Wavelet-based real-time calculation of multiple physiological parameters on an embedded platform.

Abstract:

OBJECTIVE:This paper aims to present how physiological signals can be processed based on wavelet decomposition to calculate multiple physiological parameters in real-time on an embedded platform. APPROACH:An ECG and PPG are decomposed to the appropriate scale based on a quadratic spline wavelet base in order to obtain high and narrow pulse peaks at the location of the mutation points. Based on the decomposed waveforms, feature points are positioned to calculate physiological parameters in real-time, including heart rate, pulse rate, blood oxygen, and blood pressure. The proposed algorithm has been implemented on a Texas Instruments' CC2640R2F. MAIN RESULTS:The misdetection rate of feature point location based on the square wavelet decomposition waveform is only 0.57% in the acquired ECG and 0.23% in the acquired PPG. Heart rate and pulse rate are both highly correlated with the reference, both having correlation coefficients of 0.99. The pulse rate and heart rate are 3.85% (51/1326) and 2.94% (39/1326) outside the 95% consistency limit, respectively. The systolic and diastolic blood pressures are significantly associated with standard equipment measurements, with correlation coefficients of 0.87 and 0.83. The systolic and diastolic blood pressures were 5.88% (21/357) and 5.32% (19/357) outside the 95% consistency limit, respectively. SIGNIFICANCE:The real-time calculation of multiple physiological parameters based on wavelet decomposition on an embedded platform presented here shows outstanding accuracy.

journal_name

Physiol Meas

authors

Tang M,Xia P,He Z,Zhao Z,Chen X,Yang T,Zhang Z,Zhan Q,Li X,Fang Z

doi

10.1088/1361-6579/ab6f52

subject

Has Abstract

pub_date

2020-03-06 00:00:00

pages

025010

issue

2

eissn

0967-3334

issn

1361-6579

journal_volume

41

pub_type

杂志文章
  • Time-frequency analysis of heart rate variability using short-time fourier analysis.

    abstract::This study was done to introduce new parameters derived by time frequency analysis of heart rate variability data. Four simulation experiments were carried out to compare the short-time Fourier transform (STFT) analysis method to the traditional overall spectral analysis method. Sinusoidal signals were generated with ...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/21/2/303

    authors: Elsenbruch S,Wang Z,Orr WC,Chen JD

    更新日期:2000-05-01 00:00:00

  • Thermography and colour duplex ultrasound assessments of arterio-venous fistula function in renal patients.

    abstract::Vascular and clinical assessments of arterio-venous fistula (AVF) function and access are important in patients undergoing or preparing to undergo renal dialysis. Objective assessment techniques include colour duplex ultrasound and more recently medical infrared thermography. Ideally, these should help assess problems...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/27/1/005

    authors: Allen J,Oates CP,Chishti AD,Ahmed IA,Talbot D,Murray A

    更新日期:2006-01-01 00:00:00

  • Topographic distribution of idiopathic pulmonary fibrosis: a hybrid physics- and agent-based model.

    abstract:OBJECTIVE:Idiopathic pulmonary fibrosis (IPF) is a progressive and fatal disease characterized by excessive deposition of collagen and associated stiffening of lung tissue. While it is known that inflammation and dysfunction of fibroblasts are involved in disease development, it remains poorly understood how cells and ...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/aaca86

    authors: Wellman TJ,Mondoñedo JR,Davis GS,Bates JHT,Suki B

    更新日期:2018-06-28 00:00:00

  • Development of a novel method for assessing balance: the quantitative posturography system.

    abstract::Balance is the ability to maintain equilibrium while sitting or standing. There are a number of different methods which are used to assess balance: technical methods such as sway magnetometry, ataxia meter and force platforms, and clinical methods such as the functional reach test, Berg balance test and fall risk inde...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/21/4/309

    authors: Browne J,O'Hare N

    更新日期:2000-11-01 00:00:00

  • Correlation between maximal radial muscle displacement and stiffness in gastrocnemius muscle.

    abstract:OBJECTIVE:Tensiomyography (TMG) and myotonometry (MMT) are two novel technologies that measure neuromuscular properties. These two devices measure the stiffness of the soft tissue as well as other variables. The aim of this study is to analyze if there is any correlation between maximal radial displacement (Dm) and Sti...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/abcdf4

    authors: Labata-Lezaun N,López-de-Celis C,Llurda-Almuzara L,González-Rueda V,Cadellans-Arróniz A,Pérez-Bellmunt A

    更新日期:2020-12-31 00:00:00

  • A randomized controlled trial studying the effect of maternal hyperoxygenation on fetal heart rate in suspected fetal distress.

