Understanding the impact of RapidArc therapy delivery errors for prostate cancer.

Abstract:

:The purpose of this study is to simulate random and systematic RapidArc delivery errors for external beam prostate radiotherapy plans in order to determine the dose sensitivity for each error type. Ten prostate plans were created with a single 360° arc. The DICOM files for these treatment plans were then imported into an in-house computer program that introduced delivery errors. Random and systematic gantry position (0.25°, 0.5°, 1°), monitor unit (MU) (1.25%, 2.5%, 5%), and multileaf collimator (MLC) position (0.5, 1, 2 mm) errors were introduced. The MLC errors were either random or one of three types of systematic errors, where the MLC banks moved in the same (MLC gaps remain unchanged) or opposing directions (increasing or decreasing the MLC gaps). The generalized equivalent uniform dose (gEUD) was calculated for the original plan and all treatment plans with errors introduced. The dose sensitivity for the cohort was calculated using linear regression for the gantry position, MU, and MLC position errors. Because there was a large amount of variability for systematic MLC position errors, the dose sensitivity of each plan was calculated and correlated with plan MU, mean MLC gap, and the percentage of MLC leaf gaps less than 1 and 2 cm for each individual plan. We found that random and systematic gantry position errors were relatively insignificant (< 0.1% gEUD change) for gantry errors up to 1°. Random MU errors were also insignificant, and systematic MU increases caused a systematic increase in gEUD. For MLC position errors, random MLC errors were relatively insignificant up to 2 mm as had been determined in previous IMRT studies. Systematic MLC shift errors caused a decrease of approximately -1% in the gEUD per mm. For systematic MLC gap open errors, the dose sensitivity was 8.2%/mm and for MLC gap close errors the dose sensitivity was -7.2%/mm. There was a large variability for MLC gap open/close errors for the ten RapidArc plans which correlated strongly with MU, mean gap width, and percentage of MLC gaps less than 1 or 2cm. This study evaluates the magnitude of various simulated RapidArc delivery errors by calculating gEUED on various prostate plans.

journal_name

J Appl Clin Med Phys

authors

Oliver M,Bush K,Zavgorodni S,Ansbacher W,Beckham WA

doi

10.1120/jacmp.v12i3.3409

subject

Has Abstract

pub_date

2011-05-20 00:00:00

pages

3409

issue

3

issn

1526-9914

journal_volume

12

pub_type

杂志文章
  • Role of adaptive radiation therapy for pediatric patients with diffuse pontine glioma.

    abstract::We investigate the role of adaptive radiation therapy in pediatric patients with diffuse pontine glioma and the impact of steroid-related weight gain on treatment parameters utilizing cone-beam CT. Fifteen patients with diffuse pontine glioma were treated with three-dimensional conformal radiation therapy and enrolled...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:

    authors: Beltran C,Sharma S,Merchant TE

    更新日期:2011-01-30 00:00:00

  • Case study of radiation therapy treatment of a patient with a cardiac ventricular assist device.

    abstract::A patient with a cardiac ventricular assist device (VAD) with computer-controlled driver presented to our department for radiation therapy. The treatment plan was 4500 cGy to the rectum over 25 fractions with 15MV photon beams. All beams avoided the pump and leads. The response to electromagnetic interference (EMI) wa...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v9i4.2851

    authors: Lasher DE,Wojcicka JB,Malcom R,Shears LL

    更新日期:2008-10-29 00:00:00

  • Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations.

    abstract::The purpose of this study was to implement full/half bowtie filter models in a com-mercial treatment planning system (TPS) to calculate kilovoltage (kV) cone-beam CT (CBCT) doses of Varian On-Board Imager (OBI) kV X-ray imaging system. The full/half bowtie filter models were created as compensators in Pinnacle TPS usi...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i2.5988

    authors: Kim S,Alaei P

    更新日期:2016-03-08 00:00:00

  • Dose calculation for hypofractionated volumetric-modulated arc therapy: approximating continuous arc delivery and tongue-and-groove modeling.

    abstract::Hypofractionated treatments generally increase the complexity of a treatment plan due to the more stringent constraints of normal tissues and target coverage. As a result, treatment plans contain more modulated MLC motions that may require extra efforts for accurate dose calculation. This study explores methods to min...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i2.4989

    authors: Yang J,Tang G,Zhang P,Hunt M,Lim SB,LoSasso T,Mageras G

    更新日期:2016-03-08 00:00:00

  • Improved dosimetric accuracy with semi-automatic contour propagation of organs-at-risk in glioblastoma patients undergoing chemoradiation.

