Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators.

Abstract:

:Flattening filter-free (FFF) beams are becoming the preferred beam type for stereotactic radiosurgery (SRS) and stereotactic ablative radiation therapy (SABR), as they enable an increase in dose rate and a decrease in treatment time. This work assesses the effects of the flattening filter on small field output factors for 6 MV beams generated by both Elekta and Varian linear accelerators, and determines differences between detector response in flattened (FF) and FFF beams. Relative output factors were measured with a range of detectors (diodes, ionization cham-bers, radiochromic film, and microDiamond) and referenced to the relative output factors measured with an air core fiber optic dosimeter (FOD), a scintillation dosimeter developed at Chris O'Brien Lifehouse, Sydney. Small field correction factors were generated for both FF and FFF beams. Diode measured detector response was compared with a recently published mathematical relation to predict diode response corrections in small fields. The effect of flattening filter removal on detector response was quantified using a ratio of relative detector responses in FFF and FF fields for the same field size. The removal of the flattening filter was found to have a small but measurable effect on ionization chamber response with maximum deviations of less than ± 0.9% across all field sizes measured. Solid-state detectors showed an increased dependence on the flattening filter of up to ± 1.6%. Measured diode response was within ± 1.1% of the published mathematical relation for all fields up to 30 mm, independent of linac type and presence or absence of a flattening filter. For 6 MV beams, detector correction factors between FFF and FF beams are interchangeable for a linac between FF and FFF modes, providing that an additional uncertainty of up to ± 1.6% is accepted.

journal_name

J Appl Clin Med Phys

authors

Tyler MK,Liu PZ,Lee C,McKenzie DR,Suchowerska N

doi

10.1120/jacmp.v17i3.6059

subject

Has Abstract

pub_date

2016-05-08 00:00:00

pages

223-235

issue

3

issn

1526-9914

journal_volume

17

pub_type

杂志文章
  • Optimization of intra-voxel incoherent motion measurement in diffusion-weighted imaging of breast cancer.

    abstract:PURPOSE:The purpose of this study was to optimize intra-voxel incoherent motion (IVIM) measurement in diffusion-weighted imaging (DWI) of breast cancer by separating perfusion and diffusion effects through the determination of an optimal threshold b-value, thus benign and cancerous breast tissues can be accurately diff...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12065

    authors: Chen W,Zhang J,Long D,Wang Z,Zhu JM

    更新日期:2017-05-01 00:00:00

  • An accurate algorithm to match imperfectly matched images for lung tumor detection without markers.

    abstract::In order to locate lung tumors on kV projection images without internal markers, digitally reconstructed radiographs (DRRs) are created and compared with projection images. However, lung tumors always move due to respiration and their locations change on projection images while they are static on DRRs. In addition, gl...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i3.5200

    authors: Rozario T,Bereg S,Yan Y,Chiu T,Liu H,Kearney V,Jiang L,Mao W

    更新日期:2015-05-08 00:00:00

  • Surface and build-up dose comparison between Elekta 6 MV flattening filter and flattening-filter-free beams using an advanced Markus ionization chamber and a solid water-equivalent phantom.

    abstract::Using a plane-parallel advanced Markus ionization chamber and a stack of water-equivalent solid phantom blocks, percentage surface and build-up doses of Elekta 6 MV flattening filter (FF) and flattening-filter-free (FFF) beams were measured as a function of the phantom depth for field sizes ranging from 2 × 2 to 10 × ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.13094

    authors: Imae T,Takenaka S,Watanabe Y,Aoki A,Matsuda K,Sasaki K,Saegusa S,Nawa K,Nakagawa K,Abe O

    更新日期:2020-12-01 00:00:00

  • Influence of patient's physiologic factors and immobilization choice with stereotactic body radiotherapy for upper lung tumors.

    abstract::The purpose of the present study was to compare the impact of pulmonary function, body habitus, and stereotactic body radiation therapy (SBRT) immobilization on setup and reproducibility for upper lung tumor. From 2008 through 2011, our institution's prospective SBRT database was searched for patients with upper lung ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i5.4931

    authors: Sio TT,Jensen AR,Miller RC,Fong de los Santos LE,Hallemeier CL,Foster NR,Park SS,Bauer HJ,Chang K,Garces YI,Olivier KR

    更新日期:2014-09-08 00:00:00

  • Linac-based stereotactic radiosurgery (SRS) in the treatment of refractory trigeminal neuralgia: Detailed description of SRS procedure and reported clinical outcomes.

    abstract:PURPOSE/OBJECTIVES:To present our linac-based SRS procedural technique for medically and/or surgically refractory trigeminal neuralgia (TN) treatment and simultaneously report our clinical outcomes. MATERIALS AND METHODS:Twenty-seven refractory TN patients who were treated with a single fraction of 80 Gy to TN. Treatm...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12057

    authors: Pokhrel D,Sood S,McClinton C,Saleh H,Badkul R,Jiang H,Stepp T,Camarata P,Wang F

