance image guided radiation therapy (IGRT). However, the current CBCT has poor image qualities mainly due to scatter contamination. Its current clinical application is therefore limited to patient setup based on only bony structure...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3693050

    authors: Niu T,Al-Basheer A,Zhu L

    更新日期:2012-04-01 00:00:00

  • Traceable micro-CT scaling accuracy phantom for applications requiring exact measurement of distances or volumes.

    abstract:PURPOSE:Volumetric x-ray microcomputed tomography (CT) can be employed in a variety of quantitative research applications such as image-guided interventions or characterization of medical devices. To ensure the highest geometric fidelity of images for these applications, a phantom and image processing algorithm have be...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4752083

    authors: Waring CS,Bax JS,Samarabandu A,Holdsworth DW,Fenster A,Lacefield JC

    更新日期:2012-10-01 00:00:00

  • A sparse orthogonal collimator for small animal intensity-modulated radiation therapy part I: Planning system development and commissioning.

    abstract:PURPOSE:To achieve more translatable preclinical research results, small animal irradiation needs to more closely simulate human radiotherapy. Although the clinical gold standard is intensity-modulated radiation therapy (IMRT), the direct translation of this method for small animals is impractical. In this study we des...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13872

    authors: Woods K,Nguyen D,Neph R,Ruan D,O'Connor D,Sheng K

    更新日期:2019-12-01 00:00:00

  • Predicting the noise in hybrid (phase and attenuation) x-ray images acquired with the edge illumination technique.

    abstract:PURPOSE:To analyze the noise performance of the edge illumination phase-based x-ray imaging technique when applying "single-shot" phase retrieval. The latter consists in applying a sample-specific low-pass filter to the raw data, leading to "hybrid" images in which phase and attenuation contrast are merged with each ot...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14366

    authors: Hagen CK,Roche I Morgó O,Olivo A

    更新日期:2020-09-01 00:00:00

  • Evaluation of the local dose enhancement in the combination of proton therapy and nanoparticles.

    abstract:PURPOSE:The outcome of radiotherapy can be further improved by combining irradiation with dose enhancers such as high-Z nanoparticles. Since 2004, spectacular results have been obtained when low-energy x-ray irradiations have been combined with nanoparticles. Recently, the same combination has been explored in hadron t...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4934370

    authors: Martínez-Rovira I,Prezado Y

    更新日期:2015-11-01 00:00:00

  • Accuracy of Dose-Volume Metric Calculation for Small Volume Radiosurgery Targets.

    abstract:PURPOSE:For stereotactic radiosurgery (SRS), accurate evaluation of dose-volume metrics for small structures is necessary. The purpose of this study was to compare the DVH metric capabilities of five commercially available SRS DVH analysis tools (Eclipse, Elements, Raystation, MIM, and Velocity). METHODS:DICOM RTdose ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14645

    authors: Stanley DN,Covington E,Liu H,Alexandrian A,Cardan R,Bridges D,Thomas EM,Fiveash JB,Popple R

    更新日期:2020-12-09 00:00:00

  • The generalized geometry of eye plaque therapy.

    abstract::A calculation is described that enables the rapid assessment of dose rate at various points of interest within the eye (lens, optic nerve, etc.) for the treatment of choroidal melanoma by plaque therapy. 125I seeds are used as the radiation source. The location of the plaque and its associated seeds relative to the ey...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596234

    authors: Kepka AG,Johnson PM,Kline RW

    更新日期:1988-05-01 00:00:00

  • JADA: a graphical user interface for comprehensive internal dose assessment in nuclear medicine.

    abstract:PURPOSE:The main objective of this work was to design a comprehensive dosimetry package that would keep all aspects of internal dose calculation within the framework of a single software environment and that would be applicable for a variety of dose calculation approaches. METHODS:Our MATLAB-based graphical user inter...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4810963

    authors: Grimes J,Uribe C,Celler A

    更新日期:2013-07-01 00:00:00

  • Noninvasive eye localization in ocular proton therapy through optical eye tracking: A proof of concept.

    abstract:PURPOSE:Over the last four decades, Ocular Proton Therapy has been successfully used to treat patients affected by intraocular lesions. For this, treatment geometry verification is routinely performed using radiographic images to align a configuration of fiducial radiopaque markers implanted on the sclera outer surface...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12841

    authors: Via R,Hennings F,Fattori G,Fassi A,Pica A,Lomax A,Weber DC,Baroni G,Hrbacek J

    更新日期:2018-05-01 00:00:00

  • Development of a flexible optical fiber based high resolution integrated PET∕MRI system.

    abstract:PURPOSE:The simultaneous measurement of PET and magnetic resonance imaging (MRI) is an emerging field for molecular imaging research. Although optical fiber based PET∕MRI systems have advantages on less interference between PET and MRI, there is a drawback in reducing the scintillation light due to the fiber. To reduce...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4757911

    authors: Yamamoto S,Watabe H,Kanai Y,Watabe T,Aoki M,Sugiyama E,Kato K,Hatazawa J

    更新日期:2012-11-01 00:00:00

  • Design and evaluation of an MRI-compatible linear motion stage.

