Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy.

Abstract:

:This work aims to develop, implement and validate a Monte Carlo (MC)-based independent dose calculation (IDC) framework to perform patient-specific quality assurance (QA) for multi-leaf collimator (MLC)-based CyberKnife® (Accuray Inc., Sunnyvale, CA) treatment plans. The IDC framework uses an XML-format treatment plan as exported from the treatment planning system (TPS) and DICOM format patient CT data, an MC beam model using phase spaces, CyberKnife MLC beam modifier transport using the EGS++ class library, a beam sampling and coordinate transformation engine and dose scoring using DOSXYZnrc. The framework is validated against dose profiles and depth dose curves of single beams with varying field sizes in a water tank in units of cGy/Monitor Unit and against a 2D dose distribution of a full prostate treatment plan measured with Gafchromic EBT3 (Ashland Advanced Materials, Bridgewater, NJ) film in a homogeneous water-equivalent slab phantom. The film measurement is compared to IDC results by gamma analysis using 2% (global)/2 mm criteria. Further, the dose distribution of the clinical treatment plan in the patient CT is compared to TPS calculation by gamma analysis using the same criteria. Dose profiles from IDC calculation in a homogeneous water phantom agree within 2.3% of the global max dose or 1 mm distance to agreement to measurements for all except the smallest field size. Comparing the film measurement to calculated dose, 99.9% of all voxels pass gamma analysis, comparing dose calculated by the IDC framework to TPS calculated dose for the clinical prostate plan shows 99.0% passing rate. IDC calculated dose is found to be up to 5.6% lower than dose calculated by the TPS in this case near metal fiducial markers. An MC-based modular IDC framework was successfully developed, implemented and validated against measurements and is now available to perform patient-specific QA by IDC.

journal_name

Phys Med Biol

authors

Mackeprang PH,Vuong D,Volken W,Henzen D,Schmidhalter D,Malthaner M,Mueller S,Frei D,Stampanoni MFM,Dal Pra A,Aebersold DM,Fix MK,Manser P

doi

10.1088/1361-6560/aa97f8

subject

Has Abstract

pub_date

2017-12-19 00:00:00

pages

015015

issue

1

eissn

0031-9155

issn

1361-6560

journal_volume

63

pub_type

杂志文章
  • NEMA NU 4-2008 validation and applications of the PET-SORTEO Monte Carlo simulations platform for the geometry of the Inveon PET preclinical scanner.

    abstract::Monte Carlo-based simulation of positron emission tomography (PET) data plays a key role in the design and optimization of data correction and processing methods. Our first aim was to adapt and configure the PET-SORTEO Monte Carlo simulation program for the geometry of the widely distributed Inveon PET preclinical sca...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/58/19/6749

    authors: Boisson F,Wimberley CJ,Lehnert W,Zahra D,Pham T,Perkins G,Hamze H,Gregoire MC,Reilhac A

    更新日期:2013-10-07 00:00:00

  • A three-dimensional model-based partial volume correction strategy for gated cardiac mouse PET imaging.

    abstract::Quantification in cardiac mouse positron emission tomography (PET) imaging is limited by the imaging spatial resolution. Spillover of left ventricle (LV) myocardial activity into adjacent organs results in partial volume (PV) losses leading to underestimation of myocardial activity. A PV correction method was develope...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/57/13/4309

    authors: Dumouchel T,Thorn S,Kordos M,DaSilva J,Beanlands RS,deKemp RA

    更新日期:2012-07-07 00:00:00

  • Physical characteristics of the ECAT EXACT3D positron tomograph.

    abstract::The 'EXACT3D' positron tomograph, which is now in routine clinical research use, was developed with the aim of achieving unprecedented sensitivity, high spatial and temporal resolution and simplicity of design using proven detector technology. It consists of six rings of standard detector blocks (CTI/Siemens EXACT HR+...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/45/9/313

    authors: Spinks TJ,Jones T,Bloomfield PM,Bailey DL,Miller M,Hogg D,Jones WF,Vaigneur K,Reed J,Young J,Newport D,Moyers C,Casey ME,Nutt R

    更新日期:2000-09-01 00:00:00

  • A 4D biomechanical lung phantom for joint segmentation/registration evaluation.

    abstract::At present, there exists few openly available methods for evaluation of simultaneous segmentation and registration algorithms. These methods allow for a combination of both techniques to track the tumor in complex settings such as adaptive radiotherapy. We have produced a quality assurance platform for evaluating this...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/61/19/7012

    authors: Markel D,Levesque I,Larkin J,Léger P,El Naqa I

    更新日期:2016-10-07 00:00:00

  • Online prediction of respiratory motion: multidimensional processing with low-dimensional feature learning.