    更新日期:2017-03-01 00:00:00

  • Comparative analysis for renal stereotactic body radiotherapy using Cyberknife, VMAT and proton therapy based treatment planning.

    abstract:PURPOSE:We conducted this dosimetric analysis to evaluate the feasibility of a multi-center stereotactic body radiation therapy (SBRT) trial for renal cell carcinoma (RCC) using different SBRT platforms. MATERIALS/METHODS:The computed tomography (CT) simulation images of 10 patients with unilateral RCC previously trea...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12308

    authors: Baydoun A,Vapiwala N,Ponsky LE,Awan M,Kassaee A,Sutton D,Podder TK,Zhang Y,Dobbins D,Muzic RF Jr,Traughber B,Machtay M,Ellis R

    更新日期:2018-03-14 00:00:00

  • A protocol to extend the longitudinal coverage of on-board cone-beam CT.

    abstract::The longitudinal coverage of a LINAC-mounted CBCT scan is limited to the corresponding dimensional limits of its flat panel detector, which is often shorter than the length of the treatment field. These limits become apparent when fields are designed to encompass wide regions, as when providing nodal coverage. Therefo...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v13i4.3796

    authors: Zheng D,Lu J,Jefferson A,Zhang C,Wu J,Sleeman W,Weiss E,Dogan N,Song S,Williamson J

    更新日期:2012-07-05 00:00:00

  • The performance of the progressive resolution optimizer (PRO) for RapidArc planning in targets with low-density media.

    abstract::A new version of progressive resolution optimizer (PRO) with an option of air cavity correction has been implemented for RapidArc volumetric-modulated arc therapy (RA). The purpose of this study was to compare the performance of this new PRO with the use of air cavity correction option (PRO10_air) against the one with...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i6.4382

    authors: Kan MW,Leung LH,Yu PK

    更新日期:2013-11-04 00:00:00

  • Evaluation and verification of the QFix EncompassTM couch insert for intracranial stereotactic radiosurgery.

    abstract::The QFix EncompassTM stereotactic radiosurgery (SRS) immobilization system consists of a thermoplastic mask that attaches to the couch insert to immobilize patients treated with intracranial SRS. This study evaluates the dosimetric impact and verifies a vendor provided treatment planning system (TPS) model in the Ecli...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12387

    authors: Snyder KC,Xhaferllari I,Huang Y,Siddiqui MS,Chetty IJ,Wen N

    更新日期:2018-07-01 00:00:00

  • Experience in implementing continuous arterial spin labeling on a commercial MR scanner.

    abstract::Continuous arterial spin labeling (CASL) is a technique for performing quantitative perfusion measurements without the need for exogenous contrast agent administration. This technique has seen limited use in the clinic due to problems of poor sensitivity and the potential for artifacts. In addition, CASL requires the ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 临床试验,杂志文章

    doi:10.1120/jacmp.v6i1.2068

    authors: Steger TR,Jackson EF

    更新日期:2005-01-01 00:00:00

  • Automatic calculation of patient size metrics in computed tomography: What level of computational accuracy do we need?

    abstract:OBJECTIVES:To compare the effectiveness of two different patient size metrics based on water equivalent diameter (Dw ), the mid-scan water equivalent diameter Dw_c , and the mean (average) water equivalent diameter in the imaged region, Dw_ave , for automatic detection of accidental changes in computed tomography (CT) ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12240

    authors: Sarmento S,Mendes B,Gouvêa M

    更新日期:2018-01-01 00:00:00

  • Interfraction positional variation in pancreatic tumors using daily breath-hold cone-beam computed tomography with visual feedback.

    abstract::We assessed interfraction positional variation in pancreatic tumors using daily breath-hold cone-beam computed tomography at end-exhalation (EE) with visual feedback (BH-CBCT). Eleven consecutive patients with pancreatic cancer who underwent BH intensity-modulated radiation therapy with visual feedback were enrolled. ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i2.5123

    authors: Nakamura M,Akimoto M,Ono T,Nakamura A,Yano S,Nakata M,Itasaka S,Mizowaki T,Shibuya K,Hiraoka M

    更新日期:2015-03-08 00:00:00

  • A dosimetric comparison of stereotactic body radiation therapy techniques for lung cancer: robotic versus conventional linac-based systems.

    abstract::The aim of this study is to compare the dosimetric characteristics of robotic and conventional linac-based SBRT techniques for lung cancer, and to provide planning guidance for each modality. Eight patients who received linac-based SBRT were retrospectively included in this study. A dose of 60 Gy given in three fracti...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v11i3.3223

    authors: Ding C,Chang CH,Haslam J,Timmerman R,Solberg T

    更新日期:2010-06-29 00:00:00

  • Evaluation of the MIM Symphony treatment planning system for low-dose-rate- prostate brachytherapy.

