The overshoot phenomenon in step-and-shoot IMRT delivery.

Abstract:

:The control loop in the Varian DMLC system (V4.8) requires approximately 65 msec to monitor and halt the irradiation of a segment, causing an "overshoot" effect: the segment ends on a fractional monitor unit larger than that planned. As a result, the actual MU delivered may differ from that planned. In general, for step-and-shoot treatments, the first segment receives more, the last receives less, and intermediate segments vary. The overshoot for each segment (DeltaMU) is small, approximately 0.6 MU at 600 MU/min. Our IMRT planning system (Corvus) produces plans often having more than 20% of the segments with less than 1 MU/segment. Such segments may be skipped if the DeltaMU exceeds the segments' planned MU. Furthermore, QA filming often requires reducing the total MU by a factor of 4-6, increasing the potential for dosimetric error. This study measured DeltaMU over a range of MU/min and MU/segment. At >5 MU/segment, the DeltaMU was stable, corresponding to a delay of 62 msec. DeltaMU became larger and more variable at <1 MU/segment. The behavior was modeled in a computer program that predicted the change in delivered MU/segment and total change in delivered MU to each beamlet. Beams were analyzed for patients receiving 5 field prostate or 9 field head and neck treatments. At 400 MU/min, 28% and 16%, respectively, of the planned segments were skipped. For QA filming, up to 75% of the segments were skipped. The cumulative error averaged <0.1 MU/beamlet, but individual beamlets had errors exceeding 200%. The effect is most significant for low dose regions. Recommendations are given for deciding when to treat or do QA studies with lower MU/min. In general, treatments are not significantly affected, but QA films taken at reduced MU may be improved if irradiated at lowered MU/min.

journal_name

J Appl Clin Med Phys

authors

Ezzell GA,Chungbin S

doi

10.1120/jacmp.v2i3.2607

keywords:

subject

Has Abstract

pub_date

2001-07-01 00:00:00

pages

138-48

issue

3

issn

1526-9914

journal_volume

2

pub_type

杂志文章
  • Generation and verification of QFix kVue Calypso-compatible couch top model for a dedicated stereotactic linear accelerator with FFF beams.

    abstract::This study details the generation, verification, and implementation of a treatment planning system (TPS) couch top model for patient support system used in conjunction with a dedicated stereotactic linear accelerator. Couch top model was created within the TPS using CT simulation images of the kVue Calpyso-compatible ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i4.5441

    authors: Gardner SJ,Gulam M,Song K,Li H,Huang Y,Zhao B,Qin Y,Snyder K,Kim J,Gordon J,Chetty IJ,Wen N

    更新日期:2015-07-08 00:00:00

  • Long term stability of a 50 kV X-ray unit for stereotactic irradiation.

    abstract::A low-energy (50 kV) X-ray tube used for the stereotactic irradiation of intracranial lesions has been in use since 1999. The unit is calibrated prior to every procedure and during periodic quality assurance (QA) tests. The unit uses an internal and an external scintillation detector to monitor dose as well as a conve...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v7i3.2238

    authors: Biggs PJ

    更新日期:2006-08-24 00:00:00

  • Beam-specific planning target volumes incorporating 4D CT for pencil beam scanning proton therapy of thoracic tumors.

    abstract::The purpose of this study is to determine whether organ sparing and target coverage can be simultaneously maintained for pencil beam scanning (PBS) proton therapy treatment of thoracic tumors in the presence of motion, stopping power uncertainties, and patient setup variations. Ten consecutive patients that were previ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i6.5678

    authors: Lin L,Kang M,Huang S,Mayer R,Thomas A,Solberg TD,McDonough JE,Simone CB 2nd

    更新日期:2015-11-08 00:00:00

  • An algorithm for automated modulation transfer function measurement using an edge of a PMMA phantom: Impact of field of view on spatial resolution of CT images.

    abstract:PURPOSE:The purpose of this study was to introduce a new algorithm for automated measurement of the modulation transfer function (MTF) using an edge of a readily available phantom and to evaluate the effect of reconstruction filter and field of view (FOV) on the spatial resolution in the CT images. METHODS:Our automat...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12476

    authors: Anam C,Fujibuchi T,Budi WS,Haryanto F,Dougherty G

    更新日期:2018-11-01 00:00:00

  • Feasibility of the new copper pipe method for evaluating half-value layer in computed tomography: A measurement and Monte Carlo simulation study.

