Pulmonary embolism in pregnant patients: Assessing organ dose to pregnant phantom and its fetus during lung imaging.

Abstract:

PURPOSE:The purpose of this study was to provide updated radiation dose from diagnostic exams performed for pregnant patients suspected of pulmonary embolism (PE) using the recently developed BREP phantoms of pregnant woman and the fetus. Also to challenge the validity of current recommendations suggest that ventilation/perfusion (V/Q) single photon emission computed tomography (SPECT) vs. computed tomography pulmonary angiography (CTPA) should be considered for diagnosis of PE in radiosensitive groups such as pregnant women. METHODS:The Monte Carlo calculations involving detailed geometrical simulation of pregnant women and the fetus were performed. RESULTS:The results showed that when radiation dose to the fetus is of concern, CTPA is more appropriate at early stages causes 50%-97% lower fetal doses for the first two trimesters of pregnancy. While for gestational periods more than 6 months, V/Q SPECT leads to a 15% lower fetal dose and thus, is less hazardous. The fetal dose from CTPA increases with gestational age, while that from V/Q SPECT decreases. Furthermore, the maximum amount of fetal dose is received by fetal skeleton (i.e., on average about 1.8 and 3.9 times larger dose from SPECT and CT, respectively). CONCLUSIONS:V/Q SPECT should not always be preferred for pregnant patients suspected of PE. This finding is in contrast with the guidance to choose the preferred modality based on the maternal effective dose. The reason of this issue was discussed in this paper based on chord length distributions (CLDs). The importance of considering fetal organs separately in MC calculations was also highlighted.

journal_name

Med Phys

journal_title

Medical physics

authors

Rafat Motavalli L,Hoseinian Azghadi E,Miri Hakimabad H,Akhlaghi P

doi

10.1002/mp.12558

subject

Has Abstract

pub_date

2017-11-01 00:00:00

pages

6038-6046

issue

11

eissn

0094-2405

issn

2473-4209

journal_volume

44

pub_type

杂志文章
  • Technical Note: Immunohistochemical evaluation of mouse brain irradiation targeting accuracy with 3D-printed immobilization device.

    abstract:PURPOSE:Small animal immobilization devices facilitate positioning of animals for reproducible imaging and accurate focal radiation therapy. In this study, the authors demonstrate the use of three-dimensional (3D) printing technology to fabricate a custom-designed mouse head restraint. The authors evaluate the accuracy...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4933200

    authors: Zarghami N,Jensen MD,Talluri S,Foster PJ,Chambers AF,Dick FA,Wong E

    更新日期:2015-11-01 00:00:00

  • Current status and future perspective of flattening filter free photon beams.

    abstract:PURPOSE:Flattening filters (FFs) have been considered as an integral part of the treatment head of a medical accelerator for more than 50 years. The reasons for the longstanding use are, however, historical ones. Advanced treatment techniques, such as stereotactic radiotherapy or intensity modulated radiotherapy have s...

    journal_title:Medical physics

    pub_type: 杂志文章,评审

    doi:10.1118/1.3554643

    authors: Georg D,Knöös T,McClean B

    更新日期:2011-03-01 00:00:00

  • Spectrally resolved bioluminescence tomography using the reciprocity approach.

    abstract::Spectrally resolved bioluminescence optical tomography is an approach to recover images of, for example, Luciferase activity within a volume using multiwavelength emission data from internal bioluminescence sources. The underlying problem of uniqueness associated with nonspectrally resolved intensity-based bioluminesc...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2982138

    authors: Dehghani H,Davis SC,Pogue BW

    更新日期:2008-11-01 00:00:00

  • Technical note: patient-specific quality assurance methods for TomoDirect(TM) whole breast treatment delivery.

    abstract:PURPOSE:To investigate the feasibility of implementing a novel approach for patient-specific QA of TomoDirect(TM) whole breast treatment. METHODS:The most currently used TomoTherapy DQA method, consisting in the verification of the 2D dose distribution in a coronal or sagittal plane of the Cheese Phantom by means of g...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4722967

    authors: Catuzzo P,Zenone F,Aimonetto S,Peruzzo A,Casanova Borca V,Pasquino M,Franco P,La Porta MR,Ricardi U,Tofani S

    更新日期:2012-07-01 00:00:00

  • Comparison of internal dose estimates obtained using organ-level, voxel S value, and Monte Carlo techniques.

    abstract:PURPOSE:The authors' objective was to compare internal dose estimates obtained using the Organ Level Dose Assessment with Exponential Modeling (OLINDA/EXM) software, the voxel S value technique, and Monte Carlo simulation. Monte Carlo dose estimates were used as the reference standard to assess the impact of patient-sp...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4892606

    authors: Grimes J,Celler A

    更新日期:2014-09-01 00:00:00

  • Wedge-shaped dose distributions by computer-controlled collimator motion.