    abstract:PURPOSE:To develop and evaluate a tool for accurate, reproducible, and programmable motion control of imaging phantoms for use in motion sensitive magnetic resonance imaging (MRI) appli cations. METHODS:In this paper, the authors introduce a compact linear motion stage that is made of nonmagnetic material and is actua...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4937780

    authors: Tavallaei MA,Johnson PM,Liu J,Drangova M

    更新日期:2016-01-01 00:00:00

  • Discriminating solitary cysts from soft tissue lesions in mammography using a pretrained deep convolutional neural network.

    abstract:PURPOSE:It is estimated that 7% of women in the western world will develop palpable breast cysts in their lifetime. Even though cysts have been correlated with risk of developing breast cancer, many of them are benign and do not require follow-up. We develop a method to discriminate benign solitary cysts from malignant...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12110

    authors: Kooi T,van Ginneken B,Karssemeijer N,den Heeten A

    更新日期:2017-03-01 00:00:00

  • Pulmonary nodule detection in CT images based on shape constraint CV model.

    abstract:PURPOSE:Accurate detection of pulmonary nodules remains a technical challenge in computer-aided diagnosis systems because some nodules may adhere to the blood vessels or the lung wall, which have low contrast compared to the surrounding tissues. In this paper, the analysis of typical shape features of candidate nodules...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4907961

    authors: Wang B,Tian X,Wang Q,Yang Y,Xie H,Zhang S,Gu L

    更新日期:2015-03-01 00:00:00

  • Resolution of pinhole collimators used with gamma cameras.

    abstract::The line spread function (LSF) of an imaging system incorporating a pinhole of 0.275 cm radius and a scintillation camera has been measured and its full width at half maximum (FWHM) compared to the values predicted by an expression based on the resolution index of Anger. In most cases, the actual resolution of the sys...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594755

    authors: Goulet P,Joy ML

    更新日期:1980-09-01 00:00:00

  • Fast IMRT with narrow high energy scanned photon beams.

    abstract:PURPOSE:Since the first publications on intensity modulated radiation therapy (IMRT) in the early 1980s almost all efforts have been focused on fairly time consuming dynamic or segmental multileaf collimation. With narrow fast scanned photon beams, the flexibility and accuracy in beam shaping increases, not least in co...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3615059

    authors: Andreassen B,Strååt SJ,Holmberg R,Näfstadius P,Brahme A

    更新日期:2011-08-01 00:00:00

  • Optimized low-kV spectrum of dual-energy CT equipped with high-kV tin filtration for electron density measurements.

    abstract:PURPOSE:This paper describes the low-kV spectral optimization of dual-energy CT (DECT) equipped with high-kV tin filtration for the quantitative acquisition of electron density information, which is essential for treatment planning in radiotherapy. In addition, an analytical DECT image simulation was preliminarily perf...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3584200

    authors: Saito M

    更新日期:2011-06-01 00:00:00

  • Spectra and air-kerma strength for encapsulated 192Ir sources.

    abstract::The photon spectra in vacuum around four types of 192Ir HDR brachytherapy sources are calculated using the Monte Carlo code EGS4 and the most recent spectral information on 192Ir decay. The air-kerma strengths per unit activity are calculated based on the photon fluence around a bare 192Ir source and around each of fo...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598763

    authors: Borg J,Rogers DW

    更新日期:1999-11-01 00:00:00

  • Characterization of electron contamination in megavoltage photon beams.

    abstract::The purpose of the present study is to characterize electron contamination in photon beams in different clinical situations. Variations with field size, beam modifier (tray, shaping block) and source-surface distance (SSD) were studied. Percentage depth dose measurements with and without a purging magnet and replacing...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1895793

    authors: Lopez Medina A,Teijeiro A,Garcia J,Esperon J,Terron JA,Ruiz DP,Carrion MC

    更新日期:2005-05-01 00:00:00

  • Isotope specific resolution recovery image reconstruction in high resolution PET imaging.

    abstract:PURPOSE:Measuring and incorporating a scanner-specific point spread function (PSF) within image reconstruction has been shown to improve spatial resolution in PET. However, due to the short half-life of clinically used isotopes, other long-lived isotopes not used in clinical practice are used to perform the PSF measure...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4870985

    authors: Kotasidis FA,Angelis GI,Anton-Rodriguez J,Matthews JC,Reader AJ,Zaidi H

    更新日期:2014-05-01 00:00:00

  • A Monte Carlo estimation of effective dose in chest tomosynthesis.

    abstract:PURPOSE:The recent introduction of digital tomosynthesis imaging into routine clinical use has enabled the acquisition of volumetric patient data within a standard radiographic examination. Tomosynthesis requires the acquisition of multiple projection views, requiring additional dose compared to a standard projection e...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3250907

    authors: Sabol JM

    更新日期:2009-12-01 00:00:00

  • Projection-based improvement of 3D reconstructions from motion-impaired dental cone beam CT data.

    abstract:PURPOSE:Computed tomography (CT) and, in particular, cone beam CT (CBCT) have been increasingly used as a diagnostic tool in recent years. Patient motion during acquisition is common in CBCT due to long scan times. This results in degraded image quality and may potentially increase the number of retakes. Our aim was to...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13731

    authors: Niebler S,Schömer E,Tjaden H,Schwanecke U,Schulze R

    更新日期:2019-10-01 00:00:00