Clinical commissioning and use of the Novalis Tx linear accelerator for SRS and SBRT.

Abstract:

:The purpose of this study was to perform comprehensive measurements and testing of a Novalis Tx linear accelerator, and to develop technical guidelines for com-missioning from the time of acceptance testing to the first clinical treatment. The Novalis Tx (NTX) linear accelerator is equipped with, among other features, a high-definition MLC (HD120 MLC) with 2.5 mm central leaves, a 6D robotic couch, an optical guidance positioning system, as well as X-ray-based image guidance tools to provide high accuracy radiation delivery for stereotactic radiosurgery and stereotactic body radiation therapy procedures. We have performed extensive tests for each of the components, and analyzed the clinical data collected in our clinic. We present technical guidelines in this report focusing on methods for: (1) efficient and accurate beam data collection for commissioning treatment planning systems, including small field output measurements conducted using a wide range of detectors; (2) commissioning tests for the HD120 MLC; (3) data collection for the baseline characteristics of the on-board imager (OBI) and ExacTrac X-ray (ETX) image guidance systems in conjunction with the 6D robotic couch; and (4) end-to-end testing of the entire clinical process. Established from our clinical experience thus far, recommendations are provided for accurate and efficient use of the OBI and ETX localization systems for intra- and extracranial treatment sites. Four results are presented. (1) Basic beam data measurements: Our measurements confirmed the necessity of using small detectors for small fields. Total scatter factors varied significantly (30% to approximately 62%) for small field measurements among detectors. Unshielded stereotactic field diode (SFD) overestimated dose by ~ 2% for large field sizes. Ion chambers with active diameters of 6 mm suffered from significant volume averaging. The sharpest profile penumbra was observed for the SFD because of its small active diameter (0.6 mm). (2) MLC commissioning: Winston Lutz test, light/radiation field congruence, and Picket Fence tests were performed and were within criteria established by the relevant task group reports. The measured mean MLC transmission and dynamic leaf gap of 6 MV SRS beam were 1.17% and 0.36 mm, respectively. (3) Baseline characteristics of OBI and ETX: The isocenter localization errors in the left/right, posterior/anterior, and superior/inferior directions were, respectively, -0.2 ± 0.2 mm, -0.8 ± 0.2 mm, and -0.8 ± 0.4 mm for ETX, and 0.5 ± 0.7 mm, 0.6 ± 0.5 mm, and 0.0 ± 0.5 mm for OBI cone-beam computed tomography. The registration angular discrepancy was 0.1 ± 0.2°, and the maximum robotic couch error was 0.2°. (4) End-to-end tests: The measured isocenter dose differences from the planned values were 0.8% and 0.4%, measured respectively by an ion chamber and film. The gamma pass rate, measured by EBT2 film, was 95% (3% DD and 1 mm DTA). Through a systematic series of quantitative commissioning experiments and end-to-end tests and our initial clinical experience, described in this report, we demonstrate that the NTX is a robust system, with the image guidance and MLC requirements to treat a wide variety of sites - in particular for highly accurate delivery of SRS and SBRT-based treatments.

journal_name

J Appl Clin Med Phys

authors

Kim J,Wen N,Jin JY,Walls N,Kim S,Li H,Ren L,Huang Y,Doemer A,Faber K,Kunkel T,Balawi A,Garbarino K,Levin K,Patel S,Ajlouni M,Miller B,Nurushev T,Huntzinger C,Schulz R,Chetty IJ,Movsas B,Ryu S

doi

10.1120/jacmp.v13i3.3729

subject

Has Abstract

pub_date

2012-05-10 00:00:00

pages

3729

issue

3

issn

1526-9914

journal_volume

13

pub_type

杂志文章
  • Quantitative analysis of image quality for acceptance and commissioning of an MRI simulator with a semiautomatic method.

    abstract::Magnetic Resonance Imaging (MRI) simulation differs from diagnostic MRI in purpose, technical requirements, and implementation. We propose a semiautomatic method for image acceptance and commissioning for the scanner, the radiofrequency (RF) coils, and pulse sequences for an MRI simulator. The ACR MRI accreditation la...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12311

    authors: Chen X,Dai J

    更新日期:2018-05-01 00:00:00

  • Electrocardiogram-gated coronary CT angiography dose estimates using ImPACT.

