Optimization of image quality and dose for Varian aS500 electronic portal imaging devices (EPIDs).

Abstract:

:This study was carried out to investigate whether the electronic portal imaging (EPI) acquisition process could be optimized, and as a result tolerance and action levels be set for the PIPSPro QC-3V phantom image quality assessment. The aim of the optimization process was to reduce the dose delivered to the patient while maintaining a clinically acceptable image quality. This is of interest when images are acquired in addition to the planned patient treatment, rather than images being acquired using the treatment field during a patient's treatment. A series of phantoms were used to assess image quality for different acquisition settings relative to the baseline values obtained following acceptance testing. Eight Varian aS500 EPID systems on four matched Varian 600C/D linacs and four matched Varian 2100C/D linacs were compared for consistency of performance and images were acquired at the four main orthogonal gantry angles. Images were acquired using a 6 MV beam operating at 100 MU min(-1) and the low-dose acquisition mode. Doses used in the comparison were measured using a Farmer ionization chamber placed at d(max) in solid water. The results demonstrated that the number of reset frames did not have any influence on the image contrast, but the number of frame averages did. The expected increase in noise with corresponding decrease in contrast was also observed when reducing the number of frame averages. The optimal settings for the low-dose acquisition mode with respect to image quality and dose were found to be one reset frame and three frame averages. All patients at the Northern Ireland Cancer Centre are now imaged using one reset frame and three frame averages in the 6 MV 100 MU min(-1) low-dose acquisition mode. Routine EPID QC contrast tolerance (+/-10) and action (+/-20) levels using the PIPSPro phantom based around expected values of 190 (Varian 600C/D) and 225 (Varian 2100C/D) have been introduced. The dose at dmax from electronic portal imaging has been reduced by approximately 28%, and while the image quality has been reduced, the images produced are still clinically acceptable.

journal_name

Phys Med Biol

authors

McGarry CK,Grattan MW,Cosgrove VP

doi

10.1088/0031-9155/52/23/006

subject

Has Abstract

pub_date

2007-12-07 00:00:00

pages

6865-77

issue

23

eissn

0031-9155

issn

1361-6560

pii

S0031-9155(07)60953-1

journal_volume

52

pub_type

杂志文章
  • Effects of pulmonary inhalation on hyperpolarized krypton-83 magnetic resonance T1 relaxation.

    abstract::The (83)Kr magnetic resonance (MR) relaxation time T(1) of krypton gas in contact with model surfaces was previously found to be highly sensitive to surface composition, surface-to-volume ratio, and surface temperature. The work presented here explored aspects of pulmonary (83)Kr T(1) relaxation measurements in excise...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/56/13/001

    authors: Stupic KF,Elkins ND,Pavlovskaya GE,Repine JE,Meersmann T

    更新日期:2011-07-07 00:00:00

  • Broadband ultrasound attenuation imaging: algorithm development and clinical assessment of a region growing technique.

    abstract::This paper presents a computerized method for the selection of an irregular region of interest (ROI) in broadband ultrasound attenuation (BUA) images. A region growing algorithm searches an initial region in the posterior part of the calcaneus until the pixel with the lowest attenuation value is found; this is the sta...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/47/2/310

    authors: Damilakis J,Stratakis J,Perisinakis K,Gourtsoyiannis N

    更新日期:2002-01-21 00:00:00

  • Dosimetry robustness with stochastic optimization.

    abstract::All radiation therapy treatment planning relies on accurate dose calculation. Uncertainties in dosimetric prediction can significantly degrade an otherwise optimal plan. In this work, we introduce a robust optimization method which handles dosimetric errors and warrants for high-quality IMRT plans. Unlike other dose e...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/54/11/010

    authors: Nohadani O,Seco J,Martin BC,Bortfeld T

    更新日期:2009-06-07 00:00:00

  • Head scatter modelling for irregular field shaping and beam intensity modulation.

    abstract::Scattered radiation from within the treatment head can contribute significant dose to all parts of a radiotherapy treatment field. A multileaf collimator may be used to create an arbitrarily shaped field, and may also be used, under dynamic control, to modulate the beam intensity over the field. This method of intensi...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/42/9/006

    authors: Hounsell AR,Wilkinson JM

    更新日期:1997-09-01 00:00:00

  • Experimental evaluation of fiber-interspaced antiscatter grids for large patient imaging with digital x-ray systems.