    abstract::Accurate real-time prediction of respiratory motion is desirable for effective motion management in radiotherapy for lung tumor targets. Recently, nonparametric methods have been developed and their efficacy in predicting one-dimensional respiratory-type motion has been demonstrated. To exploit the correlation among v...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/55/11/002

    authors: Ruan D,Keall P

    更新日期:2010-06-07 00:00:00

  • Implementation of pencil kernel and depth penetration algorithms for treatment planning of proton beams.

    abstract::The implementation of two algorithms for calculating dose distributions for radiation therapy treatment planning of intermediate energy proton beams is described. A pencil kernel algorithm and a depth penetration algorithm have been incorporated into a commercial three dimensional treatment planning system (Helax-TMS,...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/45/1/302

    authors: Russell KR,Isacsson U,Saxner M,Ahnesjö A,Montelius A,Grusell E,Dahlgren CV,Lorin S,Glimelius B

    更新日期:2000-01-01 00:00:00

  • Monitoring of airborne contamination during the handling of technetium-99m and radioiodine.

    abstract::Measurements have been made using an air sampler to measure airborne radioactivity produced during the routine handling of large activities of technetium-99m, iodine-125 and iodine-131. The results indicate that 99Tcm can be safely handled in environments without direct exhaust of the ventilated air but that 125I and ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/25/6/005

    authors: Eadie AS,Horton PW,Hilditch TE

    更新日期:1980-11-01 00:00:00

  • Improved iterative tomographic reconstruction for x-ray imaging with edge-illumination.

    abstract::Iterative tomographic reconstruction has been established as a viable alternative for data analysis in phase-sensitive x-ray imaging based on the edge-illumination principle. However, previously published approaches did not account for drifts of optical elements during a scan, which can lead to artefacts. Up to now, t...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab439d

    authors: Modregger P,Meganck J,Hagen CK,Massimi L,Olivo A,Endrizzi M

    更新日期:2019-10-10 00:00:00

  • The role of trapped bubbles in kidney stone detection with the color Doppler ultrasound twinkling artifact.

    abstract::The color Doppler ultrasound twinkling artifact, which highlights kidney stones with rapidly changing color, has the potential to improve stone detection; however, its inconsistent appearance has limited its clinical utility. Recently, it was proposed stable crevice bubbles on the kidney stone surface cause twinkling;...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa9a2f

    authors: Simon JC,Sapozhnikov OA,Kreider W,Breshock M,Williams JC,Bailey MR

    更新日期:2018-01-09 00:00:00

  • Spinal pedicle screw planning using deformable atlas registration.

    abstract::Spinal screw placement is a challenging task due to small bone corridors and high risk of neurological or vascular complications, benefiting from precision guidance/navigation and quality assurance (QA). Implicit to both guidance and QA is the definition of a surgical plan-i.e. the desired trajectories and device sele...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa5f42

    authors: Goerres J,Uneri A,De Silva T,Ketcha M,Reaungamornrat S,Jacobson M,Vogt S,Kleinszig G,Osgood G,Wolinsky JP,Siewerdsen JH

    更新日期:2017-04-07 00:00:00

  • Ultraviolet and visible spectroscopies for tissue diagnostics: fluorescence spectroscopy and elastic-scattering spectroscopy.

    abstract::We review the application of fluorescence spectroscopy and elastic-scattering spectroscopy, over the ultraviolet-to-visible wavelength range, to minimally invasive medical diagnostics. The promises and hopes, as well as the difficulties, of these developing techniques are discussed. ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章,评审

    doi:10.1088/0031-9155/42/5/005

    authors: Bigio IJ,Mourant JR

    更新日期:1997-05-01 00:00:00

  • Electron depth-dose dependence on energy spectral quality.

    abstract::Electron depth-dose curves and spectra have been measured from an experimental accelerator on which both energy and spectral shape could be varied independently. The measurements were made at nominal energies of 6 and 12 MeV, with the energy spread (FWHM) varied from 1.7% to over 10% of average energy. Changes in dept...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/28/12/005

    authors: Johnsen SW,LaRiviere PD,Tanabe E

    更新日期:1983-12-01 00:00:00

  • Semi-empirical formulation of multiple scattering for the Gaussian beam model of heavy charged particles stopping in tissue-like matter.