    abstract::The primary study objective was to assess radiation doses using a modified form of the Imaging Performance Assessment of Computed Tomography (CT) scanner (ImPACT) patient dosimetry for cardiac applications on an Aquilion ONE ViSION Edition scanner, including the Ca score, target computed tomography angiography (CTA), ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i4.6218

    authors: Kobayashi M,Asada Y,Matsubara K,Suzuki S,Koshida K,Matsunaga Y,Haba T,Kawaguchi A,Toyama H,Kato R

    更新日期:2016-07-08 00:00:00

  • Dose perturbation due to the polysulfone cap surrounding a Fletcher-Williamson colpostat.

    abstract::We conducted a metrological evaluation of the dosimetric impact due to the polysulfone cap used with the Fletcher-Williamson (FW) colpostat for 192Ir high-dose rate and pulsed-dose rate intracavitary brachytherapy using Monte Carlo simulations. Polysulfone caps with diameter of 30 mm, 25 mm, 20 mm, and 16 mm (mini-ovo...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v11i1.3146

    authors: Price MJ,Kry SF,Eifel PJ,Salehpour M,Mourtada F

    更新日期:2010-01-28 00:00:00

  • FDG PET/CT for rectal carcinoma radiotherapy treatment planning: comparison of functional volume delineation algorithms and clinical challenges.

    abstract::PET/CT imaging could improve delineation of rectal carcinoma gross tumor volume (GTV) and reduce interobserver variability. The objective of this work was to compare various functional volume delineation algorithms. We enrolled 31 consecutive patients with locally advanced rectal carcinoma. The FDG PET/CT and the high...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i5.4696

    authors: Withofs N,Bernard C,Van der Rest C,Martinive P,Hatt M,Jodogne S,Visvikis D,Lee JA,Coucke PA,Hustinx R

    更新日期:2014-09-08 00:00:00

  • Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations.

    abstract::The purpose of this study was to implement full/half bowtie filter models in a com-mercial treatment planning system (TPS) to calculate kilovoltage (kV) cone-beam CT (CBCT) doses of Varian On-Board Imager (OBI) kV X-ray imaging system. The full/half bowtie filter models were created as compensators in Pinnacle TPS usi...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i2.5988

    authors: Kim S,Alaei P

    更新日期:2016-03-08 00:00:00

  • Effect of electron contamination on in vivo dosimetry for lung block shielding during TBI.

    abstract::Our institution performs in vivo verification measurement for each of our total body irradiation (TBI) patients with optically stimulated luminescent dosimeters (OSLD). The lung block verification measurements were commonly higher than expected. The aim of this work is to understand this discrepancy and improve the ac...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i3.6128

    authors: Narayanasamy G,Cruz W,Saenz DL,Stathakis S,Papanikolaou N,Kirby N

    更新日期:2016-05-08 00:00:00

  • In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment.

    abstract::Intraoperative electron radiation therapy (IOERT) cannot usually benefit, as conventional external radiotherapy, from software systems of treatment planning based on computed tomography and from common dose verify procedures. For this reason, in vivo film dosimetry (IVFD) proves to be an effective methodology to evalu...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i1.5065

    authors: Severgnini M,de Denaro M,Bortul M,Vidali C,Beorchia A

    更新日期:2014-01-08 00:00:00

  • Dose calculation differences between Monte Carlo and pencil beam depend on the tumor locations and volumes for lung stereotactic body radiation therapy.

    abstract::Stereotactic body radiation therapy (SBRT) has been increasingly used as an efficacious treatment modality for early-stage non-small cell lung cancer. The accuracy of dose calculations is compromised due to the presence of inhomogeneity. For the purpose of a consistent prescription, radiation doses were calculated wit...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i2.4011

    authors: Zhuang T,Djemil T,Qi P,Magnelli A,Stephans K,Videtic G,Xia P

    更新日期:2013-03-04 00:00:00

  • Dependence of intrafraction motion on fraction duration for pediatric patients with brain tumors.

