Dosimetric comparison of preoperative single-fraction partial breast radiotherapy techniques: 3D CRT, noncoplanar IMRT, coplanar IMRT, and VMAT.

Abstract:

:The purpose of this study was to compare dosimetric parameters of treatment plans among four techniques for preoperative single-fraction partial breast radiotherapy in order to select an optimal treatment technique. The techniques evaluated were noncoplanar 3D conformal radiation therapy (3D CRT), noncoplanar intensity-modulated radiation therapy (IMRTNC), coplanar IMRT (IMRTCO), and volumetric-modulated arc therapy (VMAT). The planning CT scans of 16 patients in the prone position were used in this study, with the single-fraction prescription doses of 15 Gy for the first eight patients and 18 Gy for the remaining eight patients. Six (6) MV photon beams were designed to avoid the heart and contralateral breast. Optimization for IMRT and VMAT was performed to reduce the dose to the skin and normal breast. All plans were normalized such that 100% of the prescribed dose covered greater than 95% of the clinical target volume (CTV) consisting of gross tumor volume (GTV) plus 1.5 cm margin. Mean homogeneity index (HI) was the lowest (1.05 ± 0.02) for 3D CRT and the highest (1.11 ± 0.04) for VMAT. Mean conformity index (CI) was the lowest (1.42 ± 0.32) for IMRTNC and the highest (1.60 ± 0.32) for VMAT. Mean of the maximum point dose to skin was the lowest (73.7 ± 11.5%) for IMRTNC and the highest (86.5 ± 6.68%) for 3D CRT. IMRTCO showed very similar HI, CI, and maximum skin dose to IMRTNC (differences <1%). The estimated mean treatment delivery time, excluding the time spent for patient positioning and imaging, was 7.0 ± 1.0, 8.3 ± 1.1, 9.7 ± 1.0, and 11.0 ± 1.5min for VMAT, IMRTCO, IMRTNC and 3D CRT, respectively. In comparison of all four techniques for preoperative single-fraction partial breast radiotherapy, we can conclude that noncoplanar or coplanar IMRT were optimal in this study as IMRT plans provided homogeneous and conformal target coverage, skin sparing, and relatively short treatment delivery time.

journal_name

J Appl Clin Med Phys

authors

Yoo S,Blitzblau R,Yin FF,Horton JK

doi

10.1120/jacmp.v16i1.5126

subject

Has Abstract

pub_date

2015-01-08 00:00:00

pages

5126

issue

1

issn

1526-9914

journal_volume

16

pub_type

杂志文章
  • Stereotactic IMRT for prostate cancer: setup accuracy of a new stereotactic body localization system.

    abstract::The purpose of this work is to prospectively assess the setup accuracy that can be achieved with a stereotactic body localizer (SBL) in immobilizing patients for stereotactic intensity-modulated radiotherapy (IMRT) for prostate cancer. By quantifying this important factor and target mobility in the SBL, we expect to p...

    journal_title:Journal of applied clinical medical physics

    pub_type: 临床试验,杂志文章

    doi:10.1120/jacmp.v5i2.1947

    authors: Wang L,Jacob R,Chen L,Ma C,Movsas B,Feigenberg S,Konski A

    更新日期:2004-04-01 00:00:00

  • Effectiveness of a novel gas-release endorectal balloon in the removal of rectal gas for prostate proton radiation therapy.

    abstract::Endorectal balloons (ERBs) are routinely used in prostate proton radiation therapy to immobilize the prostate and spare the rectal wall. Rectal gas can distend the rectum and displace the prostate even in the presence of ERBs. The purpose of this work was to quantify the effects an ERB with a passive gas release condu...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v13i5.3945

    authors: Wootton LS,Kudchadker RJ,Beddar AS,Lee AK

    更新日期:2012-09-06 00:00:00

  • Estimating and reducing dose received by cardiac devices for patients undergoing radiotherapy.

    abstract::The objectives of this project are to quantify the dose reduction effect provided by a lead shield for patients with cardiac implantable electronic devices (CIED) during a clinically realistic radiation treatment on phantom and to provide a simple model of dose estimation to predict dose received by CIED in a wide ran...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i6.5317

    authors: Bourgouin A,Varfalvy N,Archambault L

    更新日期:2015-11-08 00:00:00

  • Optimization of motion management parameters in a synchrotron-based spot scanning system.

    abstract:PURPOSE:To quantify the effects of combining layer-based repainting and respiratory gating as a strategy to mitigate the dosimetric degradation caused by the interplay effect between a moving target and dynamic spot-scanning proton delivery. METHODS:An analytic routine modeled three-dimensional dose distributions of p...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12702

    authors: Johnson JE,Herman MG,Kruse JJ

    更新日期:2019-09-01 00:00:00

  • Detection of electron beam energy variations using a computed radiography system.

