Flow injection hydride generation and on-line W-coil trapping for electrothermal vaporization dielectric barrier discharge atomic emission spectrometric determination of trace cadmium.

Abstract:

:A fast, low-cost and sensitive method for the determination of trace cadmium was developed by using a miniaturized dielectric barrier discharge microplasma atomic emission spectrometer coupled with a tungsten coil (W-coil) for on-line hydride generation trapping-electrothermal vaporization. Total sample throughput can be greatly improved through the adoption of a horizontally fixed W-coil and the flow injection mode. In addition, the horizontally fixed W-coil and an inserted quartz capillary for on-line trapping contributed to stable and good signal even at a high gas flow rate when volatile cadmium species were trapped, and less sample-consuming and time-saving can be realized in this work. Compared to direct injection, the sensitivity and the LOD were improved by 29- and 38-fold, respectively. The proposed method provides a promising approach to develop a miniaturized instrumentation for highly sensitive detection of trace elements.

journal_name

Talanta

journal_title

Talanta

authors

Deng Y,Li K,Hou X,Jiang X

doi

10.1016/j.talanta.2021.122516

keywords:

["Atomic emission spectrometry","Cadmium","Dielectric barrier discharge","Electrothermal vaporization","Hydride generation"]

subject

Has Abstract

pub_date

2021-10-01 00:00:00

pages

122516

eissn

0039-9140

issn

1873-3573

pii

S0039-9140(21)00437-9

journal_volume

233

pub_type

杂志文章

相关文献