    abstract::Dose calculation for radiotherapy with protons and heavier ions deals with a large volume of path integrals involving a scattering power of body tissue. This work provides a simple model for such demanding applications. There is an approximate linearity between RMS end-point displacement and range of incident particle...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/54/5/N01

    authors: Kanematsu N

    更新日期:2009-03-07 00:00:00

  • A Monte Carlo-based method to estimate radiation dose from spiral CT: from phantom testing to patient-specific models.

    abstract::The purpose of this work is to develop and test a method to estimate the relative and absolute absorbed radiation dose from axial and spiral CT scans using a Monte Carlo approach. Initial testing was done in phantoms and preliminary results were obtained from a standard mathematical anthropomorphic model (MIRD V) and ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/48/16/306

    authors: Jarry G,DeMarco JJ,Beifuss U,Cagnon CH,McNitt-Gray MF

    更新日期:2003-08-21 00:00:00

  • Radiation binary targeted therapy for HER-2 positive breast cancers: assumptions, theoretical assessment and future directions.

    abstract::A novel radiation targeted therapy is investigated for HER-2 positive breast cancers. The proposed concept combines two known approaches, but never used together for the treatment of advanced, relapsed or metastasized HER-2 positive breast cancers. The proposed radiation binary targeted concept is based on the anti HE...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/51/6/001

    authors: Mundy DW,Harb W,Jevremovic T

    更新日期:2006-03-21 00:00:00

  • Three-dimensional optical tomography of the premature infant brain.

    abstract::For the first time, three-dimensional images of the newborn infant brain have been generated using measurements of transmitted light. A 32-channel time-resolved imaging system was employed, and data were acquired using custom-made helmets which couple source fibres and detector bundles to the infant head. Images have ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/47/23/303

    authors: Hebden JC,Gibson A,Yusof RM,Everdell N,Hillman EM,Delpy DT,Arridge SR,Austin T,Meek JH,Wyatt JS

    更新日期:2002-12-07 00:00:00

  • Image quality assessment in digital mammography: part II. NPWE as a validated alternative for contrast detail analysis.

    abstract::Assessment of image quality for digital x-ray mammography systems used in European screening programs relies mainly on contrast-detail CDMAM phantom scoring and requires the acquisition and analysis of many images in order to reduce variability in threshold detectability. Part II of this study proposes an alternative ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/56/14/003

    authors: Monnin P,Marshall NW,Bosmans H,Bochud FO,Verdun FR

    更新日期:2011-07-21 00:00:00

  • Optimization and performance study of a proton CT system for pre-clinical small animal imaging.

    abstract::Proton computed tomography (pCT) promises to reduce or even eliminate range uncertainties inherent in the conversion of Hounsfield units into relative stopping power (RSP) for proton therapy treatment planning. This is of particular interest for proton irradiation studies in animal models due to the high precision req...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab8afc

    authors: Meyer S,Bortfeldt J,Lämmer P,Englbrecht FS,Pinto M,Schnürle K,Würl M,Parodi K

    更新日期:2020-08-13 00:00:00

  • Population and patient-specific target margins for 4D adaptive radiotherapy to account for intra- and inter-fraction variation in lung tumour position.

    abstract::In this work, five 4D image-guidance strategies (two population, an offline adaptive and two online strategies) were evaluated that compensated for both inter- and intra-fraction variability such as changes to the baseline tumour position and respiratory pattern. None of the strategies required active motion compensat...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/52/1/017

    authors: Hugo GD,Yan D,Liang J

    更新日期:2007-01-07 00:00:00

  • Computed LET distributions in the build-up region of tissue for 14 MeV neutrons.

    abstract::Experimental radiobiological investigations of fast neutron beams involve irradiation of thin cell layers. The irradiation of cells is done either in free air or with a material such as graphite, TE plastic or bone-equivalent plastic covering the cells. This paper discusses the computed dose distributions in LET at sh...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/24/5/007

    authors: Bhatia DP,Nagarajan PS

    更新日期:1979-09-01 00:00:00

  • Penetration of electromagnetic fields of an open-ended coaxial probe between 1 MHz and 1 GHz in dielectric skin measurements.

    abstract::An open-ended coaxial probe is often used for investigating the dielectric properties of biological tissues. The present study indicates that in addition to the probe size, the penetration of the electromagnetic (EM) fields of an open-ended coaxial probe in contact with the skin is dependent on the applied frequency b...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/44/7/404

    authors: Alanen E,Lahtinen T,Nuutinen J

    更新日期:1999-07-01 00:00:00

  • Measurement of the absolute optical properties and cerebral blood volume of the adult human head with hybrid differential and spatially resolved spectroscopy.

