A mobile isocentric C-arm for intraoperative cone-beam CT: Technical assessment of dose and 3D imaging performance.

Abstract:

PURPOSE:To characterize the radiation dose and three-dimensional (3D) imaging performance of a recently developed mobile, isocentric C-arm equipped with a flat-panel detector (FPD) for intraoperative cone-beam computed tomography (CBCT) (Cios Spin 3D, Siemens Healthineers) and to identify potential improvements in 3D imaging protocols for pertinent imaging tasks. METHODS:The C-arm features a 30 × 30 cm2 FPD and isocentric gantry with computer-controlled motorization of rotation (0-195°), angulation (±220°), and height (0-45 cm). Geometric calibration was assessed in terms of 9 degrees of freedom of the x-ray source and detector in CBCT scans, and the reproducibility of geometric calibration was evaluated. Standard and custom scan protocols were evaluated, with variation in the number of projections (100-400) and mAs per view (0.05-1.65 mAs). Image reconstruction was based on 3D filtered backprojection using "smooth," "normal," and "sharp" reconstruction filters as well as a custom, two-dimensional 2D isotropic filter. Imaging performance was evaluated in terms of uniformity, gray value correspondence with Hounsfield units (HU), contrast, noise (noise-power spectrum, NPS), spatial resolution (modulation transfer function, MTF), and noise-equivalent quanta (NEQ). Performance tradeoffs among protocols were visualized in anthropomorphic phantoms for various anatomical sites and imaging tasks. RESULTS:Geometric calibration showed a high degree of reproducibility despite ~19 mm gantry flex over a nominal semicircular orbit. The dose for a CBCT scan varied from ~0.8-4.7 mGy for head protocols to ~6-38 mGy for body protocols. The MTF was consistent with sub-mm spatial resolution, with f10 (frequency at which MTF = 10%) equal to 0.64 mm-1 , 1.0 mm-1 , and 1.5 mm-1 for smooth, standard, and sharp filters respectively. Implementation of a custom 2D isotropic filter improved CNR ~ 50-60% for both head and body protocols and provided more isotropic resolution and noise characteristics. The NPS and NEQ quantified the 3D noise performance and provided a guide to protocol selection, confirmed in images of anthropomorphic phantoms. Alternative scan protocols were identified according to body site and task - for example, lower-dose body protocols (<3 mGy) sufficient for visualization of bone structures. CONCLUSION:The studies provided objective assessment of the dose and 3D imaging performance of a new C-arm, offering an important basis for clinical deployment and a benchmark for quality assurance. Modifications to standard 3D imaging protocols were identified that may improve performance or reduce radiation dose for pertinent imaging tasks.

journal_name

Med Phys

journal_title

Medical physics

authors

Sheth NM,De Silva T,Uneri A,Ketcha M,Han R,Vijayan R,Osgood GM,Siewerdsen JH

doi

10.1002/mp.13983

subject

Has Abstract

pub_date

2020-03-01 00:00:00

pages

958-974

issue

3

eissn

0094-2405

issn

2473-4209

journal_volume

47

pub_type

杂志文章
  • Matching photon and electron fields with dynamic intensity modulation.

    abstract::A technique was developed to reduce the size and magnitude of the hot and cold spots in the abutting regions of photon and electron fields. The photon and electron fields were set up such that the photon field extended approximately 2 cm into the electron field in the abutting region. The region of the photon beam tha...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598753

    authors: Li JG,Xing L,Boyer AL,Hamilton RJ,Spelbring DR,Turian JV

    更新日期:1999-11-01 00:00:00

  • Charged particle transport in magnetic fields in EGSnrc.

    abstract:PURPOSE:To accurately and efficiently implement charged particle transport in a magnetic field in EGSnrc and validate the code for the use in phantom and ion chamber simulations. METHODS:The effect of the magnetic field on the particle motion and position is determined using one- and three-point numerical integrations...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4954318

    authors: Malkov VN,Rogers DW

    更新日期:2016-07-01 00:00:00

  • Validation of a deformable image registration technique for cone beam CT-based dose verification.

    abstract:PURPOSE:As radiation therapy evolves toward more adaptive techniques, image guidance plays an increasingly important role, not only in patient setup but also in monitoring the delivered dose and adapting the treatment to patient changes. This study aimed to validate a method for evaluation of delivered intensity modula...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4903292

    authors: Moteabbed M,Sharp GC,Wang Y,Trofimov A,Efstathiou JA,Lu HM

    更新日期:2015-01-01 00:00:00

  • A controlled phantom study of a noise equalization algorithm for detecting microcalcifications in digital mammograms.