    abstract::Radiographic imaging of large patients is compromised by x-ray scatter. Optimization of digital x-ray imaging systems used for projection radiography requires the use of the best possible antiscatter grid. The performance of antiscatter grids used in conjunction with digital x-ray imaging systems can be characterized ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/52/16/010

    authors: Fetterly KA,Schueler BA

    更新日期:2007-08-21 00:00:00

  • Automatic bi-objective parameter tuning for inverse planning of high-dose-rate prostate brachytherapy.

    abstract::We present an automatic bi-objective parameter-tuning approach for inverse planning methods for high-dose-rate prostate brachytherapy, which aims to overcome the difficult and time-consuming manual parameter tuning that is currently required to obtain patient-specific high-quality treatment plans. We modelled treatmen...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab7362

    authors: Maree SC,Bosman PAN,van Wieringen N,Niatsetski Y,Pieters BR,Bel A,Alderliesten T

    更新日期:2020-04-02 00:00:00

  • A note on computing the derivative at a constant direction.

    abstract::The derivative at constant direction is frequently used in inversion of cone-beam data. Several algorithms for computing the derivative have been proposed in the literature. The best algorithm to date has been proposed recently by Noo et al (2007 Phys. Med. Biol. 52 5393-414). In this note we propose a new, simple and...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/56/4/N01

    authors: Katsevich A

    更新日期:2011-02-21 00:00:00

  • Simultaneous optimization of RBE-weighted dose and nanometric ionization distributions in treatment planning with carbon ions.

    abstract::Inverse treatment planning in intensity modulated particle therapy (IMPT) with scanned carbon-ion beams is currently based on the optimization of RBE-weighted dose to satisfy requirements of target coverage and limited toxicity to organs-at-risk (OARs) and healthy tissues. There are many feasible IMPT plans that meet ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aaf400

    authors: Burigo LN,Ramos-Méndez J,Bangert M,Schulte RW,Faddegon B

    更新日期:2019-01-04 00:00:00

  • PDT dose dosimetry for Photofrin-mediated pleural photodynamic therapy (pPDT).

    abstract::Photosensitizer fluorescence excited by photodynamic therapy (PDT) treatment light can be used to monitor the in vivo concentration of the photosensitizer and its photobleaching. The temporal integral of the product of in vivo photosensitizer concentration and light fluence is called PDT dose, which is an important do...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa9874

    authors: Ong YH,Kim MM,Finlay JC,Dimofte A,Singhal S,Glatstein E,Cengel KA,Zhu TC

    更新日期:2017-12-29 00:00:00

  • High-resolution fluence verification for treatment plan specific QA in ion beam radiotherapy.

    abstract::Ion beam radiotherapy exploits the finite range of ion beams and the increased dose deposition of ions toward the end of their range in material. This results in high dose conformation to the target region, which can be further increased using scanning ion beams. The standard method for patient-plan verification in io...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/58/6/1725

    authors: Martišíková M,Brons S,Hesse BM,Jäkel O

    更新日期:2013-03-21 00:00:00

  • Simulation of voltage-sensitive optical signals in three-dimensional slabs of cardiac tissue: application to transillumination and coaxial imaging methods.

    abstract::Voltage-sensitive dyes are an important tool in visualizing electrical activity in cardiac tissue. Until today, they have mainly been applied in cardiac electrophysiology to subsurface imaging. In the present study, we assess different imaging methods used in optical tomography with respect to their effectiveness in v...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/50/2/003

    authors: Bernus O,Wellner M,Mironov SF,Pertsov AM

    更新日期:2005-01-21 00:00:00

  • Investigation of the sag in linac secondary collimator and MLC carriage during arc deliveries.

    abstract::In modern radiotherapy, it is vitally important to monitor the performance of all linac components including the collimation system. In this study, a simple measurement method and accurate algorithm are introduced for investigation of the secondary and tertiary collimator sag during radiotherapy arc treatments. The me...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/57/12/N209

    authors: Rowshanfarzad P,Sabet M,O'Connor DJ,Greer PB

    更新日期:2012-06-21 00:00:00

  • An analytical reconstruction model of the spread-out Bragg peak using laser-accelerated proton beams.

    abstract::With the development of laser technology, laser-driven proton acceleration provides a new method for proton tumor therapy. However, it has not been applied in practice because of the wide and decreasing energy spectrum of laser-accelerated proton beams. In this paper, we propose an analytical model to reconstruct the ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aa6fce

    authors: Tao L,Zhu K,Zhu J,Xu X,Lin C,Ma W,Lu H,Zhao Y,Lu Y,Chen JE,Yan X

    更新日期:2017-07-07 00:00:00

  • Computational dosimetry for grounded and ungrounded human models due to contact current.