    abstract:OBJECTIVE:To investigate the effect of maternal hyperoxygenation on fetal heart rate (FHR) when applied for suspected fetal distress during the second stage of term labor. APPROACH:A single-center randomized controlled trial was conducted in a tertiary care hospital in The Netherlands. Participants were included durin...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/abc0b6

    authors: Moors S,Joshi R,Bullens LM,van Oostrum NHM,Regis M,van den Heuvel ER,Oei SG,van Laar JOEH,van der Hout-van der Jagt MB

    更新日期:2020-12-11 00:00:00

  • A comparative study of several exciting magnetic fields for induced current EIT.

    abstract::In this study, the selection of coil configuration parameters (coil radius and coil centre shift) for induced current EIT using circular coils is investigated. An alternative coil configuration is suggested, which produces approximately linear (spatially) magnetic fields in order to strengthen the currents in the cent...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/15/2a/007

    authors: Gençer NG,Ider YZ

    更新日期:1994-05-01 00:00:00

  • Magnetographic assessment of fetal hiccups and their effect on fetal heart rhythm.

    abstract::Fetal hiccups emerge as early as nine weeks post-conception, being the predominant diaphragmatic movement before 26 weeks of gestation. They are considered as a programmed isometric inspiratory muscle exercise of the fetus in preparation for the post-natal respiratory function, or a manifestation of a reflex circuitry...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/28/6/005

    authors: Popescu EA,Popescu M,Bennett TL,Lewine JD,Drake WB,Gustafson KM

    更新日期:2007-06-01 00:00:00

  • The effect of serial data collection on the accuracy of electrical impedance tomography images.

    abstract::There has been a surge of interest in using electrical impedance tomography (EIT) for monitoring regional lung ventilation, however, EIT is an ill-conditioned problem, and errors/noise in the boundary voltages can have an undesirable effect on the quality of the final image. Most EIT systems in clinical usage use seri...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/34/6/659

    authors: Yerworth R,Bayford R

    更新日期:2013-06-01 00:00:00

  • A shape-based quality evaluation and reconstruction method for electrical impedance tomography.

    abstract::Linear methods of reconstruction play an important role in medical electrical impedance tomography (EIT) and there is a wide variety of algorithms based on several assumptions. With the Graz consensus reconstruction algorithm for EIT (GREIT), a novel linear reconstruction algorithm as well as a standardized framework ...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/36/6/1161

    authors: Antink CH,Pikkemaat R,Malmivuo J,Leonhardt S

    更新日期:2015-06-01 00:00:00

  • Effect of organic and inorganic nitrates on cerebrovascular pulsatile power transmission in patients with heart failure and preserved ejection fraction.

    abstract:OBJECTIVE:Increased penetration of pulsatile power to the brain has been implicated in the pathogenesis of age-related cognitive dysfunction and dementia, a common comorbidity in patients with heart failure and preserved ejection fraction (HFpEF). However, there is a lack of knowledge on the effects of organic and inor...

    journal_title:Physiological measurement

    pub_type: 杂志文章,随机对照试验

    doi:10.1088/1361-6579/aab2ef

    authors: Londono-Hoyos F,Zamani P,Beraun M,Vasim I,Segers P,Chirinos JA

    更新日期:2018-04-20 00:00:00

  • Choice of reconstructed tissue properties affects interpretation of lung EIT images.

    abstract::Electrical impedance tomography (EIT) estimates an image of change in electrical properties within a body from stimulations and measurements at surface electrodes. There is significant interest in EIT as a tool to monitor and guide ventilation therapy in mechanically ventilated patients. In lung EIT, the EIT inverse p...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/35/6/1035

    authors: Grychtol B,Adler A

    更新日期:2014-06-01 00:00:00

  • Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes.

    abstract::A nonlinear model predictive controller has been developed to maintain normoglycemia in subjects with type 1 diabetes during fasting conditions such as during overnight fast. The controller employs a compartment model, which represents the glucoregulatory system and includes submodels representing absorption of subcut...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/25/4/010

    authors: Hovorka R,Canonico V,Chassin LJ,Haueter U,Massi-Benedetti M,Orsini Federici M,Pieber TR,Schaller HC,Schaupp L,Vering T,Wilinska ME

    更新日期:2004-08-01 00:00:00

  • A comparison of methods for measurement of spatial resolution in two-dimensional circular EIT images.