    abstract::A method to evaluate the electron beam energy constancy by employing the computed radiography (CR) system has been developed. In this method, a right triangular plastic wedge is used to produce a curve of the CR storage phosphor plate signal versus the wedge thickness. The curve, which resembles the percentage depth i...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v10i4.2911

    authors: Cai YC,Ge Y,Bernard D,Turian J,Chu JCH

    更新日期:2009-10-15 00:00:00

  • Assessment of potential jaw-tracking advantage using control point sequences of VMAT planning.

    abstract::This study aims to evaluate the potential jaw-tracking advantage using control point sequences of volume volumetric modulated arc therapy (VMAT) planning. VMAT plans for patients with prostate and head and neck (H&N) cancers were converted into new static arc (SA) plans. The SA plan consisted of a series of static fie...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i2.4625

    authors: Kim JI,Park JM,Park SY,Choi CH,Wu HG,Ye SJ

    更新日期:2014-03-06 00:00:00

  • Evaluation of an electron Monte Carlo dose calculation algorithm for treatment planning.

    abstract::The purpose of this study is to evaluate the accuracy of the electron Monte Carlo (eMC) dose calculation algorithm included in a commercial treatment planning system and compare its performance against an electron pencil beam algorithm. Several tests were performed to explore the system's behavior in simple geometries...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i3.4636

    authors: Chamberland E,Beaulieu L,Lachance B

    更新日期:2015-05-08 00:00:00

  • AAPM Task Group 103 report on peer review in clinical radiation oncology physics.

    abstract::This report provides guidelines for a peer review process between two clinical radiation oncology physicists. While the Task Group's work was primarily focused on ensuring timely and productive independent reviews for physicists in solo practice, these guidelines may also be appropriate for physicists in a group setti...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章,实务指引

    doi:10.1120/jacmp.v6i4.2142

    authors: Halvorsen PH,Das IJ,Fraser M,Freedman DJ,Rice RE 3rd,Ibbott GS,Parsai EI,Robin TT Jr,Thomadsen BR,American Association of Physicists in Medicine.

    更新日期:2005-10-01 00:00:00

  • Interfraction positional variation in pancreatic tumors using daily breath-hold cone-beam computed tomography with visual feedback.

    abstract::We assessed interfraction positional variation in pancreatic tumors using daily breath-hold cone-beam computed tomography at end-exhalation (EE) with visual feedback (BH-CBCT). Eleven consecutive patients with pancreatic cancer who underwent BH intensity-modulated radiation therapy with visual feedback were enrolled. ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i2.5123

    authors: Nakamura M,Akimoto M,Ono T,Nakamura A,Yano S,Nakata M,Itasaka S,Mizowaki T,Shibuya K,Hiraoka M

    更新日期:2015-03-08 00:00:00

  • Determination of the photon beam attenuation by the Brainlab imaging couch: angular and field size dependence.

    abstract::Highly attenuating radiation treatment couches are no longer useful in the present era of radiotherapy utilizing IMRT and IGRT. Carbon fibers couch tops with its high tensile strength and low density present a useful alternative. The objective of the current study was to quantify the attenuation of megavoltage photons...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v10i3.2979

    authors: Njeh CF,Raines TW,Saunders MW

    更新日期:2009-06-02 00:00:00

  • Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx.

    abstract::Small field dosimetry is a challenging task. The difficulties of small field measurements, particularly stereotactic field size measurements, are highlighted by the large interinstitution variability that can be observed for circular cone collimator commissioning measurements. We believe the best way to improve the co...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i6.4493

    authors: Wiant DB,Terrell JA,Maurer JM,Yount CL,Sintay BJ

    更新日期:2013-11-04 00:00:00

  • Electron energy constancy verification using a double-wedge phantom.

    abstract::Routine constancy checks of electron energy are often time consuming because of the necessity to measure a dose at two depths. A technique is described that uses a double-wedge shaped phantom positioned on a Profiler diode array for measuring an electron energy constancy metric similar to R(50). The double-wedge elect...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v4i3.2516

    authors: Wells DM,Picco PJ,Ansbacher W

    更新日期:2003-07-01 00:00:00

  • Evaluation of the influence of tumor location and size on the difference of dose calculation between Ray Tracing algorithm and Fast Monte Carlo algorithm in stereotactic body radiotherapy of non-small cell lung cancer using CyberKnife.

    abstract::This study evaluated the extent of improvement in dose predication accuracy achieved by the Fast Monte Carlo algorithm (MC) compared to the Ray Tracing algorithm (RAT) in stereotactic body radiotherapy (SBRT) of non-small cell lung cancer (NSCLC), and how their differences were influenced by the tumor site and size. T...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i5.4280

    authors: Wu VW,Tam KW,Tong SM

    更新日期:2013-09-06 00:00:00

  • A computational tool for the efficient analysis of dose-volume histograms from radiation therapy treatment plans.

