Image noise and dose performance across a clinical population: Patient size adaptation as a metric of CT performance.

Abstract:

PURPOSE:Modern CT systems adjust X-ray flux accommodating for patient size to achieve certain image noise values. The effectiveness of this adaptation is an important aspect of CT performance and should ideally be characterized in the context of real patient cases. The objective of this study was to characterize CT performance with a new metric that includes image noise and radiation dose across a clinical patient population. MATERIALS AND METHODS:The study included 1526 examinations performed by three CT scanners (one GE Healthcare Discovery CT750HD, one GE Healthcare Lightspeed VCT, and one Siemens SOMATOM definition Flash) used for two routine clinical protocols (abdominopelvic with contrast and chest without contrast). An institutional monitoring system recorded all the data involved in the study. The dose-patient size and noise-patient size dependencies were linearized by considering a first-order approximation of analytical models that describe the relationship between ionization dose and patient size, as well as image noise and patient size. A 3D-fit was performed for each protocol and each scanner with a planar function, and the root mean square error (RMSE) values were estimated as a metric of CT adaptability across the patient population. RESULTS:The data show different scanner dependencies in terms of adaptability: the RMSE values for the three scanners are between 0.0385 HU1/2 and 0.0215 HU1/2 . CONCLUSION:A theoretical relationship between image noise, CTDIvol , and patient size was determined based on real patient data. This relationship may be interpreted as a new metric related to the scanners' adaptability concerning image quality and radiation dose across a patient population. This method could be implemented to investigate the adaptability related to other image quality indexes and radiation dose in a clinical population.

journal_name

Med Phys

journal_title

Medical physics

authors

Ria F,Wilson JM,Zhang Y,Samei E

doi

10.1002/mp.12172

subject

Has Abstract

pub_date

2017-06-01 00:00:00

pages

2141-2147

issue

6

eissn

0094-2405

issn

2473-4209

journal_volume

44

pub_type

杂志文章
  • Clinical apparatus for the reduction of dose area product for patients undergoing x-ray catheterization.

    abstract:PURPOSE:The authors describe a design for prepatient region of interest attenuators (ROIAs) to reduce dose area product (DAP) for clinical use. The authors describe a model to predict DAP values from x-ray technique parameters recorded during a clinical procedure for image sequences obtained in the presence or absence ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4903898

    authors: Robert N,Watt KN,Rochette S,Desponds L,Vaillant R,Rowlands JA

    更新日期:2015-01-01 00:00:00

  • Verification of IMRT dose distributions using a water beam imaging system.

    abstract::A water beam imaging system (WBIS) has been developed and used to verify dose distributions for intensity modulated radiotherapy using dynamic multileaf collimator. This system consisted of a water container, a scintillator screen, a charge-coupled device camera, and a portable personal computer. The scintillation ima...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1413519

    authors: Li JS,Boyer AL,Ma CM

    更新日期:2001-12-01 00:00:00

  • Direct aperture optimization: a turnkey solution for step-and-shoot IMRT.

    abstract::IMRT treatment plans for step-and-shoot delivery have traditionally been produced through the optimization of intensity distributions (or maps) for each beam angle. The optimization step is followed by the application of a leaf-sequencing algorithm that translates each intensity map into a set of deliverable aperture ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1477415

    authors: Shepard DM,Earl MA,Li XA,Naqvi S,Yu C

    更新日期:2002-06-01 00:00:00

  • Accuracy, reproducibility, and uncertainty analysis of thyroid-probe-based activity measurements for determination of dose calibrator settings.

    abstract:PURPOSE:In the nuclear medicine department, the activity of radiopharmaceuticals is measured using dose calibrators (DCs) prior to patient injection. The DC consists of an ionization chamber that measures current generated by ionizing radiation (emitted from the radiotracer). In order to obtain an activity reading, the...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4966027

    authors: Esquinas PL,Tanguay J,Gonzalez M,Vuckovic M,Rodríguez-Rodríguez C,Häfeli UO,Celler A

    更新日期:2016-12-01 00:00:00

  • Tumor volume measurement errors of RECIST studied with ellipsoids.