    abstract::The literature concerning measurement of spatial resolution in electrical impedance tomography (EIT) is vague. Different groups often use their own method or a modified version of a better known method, thus hindering a generalized resolution measurement which could be useful for gauging the performance of one system ...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/23/1/316

    authors: Wheeler JL,Wang W,Tang M

    更新日期:2002-02-01 00:00:00

  • Wearable oxygen uptake and energy expenditure monitors.

    abstract::Obesity is a major health issue in both developed and developing countries. The balance between energy intake and exercise is important, and measurements of both energy intake and energy expenditure are required. Many studies have attempted to monitor energy intake via wearable technology, but no standard methods have...

    journal_title:Physiological measurement

    pub_type: 杂志文章,评审

    doi:10.1088/1361-6579/ab3827

    authors: Tamura T

    更新日期:2019-09-03 00:00:00

  • Chronodiagnostic acquisition of recovery speed of heart rate under bathing stress.

    abstract::Cycling on an ergometer is one effective means of measuring cardiovascular function while applying stress on the heart. Bathing in a hot water bath applies a low stress to the heart. The electrocardiograms of a healthy adult male (aged 35 at the start of study) were recorded while taking a hot water bath with no elect...

    journal_title:Physiological measurement

    pub_type: 临床试验,杂志文章

    doi:10.1088/0967-3334/20/4/304

    authors: Ishijima M,Togawa T

    更新日期:1999-11-01 00:00:00

  • Techniques for early detection of Alzheimer's disease using spontaneous EEG recordings.

    abstract::Alzheimer's disease (AD) is a degenerative disease which causes serious cognitive decline. Studies suggest that effective treatments for AD may be aided by the detection of the disease in its early stages, prior to extensive neuronal degeneration. In this paper, we propose a set of novel techniques which could help to...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/28/4/001

    authors: Woon WL,Cichocki A,Vialatte F,Musha T

    更新日期:2007-04-01 00:00:00

  • Empirical estimators of gamma fits to tracer-dilution curves and their technical basis and practical scope.

    abstract::A gamma fit facilitates smoothing and extrapolation of distorted tracer-dilution curves in blood flow studies. Theoretically based empirical estimators were developed as simple alternatives to direct regression approaches from simulated gamma distributions with a wide range of shape asymmetry. Key curve features of pe...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/22/4/401

    authors: Lopatatzidis A,Millard RK

    更新日期:2001-11-01 00:00:00

  • Cooperative dry-electrode sensors for multi-lead biopotential and bioimpedance monitoring.

    abstract::Cooperative sensors is a novel measurement architecture that allows the acquiring of biopotential signals on patients in a comfortable and easy-to-integrate manner. The novel sensors are defined as cooperative in the sense that at least two of them work in concert to measure a target physiological signal, such as a mu...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/36/4/767

    authors: Rapin M,Proença M,Braun F,Meier C,Solà J,Ferrario D,Grossenbacher O,Porchet JA,Chételat O

    更新日期:2015-04-01 00:00:00

  • Modelling the cerebral haemodynamic response in the physiological range of PaCO2.

    abstract:OBJECTIVE:Arterial CO2 (PaCO2) has a strong effect on cerebral blood flow (CBF), but its influence on CBF regulatory mechanisms and circulatory systemic variables has not been fully described over the entire physiological range of PaCO2. APPROACH:CBF velocity (CBFV, transcranial Doppler), blood pressure (BP, Finometer...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/aac76b

    authors: Minhas JS,Panerai RB,Robinson TG

    更新日期:2018-06-19 00:00:00

  • Non-contact video-based vital sign monitoring using ambient light and auto-regressive models.

    abstract::Remote sensing of the reflectance photoplethysmogram using a video camera typically positioned 1 m away from the patient's face is a promising method for monitoring the vital signs of patients without attaching any electrodes or sensors to them. Most of the papers in the literature on non-contact vital sign monitoring...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/35/5/807

    authors: Tarassenko L,Villarroel M,Guazzi A,Jorge J,Clifton DA,Pugh C

    更新日期:2014-05-01 00:00:00

  • Preliminary static EIT images of the thorax in health and disease.

    abstract::The results of a preliminary clinical evaluation of a one-frequency electrical impedance tomography (EIT) system enabling static in vivo imaging are presented. The design of the measuring system and image reconstruction software are described. Thirty-one subjects were examined and divided into four clinical groups. Th...

    journal_title:Physiological measurement

    pub_type: 临床试验,杂志文章

    doi:10.1088/0967-3334/23/1/304

    authors: Cherepenin V,Karpov A,Korjenevsky A,Kornienko V,Kultiasov Y,Mazaletskaya A,Mazourov D

    更新日期:2002-02-01 00:00:00

  • In vivo quantification of accumulating abdominal fluid using an electrical impedance tomography hemiarray.