    abstract::This study aimed to verify the accuracy of half-value layer (HVL) measured using the new copper pipe method with the CT ionization chamber while the X-ray tube is rotating and to compare it with the conventional nonrotating method and Monte Carlo simulation method based on the actual measurement and geometry of the ne...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12780

    authors: Okubo R,Matsubara K,Chusin T,Hibino T,Ito Y

    更新日期:2019-12-01 00:00:00

  • Automated phantom analysis for gamma cameras and SPECT: A methodology for use in a clinical setting.

    abstract:PURPOSE:We introduce an automated, quantitative image analysis package for gamma camera and single photon emission computed tomography quality control. Our focus was to produce consistent methods that are feasible in clinical settings and use standard phantoms. METHODS:Four gamma cameras were used to acquire planar im...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.13057

    authors: Tazegul TE,Polemi AM,Snyder A,Snyder C,Collins PG

    更新日期:2020-11-03 00:00:00

  • Quality assurance for a six degrees-of-freedom table using a 3D printed phantom.

    abstract:PURPOSE:To establish a streamlined end-to-end test of a 6 degrees-of-freedom (6DoF) robotic table using a 3D printed phantom for periodic quality assurance. METHODS:A 3D printed phantom was fabricated with translational and rotational offsets and an imbedded central ball-bearing (BB). The phantom underwent each step o...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12227

    authors: Woods K,Ayan AS,Woollard J,Gupta N

    更新日期:2018-01-01 00:00:00

  • A new homogeneity index based on statistical analysis of the dose-volume histogram.

    abstract::The goal of the present study was to develop a new dose-volume histogram (DVH)- based homogeneity index for effectively evaluating the dose homogeneity of intensity-modulated radiotherapy plans. The new index, called the sigma-index ("S-index") is defined as the standard deviation of the normalized differential DVH cu...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v8i2.2390

    authors: Yoon M,Park SY,Shin D,Lee SB,Pyo HR,Kim DY,Cho KH

    更新日期:2007-03-20 00:00:00

  • Mechanisms and prevention of thermal injury from gamma radiosurgery headframes during 3T MR imaging.

    abstract::Magnetic resonance imaging (MRI) is regularly used for stereotactic imaging of Gamma Knife (GK) radiosurgery patients for GK treatment planning. MRI-induced thermal injuries have occurred and been reported for GK patients with attached metallic headframes. Depending on the specific MR imaging and headframe conditions,...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v13i4.3613

    authors: Bennett MC,Wiant DB,Gersh JA,Dolesh W,Ding X,Best RC,Bourland JD

    更新日期:2012-07-05 00:00:00

  • Preventative maintenance and unscheduled downtime from an economic perspective.

    abstract::A spreadsheet-based model for economically characterizing the operation of a radiation treatment program has been used to perform a quantitative financial analysis of scheduled and unscheduled downtime. The incremental cost of downtime is broken down into three categories: remuneration of in-house or third party servi...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v1i2.2647

    authors: Dunscombe P,Roberts G,Valiquette L

    更新日期:2000-04-01 00:00:00

  • Development of a stereoscopic CT metal artifact management algorithm using gantry angle tilts for head and neck patients.

    abstract::Dental amalgams are a common source of artifacts in head and neck (HN) images. Commercial artifact reduction techniques have been offered, but are substantially ineffectual at reducing artifacts from dental amalgams, can produce additional artifacts, provide inaccurate HU information, or require extensive computation ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12922

    authors: Branco D,Kry S,Taylor P,Rong J,Zhang X,Peterson C,Frank S,Followill D

    更新日期:2020-08-01 00:00:00

  • Experience in implementing continuous arterial spin labeling on a commercial MR scanner.

    abstract::Continuous arterial spin labeling (CASL) is a technique for performing quantitative perfusion measurements without the need for exogenous contrast agent administration. This technique has seen limited use in the clinic due to problems of poor sensitivity and the potential for artifacts. In addition, CASL requires the ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 临床试验,杂志文章

    doi:10.1120/jacmp.v6i1.2068

    authors: Steger TR,Jackson EF

    更新日期:2005-01-01 00:00:00

  • Radiation therapy plan checks in a paperless clinic.

    abstract::Traditional quality assurance checks of a patient's radiation therapy plan involve printing out treatment parameters from the treatment planning system and the "record and verify" (R&V) system and visually checking the information for one-to-one correspondence. In a paperless environment, one can automate this process...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v10i1.2905

    authors: Siochi RA,Pennington EC,Waldron TJ,Bayouth JE

    更新日期:2009-01-27 00:00:00

  • Assessment of PlanIQ Feasibility DVH for head and neck treatment planning.