    abstract::A hybrid differential and spatially resolved spectroscopy (SRS) technique has been developed to measure absolute absorption coefficient (mu(a)), reduced scattering coefficient (mu'(s)) and cerebral blood volume (CBV) in the adult human head. A spectrometer with both differential and SRS capabilities has been used to c...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/51/3/015

    authors: Leung TS,Tachtsidis I,Smith M,Delpy DT,Elwell CE

    更新日期:2006-02-07 00:00:00

  • Anatomically-aided PET reconstruction using the kernel method.

    abstract::This paper extends the kernel method that was proposed previously for dynamic PET reconstruction, to incorporate anatomical side information into the PET reconstruction model. In contrast to existing methods that incorporate anatomical information using a penalized likelihood framework, the proposed method incorporate...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/61/18/6668

    authors: Hutchcroft W,Wang G,Chen KT,Catana C,Qi J

    更新日期:2016-09-21 00:00:00

  • A dedicated phantom design for positron emission mammography performance evaluation.

    abstract::A standard protocol for performance evaluation of positron emission mammography (PEM) systems has not yet been established. In this work we propose a methodology based on the design of specific phantoms for this imaging modality with component dimensions in accordance with typical breast lesion sizes together with the...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aba7d1

    authors: Torres-Urzúa LF,Alva-Sanchez H,Martinez-Davalos A,García-Pérez FO,Peruyero-Rivas RM,Rodriguez-Villafuerte M

    更新日期:2020-07-21 00:00:00

  • Monte Carlo and experimental evaluation of accuracy and noise properties of two scatter correction methods for SPECT.

    abstract::Scatter correction is a prerequisite for quantitative SPECT, but potentially increases noise. Monte Carlo simulations (EGS4) and physical phantom measurements were used to compare accuracy and noise properties of two scatter correction techniques: the triple-energy window (TEW), and the transmission dependent convolut...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/41/11/017

    authors: Narita Y,Eberl S,Iida H,Hutton BF,Braun M,Nakamura T,Bautovich G

    更新日期:1996-11-01 00:00:00

  • Dose effect of guidewire position in intravascular brachytherapy.

    abstract::It has been reported that the dose effects of metallic guidewires are significant in intravascular brachytherapy (IVBT) using a beta source. The purpose of this work is to investigate the dependence of these dose effects on guidewire position. The EGS4 Monte Carlo codes were used to perform the dose calculations for t...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/47/10/309

    authors: Shih R,Hsu WL,Li XA

    更新日期:2002-05-21 00:00:00

  • Stopping power and ranges of low-energy protons in tissue-equivalent gas.

    abstract::Stopping power data for protons in tissue-equivalent gas (64.4% CH4 + 32.4% CO2 + 3.2% N2, partial pressures) were derived from experimentally determined ranges and from differential ionisation distributions using the differential energy required per ion pair formed to convert ionisation to energy loss. The proton ene...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/32/3/006

    authors: Waibel E,Willems G

    更新日期:1987-03-01 00:00:00

  • The radiation dose from a proposed measurement of arsenic and selenium in human skin.

    abstract::Dose measurements following 10 min irradiations with a portable x-ray fluorescence spectrometer composed of a miniature x-ray tube and a silicon PiN diode detector were performed using thermoluminescent dosimeters consisting of LiF:Mg,Ti chips of 3 mm diameter and 0.4 mm thickness. The table-top setup of the spectrome...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/55/18/015

    authors: Gherase MR,Mader JE,Fleming DE

    更新日期:2010-09-21 00:00:00

  • Measurement of scattered radiation in a volumetric 64-slice CT scanner using three experimental techniques.

    abstract::Compton scatter poses a significant threat to volumetric x-ray computed tomography, bringing cupping and streak artefacts thus impacting qualitative and quantitative imaging procedures. To perform appropriate scatter compensation, it is necessary to estimate the magnitude and spatial distribution of x-ray scatter. The...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/55/8/010

    authors: Akbarzadeh A,Ay MR,Ghadiri H,Sarkar S,Zaidi H

    更新日期:2010-04-21 00:00:00

  • The exponential edge-gradient effect in x-ray computed tomography.

    abstract::The exponential edge-gradient effect must arise in any x-ray transmission CT scanner whenever long sharp edges of high contrast are encountered. The effect is non-linear and is due to the interaction of the exponential law of x-ray attenuation and the finite width of the scanning beam in the x-y plane. The error induc...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/26/3/010

    authors: Joseph PM,Spital RD

    更新日期:1981-05-01 00:00:00