    abstract::A Histogram Analysis in Radiation Therapy (HART) program was primarily developed to increase the efficiency and accuracy of dose-volume histogram (DVH) analysis of large quantities of patient data in radiation therapy research. The program was written in MATLAB to analyze patient plans exported from the treatment plan...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:

    authors: Pyakuryal A,Myint WK,Gopalakrishnan M,Jang S,Logemann JA,Mittal BB

    更新日期:2010-01-28 00:00:00

  • Dose calculation for hypofractionated volumetric-modulated arc therapy: approximating continuous arc delivery and tongue-and-groove modeling.

    abstract::Hypofractionated treatments generally increase the complexity of a treatment plan due to the more stringent constraints of normal tissues and target coverage. As a result, treatment plans contain more modulated MLC motions that may require extra efforts for accurate dose calculation. This study explores methods to min...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i2.4989

    authors: Yang J,Tang G,Zhang P,Hunt M,Lim SB,LoSasso T,Mageras G

    更新日期:2016-03-08 00:00:00

  • Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification.

    abstract::Radiochromic film has become an important tool to verify dose distributions in highly conformal radiation therapy such as IMRT. Recently, a new generation of these films, EBT3, has become available. EBT3 has the same composition and thickness of the sensitive layer of the previous EBT2 films, but its symmetric layer c...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i2.4111

    authors: Casanova Borca V,Pasquino M,Russo G,Grosso P,Cante D,Sciacero P,Girelli G,La Porta MR,Tofani S

    更新日期:2013-03-04 00:00:00

  • Radiation therapy plan checks in a paperless clinic.

    abstract::Traditional quality assurance checks of a patient's radiation therapy plan involve printing out treatment parameters from the treatment planning system and the "record and verify" (R&V) system and visually checking the information for one-to-one correspondence. In a paperless environment, one can automate this process...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v10i1.2905

    authors: Siochi RA,Pennington EC,Waldron TJ,Bayouth JE

    更新日期:2009-01-27 00:00:00

  • Gemstone spectral imaging: determination of CT to ED conversion curves for radiotherapy treatment planning.

    abstract::The monochromatic images acquired by Gemstone spectral imaging (GSI) mode on the GE CT750 HD theoretically determines the computed tomography (CT) number more accurately than that of conventional scanner. Using the former, the CT number is calculated from (synthesized) monoenergetic X-ray data. We reasoned that the mo...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i5.4335

    authors: Yagi M,Ueguchi T,Koizumi M,Ogata T,Yamada S,Takahashi Y,Sumida I,Akino Y,Konishi K,Isohashi F,Tomiyama N,Yoshioka Y,Ogawa K

    更新日期:2013-09-06 00:00:00

  • Radiographic film dosimetry for IMRT fields in the nearsurface buildup region.

    abstract::Radiographic film dosimetry provides fast, convenient 2-D dose distributions, but is challenged by the dependence of film response on scatter conditions (i.e., energy dependence). Verification of delivered dose in the surface buildup region is important for intensity modulated radiation therapy (IMRT) when volumes of ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v9i4.2782

    authors: Roberson PL,Moran JM,Kulasekere R

    更新日期:2008-10-24 00:00:00

  • Feasibility of a GATE Monte Carlo platform in a clinical pretreatment QA system for VMAT treatment plans using TrueBeam with an HD120 multileaf collimator.

    abstract:PURPOSE:To evaluate the quality of patient-specific complicated treatment plans, including commercialized treatment planning systems (TPS) and commissioned beam data, we developed a process of quality assurance (QA) using a Monte Carlo (MC) platform. Specifically, we constructed an interface system that automatically c...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12718

    authors: Lee B,Jeong S,Chung K,Yoon M,Park HC,Han Y,Jung SH

    更新日期:2019-10-01 00:00:00

  • "Dose of the day" based on cone beam computed tomography and deformable image registration for lung cancer radiotherapy.

    abstract:PURPOSE:Adaptive radiotherapy (ART) has potential to reduce toxicity and facilitate safe dose escalation. Dose calculations with the planning CT deformed to cone beam CT (CBCT) have shown promise for estimating the "dose of the day". The purpose of this study is to investigate the "dose of the day" calculation accuracy...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12793

    authors: Yuan Z,Rong Y,Benedict SH,Daly ME,Qiu J,Yamamoto T

    更新日期:2020-01-01 00:00:00

  • On the selection of gantry and collimator angles for isocenter localization using Winston-Lutz tests.