    abstract::We have recently installed a linear accelerator, modified to allow computer control of several machine parameters during irradiation of the patient. As an initial feasibility study of computer-controlled radiation therapy, its application to produce wedge-shaped dose distributions by moving the collimator jaws has bee...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594440

    authors: Kijewski PK,Chin LM,Bjärngard BE

    更新日期:1978-09-01 00:00:00

  • 3T MRI evaluation of the accuracy of atlas-based subthalamic nucleus identification.

    abstract::Modulation of the activity of the subthalamic nucleus (STN) using deep brain stimulation (DBS) in patients with advanced Parkinson's disease is the most common procedure performed today by functional neurosurgeons. The STN contours cannot be entirely identified on common 1.5 T images; in particular, the ventromedial b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2936229

    authors: Stancanello J,Muacevic A,Sebastiano F,Modugno N,Cerveri P,Ferrigno G,Uggeri F,Romanelli P

    更新日期:2008-07-01 00:00:00

  • System considerations in CCD-based x-ray imaging for digital chest radiography and digital mammography.

    abstract::System modeling is used to investigate the effect of various system parameters on the image quality in CCD-based x-ray imaging systems. The systems considered consist of a typical phosphor-based scintillating screen coupled to a CCD through lens or fiberoptic taper. Two applications, chest radiography and mammography,...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598072

    authors: Hejazi S,Trauernicht DP

    更新日期:1997-02-01 00:00:00

  • A new technique to characterize CT scanner bow-tie filter attenuation and applications in human cadaver dosimetry simulations.

    abstract:PURPOSE:To present a noninvasive technique for directly measuring the CT bow-tie filter attenuation with a linear array x-ray detector. METHODS:A scintillator based x-ray detector of 384 pixels, 307 mm active length, and fast data acquisition (model X-Scan 0.8c4-307, Detection Technology, FI-91100 Ii, Finland) was use...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4932364

    authors: Li X,Shi JQ,Zhang D,Singh S,Padole A,Otrakji A,Kalra MK,Xu XG,Liu B

    更新日期:2015-11-01 00:00:00

  • Fast iterative algorithms for three-dimensional inverse treatment planning.

    abstract::Three types of iterative algorithms, algebraic inverse treatment planning (AITP), simultaneous iterative inverse treatment planning (SIITP), and iterative least-square inverse treatment planning (ILSITP), differentiated according to their updating sequences, were generalized to three dimension with true beam geometry ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598374

    authors: Xing L,Hamilton RJ,Spelbring D,Pelizzari CA,Chen GT,Boyer AL

    更新日期:1998-10-01 00:00:00

  • Comparison of action levels for patient-specific quality assurance of intensity modulated radiation therapy and volumetric modulated arc therapy treatments.

    abstract:PURPOSE:To perform a comprehensive and systematic comparison of fixed-beam IMRT and volumetric modulated arc therapy (VMAT) patient-specific QA measurements for a common set of geometries using typical measurement methods. METHODS:Fixed-beam IMRT and VMAT plans were constructed for structure set geometries provided by...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4729738

    authors: Mancuso GM,Fontenot JD,Gibbons JP,Parker BC

    更新日期:2012-07-01 00:00:00

  • Measurement of energy dependence for XRCT radiochromic film.

    abstract::Gafchromic XRCT, radiochromic film is assessed over a broad energy range, from kilovoltage to megavoltage x rays for variations in reflected optical density to dose response. A large energy dependence was found with reflected optical density output for the same delivered dose varying from 7.8 +/- 0.35 at 25.5 keV (50 ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2219330

    authors: Butson MJ,Cheung T,Yu PK

    更新日期:2006-08-01 00:00:00

  • 3D image-guided robotic needle positioning system for small animal interventions.

    abstract:PURPOSE:This paper presents the design of a micro-CT guided small animal robotic needle positioning system. In order to simplify the robotic design and maintain a small targeting error, a novel implementation of the remote center of motion is used in the system. The system has been developed with the objective of achie...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4771958

    authors: Bax JS,Waring CS,Sherebrin S,Stapleton S,Hudson TJ,Jaffray DA,Lacefield JC,Fenster A

    更新日期:2013-01-01 00:00:00

  • A treatment planning study comparing HDR and AGIMRT for cervical cancer.