    abstract::We report on some extensions and further developments of a well-known microcalcification detection algorithm based on adaptive noise equalization. Tissue equivalent phantom images with and without labeled microcalcifications were subjected to this algorithm, and analyses of results revealed some shortcomings in the ap...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1359245

    authors: Gürün OO,Fatouros PP,Kuhn GM,de Paredes ES

    更新日期:2001-04-01 00:00:00

  • A neural network approach for fast, automated quantification of DIR performance.

    abstract:PURPOSE:A critical step in adaptive radiotherapy (ART) workflow is deformably registering the simulation CT with the daily or weekly volumetric imaging. Quantifying the deformable image registration accuracy under these circumstances is a complex task due to the lack of known ground-truth landmark correspondences betwe...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12321

    authors: Neylon J,Min Y,Low DA,Santhanam A

    更新日期:2017-08-01 00:00:00

  • Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning.

    abstract::The Monte Carlo (MC) method has been shown through many research studies to calculate accurate dose distributions for clinical radiotherapy, particularly in heterogeneous patient tissues where the effects of electron transport cannot be accurately handled with conventional, deterministic dose algorithms. Despite its p...

    journal_title:Medical physics

    pub_type: 杂志文章,评审

    doi:10.1118/1.2795842

    authors: Chetty IJ,Curran B,Cygler JE,DeMarco JJ,Ezzell G,Faddegon BA,Kawrakow I,Keall PJ,Liu H,Ma CM,Rogers DW,Seuntjens J,Sheikh-Bagheri D,Siebers JV

    更新日期:2007-12-01 00:00:00

  • Dosimetric comparison of helical tomotherapy treatment plans for total marrow irradiation created using GPU and CPU dose calculation engines.

    abstract:PURPOSE:To compare optimization characteristics, plan quality, and treatment delivery efficiency between total marrow irradiation (TMI) plans using the new TomoTherapy graphic processing unit (GPU) based dose engine and CPU/cluster based dose engine. METHODS:Five TMI plans created on an anthropomorphic phantom were op...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4810936

    authors: Nalichowski A,Burmeister J

    更新日期:2013-07-01 00:00:00

  • Dynamic detector offsets for field of view extension in C-arm computed tomography with application to weight-bearing imaging.

    abstract:PURPOSE:In C-arm computed tomography (CT), the field of view (FOV) is often not sufficient to acquire certain anatomical structures, e.g., a full hip or thorax. Proposed methods to extend the FOV use a fixed detector displacement and a 360° scan range to double the radius of the FOV. These trajectories are designed for...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4915542

    authors: Herbst M,Schebesch F,Berger M,Choi JH,Fahrig R,Hornegger J,Maier A

    更新日期:2015-05-01 00:00:00

  • Investigating energy deposition in glandular tissues for mammography using multiscale Monte Carlo simulations.

    abstract:PURPOSE:To investigate energy deposition in glandular tissues of the breast on macro- and microscopic length scales in the context of mammography. METHODS:Multiscale mammography models of breasts are developed, which include segmented, voxelized macroscopic tissue structure as well as nine regions of interest (ROIs) e...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13372

    authors: Oliver PAK,Thomson RM

    更新日期:2019-03-01 00:00:00

  • Individually optimized contrast-enhanced 4D-CT for radiotherapy simulation in pancreatic ductal adenocarcinoma.

    abstract:PURPOSE:To develop an individually optimized contrast-enhanced (CE) 4D-computed tomography (CT) for radiotherapy simulation in pancreatic ductal adenocarcinomas (PDA). METHODS:Ten PDA patients were enrolled. Each underwent three CT scans: a 4D-CT immediately following a CE 3D-CT and an individually optimized CE 4D-CT ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4963213

    authors: Choi W,Xue M,Lane BF,Kang MK,Patel K,Regine WF,Klahr P,Wang J,Chen S,D'Souza W,Lu W

    更新日期:2016-10-01 00:00:00

  • Penumbral measurements in water for high-energy x rays.

    abstract::Ionization chambers of varying inside diameter have been used to investigate the penumbral region of 60Co, 6-MV, and 31-MV x-ray beams. Measurements were made in water at varying depths up to 25 cm for a square field of side length 10 cm. The dependence of the penumbral widths on both the inside diameter of the ioniza...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595963

    authors: Dawson DJ,Schroeder NJ,Hoya JD

    更新日期:1986-01-01 00:00:00

  • Computerized detection of masses in digital mammograms: analysis of bilateral subtraction images.

    abstract::A computerized scheme is being developed for the detection of masses in digital mammograms. Based on the deviation from the normal architectural symmetry of the right and left breasts, a bilateral subtraction technique is used to enhance the conspicuity of possible masses. The scheme employs two pairs of conventional ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596610

    authors: Yin FF,Giger ML,Doi K,Metz CE,Vyborny CJ,Schmidt RA

    更新日期:1991-09-01 00:00:00

  • Design and experimental testing of air slab caps which convert commercial electron diodes into dual purpose, correction-free diodes for small field dosimetry.