    abstract:PURPOSE:The authors investigate the extent to which Response Evaluation Criteria in Solid Tumors (RECIST) can predict tumor volumes in ideal geometric settings and using clinical data. METHODS:The authors consider a hierarchy of models including uniaxial ellipsoids, general ellipsoids, and composites of ellipsoids, us...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3577602

    authors: Levine ZH,Galloway BR,Peskin AP,Heussel CP,Chen JJ

    更新日期:2011-05-01 00:00:00

  • Dual-storage phosphor proton therapy dosimetry: simultaneous quantification of dose and linear energy transfer.

    abstract:PURPOSE:To investigate the feasibility of using the high Zeff storage phosphor material BaFBrI:Eu2+ in conjunction with the low Zeff storage phosphor material KCl:Eu2+ for simultaneous proton dose and linear energy transfer (LET) measurements by 1) measuring the fundamental optical and dosimetric properties of BaFBrI:E...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14748

    authors: Setianegara J,Mazur TR,Yang D,Li HH

    更新日期:2021-02-01 00:00:00

  • A methodology to determine margins by EPID measurements of patient setup variation and motion as applied to immobilization devices.

    abstract::Assessment of clinic and site specific margins are essential for the effective use of three-dimensional and intensity modulated radiation therapy. An electronic portal imaging device (EPID) based methodology is introduced which allows individual and population based CTV-to-PTV margins to be determined and compared wit...

    journal_title:Medical physics

    pub_type: 临床试验,杂志文章

    doi:10.1118/1.1800712

    authors: Prisciandaro JI,Frechette CM,Herman MG,Brown PD,Garces YI,Foote RL

    更新日期:2004-11-01 00:00:00

  • 3D MTF estimation using sphere phantoms for cone-beam computed tomography systems.

    abstract:PURPOSE:For cone-beam computed tomography (CBCT) systems, we propose a sphere phantom based method to estimate the full three-dimensional (3D) modulation transfer function (MTF). METHODS:The FDK reconstruction of CBCT system in a local region was modeled by a triple convolution operator. Afterward, we calculated the d...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14147

    authors: Lee C,Song HD,Baek J

    更新日期:2020-07-01 00:00:00

  • Technical Note: Experimental verification of magnetic field-induced beam deflection and Bragg peak displacement for MR-integrated proton therapy.

    abstract:PURPOSE:Given its sensitivity to anatomical variations, proton therapy is expected to benefit greatly from integration with magnetic resonance imaging for online anatomy monitoring during irradiation. Such an integration raises several challenges, as both systems mutually interact. The proton beam will experience quasi...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12961

    authors: Schellhammer SM,Gantz S,Lühr A,Oborn BM,Bussmann M,Hoffmann AL

    更新日期:2018-07-01 00:00:00

  • Branch order regression for modeling brain vasculature.

    abstract:PURPOSE:Many biological objects, including neuronal dendrites, blood vasculature, airways, phylogenetic trees, produce tree structured data. Current methods of analysis either ignore the complex structure of trees or use distance-based methods which limit the scope of multivariate modeling. METHODS:We propose a branch...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12751

    authors: Roy Choudhury K,Skwerer S

    更新日期:2018-03-01 00:00:00

  • A topographic leaf-sequencing algorithm for delivering intensity modulated radiation therapy.

    abstract::Topographic treatment is a radiation therapy delivery technique for fixed-gantry (nonrotational) treatments on a helical tomotherapy system. The intensity-modulated fields are created by moving the treatment couch relative to a fan-beam positioned at fixed gantry angles. The delivered dose distribution is controlled b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2216876

    authors: Desai D,Ramsey CR,Breinig M,Mahan SL

    更新日期:2006-08-01 00:00:00

  • Sheet beam x-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticles.

    abstract:PURPOSE:X-ray fluorescence computed tomography (XFCT) experiments have typically used pencil beams for data acquisition, which yielded good quality images of gold nanoparticles (AuNP) but prolonged the imaging time. Here we propose three novel collimator geometries for use with faster sheet beam XFCT data acquisition. ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12893

    authors: Dunning CAS,Bazalova-Carter M

    更新日期:2018-06-01 00:00:00

  • The effect of pitch in multislice spiral/helical CT.

    abstract::The purpose of this study is to understand the effect of pitch on raw data interpolation in multislice spiral/helical computed tomography (CT) and provide guidelines for scanner design and protocol optimization. Multislice spiral CT is mainly characterized by the three parameters: the number of detector arrays, the de...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598804

    authors: Wang G,Vannier MW

    更新日期:1999-12-01 00:00:00

  • Compensator quality control with an amorphous silicon EPID.