    abstract::The customization of brachytherapy dose distributions for gynecologic malignancies is limited by the spatial positioning of the applicators. We tested the hypothesis that applicator-guided intensity modulated radiation therapy (AGIMRT) has the potential to deliver highly conformal dose distributions to cervical tumors...

    journal_title:Medical physics

    pub_type: 临床试验,杂志文章

    doi:10.1118/1.1645678

    authors: Wahab SH,Malyapa RS,Mutic S,Grigsby PW,Deasy JO,Miller TR,Zoberi I,Low DA

    更新日期:2004-04-01 00:00:00

  • Use of proximal operator graph solver for radiation therapy inverse treatment planning.

    abstract:PURPOSE:Most radiation therapy optimization problems can be formulated as an unconstrained problem and solved efficiently by quasi-Newton methods such as the Limited-memory Broyden-Fletcher-Goldfarb-Shanno (L-BFGS) algorithm. However, several next generation planning techniques such as total variation regularization- b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12165

    authors: Liu X,Pelizzari C,Belcher AH,Grelewicz Z,Wiersma RD

    更新日期:2017-04-01 00:00:00

  • Technical Note: An investigation of polarity effects for wide-angle free-air chambers.

    abstract:PURPOSE:Wide-angle free-air chambers (WAFACs) are used as primary standard measurement devices for establishing the air-kerma strength of low-energy, low-dose rate brachytherapy seeds. The National Research Council of Canada (NRC) is commissioning a primary standard wide-angle free-air chamber (NRC WAFAC) to serve the ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4951727

    authors: Shen H,Culberson WS,Ross CK

    更新日期:2016-07-01 00:00:00

  • Future of medical physics: Real-time MRI-guided proton therapy.

    abstract::With the recent clinical implementation of real-time MRI-guided x-ray beam therapy (MRXT), attention is turning to the concept of combining real-time MRI guidance with proton beam therapy; MRI-guided proton beam therapy (MRPT). MRI guidance for proton beam therapy is expected to offer a compelling improvement to the c...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12371

    authors: Oborn BM,Dowdell S,Metcalfe PE,Crozier S,Mohan R,Keall PJ

    更新日期:2017-08-01 00:00:00

  • On the speckle-free nature of photoacoustic tomography.

    abstract:PURPOSE:A long-standing conundrum is why photoacoustic tomography (PAT) possesses the unique ability to produce images devoid of speckle artifacts while all other coherent imaging technologies do not. METHODS:In this paper, we explain the inherent mechanism that suppresses speckle in PAT, and the analysis was validate...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3187231

    authors: Guo Z,Li L,Wang LV

    更新日期:2009-09-01 00:00:00

  • Suitability of microDiamond detectors for the determination of absorbed dose to water around high-dose-rate 192 Ir brachytherapy sources.

    abstract:PURPOSE:Experimental dosimetry of high-dose-rate (HDR) 192 Ir brachytherapy (BT) sources is complicated due to high dose and dose-rate gradients, and softening of photon energy spectrum with depth. A single crystal synthetic diamond detector microDiamond (PTW 60019, Freiburg, Germany) has a small active volume, high se...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12694

    authors: Kaveckyte V,Malusek A,Benmakhlouf H,Alm Carlsson G,Carlsson Tedgren Å

    更新日期:2018-01-01 00:00:00

  • Technical Note: Gray tracking in medical color displays-A report of Task Group 196.

    abstract:PURPOSE:The authors discuss measurement methods and instrumentation useful for the characterization of the gray tracking performance of medical color monitors for diagnostic applications. The authors define gray tracking as the variability in the chromaticity of the gray levels in a color monitor. METHODS:The authors ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4953186

    authors: Badano A,Wang J,Boynton P,Le Callet P,Cheng WC,Deroo D,Flynn MJ,Matsui T,Penczek J,Revie C,Samei E,Steven PM,Swiderski S,Van Hoey G,Yamaguchi M,Hasegawa M,Nagy BV

    更新日期:2016-07-01 00:00:00

  • Half-scan cone-beam CT fluoroscopy with multiple x-ray sources.

    abstract::To develop volumetric micro-CT fluoroscopy for small animal imaging, we have proposed a cone-beam system with multiple x-ray sources. In this paper, we extend Parker's single-source half-scan weighting scheme to the case of an odd number of x-ray sources that are equiangularly distributed, and apply it for half-scan F...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1381549

    authors: Liu Y,Liu H,Wang Y,Wang G

    更新日期:2001-07-01 00:00:00

  • Age-dependent dose calculations for common PET radionuclides and brain radiotracers in nonhuman primate computational models.

