Comparison of handheld rebound tonometry with Goldmann applanation tonometry in children with glaucoma: a cohort study.

Abstract:

OBJECTIVE:To test agreement of two methods to measure intraocular pressure (IOP): rebound tonometry (RBT) and gold standard Goldmann applanation tonometry (GAT) in children with glaucoma. DESIGN:Observational prospective cohort study. SETTING:Tertiary paediatric glaucoma clinic at a single centre. PARTICIPANTS:102 individuals attending a paediatric glaucoma clinic, mean (SD) age 11.85 (3.17), of whom 53 were male. PRIMARY AND SECONDARY OUTCOME MEASURES:Intraocular pressure, central corneal thickness, child preference for measurement method. RESULTS:Limits of agreement for intraobserver and interobserver were, respectively, (-2.71, 2.98) mm Hg and (-5.75, 5.97) mm Hg. RBT frequently gave higher readings than GAT and the magnitude of disagreement depend on the level of IOP being assessed. Differences of 10 mm Hg were not uncommon. RBT was the preferred method for 70% of children. CONCLUSIONS:There is poor agreement between RBT and GAT in children with glaucoma. RBT frequently and significantly overestimates IOP. However, 'normal' RBT readings are likely to be accurate and may spare children an examination under anaesthesia (EUA). High RBT readings should prompt the practitioner to use another standard method of IOP measurement if possible, or consider the RBT measurement in the context of clinical findings before referring the child to a specialist clinic or considering EUA.

journal_name

BMJ Open

journal_title

BMJ open

authors

Dahlmann-Noor AH,Puertas R,Tabasa-Lim S,El-Karmouty A,Kadhim M,Wride NK,Lewis A,Grosvenor D,Rai P,Papadopoulos M,Brookes J,Bunce C,Khaw PT

doi

10.1136/bmjopen-2012-001788

subject

Has Abstract

pub_date

2013-04-02 00:00:00

issue

4

issn

2044-6055

pii

bmjopen-2012-001788

journal_volume

3

pub_type

杂志文章

相关文献

BMJ Open文献大全