    abstract:PURPOSE:Two diodes which do not require correction factors for small field relative output measurements are designed and validated using experimental methodology. This was achieved by adding an air layer above the active volume of the diode detectors, which canceled out the increase in response of the diodes in small f...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4894728

    authors: Charles PH,Cranmer-Sargison G,Thwaites DI,Kairn T,Crowe SB,Pedrazzini G,Aland T,Kenny J,Langton CM,Trapp JV

    更新日期:2014-10-01 00:00:00

  • Broadband lung dielectric properties over the ablative temperature range: Experimental measurements and parametric models.

    abstract:PURPOSE:Computational models of microwave tissue ablation are widely used to guide the development of ablation devices, and are increasingly being used for the development of treatment planning and monitoring platforms. Knowledge of temperature-dependent dielectric properties of lung tissue is essential for accurate mo...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13704

    authors: Sebek J,Bortel R,Prakash P

    更新日期:2019-10-01 00:00:00

  • Radiative transport in the delta-P1 approximation for semi-infinite turbid media.

    abstract::We have developed an analytic solution for spatially resolved diffuse reflectance within the deltaP1 approximation to the radiative transport equation for a semi-infinite homogeneous turbid medium. We evaluate the performance of this solution by comparing its predictions with those provided by Monte Carlo simulations ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2828184

    authors: Seo I,Hayakawa CK,Venugopalan V

    更新日期:2008-02-01 00:00:00

  • Motion estimation for cardiac functional analysis using two x-ray computed tomography scans.

    abstract:PURPOSE:This work concerns computed tomography (CT)-based cardiac functional analysis (CFA) with a reduced radiation dose. As CT-CFA requires images over the entire heartbeat, the scans are often performed at 10-20% of the tube current settings that are typically used for coronary CT angiography. A large image noise th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12425

    authors: Fung GSK,Ciuffo L,Ashikaga H,Taguchi K

    更新日期:2017-09-01 00:00:00

  • Robust optimization in IMPT using quadratic objective functions to account for the minimum MU constraint.

    abstract:PURPOSE:Currently, in clinical practice of intensity-modulated proton therapy (IMPT), the influence of the minimum monitor unit (MU) constraint is taken into account through postprocessing after the optimization is completed. This may degrade the plan quality and plan robustness. This study aims to mitigate the impact ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12677

    authors: Shan J,An Y,Bues M,Schild SE,Liu W

    更新日期:2018-01-01 00:00:00

  • In vitro in-stent restenoses evaluated by 3D ultrasound.

    abstract::The purpose of this study was to quantify in-stent restenoses with 3D B mode and power Doppler ultrasound (U.S.) imaging. In-stent restenoses were mimicked with vascular phantoms in which a nonferromagnetic prototype stent (Boston Scientific) and a ferromagnetic clinical stainless steel stent (Palmaz P295) were embedd...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3062944

    authors: Lécart M,Cardinal MH,Qin Z,Soulez G,Cloutier G

    更新日期:2009-02-01 00:00:00

  • Modification of the TASMIP x-ray spectral model for the simulation of microfocus x-ray sources.

    abstract:PURPOSE:The availability of accurate and simple models for the estimation of x-ray spectra is of great importance for system simulation, optimization, or inclusion of photon energy information into data processing. There is a variety of publicly available tools for estimation of x-ray spectra in radiology and mammograp...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4837220

    authors: Sisniega A,Desco M,Vaquero JJ

    更新日期:2014-01-01 00:00:00

  • SU-E-T-203: Development of a QA Software Tool for Automatic Verification of Plan Data Transfer and Delivery.

    abstract:PURPOSE:Consistency verification between the data from treatment planning system (TPS), record and verification system (R&V), and delivered recorder with visual inspection is time consuming and subject to human error. The purpose of this work is to develop a software tool to automatically perform such verifications. M...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735263

    authors: Chen G,Li X

    更新日期:2012-06-01 00:00:00

  • Investigation of gamma knife image registration errors resulting from misalignment between the patient and the imaging axis.

    abstract::The ability of Leksell GammaPlan to perform stereotactic space localizations with image sets where there is misalignment of the patient's head (stereotactic frame and fiducial apparatus) relative to the computed tomography (CT) scanner coordinate system was studied. Misalignment is sometimes necessary for patient comf...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2179751

    authors: Cernica G,Wang Z,Malhotra H,de Boer S,Podgorsak MB

    更新日期:2006-04-01 00:00:00

  • Development and evaluation of a case-based reasoning classifier for prediction of breast biopsy outcome with BI-RADS lexicon.