    abstract::In Winston-Lutz (WL) tests, the isocenter of a linear accelerator (linac) is determined as the intersection of radiation central axes (CAX) from multiple gantry, collimator, and couch angles. It is well known that the CAX can wobble due to mechanical imperfections of the linac. Previous studies suggested that the wobb...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i1.5792

    authors: Du W,Johnson JL,Jiang W,Kudchadker RJ

    更新日期:2016-01-08 00:00:00

  • Peer support: A needs assessment for social support from trained peers in response to stress among medical physicists.

    abstract:PURPOSE:Previous studies suggest that within radiation oncology, medical physicists (MP) experience high workloads. Little is known about how MPs use social support (SS) in times of stress. METHODS:In collaboration with the Workgroup on Prevention of Medical Error, the American Association of Physicists in Medicine ad...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12675

    authors: Johnson J,Ford E,Yu J,Buckey C,Fogh S,Evans SB

    更新日期:2019-09-01 00:00:00

  • Frameless stereotactic body radiation therapy for multiple lung metastases.

    abstract::Two patients with multiple lung metastases (≥ 5) were treated using frameless stereotactic body radiation therapy (SBRT) on an Elekta Axesse linear accelerator equipped with an interdigitation-capable multileaf collimator and four-dimensional cone-beam CT (4D CBCT). The technique and the early clinical outcomes were e...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i4.4737

    authors: Li Q,Mu J,Gu W,Chen Y,Ning Z,Jin J,Pei H

    更新日期:2014-07-08 00:00:00

  • Applying three different methods of measuring CTDIfree air to the extended CTDI formalism for wide-beam scanners (IEC 60601-2-44): A comparative study.

    abstract:PURPOSE:The weighted CT dose index (CTDIw ) has been extended for a nominal total collimation width (nT) greater than 40 mm and relies on measurements of CTDIfreeair. The purpose of this work was to compare three methods of measuring CTDIfreeair and subsequent calculations of CTDIw to investigate their clinical appropr...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12363

    authors: Bujila R,Kull L,Danielsson M,Andersson J

    更新日期:2018-07-01 00:00:00

  • FDG PET/CT for rectal carcinoma radiotherapy treatment planning: comparison of functional volume delineation algorithms and clinical challenges.

    abstract::PET/CT imaging could improve delineation of rectal carcinoma gross tumor volume (GTV) and reduce interobserver variability. The objective of this work was to compare various functional volume delineation algorithms. We enrolled 31 consecutive patients with locally advanced rectal carcinoma. The FDG PET/CT and the high...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i5.4696

    authors: Withofs N,Bernard C,Van der Rest C,Martinive P,Hatt M,Jodogne S,Visvikis D,Lee JA,Coucke PA,Hustinx R

    更新日期:2014-09-08 00:00:00

  • Small field electron beam dosimetry using MOSFET detector.

    abstract::The dosimetry of very small electron fields can be challenging due to relative shifts in percent depth-dose curves, including the location of dmax, and lack of lateral electronic equilibrium in an ion chamber when placed in the beam. Conventionally a small parallel plate chamber or film is utilized to perform small fi...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v12i1.3267

    authors: Amin MN,Heaton R,Norrlinger B,Islam MK

    更新日期:2010-10-04 00:00:00

  • Rounded leaf end modeling in Pinnacle VMAT treatment planning for fixed jaw linacs.

    abstract::During volume-modulated arc therapies (VMAT), dosimetric errors are introduced by multiple open dynamic leaf gaps that are present in fixed diaphragm linear accelerators. The purpose of this work was to develop a methodology for adjusting the rounded leaf end modeling parameters to improve out-of-field dose agreement ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i6.6343

    authors: Young LA,Yang F,Cao N,Meyer J

    更新日期:2016-11-08 00:00:00

  • Dose variations in tumor volumes and organs at risk during IMRT for head-and-neck cancer.

    abstract::Many head-and-neck cancer (HNC) patients treated with radiotherapy suffer significant anatomical changes due to tumor shrinkage or weight loss. The purpose of this study was to assess dose changes over target volumes and organs at risk during intensity-modulated radiotherapy for HNC patients. Sixteen HNC IMRT patients...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v13i6.3723

    authors: Beltran M,Ramos M,Rovira JJ,Perez-Hoyos S,Sancho M,Puertas E,Benavente S,Ginjaume M,Giralt J

    更新日期:2012-11-08 00:00:00

  • Initial clinical experience with a radiation oncology dedicated open 1.0T MR-simulation.

    abstract::The purpose of this study was to describe our experience with 1.0T MR-SIM including characterization, quality assurance (QA) program, and features necessary for treatment planning. Staffing, safety, and patient screening procedures were developed. Utilization of an external laser positioning system (ELPS) and MR-compa...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i2.5201

    authors: Glide-Hurst CK,Wen N,Hearshen D,Kim J,Pantelic M,Zhao B,Mancell T,Levin K,Movsas B,Chetty IJ,Siddiqui MS

    更新日期:2015-03-08 00:00:00