    abstract::The purpose of this study was to quantify the intrafraction motion of pediatric patients with brain tumors during radiation therapy and investigate any correlation between motion, use of general anesthesia, and daily treatment duration. 100 pediatric patients with a mean age of 8.5 years (range: 1.0 to 17.8) were incl...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v12i4.3609

    authors: Beltran C,Merchant TE

    更新日期:2011-11-15 00:00:00

  • A robust VMAT delivery solution for single-fraction lung SABR utilizing FFF beams minimizing dosimetric compromise.

    abstract::Peripheral lung lesions treated with a single fraction of stereotactic ablative body radiotherapy (SABR) utilizing volumetric modulated arc therapy (VMAT) delivery and flattening filter-free (FFF) beams represent a potentially high-risk scenario for clinically significant dose blurring effects due to interplay between...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12919

    authors: Burton A,Offer K,Hardcastle N

    更新日期:2020-08-01 00:00:00

  • Assessment of PlanIQ Feasibility DVH for head and neck treatment planning.

    abstract:INTRODUCTION:Designing a radiation plan that optimally delivers both target coverage and normal tissue sparing is challenging. There are limited tools to determine what is dosimetrically achievable and frequently the experience of the planner/physician is relied upon to make these determinations. PlanIQ software provid...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12165

    authors: Fried DV,Chera BS,Das SK

    更新日期:2017-09-01 00:00:00

  • A line-profile based double partial fusion method for acquiring planning CT of oversized patients in radiation treatment.

    abstract::True 3D CT dataset for treatment planning of an oversized patient is difficult to acquire due to the bore size and field of view (FOV) reconstruction. This project aims to provide a simple approach to reconstruct true CT data for oversize patients using CT scanner with limited FOV by acquiring double partial CT (left ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v13i2.3629

    authors: Wu H,Zhao Q,Cao M,Das I

    更新日期:2012-03-08 00:00:00

  • Gemstone spectral imaging: determination of CT to ED conversion curves for radiotherapy treatment planning.

    abstract::The monochromatic images acquired by Gemstone spectral imaging (GSI) mode on the GE CT750 HD theoretically determines the computed tomography (CT) number more accurately than that of conventional scanner. Using the former, the CT number is calculated from (synthesized) monoenergetic X-ray data. We reasoned that the mo...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i5.4335

    authors: Yagi M,Ueguchi T,Koizumi M,Ogata T,Yamada S,Takahashi Y,Sumida I,Akino Y,Konishi K,Isohashi F,Tomiyama N,Yoshioka Y,Ogawa K

    更新日期:2013-09-06 00:00:00

  • Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets.

    abstract::Studies have shown that source dwells within Varian's HDR CT/MR compatible ring applicators can deviate from intended positions by several millimeters. Quantifying this offset is an important part of commissioning. The aims of this study were to: 1) determine the reproducibility of the offset, 2) study the interobserv...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v12i4.3447

    authors: McMahon R,Zhuang T,Steffey BA,Song H,Craciunescu OI

    更新日期:2011-11-15 00:00:00

  • Evaluating the positional uncertainty of intrafraction, adjacent fields, and daily setup with the BrainLAB ExacTrac system in patients who are receiving craniospinal irradiation.

    abstract:PURPOSE:To investigate the daily setup, interfraction motion, variability in the junction areas, and dosimetric effect in craniospinal irradiation (CSI) patients. METHODS:Fifteen CSI patients who had undergone split-field IMRT were followed in the study. Previous, middle, and posttreatment, each target volume position...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12909

    authors: Duan X,Zhou Y,Dai H,Zhao L,Qian J,Yang D,Zhang L,Luo C,Li G

    更新日期:2020-08-01 00:00:00

  • Voxel-based supervised machine learning of peripheral zone prostate cancer using noncontrast multiparametric MRI.