    abstract::This study presents the computational dosimetry of contact currents for grounded and ungrounded human models. The uncertainty of the quasi-static (QS) approximation of the in situ electric field induced in a grounded/ungrounded human body due to the contact current is first estimated. Different scenarios of cylindrica...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/58/15/5153

    authors: Chan KH,Hattori J,Laakso I,Hirata A,Taki M

    更新日期:2013-08-07 00:00:00

  • A framework for organ dose estimation in x-ray angiography and interventional radiology based on dose-related data in DICOM structured reports.

    abstract::Although interventional x-ray angiography (XA) procedures involve relatively high radiation doses that can lead to deterministic tissue reactions in addition to stochastic effects, convenient and accurate estimation of absorbed organ doses has traditionally been out of reach. This has mainly been due to the absence of...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/61/8/3063

    authors: Omar A,Bujila R,Fransson A,Andreo P,Poludniowski G

    更新日期:2016-04-21 00:00:00

  • Exploration and application of phenomenological RBE models for proton therapy.

    abstract::The relative biological effectiveness (RBE) of protons varies with multiple physical and biological factors. Phenomenological RBE models have been developed to include such factors in the estimation of a variable RBE, in contrast to the clinically applied constant RBE of 1.1. In this study, eleven published phenomenol...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aad9db

    authors: Rørvik E,Fjæra LF,Dahle TJ,Dale JE,Engeseth GM,Stokkevåg CH,Thörnqvist S,Ytre-Hauge KS

    更新日期:2018-09-13 00:00:00

  • Quantitation in SPECT using an effective model of the scattering.

    abstract::A new method for correcting simultaneously the attenuation, scatter and resolution effects in SPECT has been developed for the case of a homogeneous medium. It is based on an effective model of the scatter process. This model depends on only four parameters which are determined experimentally and remain independent of...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/39/4/005

    authors: Walrand SH,van Elmbt LR,Pauwels S

    更新日期:1994-04-01 00:00:00

  • An in silico framework to analyze the anisotropic shear wave mechanics in cardiac shear wave elastography.

    abstract::Shear wave elastography (SWE) is a potential tool to non-invasively assess cardiac muscle stiffness. This study focused on the effect of the orthotropic material properties and mechanical loading on the performance of cardiac SWE, as it is known that these factors contribute to complex 3D anisotropic shear wave propag...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aaaffe

    authors: Caenen A,Pernot M,Peirlinck M,Mertens L,Swillens A,Segers P

    更新日期:2018-03-23 00:00:00

  • Modelling the electrical properties of tissue as a porous medium.

    abstract::Models of the electrical properties of biological tissue have been the subject of many studies. These models have sought to explain aspects of the dielectric dispersion of tissue. This paper develops a mathematical model of the complex permittivity of tissue as a function of frequency f, in the range 10(4) < f < 10(7)...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/52/23/016

    authors: Smye SW,Evans CJ,Robinson MP,Sleeman BD

    更新日期:2007-12-07 00:00:00

  • Extension of the NCAT phantom for the investigation of intra-fraction respiratory motion in IMRT using 4D Monte Carlo.

    abstract::The purpose of this work was to create a computational platform for studying motion in intensity modulated radiotherapy (IMRT). Specifically, the non-uniform rational B-spline (NURB) cardiac and torso (NCAT) phantom was modified for use in a four-dimensional Monte Carlo (4D-MC) simulation system to investigate the eff...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/55/5/014

    authors: McGurk R,Seco J,Riboldi M,Wolfgang J,Segars P,Paganetti H

    更新日期:2010-03-07 00:00:00

  • Accuracy of Monte Carlo photon transport simulation in characterizing brachytherapy dosimeter energy-response artefacts.

    abstract::Practical dosimeters in brachytherapy, such as thermoluminescent dosimeters (TLD) and diodes, are usually calibrated against low-energy megavoltage beams. To measure absolute dose rate near a brachytherapy source, it is necessary to establish the energy response of the detector relative to that of the calibration ener...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/41/6/004

    authors: Das RK,Li Z,Perera H,Williamson JF

    更新日期:1996-06-01 00:00:00

  • On the total system error of a robotic radiosurgery system: phantom measurements, clinical evaluation and long-term analysis.

    abstract::The total system error (TSE) of a CyberKnife® system was measured using two phantom-based methods and one patient-based method. The standard radiochromic film (RCF) end-to-end (E2E) test using an anthropomorphic head and neck phantom and isocentric treatment delivery was used with the 6Dskull, Fiducial and Xsight® spi...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aad516

    authors: Pantelis E,Moutsatsos A,Antypas C,Zoros E,Pantelakos P,Lekas L,Romanelli P,Zourari K,Hourdakis CJ