    abstract:BACKGROUND:We study the changes in organs-at-risk (OARs) morphology as contoured on serial MRIs during chemoradiation therapy (CRT) of glioblastoma (GBM). The dosimetric implication of assuming non-deformable OAR changes and the accuracy and feasibility of semi-automatic OAR contour propagation are investigated. METHO...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12758

    authors: Lee S,Stewart J,Lee Y,Myrehaug S,Sahgal A,Ruschin M,Tseng CL

    更新日期:2019-12-01 00:00:00

  • Estimating and reducing dose received by cardiac devices for patients undergoing radiotherapy.

    abstract::The objectives of this project are to quantify the dose reduction effect provided by a lead shield for patients with cardiac implantable electronic devices (CIED) during a clinically realistic radiation treatment on phantom and to provide a simple model of dose estimation to predict dose received by CIED in a wide ran...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i6.5317

    authors: Bourgouin A,Varfalvy N,Archambault L

    更新日期:2015-11-08 00:00:00

  • Cross verification of independent dose recalculation, log files based, and phantom measurement-based pretreatment quality assurance for volumetric modulated arc therapy.

    abstract::Independent treatment planning system (TPS) check with Mobius3D software, log files based quality assurance (QA) with MobiusFX, and phantom measurement-based QA with ArcCHECK were performed and cross verified for head-and-neck (17 patients), chest (16 patients), and abdominal (19 patients) cancer patients who underwen...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.13036

    authors: Han C,Yi J,Zhu K,Zhou Y,Ai Y,Zheng X,Xie C,Jin X

    更新日期:2020-10-01 00:00:00

  • Preclinical assessment of comfort and secure fit of thermobrachytherapy surface applicator (TBSA) on volunteer subjects.

    abstract::A thermobrachytherapy surface applicator (TBSA) was developed for simultaneous heat and brachytherapy treatment of chest wall (CW) recurrence of breast cancer. The ability to comfortably secure the applicator over the upper torso relative to the CW target throughout treatment is assessed on volunteers. Male and postma...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v13i5.3845

    authors: Arunachalam K,Craciunescu OI,Markewitz EJ,Maccarini PF,Schlorff JL,Stauffer PR

    更新日期:2012-09-06 00:00:00

  • Monte Carlo calculation based on hydrogen composition of the tissue for MV photon radiotherapy.

    abstract::The purpose of this study was to demonstrate that Monte Carlo treatment planning systems require tissue characterization (density and composition) as a function of CT number. A discrete set of tissue classes with a specific composition is introduced. In the current work we demonstrate that, for megavoltage photon radi...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i5.5586

    authors: Demol B,Viard R,Reynaert N

    更新日期:2015-09-08 00:00:00

  • Quantitative evaluation of breast density using a dual-energy technique on a digital breast tomosynthesis system.

    abstract:PURPOSE:Although breast density is considered a strong risk factor of breast cancer, its quantitative assessment is difficult. To investigate a quantitative method of measuring breast density using dual-energy mammographic imaging with central digital breast tomosynthesis in physically uniform and nonuniform phantoms. ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12618

    authors: Lu KM,Yeh DM,Cao BH,Lin CY,Liang CY,Zhou YB,Tsai CJ

    更新日期:2019-06-01 00:00:00

  • Assessment of error in the MV radiation isocenter position calculated with the Elekta XVI software.

    abstract::The assessment of the coincidence of imaging and radiation isocenters is an important task of regular quality assurance of medical linear accelerators (linacs) as recommended in national and international quality assurance guidelines. A previously reported investigation of the accuracy of the Elekta XVI software to lo...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12861

    authors: Chojnowski JM,Warr GB,Sykes JR,Thwaites DI

    更新日期:2020-05-01 00:00:00

  • Development of a stereoscopic CT metal artifact management algorithm using gantry angle tilts for head and neck patients.

    abstract::Dental amalgams are a common source of artifacts in head and neck (HN) images. Commercial artifact reduction techniques have been offered, but are substantially ineffectual at reducing artifacts from dental amalgams, can produce additional artifacts, provide inaccurate HU information, or require extensive computation ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12922

    authors: Branco D,Kry S,Taylor P,Rong J,Zhang X,Peterson C,Frank S,Followill D

    更新日期:2020-08-01 00:00:00

  • Intrafraction 4D-cone beam CT acquired during volumetric arc radiotherapy delivery: kV parameter optimization and 4D motion accuracy for lung stereotactic body radiotherapy (SBRT) patients.

    abstract:PURPOSE:Elekta XVI 5.0 allows for four-dimensional cone beam computed tomography (4D CBCT) image acquisition during treatment delivery to monitor intrafraction motion. These images can have poorer image quality due to undersampling of kV projections and treatment beam MV scatter effects. We determine if a universal int...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12755

    authors: Liang J,Lack D,Zhou J,Liu Q,Grills I,Yan D

    更新日期:2019-12-01 00:00:00

  • Image texture and radiation dose properties in CT.

