A novel method for continuous blood pressure estimation based on a single-channel photoplethysmogram signal.

Abstract:

OBJECTIVE:Currently, continuous blood pressure (BP) measurements are mostly based on multi-sensor combinations and datasets with limited BP ranges. Besides, most BP-related features derive from the photoplethysmogram (PPG) signal. The mechanism of PPG signal formation is not considered. We aimed to design a noninvasive and continuous method for estimation of BP using a single PPG sensor, which takes the mechanism of PPG signal formation into account. APPROACH:We prepared a dataset containing PPG signals for 294 patients from three public databases for constructing the BP estimation model. The features used in the model consisted of two types: novel features based on a multi-Gaussian model and existing features. The multi-Gaussian model fitted the different components (i.e. the main wave, the dicrotic wave and the tidal wave) of the PPG signal. Ensemble machine learning algorithms were applied to estimate systolic blood pressure (SBP) and diastolic blood pressure (DBP). When partitioning the dataset, there was an overlap between the training set and the testing set. MAIN RESULTS:Datasets with a wide-range of SBP and DBP values (SBP ranging from 74 to 229 mmHg and DBP ranging from 26 to 141 mmHg) were used to evaluate our method. The mean and standard deviation of error for SBP and DBP estimations were -0.21 ± 5.21 mmHg and -0.19 ± 3.37 mmHg, respectively. The model performance fully met the Association for the Advancement of Medical Instrumentation standard and was grade 'A' on the British Hypertension Society standard. SIGNIFICANCE:The multi-Gaussian model could be used to estimate BP, and our method was able to track a wide range of BP accurately. In addition our method is based on a single PPG sensor, making it very convenient.

journal_name

Physiol Meas

authors

Hu Q,Deng X,Wang A,Yang C

doi

10.1088/1361-6579/abc8dd

subject

Has Abstract

pub_date

2021-01-01 00:00:00

pages

125009

issue

12

eissn

0967-3334

issn

1361-6579

journal_volume

41

pub_type

杂志文章
  • In vivo measurement of human skin absorption of topically applied substances by a photoacoustic technique.

    abstract::A photoacoustic technique is used for studying topically applied substance absorption in human skin. The proposed method utilizes a double-chamber PA cell. The absorption determination was obtained through the measurement of the thermal effusivity of the binary system substance-skin. The theoretical model assumes that...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/23/3/304

    authors: Gutiérrez-Juárez G,Vargas-Luna M,Córdova T,Varela JB,Bernal-Alvarado JJ,Sosa M

    更新日期:2002-08-01 00:00:00

  • Life-threatening false alarm rejection in ICU: using the rule-based and multi-channel information fusion method.

    abstract::False alarm (FA) rates as high as 86% have been reported in intensive care unit monitors. High FA rates decrease quality of care by slowing staff response times while increasing patient burdens and stresses. In this study, we proposed a rule-based and multi-channel information fusion method for accurately classifying ...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/37/8/1298

    authors: Liu C,Zhao L,Tang H,Li Q,Wei S,Li J

    更新日期:2016-08-01 00:00:00

  • Non-invasive measurements of respiratory system mechanical properties by the forced oscillation technique in spontaneously breathing, mixed-breed, normal term lambs from birth to five months of age.

    abstract:OBJECTIVE:To provide a non-invasive approach to monitoring lung function in spontaneously breathing lambs, from birth to five months of life, by the forced oscillation technique (FOT). This report describes the experimental set-up, data processing, and identification of normal predicted values of resistance (Rrs) and r...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/ab3493

    authors: Veneroni C,Dahl MJ,Lavizzari A,Dawson E,Rebentisch A,Studstill S,Bowen S,Albertine KH,Dellaca RL

    更新日期:2019-11-04 00:00:00

  • Tracking time-varying cerebral autoregulation in response to changes in respiratory PaCO2.

    abstract::Cerebral autoregulation has been studied by linear filter systems, with arterial blood pressure (ABP) as the input and cerebral blood flow velocity (CBFV-from transcranial Doppler Ultrasound) as the output. The current work extends this by using adaptive filters to investigate the dynamics of time-varying cerebral aut...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/31/10/001

    authors: Liu J,Simpson MD,Yan J,Allen R

    更新日期:2010-10-01 00:00:00

  • Home monitoring of within-breath respiratory mechanics by a simple and automatic forced oscillation technique device.

    abstract::Spirometry is the gold standard to determine the presence of airflow obstruction, but it requires volitional participation and needs qualified supervision. The forced oscillation technique (FOT) measures respiratory input impedance (Zrs) during spontaneous breathing and it could be useful for unsupervised monitoring o...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/31/4/N01

    authors: Dellacà RL,Gobbi A,Pastena M,Pedotti A,Celli B

    更新日期:2010-04-01 00:00:00

  • Masticatory muscle activity during deliberately performed oral tasks.