    abstract:INTRODUCTION:Designing a radiation plan that optimally delivers both target coverage and normal tissue sparing is challenging. There are limited tools to determine what is dosimetrically achievable and frequently the experience of the planner/physician is relied upon to make these determinations. PlanIQ software provid...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12165

    authors: Fried DV,Chera BS,Das SK

    更新日期:2017-09-01 00:00:00

  • Automatic calculation of patient size metrics in computed tomography: What level of computational accuracy do we need?

    abstract:OBJECTIVES:To compare the effectiveness of two different patient size metrics based on water equivalent diameter (Dw ), the mid-scan water equivalent diameter Dw_c , and the mean (average) water equivalent diameter in the imaged region, Dw_ave , for automatic detection of accidental changes in computed tomography (CT) ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12240

    authors: Sarmento S,Mendes B,Gouvêa M

    更新日期:2018-01-01 00:00:00

  • Fluoroscopic evaluation of diaphragmatic motion reduction with a respiratory gated radiotherapy system.

    abstract::We report on initial patient studies to evaluate the performance of a commercial respiratory gating radiotherapy system. The system uses a breathing monitor, consisting of a video camera and passive infrared reflective markers placed on the patient's thorax, to synchronize radiation from a linear accelerator with the ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v2i4.2596

    authors: Mageras GS,Yorke E,Rosenzweig K,Braban L,Keatley E,Ford E,Leibel SA,Ling CC

    更新日期:2001-10-01 00:00:00

  • Using an onboard kilovoltage imager to measure setup deviation in intensity-modulated radiation therapy for head-and-neck patients.

    abstract::The purpose of the present study was to use a kilovoltage imaging device to measure interfractional and intrafractional setup deviations in patients with head-and-neck or brain cancers receiving intensity-modulated radiotherapy (IMRT) treatment. Before and after IMRT treatment, approximately 3 times weekly, 7 patients...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v8i4.2439

    authors: Mechalakos JG,Hunt MA,Lee NY,Hong LX,Ling CC,Amols HI

    更新日期:2007-09-24 00:00:00

  • Surface and build-up dose comparison between Elekta 6 MV flattening filter and flattening-filter-free beams using an advanced Markus ionization chamber and a solid water-equivalent phantom.

    abstract::Using a plane-parallel advanced Markus ionization chamber and a stack of water-equivalent solid phantom blocks, percentage surface and build-up doses of Elekta 6 MV flattening filter (FF) and flattening-filter-free (FFF) beams were measured as a function of the phantom depth for field sizes ranging from 2 × 2 to 10 × ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.13094

    authors: Imae T,Takenaka S,Watanabe Y,Aoki A,Matsuda K,Sasaki K,Saegusa S,Nawa K,Nakagawa K,Abe O

    更新日期:2020-12-01 00:00:00

  • Defining internal target volume using positron emission tomography for radiation therapy planning of moving lung tumors.

    abstract::Substantial disagreement exists over appropriate PET segmentation techniques for non-small cell lung cancer. Currently, no segmentation algorithm explicitly considers tumor motion in determining tumor borders. We developed an automatic PET segmentation model as a function of target volume, motion extent, and source-to...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i1.4600

    authors: Riegel AC,Bucci MK,Mawlawi OR,Ahmad M,Luo D,Chandler A,Pan T

    更新日期:2014-01-06 00:00:00

  • A new frontier of image guidance: Organs at risk avoidance with MRI-guided respiratory-gated intensity modulated radiotherapy: Technical note and report of a case.

    abstract::The case of a 50-year-old man affected by a rhabdomiosarcoma metastatic lesion in the left flank Is reported. The patient was addressed to 50.4 Gy radiotherapy with concomitant chemotherapy in order to locally control the lesion. A Tri-60-Co magnetic resonance hybrid radiotherapy unit was used for treatment delivery a...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12575

    authors: Massaccesi M,Cusumano D,Boldrini L,Dinapoli N,Fionda B,Teodoli S,Azario L,Mattiucci GC,Balducci M,Cellini F,Valentini V

    更新日期:2019-06-01 00:00:00

  • Surface imaging-based analysis of intrafraction motion for breast radiotherapy patients.

    abstract::Breast treatments are becoming increasingly complex as the use of modulated and partial breast therapies becomes more prevalent. These methods are predicated on accurate and precise positioning for treatment. However, the ability to quantify intrafraction motion has been limited by the excessive dose that would result...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i6.4957

    authors: Wiant DB,Wentworth S,Maurer JM,Vanderstraeten CL,Terrell JA,Sintay BJ

    更新日期:2014-11-08 00:00:00

  • Dose differences in intensity-modulated radiotherapy plans calculated with pencil beam and Monte Carlo for lung SBRT.