    abstract::Approximately 70-85% of breast biopsies are performed on benign lesions. To reduce this high number of biopsies performed on benign lesions, a case-based reasoning (CBR) classifier was developed to predict biopsy results from BI-RADS findings. We used 1433 (931 benign) biopsy-proven mammographic cases. CBR similarity ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1501140

    authors: Bilska-Wolak AO,Floyd CE Jr

    更新日期:2002-09-01 00:00:00

  • Temporal stability assessment in shear wave elasticity images validated by deep learning neural network for chronic liver disease fibrosis stage assessment.

    abstract:PURPOSE:To automatically detect and isolate areas of low and high stiffness temporal stability in shear wave elastography (SWE) image sequences and define their impact in chronic liver disease (CLD) diagnosis improvement by means of clinical examination study and deep learning algorithm employing convolutional neural n...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13521

    authors: Gatos I,Tsantis S,Spiliopoulos S,Karnabatidis D,Theotokas I,Zoumpoulis P,Loupas T,Hazle JD,Kagadis GC

    更新日期:2019-05-01 00:00:00

  • Patient dose simulations for scanning-beam digital x-ray tomosynthesis of the lungs.

    abstract:PURPOSE:An improved method of image guidance for lung tumor biopsies could help reduce the high rate of false negatives. The aim of this work is to optimize the geometry of the scanning-beam digital tomography system (SBDX) for providing real-time 3D tomographic reconstructions for target verification. The unique geome...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4826159

    authors: Nelson G,Yoon S,Krishna G,Wilfley B,Fahrig R

    更新日期:2013-11-01 00:00:00

  • Hybrid ZTE/Dixon MR-based attenuation correction for quantitative uptake estimation of pelvic lesions in PET/MRI.

    abstract:PURPOSE:This study introduces a new hybrid ZTE/Dixon MR-based attenuation correction (MRAC) method including bone density estimation for PET/MRI and quantifies the effects of bone attenuation on metastatic lesion uptake in the pelvis. METHODS:Six patients with pelvic lesions were scanned using fluorodeoxyglucose (18F-...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12122

    authors: Leynes AP,Yang J,Shanbhag DD,Kaushik SS,Seo Y,Hope TA,Wiesinger F,Larson PE

    更新日期:2017-03-01 00:00:00

  • Image reconstruction for positron emission tomography based on patch-based regularization and dictionary learning.

    abstract:PURPOSE:Positron emission tomography (PET) is an important tool for nuclear medical imaging. It has been widely used in clinical diagnosis, scientific research, and drug testing. PET is a kind of emission computed tomography. Its basic imaging principle is to use the positron annihilation radiation generated by radionu...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13804

    authors: Zhang W,Gao J,Yang Y,Liang D,Liu X,Zheng H,Hu Z

    更新日期:2019-11-01 00:00:00

  • Resampling: an optimization method for inverse planning in robotic radiosurgery.

    abstract::By design, the range of beam directions in conventional radiosurgery are constrained to an isocentric array. However, the recent introduction of robotic radiosurgery dramatically increases the flexibility of targeting, and as a consequence, beams need be neither coplanar nor isocentric. Such a nonisocentric design per...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2357020

    authors: Schweikard A,Schlaefer A,Adler JR Jr

    更新日期:2006-11-01 00:00:00

  • Intensity modulated arc deliveries approximated by a large number of fixed gantry position sliding window dynamic multileaf collimator fields.

    abstract::The intensity modulated arc has been proposed as an alternative to tomotherapy. Treatment planing systems more typically model the conventional step and shoot or sliding window dynamic multileaf collimator (DMLC) deliveries, and may not support intensity modulated arc therapy (IMAT). As well, another potential drawbac...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1508110

    authors: MacKenzie MA,Robinson DM

    更新日期:2002-10-01 00:00:00

  • Penetration quality measurement for standardization of radiographic image quality.

    abstract::Interchange or transfer of radiographic techniques through measurement of penetration quality is discussed. The usefulness of this achievement is verified using an aluminum step wedge. Also, the procedure described provides a method for performance evaluation of kVp compensation on diagnostic x-ray installations. ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594156

    authors: Lin PJ

    更新日期:1975-01-01 00:00:00

  • MO-C-BRCD-03: The Role of Informatics in Medical Physics and Vice Versa.

    abstract::Like Medical Physics, Imaging Informatics encompasses concepts touching every aspect of the imaging chain from image creation, acquisition, management and archival, to image processing, analysis, display and interpretation. The two disciplines are in fact quite complementary, with similar goals to improve the quality ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735777

    authors: Andriole K

    更新日期:2012-06-01 00:00:00