    abstract::The aim of this study was to investigate masticatory muscle activity during deliberately performed functional and non-functional oral tasks. Electromyographic (EMG) surface activity was recorded unilaterally from the masseter, anterior temporalis and suprahyoid muscles in 11 subjects (5 men, 6 women; age = 34.6 +/- 10...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/29/12/004

    authors: Farella M,Palla S,Erni S,Michelotti A,Gallo LM

    更新日期:2008-12-01 00:00:00

  • In vivo quantification of accumulating abdominal fluid using an electrical impedance tomography hemiarray.

    abstract::A new method to image and quantify intra-abdominal haemorrhage using electrical impedance tomography (EIT) was tested in vivo. Supine peritoneal dialysis patients were monitored using an 8-electrode hemiarray placed on the anterior abdomen. EIT measurements were recorded using the EPack II data acquisition system befo...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/32/2/001

    authors: Tucker AS,Ross EA,Paugh-Miller J,Sadleir RJ

    更新日期:2011-02-01 00:00:00

  • Chasing the evidence: the influence of data segmentation on estimates of dynamic cerebral autoregulation.

    abstract:OBJECTIVE:Transfer function analysis (TFA) of dynamic cerebral autoregulation (dCA) requires smoothing of spectral estimates using segmentation of the data (SD). Systematic studies are required to elucidate the potential influence of SD on dCA parameters. APPROACH:Healthy subjects (HS, n = 237) and acute ischaemic str...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/ab7ddf

    authors: Panerai RB,Intharakham K,Haunton V,Minhas JS,Llwyd O,Lam M,Salinet ASM,Nogueira RC,Katsogridakis E,Maggio P,Robinson TG

    更新日期:2020-04-17 00:00:00

  • Evaluation of an algorithm based on single-condition decision rules for binary classification of 12-lead ambulatory ECG recording quality.

    abstract::A new algorithm for classifying ECG recording quality based on the detection of commonly observed ECG contaminants which often render the ECG unusable for diagnostic purposes was evaluated. Contaminants (baseline drift, flat line, QRS-artefact, spurious spikes, amplitude stepwise changes, noise) were detected on indiv...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/33/9/1435

    authors: Di Marco LY,Duan W,Bojarnejad M,Zheng D,King S,Murray A,Langley P

    更新日期:2012-09-01 00:00:00

  • The assessment of acute load and contractility changes by left ventricular pressure measurements.

    abstract::The aim of this study was to establish whether analysis of the left ventricular pressure waveform provides indicative information about cardiac load and contractility and to develop an algorithm for computer-based assessment of changes in these variables. In eight healthy standard breed anaesthetized open-chest pigs, ...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/27/12/003

    authors: Söderqvist E,Hultman J,Källner G,Nowak J,Brodin LA

    更新日期:2006-12-01 00:00:00

  • Time-related heart autonomic balance characteristics in healthy subjects.

    abstract::Changes of the heart autonomic balance between morning (8-9 h) and afternoon (14-15 h) measurements were studied in 22 healthy subjects. The selection of these two daytime periods was substantiated by the established higher risk of cardiovascular incidents in the morning and the relative balance of the vegetative nerv...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/24/3/309

    authors: Matveev M,Prokopova R,Nachev Ch

    更新日期:2003-08-01 00:00:00

  • Wavelet-based real-time calculation of multiple physiological parameters on an embedded platform.

    abstract:OBJECTIVE:This paper aims to present how physiological signals can be processed based on wavelet decomposition to calculate multiple physiological parameters in real-time on an embedded platform. APPROACH:An ECG and PPG are decomposed to the appropriate scale based on a quadratic spline wavelet base in order to obtain...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/ab6f52

    authors: Tang M,Xia P,He Z,Zhao Z,Chen X,Yang T,Zhang Z,Zhan Q,Li X,Fang Z

    更新日期:2020-03-06 00:00:00

  • Adipose tissue thickness does not affect the electromechanical delay.

    abstract::During voluntary contractions in humans, the subcutaneous tissues between surface electrodes and active motor units have been shown to attenuate surface electromyographic (EMG) signal amplitude. The purpose of this investigation was to examine the relationship between adipose tissue thickness and the electromechnical ...

    journal_title:Physiological measurement

    pub_type: 临床试验,杂志文章

    doi:10.1088/0967-3334/37/3/418

    authors: Stock MS,Thompson BJ

    更新日期:2016-03-01 00:00:00

  • Cardiorespiratory coupling in young healthy subjects.