    abstract::The aim of this study was to compare image noise properties of GE Discovery HD 750 and Toshiba Aquilion ONE. The uniformity section of a Catphan 600 image quality assurance phantom was scanned with both scanners, at different dose levels and with extension rings simulating patients of different sizes. 36 datasets were...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i3.5900

    authors: Mozejko D,Andersen HK,Pedersen M,Waaler D,Martinsen AC

    更新日期:2016-05-08 00:00:00

  • A protocol to extend the longitudinal coverage of on-board cone-beam CT.

    abstract::The longitudinal coverage of a LINAC-mounted CBCT scan is limited to the corresponding dimensional limits of its flat panel detector, which is often shorter than the length of the treatment field. These limits become apparent when fields are designed to encompass wide regions, as when providing nodal coverage. Therefo...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v13i4.3796

    authors: Zheng D,Lu J,Jefferson A,Zhang C,Wu J,Sleeman W,Weiss E,Dogan N,Song S,Williamson J

    更新日期:2012-07-05 00:00:00

  • Hepatic encephalopathy: a neurochemical, neuroanatomical, and neuropsychological study.

    abstract::Hepatic encephalopathy (HE) is normally diagnosed by neuropsychological (NP) tests, which are not very specific and do not reveal the underlying pathology. Magnetic resonance imaging (MRI) and spectroscopy (MRS) of the brain offer alternative and possibly more specific markers for HE. These methods were applied in con...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v7i1.2151

    authors: Binesh N,Huda A,Thomas MA,Wyckoff N,Bugbee M,Han S,Rasgon N,Davanzo P,Sayre J,Guze B,Martin P,Fawzy F

    更新日期:2006-01-01 00:00:00

  • Development of a treatment planning protocol for prostate treatments using intensity modulated radiotherapy.

    abstract::We have developed a treatment planning protocol for intensity-modulated radiation therapy of the prostate using commercially available inverse planning software. Treatment plans were developed for ten patients using the Corvus version 3.8 planning system, testing various prescription options, including tissue types, d...

    journal_title:Journal of applied clinical medical physics

    pub_type: 临床试验,杂志文章

    doi:10.1120/jacmp.v2i2.2614

    authors: Ezzell GA,Schild SE,Wong WW

    更新日期:2001-04-01 00:00:00

  • Comparative analysis for renal stereotactic body radiotherapy using Cyberknife, VMAT and proton therapy based treatment planning.

    abstract:PURPOSE:We conducted this dosimetric analysis to evaluate the feasibility of a multi-center stereotactic body radiation therapy (SBRT) trial for renal cell carcinoma (RCC) using different SBRT platforms. MATERIALS/METHODS:The computed tomography (CT) simulation images of 10 patients with unilateral RCC previously trea...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12308

    authors: Baydoun A,Vapiwala N,Ponsky LE,Awan M,Kassaee A,Sutton D,Podder TK,Zhang Y,Dobbins D,Muzic RF Jr,Traughber B,Machtay M,Ellis R

    更新日期:2018-03-14 00:00:00

  • Beam-specific planning target volumes incorporating 4D CT for pencil beam scanning proton therapy of thoracic tumors.

    abstract::The purpose of this study is to determine whether organ sparing and target coverage can be simultaneously maintained for pencil beam scanning (PBS) proton therapy treatment of thoracic tumors in the presence of motion, stopping power uncertainties, and patient setup variations. Ten consecutive patients that were previ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i6.5678

    authors: Lin L,Kang M,Huang S,Mayer R,Thomas A,Solberg TD,McDonough JE,Simone CB 2nd

    更新日期:2015-11-08 00:00:00

  • Use of a radioactive check device for redundancy check of ionization chambers.

    abstract::The use of a radioactive check device containing Strontium 90 was investigated to carry out the redundancy checks of Farmer-type ionization chambers. It was not possible to meet the recommended tolerance limits of the redundancy checks without taking into account the angular response of the ionization chambers. The io...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v1i4.2636

    authors: Sidhu NP,Rouvas A,Cadman P

    更新日期:2000-10-01 00:00:00

  • Retrospective analysis of linear accelerator output constancy checks using process control techniques.

    abstract::Shewhart control charts have previously been suggested as a process control tool for use in routine linear accelerator (linac) output verifications. However, a comprehensive approach to process control has not been investigated for linac output verifications. The purpose of this work is to investigate a comprehensive ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v14i1.4032

    authors: Sanghangthum T,Suriyapee S,Srisatit S,Pawlicki T

    更新日期:2013-01-07 00:00:00

  • Evaluation of the combined use of two different respiratory monitoring systems for 4D CT simulation and gated treatment.

