Abstract:
PURPOSE:To investigate the effect of energy (kVp) and filters (no filter, half Bowtie, and full Bowtie) on the dose response curves of the Gafchromic XRQA2 film and nanoDot optical stimulated luminescence dosimeters (OSLDs) in CBCT dose fields. To measure surface and internal doses received during x-ray volume imager (XVI) (Version R4.5) and on board imager (OBI) (Version 1.5) CBCT imaging protocols using these two types of dosimeters. METHODS:Gafchromic XRQA2 film and nanoDot OSLD dose response curves were generated at different kV imaging settings used by XVI (software version R4.5) and OBI (software version 1.5) CBCT systems. The settings for the XVI system were: 100 kVp∕F0 (no filter), 120 kVp∕F0, and 120 kVp∕F1 (Bowtie filter), and for the OBI system were: 100 kVp∕full fan, 125 kVp∕full fan, and 125 kVp∕half fan. XRQA2 film was calibrated in air to air kerma levels between 0 and 11 cGy and scanned using reflection scanning mode with the Epson Expression 10000 XL flat-bed document scanner. NanoDot OSLDs were calibrated on phantom to surface dose levels between 0 and 14 cGy and read using the inLight(TM) MicroStar reader. Both dosimeters were used to measure in field surface and internal doses in a male Alderson Rando Phantom. RESULTS:Dose response curves of XRQA2 film and nanoDot OSLDs at different XVI and OBI CBCT settings were reported. For XVI system, the surface dose ranged between 0.02 cGy in head region during fast head and neck scan and 4.99 cGy in the chest region during symmetry scan. On the other hand, the internal dose ranged between 0.02 cGy in the head region during fast head and neck scan and 3.17 cGy in the chest region during chest M20 scan. The average (internal and external) dose ranged between 0.05 cGy in the head region during fast head and neck scan and 2.41 cGy in the chest region during chest M20 scan. For OBI system, the surface dose ranged between 0.19 cGy in head region during head scan and 4.55 cGy in the pelvis region during spot light scan. However, the internal dose ranged between 0.47 cGy in the head region during head scan and 5.55 cGy in the pelvis region during spot light scan. The average (internal and external) dose ranged between 0.45 cGy in the head region during head scan and 3.59 cGy in the pelvis region during spot light scan. Both Gafchromic XRQA2 film and nanoDot OSLDs gave close estimation of dose (within uncertainties) in many cases. Though, discrepancies of up to 20%-30% were observed in some cases. CONCLUSIONS:Dose response curves of Gafchromic XRQA2 film and nanoDot OSLDs indicated that the dose responses of these two dosimeters were different even at the same photon energy when different filters were used. Uncertainty levels of both dosimetry systems were below 6% at doses above 1 cGy. Both dosimetry systems gave almost similar estimation of doses (within uncertainties) in many cases, with exceptions of some cases when the discrepancy was around 20%-30%. New versions of the CBCT systems (investigated in this study) resulted in lower imaging doses compared with doses reported on earlier versions in previous studies.
journal_name
Med Physjournal_title
Medical physicsauthors
Giaddui T,Cui Y,Galvin J,Yu Y,Xiao Ydoi
10.1118/1.4803466subject
Has Abstractpub_date
2013-06-01 00:00:00pages
062102issue
6eissn
0094-2405issn
2473-4209journal_volume
40pub_type
杂志文章相关文献
MEDICAL PHYSICS文献大全abstract:PURPOSE:In Stereotactic body radiation therapy (SBRT) higher dose per fraction is delivered to patients and thus the need for an accurate dose computation is further elevated. However under such conditions of small field geometries, the electronic equilibrium can be lost, making it challenging for the dose-calculation ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4735539
更新日期:2012-06-01 00:00:00
abstract:PURPOSE:One of the promising options for motion management in radiation therapy (RT) is the use of LINAC-compatible robotic-arm-mounted ultrasound imaging system due to its high soft tissue contrast, real-time capability, absence of ionizing radiation, and low cost. The purpose of this work is to develop a novel deep l...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13510
更新日期:2019-05-01 00:00:00
abstract:PURPOSE:The purpose of this study is to investigate some characteristics of the beam delivery system and their effects on the dose distribution of intensity-modulated radiation therapy (IMRT) and the results of patient-specific IMRT quality assurance (QA). These characteristics include the accelerator source distributi...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3413997
更新日期:2010-06-01 00:00:00
abstract:PURPOSE:To develop a pencil beam dose calculation algorithm for scanned proton beams that improves modeling of scatter events. METHODS:Our pencil beam algorithm (PBA) was developed for calculating dose from monoenergetic, parallel proton beams in homogeneous media. Fermi-Eyges theory was implemented for pencil beam tr...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4735624
更新日期:2012-06-01 00:00:00
abstract::The contribution of a commercially available diode matrix (MapCHECK, provided by Sun Nuclear, Melbourne, FL) for the commissioning procedures of the voxel based Monte Carlo (VMC++) algorithm for electron beams of MasterPlan treatment planning system was investigated. The attention is mainly focused on the calculation ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3081414
更新日期:2009-04-01 00:00:00
abstract:PURPOSE:Computational models of microwave tissue ablation are widely used to guide the development of ablation devices, and are increasingly being used for the development of treatment planning and monitoring platforms. Knowledge of temperature-dependent dielectric properties of lung tissue is essential for accurate mo...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13704
更新日期:2019-10-01 00:00:00
abstract::Respiratory motion can introduce significant errors in radiotherapy. Conventional CT scans as commonly used for treatment planning can include severe motion artifacts that result from interplay effects between the advancing scan plane and object motion. To explicitly include organ/target motion in treatment planning a...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1869852