    abstract::The calibration and quality control of compensators is conventionally performed with an ion chamber in a water-equivalent phantom. In our center, the compensator factor and four off-axis fluence ratios are measured to verify the central axis beam modulation and orientation of the compensator. Here we report the invest...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1584040

    authors: Menon GV,Sloboda RS

    更新日期:2003-07-01 00:00:00

  • Eye tracking and gating system for proton therapy of orbital tumors.

    abstract:PURPOSE:A new motion-based gated proton therapy for the treatment of orbital tumors using real-time eye-tracking system was designed and evaluated. METHODS:We developed our system by image-pattern matching, using a normalized cross-correlation technique with LabVIEW 8.6 and Vision Assistant 8.6 (National Instruments, ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4729708

    authors: Shin D,Yoo SH,Moon SH,Yoon M,Lee SB,Park SY

    更新日期:2012-07-01 00:00:00

  • Fluid-filled dynamic bowtie filter: Description and comparison with other modulators.

    abstract:PURPOSE:A dynamic bowtie filter can modulate flux along both fan and view angles for reduced patient dose, scatter, and required photon flux, which is especially important for photon counting detectors (PCDs). Among the proposed dynamic bowtie designs, the piecewise-linear attenuator (Hsieh and Pelc, Med Phys. 2013;40:...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13272

    authors: Shunhavanich P,Hsieh SS,Pelc NJ

    更新日期:2019-01-01 00:00:00

  • SU-E-J-194: The Impact of Different Imaging Measures on Treatment Response Assessment of Bone Metastases.

    abstract:PURPOSE:Several treatment response metrics, such as RECIST and PERCIST, have been established for assessing individual solid tumors. However, metastatic prostate cancer poses a unique challenge to these metrics because bone lesions are often numerous and non-measureable. This study investigated the impact of using diff...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735035

    authors: Morrison C,Liu G,Jeraj R

    更新日期:2012-06-01 00:00:00

  • Assessment of a new multileaf collimator concept using GEANT4 Monte Carlo simulations.

    abstract::The aim of the work was to investigate in advance the dosimetric properties of a new multileaf collimator (MLC) concept with the help of Monte Carlo (MC) simulations prior to the production of a prototype. The geometrical design of the MLC was implemented in the MC code GEANT4. For the simulation of a 6 MV treatment b...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2181298

    authors: Tacke MB,Szymanowski H,Oelfke U,Schulze C,Nuss S,Wehrwein E,Leidenberger S

    更新日期:2006-04-01 00:00:00

  • Unwanted radiation produced by leakage neutrons from medical electron accelerators.

    abstract::Monte Carlo calculations have been made to determine the energy delivered to a phantom by neutrons escaping from the head of a Varian Clinac 35 medical accelerator at an SSD of 1 m. The energy was sorted into two regions: inside and outside a volume defined by a circular beam of area 100 and 600 cm2. For the two beam ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595067

    authors: Ing H,Shore RA

    更新日期:1982-01-01 00:00:00

  • Effects of protocol and obesity on dose conversion factors in adult body CT.

    abstract:PURPOSE:In computed tomography (CT), organ dose, effective dose, and risk index can be estimated from volume-weighted CT dose index (CTDI(vol)) or dose-length product (DLP) using conversion coefficients. Studies have investigated how these coefficients vary across scanner models, scan parameters, and patient size. Howe...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4754584

    authors: Li X,Samei E,Williams CH,Segars WP,Tward DJ,Miller MI,Ratnanather JT,Paulson EK,Frush DP

    更新日期:2012-11-01 00:00:00

  • Estimation of organ doses from kilovoltage cone-beam CT imaging used during radiotherapy patient position verification.

    abstract:PURPOSE:The purpose of this study was to develop a practical method for estimating organ doses from kilovoltage cone-beam CT (CBCT) that can be performed with readily available phantoms and dosimeters. The accuracy of organ dose estimates made using the ImPACT patient dose calculator was also evaluated. METHODS:A 100 ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3476459

    authors: Hyer DE,Hintenlang DE

    更新日期:2010-09-01 00:00:00

  • Dosimetry for 252Cf neutron emitting brachytherapy sources: protocol, measurements, and calculations.

    abstract::The mixed-field dosimetry for 252Cf Applicator Tube (AT) type medical sources available from Oak Ridge National Laboratory (ORNL) has been characterized using ionization chambers, a GM counter, and Monte Carlo methods. Unlike the AAPM Task Group No. 43 (TG-43), specification of dose to muscle instead of water is recom...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598646

    authors: Rivard MJ

    更新日期:1999-08-01 00:00:00

  • Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy.

