Mediated xanthine oxidase potentiometric biosensors for hypoxanthine based on ferrocene carboxylic acid modified electrode.

Abstract:

:A potentiometric enzyme electrode for detection of hypoxanthine (Hx) in fish meat is described. The sensor was developed by entrapment of xanthine oxidase (XOD) and ferrocene carboxylic acid (Fc) into polypyrrole (PPy) film during galvanostatic polymerisation film formation. The responses for Hx were obtained in 0-05 M phosphate buffer (pH 7.1) at 0.0 mV vs Ag/AgCl (3M KCl). The optimum condition for the formation of PPy-XOD-Fc film include 0.4M PPy, 6.2U/mL XOD, 40 mM Fc, polymerisation time of 200 s and applied current density of 0-5 mA cm(-2). The sensor provides a linear response to Hx in concentration range of 5-20 μM, (r=0.998) and was successfully used for determination of Hx in fish.

journal_name

Food Chem

journal_title

Food chemistry

authors

Lawal AT,Adeloju SB

doi

10.1016/j.foodchem.2012.07.052

subject

Has Abstract

pub_date

2012-12-15 00:00:00

pages

2982-7

issue

4

eissn

0308-8146

issn

1873-7072

pii

S0308-8146(12)01166-1

journal_volume

135

pub_type

杂志文章