    abstract:PURPOSE:The aim of this study was to develop and assess the performance of supervised machine learning technique to classify magnetic resonance imaging (MRI) voxels as cancerous or noncancerous using noncontrast multiparametric MRI (mp-MRI), comprised of T2-weighted imaging (T2WI), diffusion-weighted imaging (DWI), and...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12992

    authors: Gholizadeh N,Simpson J,Ramadan S,Denham J,Lau P,Siddique S,Dowling J,Welsh J,Chalup S,Greer PB

    更新日期:2020-10-01 00:00:00

  • Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification.

    abstract::Radiochromic film has become an important tool to verify dose distributions in highly conformal radiation therapy such as IMRT. Recently, a new generation of these films, EBT3, has become available. EBT3 has the same composition and thickness of the sensitive layer of the previous EBT2 films, but its symmetric layer c...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i2.4111

    authors: Casanova Borca V,Pasquino M,Russo G,Grosso P,Cante D,Sciacero P,Girelli G,La Porta MR,Tofani S

    更新日期:2013-03-04 00:00:00

  • Effectiveness of a novel gas-release endorectal balloon in the removal of rectal gas for prostate proton radiation therapy.

    abstract::Endorectal balloons (ERBs) are routinely used in prostate proton radiation therapy to immobilize the prostate and spare the rectal wall. Rectal gas can distend the rectum and displace the prostate even in the presence of ERBs. The purpose of this work was to quantify the effects an ERB with a passive gas release condu...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v13i5.3945

    authors: Wootton LS,Kudchadker RJ,Beddar AS,Lee AK

    更新日期:2012-09-06 00:00:00

  • The development of a 4D treatment planning methodology to simulate the tracking of central lung tumors in an MRI-linac.

    abstract:PURPOSE:Targeting and tracking of central lung tumors may be feasible on the Elekta MRI-linac (MRL) due to the soft-tissue visualization capabilities of MRI. The purpose of this work is to develop a novel treatment planning methodology to simulate tracking of central lung tumors with the MRL and to quantify the benefit...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12233

    authors: Al-Ward SM,Kim A,McCann C,Ruschin M,Cheung P,Sahgal A,Keller BM

    更新日期:2018-01-01 00:00:00

  • AAPM Task Group 103 report on peer review in clinical radiation oncology physics.

    abstract::This report provides guidelines for a peer review process between two clinical radiation oncology physicists. While the Task Group's work was primarily focused on ensuring timely and productive independent reviews for physicists in solo practice, these guidelines may also be appropriate for physicists in a group setti...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章,实务指引

    doi:10.1120/jacmp.v6i4.2142

    authors: Halvorsen PH,Das IJ,Fraser M,Freedman DJ,Rice RE 3rd,Ibbott GS,Parsai EI,Robin TT Jr,Thomadsen BR,American Association of Physicists in Medicine.

    更新日期:2005-10-01 00:00:00

  • Quantitative variations in texture analysis features dependent on MRI scanning parameters: A phantom model.

    abstract:OBJECTIVES:To evaluate the influence of MRI scanning parameters on texture analysis features. METHODS:Publicly available data from the Reference Image Database to Evaluate Therapy Response (RIDER) project sponsored by The Cancer Imaging Archive included MRIs on a phantom comprised of 18 25-mm doped, gel-filled tubes, ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12482

    authors: Buch K,Kuno H,Qureshi MM,Li B,Sakai O

    更新日期:2018-11-01 00:00:00

  • CT and MR image fusion of tandem and ring applicator using rigid registration in intracavitary brachytherapy planning.

    abstract::The purpose of this study is to find the uncertainties in the reconstruction of MR compatible ring-tandem intracavitary applicators of high-dose rate image-based brachytherapy treatment planning using rigid registration of 3D MR and CT image fusion. Tandem and ring reconstruction in MR image based brachytherapy planni...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i2.4206

    authors: Oinam AS,Tomar P,Patel FD,Singh L,Rai B,Bahl A

    更新日期:2014-03-06 00:00:00

  • Method and phantom to study combined effects of in-plane (x,y) and z-axis resolution for 3D CT imaging.