    abstract:PURPOSE:The combination of nonhuman primates (NHPs) with the state-of-the-art molecular imaging technologies allows for within-subject longitudinal research aiming at gaining new insights into human normal and disease conditions and provides an ideal foundation for future translational studies of new diagnostic tools, ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14333

    authors: Xie T,Chen X,Zaidi H

    更新日期:2020-09-01 00:00:00

  • Deformable mapping technique to correlate lesions in digital breast tomosynthesis and automated breast ultrasound images.

    abstract:PURPOSE:To develop a deformable mapping technique to match corresponding lesions between digital breast tomosynthesis (DBT) and automated breast ultrasound (ABUS) images. METHODS:External fiducial markers were attached to the surface of two CIRS multi-modality compressible breast phantoms (A and B) containing multiple...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13113

    authors: Green CA,Goodsitt MM,Brock KK,Davis CL,Larson ED,Lau JH,Carson PL

    更新日期:2018-10-01 00:00:00

  • Comparison of measured and estimated maximum skin doses during CT fluoroscopy lung biopsies.

    abstract:PURPOSE:To measure patient-specific maximum skin dose (MSD) associated with CT fluoroscopy (CTF) lung biopsies and to compare measured MSD with the MSD estimated from phantom measurements, as well as with the CTDIvol of patient examinations. METHODS:Data from 50 patients with lung lesions who underwent a CT fluoroscop...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4884231

    authors: Zanca F,Jacobs A,Crijns W,De Wever W

    更新日期:2014-07-01 00:00:00

  • Combined use of transverse and scout computed tomography scans to localize radioactive seeds in an interstitial brachytherapy implant.

    abstract::Various techniques have been developed to localize radioactive sources in brachytherapy implants. The most common methods include the orthogonal film method, the stereo-shift film method, and recently, direct localization from a series of contiguous CT transverse images. The major advantage of the CT method is that it...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598548

    authors: Yue N,Chen Z,Bond JE,Son YH,Nath R

    更新日期:1999-04-01 00:00:00

  • Evaluating the impact of extended field-of-view CT reconstructions on CT values and dosimetric accuracy for radiation therapy.

    abstract:PURPOSE:Wide bore CT scanners use extended field-of-view (eFOV) reconstruction algorithms to attempt to recreate tissue truncated due to large patient habitus. Radiation therapy planning systems rely on accurate CT numbers in order to correctly plan and calculate radiation dose. This study looks at the impact of eFOV r...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13299

    authors: Cheung JP,Shugard E,Mistry N,Pouliot J,Chen J

    更新日期:2019-02-01 00:00:00

  • A Monte Carlo comparison of the response of the PTW-diamond and the TL-diamond detectors in megavoltage photon beams.

    abstract::A detailed Monte Carlo study of the PTW-diamond solid state detector response in megavoltage photon beams (60Co gamma rays to 25 MV x rays) has been performed with the EGS4 Monte Carlo Code. The sensitive volume of the diamond detector is a disk of diameter 4.4 mm and thickness 0.40 mm. The phantom material was water ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598771

    authors: Mobit PN,Sandison GA

    更新日期:1999-11-01 00:00:00

  • Factors influencing the performance of patient specific quality assurance for pencil beam scanning IMPT fields.

    abstract:PURPOSE:A detailed analysis of 2728 intensity modulated proton therapy (IMPT) fields that were clinically delivered to patients between 2007 and 2013 at Paul Scherrer Institute (PSI) was performed. The aim of this study was to analyze the results of patient specific dosimetric verifications and to assess possible corre...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4964449

    authors: Trnková P,Bolsi A,Albertini F,Weber DC,Lomax AJ

    更新日期:2016-11-01 00:00:00

  • WE-D-213AB-03: Preparing for the ABR Diagnostic Medical Physics Board Exams - Oral Exam.

    abstract::Part III (oral exam) is the final part of ABR board exam for Diagnostic Medical Physics. In this exam, each of five oral examiners questions the candidate in each of five question categories. Oral exam has unique challenges to the candidate compared with the written exams. The candidate is expected to have not only ad...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4736136

    authors: Leng S,Chen L,Shu Y,McCollough C

    更新日期:2012-06-01 00:00:00

  • Relaxing leaf-motion restrictions in dynamic multileaf collimator leaf sequencing.

    abstract:PURPOSE:Traditionally, unidirectional leaf-sweeping schemes have been employed to deliver IMRT plans using the dynamic multileaf collimator (DMLC) technique. The goal of this research is to investigate the potential impact of relaxing the leaf-motion restrictions in DMLC IMRT on the beam-modulation quality and the deli...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13158

    authors: Mirzapour SA,Salari E

    更新日期:2018-11-01 00:00:00