    abstract::The objective of this work is to assess the suitability and performance of a new dosimeter system with a novel geometry for the quality assurance (QA) of volumetric modulated arc therapy (VMAT). The new dosimeter system consists of a hollow cylinder (15 and 25 cm inner and outer diameters) with 124 diodes embedded in ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3117563

    authors: Létourneau D,Publicover J,Kozelka J,Moseley DJ,Jaffray DA

    更新日期:2009-05-01 00:00:00

  • Radiation attenuation by lead and nonlead materials used in radiation shielding garments.

    abstract::The attenuating properties of several types of lead (Pb)-based and non-Pb radiation shielding materials were studied and a correlation was made of radiation attenuation, materials properties, calculated spectra and ambient dose equivalent. Utilizing the well-characterized x-ray and gamma ray beams at the National Rese...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2426404

    authors: McCaffrey JP,Shen H,Downton B,Mainegra-Hing E

    更新日期:2007-02-01 00:00:00

  • Validation for 2D/3D registration. I: A new gold standard data set.

    abstract:PURPOSE:In this article, the authors propose a new gold standard data set for the validation of two-dimensional/three-dimensional (2D/3D) and 3D/3D image registration algorithms. METHODS:A gold standard data set was produced using a fresh cadaver pig head with attached fiducial markers. The authors used several imagin...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3553402

    authors: Pawiro SA,Markelj P,Pernus F,Gendrin C,Figl M,Weber C,Kainberger F,Nöbauer-Huhmann I,Bergmeister H,Stock M,Georg D,Bergmann H,Birkfellner W

    更新日期:2011-03-01 00:00:00

  • SU-E-J-139: Feasibility of Using EPID for Real-Time Target Localization during Treatment.

    abstract:PURPOSE:This study aims to investigate the feasibility of using the images of the treatment fields acquired by an electronic portal imaging device (EPID) for real-time target localization. METHODS:Forty one patients treated with IMRT and RapidArc were recruited in this study including 37 prostate patients and 4 lung p...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734975

    authors: Xu Q,Lin M,Chen X,Tong X,Fan J,Dong Z,Chen L,Ma C

    更新日期:2012-06-01 00:00:00

  • Biomechanical modeling constrained surface-based image registration for prostate MR guided TRUS biopsy.

    abstract:PURPOSE:Adding magnetic resonance (MR)-derived information to standard transrectal ultrasound (TRUS) images for guiding prostate biopsy is of substantial clinical interest. A tumor visible on MR images can be projected on ultrasound (US) by using MR-US registration. A common approach is to use surface-based registratio...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4917481

    authors: van de Ven WJ,Hu Y,Barentsz JO,Karssemeijer N,Barratt D,Huisman HJ

    更新日期:2015-05-01 00:00:00

  • A novel 3D printing method for accurate anatomy replication in patient-specific phantoms.

    abstract:PURPOSE:A novel 3D printing method has been developed, which employs human CT images to construct patient specific phantoms by controlling the filament extrusion rate. METHODS:An in-house software was developed comprising pixel-by-pixel (PbP) reading of the Hounsfield Units (HU) values in the original patient DICOM im...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.13154

    authors: Okkalidis N

    更新日期:2018-10-01 00:00:00

  • The impact of small motion on the visualization of coronary vessels and lesions in cardiac CT: A simulation study.

    abstract:PURPOSE:Coronary x-ray computed tomography angiography (CCTA) is used to non-invasively assess coronary artery geometry and has, combined with computational modeling, demonstrated the potential to identify physiologically significant lesions. These measurements require robust and accurate coronary imaging and delineati...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12295

    authors: Contijoch F,Stayman JW,McVeigh ER

    更新日期:2017-07-01 00:00:00

  • Matching photon and electron fields with dynamic intensity modulation.

    abstract::A technique was developed to reduce the size and magnitude of the hot and cold spots in the abutting regions of photon and electron fields. The photon and electron fields were set up such that the photon field extended approximately 2 cm into the electron field in the abutting region. The region of the photon beam tha...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598753

    authors: Li JG,Xing L,Boyer AL,Hamilton RJ,Spelbring DR,Turian JV

    更新日期:1999-11-01 00:00:00