    abstract::MIM Symphony is a recently introduced low-dose-rate prostate brachytherapy treatment planning system (TPS). We evaluated the dosimetric and planning accuracy of this new TPS compared to the universally used VariSeed TPS. For dosimetric evaluation of the MIM Symphony version 5.4 TPS, we compared dose calculations from ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i5.5057

    authors: Dhanesar SK,Lim TY,Du W,Bruno TL,Frank SJ,Kudchadker RJ

    更新日期:2015-09-08 00:00:00

  • Dose deviations induced by respiratory motion for radiotherapy of lung tumors: Impact of CT reconstruction, plan complexity, and fraction size.

    abstract::A thorax phantom was used to assess radiotherapy dose deviations induced by respiratory motion of the target volume. Both intensity modulated and static, non-modulated treatment plans were planned on CT scans of the phantom. The plans were optimized using various CT reconstructions, to investigate whether they had an ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12847

    authors: Sande EPS,Acosta Roa AM,Hellebust TP

    更新日期:2020-04-01 00:00:00

  • The impact of isocentric shifts on delivery accuracy during the irradiation of small cerebral targets-Quantification and possible corrections.

    abstract:PURPOSE:To assess the impact of isocenter shifts due to linac gantry and table rotation during cranial stereotactic radiosurgery on D98 , target volume coverage (TVC), conformity (CI), and gradient index (GI). METHODS:Winston-Lutz (WL) checks were performed on two Elekta Synergy linacs. A stereotactic quality assuranc...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12854

    authors: Wack LJ,Exner F,Wegener S,Sauer OA

    更新日期:2020-05-01 00:00:00

  • Frameless stereotactic body radiation therapy for multiple lung metastases.

    abstract::Two patients with multiple lung metastases (≥ 5) were treated using frameless stereotactic body radiation therapy (SBRT) on an Elekta Axesse linear accelerator equipped with an interdigitation-capable multileaf collimator and four-dimensional cone-beam CT (4D CBCT). The technique and the early clinical outcomes were e...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i4.4737

    authors: Li Q,Mu J,Gu W,Chen Y,Ning Z,Jin J,Pei H

    更新日期:2014-07-08 00:00:00

  • Gamma Knife® icon CBCT offers improved localization workflow for frame-based treatment.

    abstract:OBJECT:The purpose of this study was to compare two methods of stereotactic localization in Gamma Knife treatment planning: cone beam computed tomography (CBCT) or fiducial. While the fiducial method is the traditional method of localization, CBCT is now available for use with the Gamma Knife Icon. This study seeks to ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12745

    authors: Duggar WN,Morris B,Fatemi A,Bonds J,He R,Kanakamedala M,Rey-Dios R,Vijayakumar S,Yang C

    更新日期:2019-11-01 00:00:00

  • Geometric accuracy of the MR imaging techniques in the presence of motion.

    abstract::Magnetic Resonance Imaging (MRI) is increasingly being used for improving tumor delineation and tumor tracking in the presence of respiratory motion. The purpose of this work is to design and build an MR compatible motion platform and to use it for evaluating the geometric accuracy of MR imaging techniques during resp...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12274

    authors: Torfeh T,Hammoud R,El Kaissi T,McGarry M,Aouadi S,Fayad H,Al-Hammadi N

    更新日期:2018-03-01 00:00:00

  • Dose calculation differences between Monte Carlo and pencil beam depend on the tumor locations and volumes for lung stereotactic body radiation therapy.

    abstract::Stereotactic body radiation therapy (SBRT) has been increasingly used as an efficacious treatment modality for early-stage non-small cell lung cancer. The accuracy of dose calculations is compromised due to the presence of inhomogeneity. For the purpose of a consistent prescription, radiation doses were calculated wit...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i2.4011

    authors: Zhuang T,Djemil T,Qi P,Magnelli A,Stephans K,Videtic G,Xia P

    更新日期:2013-03-04 00:00:00

  • Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets.

    abstract::Studies have shown that source dwells within Varian's HDR CT/MR compatible ring applicators can deviate from intended positions by several millimeters. Quantifying this offset is an important part of commissioning. The aims of this study were to: 1) determine the reproducibility of the offset, 2) study the interobserv...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v12i4.3447

    authors: McMahon R,Zhuang T,Steffey BA,Song H,Craciunescu OI

    更新日期:2011-11-15 00:00:00

  • Voxel-based supervised machine learning of peripheral zone prostate cancer using noncontrast multiparametric MRI.