    abstract::A new method to image and quantify intra-abdominal haemorrhage using electrical impedance tomography (EIT) was tested in vivo. Supine peritoneal dialysis patients were monitored using an 8-electrode hemiarray placed on the anterior abdomen. EIT measurements were recorded using the EPack II data acquisition system befo...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/32/2/001

    authors: Tucker AS,Ross EA,Paugh-Miller J,Sadleir RJ

    更新日期:2011-02-01 00:00:00

  • Coupling between arterial pressure, cerebral blood velocity, and cerebral tissue oxygenation with spontaneous and forced oscillations.

    abstract::We tested the hypothesis that transmission of arterial pressure to brain tissue oxygenation is low under conditions of arterial pressure instability. Two experimental models of hemodynamic instability were used in healthy human volunteers; (1) oscillatory lower body negative pressure (OLBNP) (N = 8; 5 male, 3 female),...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/36/4/785

    authors: Rickards CA,Sprick JD,Colby HB,Kay VL,Tzeng YC

    更新日期:2015-04-01 00:00:00

  • The effect of fracture and fracture fixation on ultrasonic velocity and attenuation.

    abstract::Measurement of the velocity of propagation and attenuation of ultrasound (200 kHz) is believed to be a useful non-invasive technique for assessing the mechanical properties of bone. A new method for the determination of ultrasound velocity and attenuation of longitudinal waves in cortical bone was used in vivo and in ...

    journal_title:Physiological measurement

    pub_type: 临床试验,杂志文章

    doi:10.1088/0967-3334/17/3/006

    authors: Saulgozis J,Pontaga I,Lowet G,Van der Perre G

    更新日期:1996-08-01 00:00:00

  • Cardiac eigen imaging: a novel method to isolate cardiac activity in thoracic electrical impedance tomography.

    abstract:OBJECTIVE:As the global burden of cardiovascular disease increases, proactive cardiovascular healthcare by means of accurate, precise, continuous, and non-invasive monitoring is becoming crucial. However, no current device is able to provide cardiac hemodynamic monitoring with the aforementioned criterion. Electrical i...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/abb141

    authors: Arshad SH,Murphy EK,Callahan JM,DeVries JT,Odame KM,Halter RJ

    更新日期:2020-10-06 00:00:00

  • Chasing the evidence: the influence of data segmentation on estimates of dynamic cerebral autoregulation.

    abstract:OBJECTIVE:Transfer function analysis (TFA) of dynamic cerebral autoregulation (dCA) requires smoothing of spectral estimates using segmentation of the data (SD). Systematic studies are required to elucidate the potential influence of SD on dCA parameters. APPROACH:Healthy subjects (HS, n = 237) and acute ischaemic str...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/ab7ddf

    authors: Panerai RB,Intharakham K,Haunton V,Minhas JS,Llwyd O,Lam M,Salinet ASM,Nogueira RC,Katsogridakis E,Maggio P,Robinson TG

    更新日期:2020-04-17 00:00:00

  • The influence of dilution on the offline measurement of exhaled nitric oxide.

    abstract:OBJECTIVE:Measurement of fractional exhaled nitric oxide (FeNO) is used to determine the presence and severity of eosinophilic airway inflammation in asthma and other wheezing illnesses. The gold standard of online measurement during a single prolonged exhalation is not suitable for use in young children. The internati...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/aaa455

    authors: MacBean V,Pooranampillai D,Howard C,Lunt A,Greenough A

    更新日期:2018-02-26 00:00:00

  • Pulmonary perfusion measured by means of electrical impedance tomography.

    abstract::Electrical impedance tomography (EIT) is a recent imaging technique based on electrical impedance, offering the possibility of measuring pulmonary perfusion. In the present study the influence of several pulmonary haemodynamical parameters on the EIT signal were investigated. First, the influence on the systolic wave ...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/19/2/013

    authors: Vonk Noordegraaf A,Kunst PW,Janse A,Marcus JT,Postmus PE,Faes TJ,de Vries PM

    更新日期:1998-05-01 00:00:00

  • Multichannel simultaneous magnetic induction measurement system (MUSIMITOS).

    abstract::Non-contact heart and lung activity monitoring would be a desirable supplement to conventional monitoring techniques. Based on the potential of non-contact magnetic induction measurements, requirements for an adequate monitoring system were estimated. This formed the basis for the development of the presented extendab...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/29/6/S25

    authors: Steffen M,Heimann K,Bernstein N,Leonhardt S

    更新日期:2008-06-01 00:00:00