    abstract:OBJECTIVE:To quantify the presence of cardiorespiratory interaction in a group of 41 healthy subjects performing a subset of the Ewing test battery. APPROACH:We measure the empirical distribution of the cardiorespiratory coupling time (RI), defined as the time from inspiration onset to R peaks in the ECG. The study pr...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/aa9693

    authors: Sobiech T,Buchner T,Krzesiński P,Gielerak G

    更新日期:2017-11-30 00:00:00

  • Fetal beat detection in abdominal ECG recordings: global and time adaptive approaches.

    abstract::We present a method for location of fetal QRS in maternal abdominal ECG recordings. This method's initial, global approach was proposed in the context of the 2013 PhysioNet/Computing in Cardiology Challenge where it was tested on the 447 four channel one-minute recordings.The first step is filtering to eliminate basel...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/35/8/1699

    authors: Rodrigues R

    更新日期:2014-08-01 00:00:00

  • Age-related changes in pulse risetime measured by multi-site photoplethysmography.

    abstract:OBJECTIVE:It is accepted that changes in the peripheral pulse waveform characteristics occur with ageing. Pulse risetime is one important feature which has clinical value. However, it is unclear how it varies across the full age spectrum from child to senior and for different peripheral measurement sites. The objective...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/ab9b67

    authors: Allen J,O'Sullivan J,Stansby G,Murray A

    更新日期:2020-08-11 00:00:00

  • Phase shift between respiratory oscillations in cerebral blood flow velocity and arterial blood pressure.

    abstract::We aim to investigate whether phase shift between respiratory oscillations in cerebral blood flow velocity (CBFV) and arterial blood pressure (ABP) is associated with changes in cerebral autoregulation (CA) or reflects the mechano-elastic properties of the cerebrovascular bed. The relationships between respiratory pha...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/38/2/310

    authors: Uryga A,Placek MM,Wachel P,Szczepański T,Czosnyka M,Kasprowicz M

    更新日期:2017-02-01 00:00:00

  • Hidden Markov model-based activity recognition for toddlers.

    abstract:OBJECTIVE:Physical activity has been shown to impact future health outcomes in adults, but little is known about the long-term impact of physical activity in toddlers. Accurately measuring the specific types and amounts of physical activity in toddlers will help us to understand, predict, and better affect their future...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/ab6ebb

    authors: Albert MV,Sugianto A,Nickele K,Zavos P,Sindu P,Ali M,Kwon S

    更新日期:2020-03-06 00:00:00

  • Reference intervals for the evoked tympanic membrane displacement measurement: a non-invasive measure of intracranial pressure.

    abstract:OBJECTIVE:Evoked tympanic membrane displacement (TMD) is a non-invasive technique for assessing intracranial pressure (ICP). The aim of this study was to define reference intervals (RIs) in the healthy population. APPROACH:Measurements were made in 154 healthy adults. Results were quantified by V m, which is the most ...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/aaa1d3

    authors: Campbell-Bell CM,Birch AA,Vignali D,Bulters D,Marchbanks RJ

    更新日期:2018-01-31 00:00:00

  • Dynamic data monitoring improves predictive analytics for failed extubation in the ICU.

    abstract:OBJECTIVE:Predictive analytics monitoring that informs clinicians of the risk for failed extubation would help minimize both the duration of mechanical ventilation and the risk of emergency re-intubation in ICU patients. We hypothesized that dynamic monitoring of cardiorespiratory data, vital signs, and lab test result...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/aace95

    authors: Keim-Malpass J,Enfield KB,Calland JF,Lake DE,Clark MT

    更新日期:2018-07-16 00:00:00

  • In vitro measurements of respiratory mechanics during HFPV using a mechanical lung model.

    abstract::High-frequency percussive ventilation (HFPV) may be defined as flow-regulated time-cycled ventilation that creates controlled pressure and delivers a series of high-frequency subtidal volumes in combination with low-frequency breathing cycles. In recent years, the usefulness of HFPV has been clinically assessed as an ...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/32/6/002

    authors: Riscica F,Lucangelo U,Ferluga M,Accardo A

    更新日期:2011-06-01 00:00:00

  • Assessment of non-invasive time and frequency atrial fibrillation organization markers with unipolar atrial electrograms.

    abstract::The standard electrocardiogram (ECG) is the most common non-invasive way to study atrial fibrillation (AF). In this respect, previous works have shown that the surface lead V(1) reflects mainly the dominant atrial frequency (DAF) of the right atrium (RA), which has been widely used to study AF. In a similar way, AF or...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/32/1/007

    authors: Alcaraz R,Hornero F,Rieta JJ

    更新日期:2011-01-01 00:00:00

  • Validity and reliability of a timed 5 km cycle ergometer ride to predict maximum oxygen uptake.