    abstract:PURPOSE:Two different respiratory monitoring systems (Varian's Real-Time Position Management (RPM) System and Siemens' ANZAI belt Respiratory Gating System) are compared in the context of respiratory signals and 4D CT images that are accordingly reconstructed. This study aims to evaluate the feasibility of combined use...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12434

    authors: Liu J,Lin T,Fan J,Chen L,Price R,Ma CC

    更新日期:2018-09-01 00:00:00

  • Dosimetric characteristics of a cubic-block-piled compensator for intensity-modulated radiation therapy in the Pinnacle radiotherapy treatment planning system.

    abstract::We examined the dose distributions generated by Pinnacle3 (Philips Radiation Oncology Systems, Milpitas, CA) for intensity-modulated radiotherapy (IMRT) plans using a cubic-block-piled compensator as the intensity modulator for 4-MV and 10-MV photon beams. The Pinnacle treatment planning system (TPS) uses an algorithm...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v8i1.2323

    authors: Sasaki K,Obata Y

    更新日期:2006-06-16 00:00:00

  • Rounded leaf end modeling in Pinnacle VMAT treatment planning for fixed jaw linacs.

    abstract::During volume-modulated arc therapies (VMAT), dosimetric errors are introduced by multiple open dynamic leaf gaps that are present in fixed diaphragm linear accelerators. The purpose of this work was to develop a methodology for adjusting the rounded leaf end modeling parameters to improve out-of-field dose agreement ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v17i6.6343

    authors: Young LA,Yang F,Cao N,Meyer J

    更新日期:2016-11-08 00:00:00

  • Detection of electron beam energy variations using a computed radiography system.

    abstract::A method to evaluate the electron beam energy constancy by employing the computed radiography (CR) system has been developed. In this method, a right triangular plastic wedge is used to produce a curve of the CR storage phosphor plate signal versus the wedge thickness. The curve, which resembles the percentage depth i...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v10i4.2911

    authors: Cai YC,Ge Y,Bernard D,Turian J,Chu JCH

    更新日期:2009-10-15 00:00:00

  • "Dose of the day" based on cone beam computed tomography and deformable image registration for lung cancer radiotherapy.

    abstract:PURPOSE:Adaptive radiotherapy (ART) has potential to reduce toxicity and facilitate safe dose escalation. Dose calculations with the planning CT deformed to cone beam CT (CBCT) have shown promise for estimating the "dose of the day". The purpose of this study is to investigate the "dose of the day" calculation accuracy...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1002/acm2.12793

    authors: Yuan Z,Rong Y,Benedict SH,Daly ME,Qiu J,Yamamoto T

    更新日期:2020-01-01 00:00:00

  • Effect of skull shape approximations in Gamma Knife dose calculations.

    abstract::Doses in Leksell GammaPlan (Elekta, Stockholm, Sweden) are calculated for each shot by summing the contribution from the 201 radiation beams emitted by the Gamma Knife (Elekta), weighted for attenuation in the tissue traversed. The patient's head is modeled based on 24 skull measurements, from which the depth to the c...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v8i3.2377

    authors: Berndt A,Beck J

    更新日期:2007-07-17 00:00:00

  • Simplified "on-couch" daily quality assurance procedure for CT simulators.

    abstract::For most computed tomography (CT) simulators, radiation therapists must remove the flat couch top to perform daily CT quality assurance (QA), and then use separate tools to perform localization-laser QA. This process wastes time and effort and creates the opportunity for accidents to occur. In this study, we tested a ...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v10i3.2844

    authors: Liu RR,Prado K,Gillin M

    更新日期:2009-06-17 00:00:00

  • CT and MR image fusion of tandem and ring applicator using rigid registration in intracavitary brachytherapy planning.

    abstract::The purpose of this study is to find the uncertainties in the reconstruction of MR compatible ring-tandem intracavitary applicators of high-dose rate image-based brachytherapy treatment planning using rigid registration of 3D MR and CT image fusion. Tandem and ring reconstruction in MR image based brachytherapy planni...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v15i2.4206

    authors: Oinam AS,Tomar P,Patel FD,Singh L,Rai B,Bahl A

    更新日期:2014-03-06 00:00:00

  • Accelerated tomotherapy delivery with TomoEdge technique.

    abstract::TomoEDGE is an advanced delivery form of tomotherapy which uses a dynamic secondary collimator. This plan comparison study describes the new features, their clinical applicability, and their effect on plan quality and treatment speed. For the first 45 patients worldwide that were scheduled for a treatment with TomoEdg...

    journal_title:Journal of applied clinical medical physics

    pub_type: 杂志文章

    doi:10.1120/jacmp.v16i2.4964

    authors: Katayama S,Haefner MF,Mohr A,Schubert K,Oetzel D,Debus J,Sterzing F

    更新日期:2015-03-08 00:00:00