    abstract::Increasingly, the advent of multislice CT scanners, volume CT scanners, and total body spiral acquisition modes has led to the use of Multi Planar Reconstruction and 3D datasets. In considering 3D resolution properties of a CT system it is important to note that both the in-plane (x,y) and z-axis (slice thickness) inf...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i5.6294

    authors: Goodenough D,Levy J,Kristinsson S,Fredriksson J,Olafsdottir H,Healy A

    更新日期:2016-09-08 00:00:00

  • Simplified "on-couch" daily quality assurance procedure for CT simulators.

    abstract::For most computed tomography (CT) simulators, radiation therapists must remove the flat couch top to perform daily CT quality assurance (QA), and then use separate tools to perform localization-laser QA. This process wastes time and effort and creates the opportunity for accidents to occur. In this study, we tested a ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v10i3.2844

    authors: Liu RR,Prado K,Gillin M

    更新日期:2009-06-17 00:00:00

  • Evaluation of an electron Monte Carlo dose calculation algorithm for treatment planning.

    abstract::The purpose of this study is to evaluate the accuracy of the electron Monte Carlo (eMC) dose calculation algorithm included in a commercial treatment planning system and compare its performance against an electron pencil beam algorithm. Several tests were performed to explore the system's behavior in simple geometries...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i3.4636

    authors: Chamberland E,Beaulieu L,Lachance B

    更新日期:2015-05-08 00:00:00

  • Preventative maintenance and unscheduled downtime from an economic perspective.

    abstract::A spreadsheet-based model for economically characterizing the operation of a radiation treatment program has been used to perform a quantitative financial analysis of scheduled and unscheduled downtime. The incremental cost of downtime is broken down into three categories: remuneration of in-house or third party servi...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v1i2.2647

    authors: Dunscombe P,Roberts G,Valiquette L

    更新日期:2000-04-01 00:00:00

  • Evaluation of the influence of tumor location and size on the difference of dose calculation between Ray Tracing algorithm and Fast Monte Carlo algorithm in stereotactic body radiotherapy of non-small cell lung cancer using CyberKnife.

    abstract::This study evaluated the extent of improvement in dose predication accuracy achieved by the Fast Monte Carlo algorithm (MC) compared to the Ray Tracing algorithm (RAT) in stereotactic body radiotherapy (SBRT) of non-small cell lung cancer (NSCLC), and how their differences were influenced by the tumor site and size. T...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i5.4280

    authors: Wu VW,Tam KW,Tong SM

    更新日期:2013-09-06 00:00:00

  • A dosimetric comparison of stereotactic body radiation therapy techniques for lung cancer: robotic versus conventional linac-based systems.

    abstract::The aim of this study is to compare the dosimetric characteristics of robotic and conventional linac-based SBRT techniques for lung cancer, and to provide planning guidance for each modality. Eight patients who received linac-based SBRT were retrospectively included in this study. A dose of 60 Gy given in three fracti...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v11i3.3223

    authors: Ding C,Chang CH,Haslam J,Timmerman R,Solberg T

    更新日期:2010-06-29 00:00:00

  • Peripheral dose distributions for a linear accelerator equipped with a secondary multileaf collimator and universal wedge.

    abstract::The American Association of Physicists in Medicine Task Group 36 (AAPM TG-36) data can be used to estimate peripheral dose (PD) distributions outside the primary radiation field. However, the report data does not apply to linear accelerators equipped with a multileaf collimator (MLC) and universal wedge (UW). Tertiary...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v3i4.2554

    authors: Mutic S,Esthappan J,Klein EE

    更新日期:2002-10-01 00:00:00

  • Translucent poly(vinyl alcohol) cryogel dosimeters for simultaneous dose buildup and monitoring during chest wall radiation therapy.

    abstract::Chest wall radiation therapy treatment delivery was monitored using a 5 mm thick radiochromic poly(vinyl alcohol) cryogel that also provided buildup material. The cryogels were used to detect positioning errors and measure the impact of shifts for a chest wall treatment that was delivered to a RANDO phantom. The phant...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i5.6148

    authors: Eyadeh MM,Weston MA,Juhasz J,Diamond KR

    更新日期:2016-09-08 00:00:00