    abstract:PURPOSE:The aim of this study was to develop and assess the performance of supervised machine learning technique to classify magnetic resonance imaging (MRI) voxels as cancerous or noncancerous using noncontrast multiparametric MRI (mp-MRI), comprised of T2-weighted imaging (T2WI), diffusion-weighted imaging (DWI), and...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12992

    authors: Gholizadeh N,Simpson J,Ramadan S,Denham J,Lau P,Siddique S,Dowling J,Welsh J,Chalup S,Greer PB

    更新日期:2020-10-01 00:00:00

  • New adaptive statistical iterative reconstruction ASiR-V: Assessment of noise performance in comparison to ASiR.

    abstract:PURPOSE:To assess the noise characteristics of the new adaptive statistical iterative reconstruction (ASiR-V) in comparison to ASiR. METHODS:A water phantom was acquired with common clinical scanning parameters, at five different levels of CTDIvol . Images were reconstructed with different kernels (STD, SOFT, and BONE...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12253

    authors: De Marco P,Origgi D

    更新日期:2018-03-01 00:00:00

  • Dosimetric comparison of preoperative single-fraction partial breast radiotherapy techniques: 3D CRT, noncoplanar IMRT, coplanar IMRT, and VMAT.

    abstract::The purpose of this study was to compare dosimetric parameters of treatment plans among four techniques for preoperative single-fraction partial breast radiotherapy in order to select an optimal treatment technique. The techniques evaluated were noncoplanar 3D conformal radiation therapy (3D CRT), noncoplanar intensit...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i1.5126

    authors: Yoo S,Blitzblau R,Yin FF,Horton JK

    更新日期:2015-01-08 00:00:00

  • Comparison of two algorithms for determining beam weights and wedge filters.

    abstract::This article compares two algorithms for determining beam weights and wedge filters for conformal treatment planning. One algorithm, which is based on dose-gradient analysis, provides analytic formulas for determining beam weights, wedge angles, and collimator angles (i.e., wedge orientations) so that the dose distrib...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v3i3.2562

    authors: Dai J,Zhu Y

    更新日期:2002-07-01 00:00:00

  • Two-dimensional solid-state array detectors: A technique for in vivo dose verification in a variable effective area.

    abstract:PURPOSE:We introduce a technique that employs a 2D detector in transmission mode (TM) to verify dose maps at a depth of dmax in Solid Water. TM measurements, when taken at a different surface-to-detector distance (SDD), allow for the area at dmax (in which the dose map is calculated) to be adjusted. METHODS:We conside...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12744

    authors: Utitsarn K,Biasi G,Stansook N,Alrowaili ZA,Petasecca M,Carolan M,Perevertaylo VL,Tomé WA,Kron T,Lerch MLF,Rosenfeld AB

    更新日期:2019-11-01 00:00:00

  • Measurement of gantry rotation time in modern ct.

    abstract::The purpose of this study was to develop and evaluate a noninvasive method to assess rotation time in modern commercial computed tomography (CT) systems. The rotation time was measured at a selected nominal rotation time (400 ms) utilizing two types of solid-state detectors: the RTI's CT Dose Profiler (CTDP) and Unfor...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i1.4517

    authors: Fukuda A,Lin PJ,Matsubara K,Miyati T

    更新日期:2014-01-06 00:00:00

  • The overshoot phenomenon in step-and-shoot IMRT delivery.

    abstract::The control loop in the Varian DMLC system (V4.8) requires approximately 65 msec to monitor and halt the irradiation of a segment, causing an "overshoot" effect: the segment ends on a fractional monitor unit larger than that planned. As a result, the actual MU delivered may differ from that planned. In general, for st...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v2i3.2607

    authors: Ezzell GA,Chungbin S

    更新日期:2001-07-01 00:00:00

  • Volumetric-modulated arc therapy planning using multicriteria optimization for localized prostate cancer.

    abstract::The purpose of this work is to evaluate the volumetric-modulated arc therapy (VMAT) multicriteria optimization (MCO) algorithm clinically available in the RayStation treatment planning system (TPS) and its ability to reduce treatment planning time while providing high dosimetric plan quality. Nine patients with locali...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i3.5410

    authors: Ghandour S,Matzinger O,Pachoud M

    更新日期:2015-05-08 00:00:00

  • Electrocardiogram-gated coronary CT angiography dose estimates using ImPACT.

    abstract::The primary study objective was to assess radiation doses using a modified form of the Imaging Performance Assessment of Computed Tomography (CT) scanner (ImPACT) patient dosimetry for cardiac applications on an Aquilion ONE ViSION Edition scanner, including the Ca score, target computed tomography angiography (CTA), ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i4.6218

    authors: Kobayashi M,Asada Y,Matsubara K,Suzuki S,Koshida K,Matsunaga Y,Haba T,Kawaguchi A,Toyama H,Kato R

    更新日期:2016-07-08 00:00:00