    更新日期:2018-08-20 00:00:00

  • The quality of ionising radiations emitted by radionuclides incorporated into mammalian cells.

    abstract::Radionuclides which decay by electron capture accompanied by Auger electron cascades, or by beta emission are thought to be excessively damaging when incorporated into the molecular structure of mammalian cell nuclei and thereby pose a radiation hazard which is not assessable by conventional dosimetry. Survival data, ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/34/7/003

    authors: Younis AR,Watt DE

    更新日期:1989-07-01 00:00:00

  • Volumetric CT with sparse detector arrays (and application to Si-strip photon counters).

    abstract::Novel x-ray medical imaging sensors, such as photon counting detectors (PCDs) and large area CCD and CMOS cameras can involve irregular and/or sparse sampling of the detector plane. Application of such detectors to CT involves undersampling that is markedly different from the commonly considered case of sparse angular...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/61/1/90

    authors: Sisniega A,Zbijewski W,Stayman JW,Xu J,Taguchi K,Fredenberg E,Lundqvist M,Siewerdsen JH

    更新日期:2016-01-07 00:00:00

  • Comment on 'Proton beam monitor chamber calibration'.

    abstract::We comment on a recent article (Gomà et al 2014 Phys. Med. Biol. 59 4961-71) which compares different routes of reference dosimetry for the energy dependent beam monitor calibration in scanned proton beams. In this article, a 3% discrepancy is reported between a Faraday cup and a plane-parallel ionization chamber in t...

    journal_title:Physics in medicine and biology

    pub_type: 评论,杂志文章

    doi:10.1088/0031-9155/61/17/6585

    authors: Palmans H,Vatnitsky SM

    更新日期:2016-09-07 00:00:00

  • Array detection for speckle reduction in optical coherence microscopy.

    abstract::This paper introduces a spatial-diversity method for speckle suppression in optical coherence microscopy. The method is based on combining interference signals from an array of detectors placed in the back focal plane of the objective lens, such that elements receive light backscattered from the sample volume at diffe...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/42/7/015

    authors: Schmitt JM

    更新日期:1997-07-01 00:00:00

  • Particle beam microstructure reconstruction and coincidence discrimination in PET monitoring for hadron therapy.

    abstract::Positron emission tomography is one of the most mature techniques for monitoring the particles range in hadron therapy, aiming to reduce treatment uncertainties and therefore the extent of safety margins in the treatment plan. In-beam PET monitoring has been already performed using inter-spill and post-irradiation dat...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aafa28

    authors: Kostara E,Sportelli G,Belcari N,Camarlinghi N,Cerello P,Del Guerra A,Ferrero V,Fiorina E,Giraudo G,Morrocchi M,Pennazio F,Pullia M,Rivetti A,Rolo MD,Rosso V,Wheadon R,Bisogni MG

    更新日期:2019-01-21 00:00:00

  • Trimmer sequencing time minimization during dynamically collimated proton therapy using a colony of cooperating agents.

    abstract::The dynamic collimation system (DCS) can be combined with pencil beam scanning proton therapy to deliver highly conformal treatment plans with unique collimation at each energy layer. This energy layer-specific collimation is accomplished through the synchronized motion of four trimmer blades that intercept the proton...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/ab416d

    authors: Smith BR,Hyer DE,Flynn RT,Hill PM,Culberson WS

    更新日期:2019-10-21 00:00:00

  • An automatic algorithm to exploit the symmetries of the system response matrix in positron emission tomography iterative reconstruction.

    abstract::Positron emission tomography (PET) iterative 3D reconstruction is a very computational demanding task. One of the main issues of the iterative reconstruction concerns the management of the system response matrix (SRM). The SRM models the relationship between the projection and the voxel space and its memory footprint ...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/1361-6560/aae12b

    authors: Camarlinghi N,Sportelli G,Guerra AD,Belcari N

    更新日期:2018-09-28 00:00:00

  • Experimental study on the influence of the central electrode in Farmer-type ionization chambers.

    abstract::In the IAEA TRS-381 protocol, k(cel) and p(cel) account for the central electrode perturbation during the air kerma chamber calibration and the in-phantom measurements. The values of these correction factors are based mainly on Monte Carlo simulations. In the present work experimental data on k(cel) and p(cel) for the...

    journal_title:Physics in medicine and biology

    pub_type: 杂志文章

    doi:10.1088/0031-9155/44/5/315

    authors: Palm A,Mattsson O

    更新日期:1999-05-01 00:00:00