    abstract::For patients with medically inoperable early-stage non-small cell lung cancer (NSCLC) treated with stereotactic body radiation therapy, early treatment plans were based on a simpler dose calculation algorithm, the pencil beam (PB) calculation. Because these patients had the longest treatment follow-up, identifying dos...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i6.5514

    authors: Liu H,Zhuang T,Stephans K,Videtic G,Raithel S,Djemil T,Xia P

    更新日期:2015-11-08 00:00:00

  • A novel and clinically useful dynamic conformal arc (DCA)-based VMAT planning technique for lung SBRT.

    abstract:PURPOSE:Volumetric modulated arc therapy (VMAT) is gaining popularity for stereotactic treatment of lung lesions for medically inoperable patients. Due to multiple beamlets in delivery of highly modulated VMAT plans, there are dose delivery uncertainties associated with small-field dosimetry error and interplay effects...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12878

    authors: Pokhrel D,Visak J,Sanford L

    更新日期:2020-07-01 00:00:00

  • Effect of RTOG breast/chest wall guidelines on dose-volume histogram parameters.

    abstract::Treatment planning for breast cancer has been traditionally based on clinical landmarks. The Radiation Therapy Oncology Group (RTOG) published consensus guidelines on contouring target volumes (TV) for the breast/chest wall and draining lymphatics. The effect of these guidelines on dosimetric parameters in surrounding...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i2.4547

    authors: Rudra S,Al-Hallaq HA,Feng C,Chmura SJ,Hasan Y

    更新日期:2014-03-06 00:00:00

  • Bleeding detection in wireless capsule endoscopy videos - Color versus texture features.

    abstract::Wireless capsule endoscopy (WCE) is an effective technology that can be used to make a gastrointestinal (GI) tract diagnosis of various lesions and abnormalities. Due to a long time required to pass through the GI tract, the resulting WCE data stream contains a large number of frames which leads to a tedious job for c...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12662

    authors: Pogorelov K,Suman S,Azmadi Hussin F,Saeed Malik A,Ostroukhova O,Riegler M,Halvorsen P,Hooi Ho S,Goh KL

    更新日期:2019-08-01 00:00:00

  • On correlations in IMRT planning aims.

    abstract::The purpose was to study correlations amongst IMRT DVH evaluation points and how their relaxation impacts the overall plan. 100 head-and-neck cancer cases, using the Eclipse treatment planning system with the same protocol, are statisti-cally analyzed for PTV, brainstem, and spinal cord. To measure variations amongst ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i6.6411

    authors: Roy A,Das IJ,Nohadani O

    更新日期:2016-11-08 00:00:00

  • Simplified "on-couch" daily quality assurance procedure for CT simulators.

    abstract::For most computed tomography (CT) simulators, radiation therapists must remove the flat couch top to perform daily CT quality assurance (QA), and then use separate tools to perform localization-laser QA. This process wastes time and effort and creates the opportunity for accidents to occur. In this study, we tested a ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v10i3.2844

    authors: Liu RR,Prado K,Gillin M

    更新日期:2009-06-17 00:00:00

  • The accuracy of dynamic wedge dose computation in the ADAC Pinnacle RTP system.

    abstract::The nonphysical wedge is a modality that uses computer-controlled jaw motion to generate wedge-shaped dose distributions. There are Varian enhanced dynamic wedges (EDWs) and Siemens virtual wedges (VWs). We recently commissioned dynamic wedges on both Varian and Siemens LINACs. The beam data, acquired with a Wellhofer...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v5i4.1964

    authors: Shao H,Wu X,Luo C,Crooks A,Bernstein A,Markoe A

    更新日期:2004-10-01 00:00:00

  • Calculating the peak skin dose resulting from fluoroscopically-guided interventions. Part II: case studies.

    abstract::Part II of this review article presents four case studies for which peak skin doses are calculated following the methods outlined in Part I of this review. The data available in the cases ranges from proprietary dose reports to fluoroscopy time and number of digital acquisition frames only. Flowcharts are provided for...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v13i1.3693

    authors: Jones AK,Pasciak AS

    更新日期:2012-01-05 00:00:00

  • Dosimetric comparison of three intensity-modulated radiation therapies for left breast cancer after breast-conserving surgery.

    abstract:PURPOSE:This study aimed to evaluate dosimetric differences of intensity-modulated radiation therapy (IMRT) in target and normal tissues after breast-conserving surgery. METHODS:IMRT five-field plan I, IMRT six-field plan II, and field-in-field-direct machine parameter optimization-IMRT plan III were designed for each...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12287

    authors: Zhang HW,Hu B,Xie C,Wang YL

    更新日期:2018-05-01 00:00:00