    abstract::The purpose of this study was to determine the validity and reliability of a timed 5 km cycle ergometer ride to predict maximum oxygen uptake (VO2max). Fifty-two healthy volunteers had their VO2max determined during a maximal treadmill run. They also performed a timed 5 km cycle ergometer ride. The resistance setting ...

    journal_title:Physiological measurement

    pub_type: 临床试验,杂志文章,随机对照试验

    doi:10.1088/0967-3334/17/4/009

    authors: Buono MJ,Borin TL,Sjoholm NT,Hodgdon JA

    更新日期:1996-11-01 00:00:00

  • Within-breath measurement of respiratory impedance.

    abstract::The measurement of the complex respiratory impedance at a high temporal resolution is described. The measurement of impedance is correlated with the breathing cycle using electrical transthoracic impedance. This allows the determination of the complex impedance at any fixed point in the respiratory cycle, which result...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/14/3/019

    authors: van der Putten WJ,MacLeod D,Prichard JS

    更新日期:1993-08-01 00:00:00

  • Monitoring of heart rate and inter-beat intervals with wrist plethysmography in patients with atrial fibrillation.

    abstract:OBJECTIVE:Atrial fibrillation (AF) causes marked risk for patients, while silent fibrillation may remain unnoticed if not suspected and screened. Development of comfortable yet accurate beat-to-beat heart rate (HR) monitoring with good AF detection sensitivity would facilitate screening and improve treatment. The purpo...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/1361-6579/aac9a9

    authors: Harju J,Tarniceriu A,Parak J,Vehkaoja A,Yli-Hankala A,Korhonen I

    更新日期:2018-06-27 00:00:00

  • Localization of sources of brain alpha/theta/delta activity and the influence of the mode of spontaneous mentation.

    abstract::A method is described that accounts for multichannel brain field data (EEG) epochs, after transformation into the frequency domain, by a single oscillating dipole source in terms of phase angles. The method produces a potential distribution for each frequency point ('FFT dipole approximation'). These maps can be subje...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/14/4a/004

    authors: Michel CM,Henggeler B,Brandeis D,Lehmann D

    更新日期:1993-11-01 00:00:00

  • Real-time arrhythmia detection with supplementary ECG quality and pulse wave monitoring for the reduction of false alarms in ICUs.

    abstract::False intensive care unit (ICU) alarms induce stress in both patients and clinical staff and decrease the quality of care, thus significantly increasing both the hospital recovery time and rehospitalization rates. In the PhysioNet/CinC Challenge 2015 for reducing false arrhythmia alarms in ICU bedside monitor data, th...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/37/8/1273

    authors: Krasteva V,Jekova I,Leber R,Schmid R,Abächerli R

    更新日期:2016-08-01 00:00:00

  • Circuit modeling of the electrical impedance: II. Normal subjects and system reproducibility.

    abstract::Part I of this series showed that the five-element circuit model accurately mimics impedances measured using multi-frequency electrical impedance myography (MFEIM), focusing on changes brought on by disease. This paper addresses two requirements which must be met if the method is to qualify for clinical use. First, th...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/34/2/223

    authors: Shiffman CA,Rutkove SB

    更新日期:2013-02-01 00:00:00

  • Obstructive sleep apnea-hypopnea results in significant variations in cerebral hemodynamics detected by diffuse optical spectroscopies.

    abstract::The objective of this study was to adapt a novel near-infrared diffuse correlation spectroscopy (DCS) flow-oximeter for simultaneous and continuous monitoring of relative changes in cerebral blood flow (rCBF) and cerebral oxygenation (i.e. oxygenated/deoxygenated/total hemoglobin concentration: Δ[HbO2]/Δ[Hb]/ΔTHC) dur...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/35/10/2135

    authors: Hou Y,Shang Y,Cheng R,Zhao Y,Qin Y,Kryscio R,Rayapati A,Hayes D,Yu G

    更新日期:2014-10-01 00:00:00

  • Infrared thermography for indirect assessment of activation of brown adipose tissue in lean and obese male subjects.

    abstract::Brown adipose tissue (BAT) plays a key role in adaptive thermogenesis in mammals, and it has recently been considered as an attractive therapeutic target for tackling human obesity by increasing energy expenditure. Thermal imaging using infrared thermography (IRT) has emerged as a potential safe, rapid and inexpensive...

    journal_title:Physiological measurement

    pub_type: 杂志文章

    doi:10.1088/0967-3334/37/12/N118

    authors: El Hadi H,Frascati A,Granzotto M,Silvestrin V,Ferlini E,Vettor R,Rossato M

    更新日期:2016-12-01 00:00:00