更新日期:2005-04-01 00:00:00
abstract::Weighting of given doses from parallel opposed fields is a widespread practice in radiation therapy of cancer in the tonsillar region. In order to determine when weighting is useful, dose distributions on the central axis were calculated for various energies (60Co, 4-, 8-, 15-, and MVp photons), weighting factors, and...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.594276
更新日期:1976-03-01 00:00:00
abstract:PURPOSE:Until recently the treatment verification for patients undergoing fast neutron therapy at our facility was accomplished through a combination of neutron beam portal films aligned with a graticule mounted on an orthronormal x-ray tube. To eliminate uncertainty with respect to the relative positions of the x-ray ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4734951
更新日期:2012-06-01 00:00:00
abstract::IMRT treatment plans for step-and-shoot delivery have traditionally been produced through the optimization of intensity distributions (or maps) for each beam angle. The optimization step is followed by the application of a leaf-sequencing algorithm that translates each intensity map into a set of deliverable aperture ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1477415
更新日期:2002-06-01 00:00:00
abstract:PURPOSE:To provide a fast computational method, based on the proximal graph solver (POGS) - A convex optimization solver using the alternating direction method of multipliers (ADMM), for calculating an optimal treatment plan in rotating shield brachytherapy (RSBT). RSBT treatment planning has more degrees of freedom th...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.12486
更新日期:2017-10-01 00:00:00
abstract::Conventional chest radiography is limited by the presence of scattered radiation and the small useful exposure range of radiographic film. A computer-assisted scanning system to minimize these two effects is outlined. The system uses a small beam of radiation swept over the patient's chest in a raster pattern to expos...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.595327
更新日期:1983-09-01 00:00:00
abstract:BACKGROUND:The rapid development and complexity of new x-ray computed tomography (CT) technologies and the need for evidence-based optimization of image quality with respect to radiation and contrast media dose call for an updated approach towards CT performance evaluation. AIMS:This report offers updated testing guid...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13763
更新日期:2019-11-01 00:00:00
abstract::Performance characteristics sufficient to provide physical data base specific to the Siemen's Stabilipan 2 orthovoltage x-ray therapy machine are presented. Operating conditions covering the working range of the unit from 100 to 300 kVp are selected. Beam quality, output, the central axis depth dose, relative output f...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.596016
更新日期:1987-09-01 00:00:00
abstract::An energy discriminating detector for dual-energy radiography can be configured as a two-layer sandwich, where the mean energy of photons detected by the two layers differs. To characterize the quantum noise of such a detector, the noise covariance between the two layers must be known in addition to the noise variance...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.596311
更新日期:1989-11-01 00:00:00
abstract:PURPOSE:We have investigated the dosimetric errors caused by the interplay between the motions of the LINAC and the tumor during the delivery of a volume modulated arc therapy treatment. This includes the development of an IMRT QA technique, applied here to evaluate RapidArc plans of varying complexity. METHODS:An IMR...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3263614
更新日期:2010-01-01 00:00:00
abstract::In this paper, the authors review the field of resolution modeling in positron emission tomography (PET) image reconstruction, also referred to as point-spread-function modeling. The review includes theoretical analysis of the resolution modeling framework as well as an overview of various approaches in the literature...
journal_title:Medical physics
pub_type: 杂志文章,评审
doi:10.1118/1.4800806
更新日期:2013-06-01 00:00:00
abstract:PURPOSE:To design an efficient method for utilizing phase-space source models in the GPU-based Monte Carlo (MC) dose calculation engine gDPM. METHODS:In GPU-based MC algorithms, particles are transported in parallel on different threads. Particles of different types and energies can require significantly different exe...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4735565
更新日期:2012-06-01 00:00:00
abstract::A method is presented for calculation of transmission functions for high energy photon beams through patients. These functions are being used in our clinic for prediction of portal dose images (PDIs) which are compared with PDIs measured with an electronic portal imaging device (EPID). The calculations are based on th...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.598293
更新日期:1998-06-01 00:00:00
abstract:PURPOSE:Patient-specific QA of highly conformal radiotherapy treatments are usually conducted using 2D or 3D dosimetry of the incident dose distribution in a water-equivalent phantom. However, dosimeters typically used for this task usually lack in either spatial resolution or dose accuracy. The purpose of this work is...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4736526
更新日期:2012-08-01 00:00:00
abstract:PURPOSE:The aim of this work is to investigate the predictive power of a common conventional intensity modulated radiation therapy (IMRT) quality assurance (QA) performance metric, the gamma passing rate (%GP), through the analysis of the sensitivity and of the correlation between %GP and different dose discrepancies b...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4767763
更新日期:2012-12-01 00:00:00
abstract:PURPOSE:To evaluate the treatment planning system (TPS) performance of the ZAP-X stereotactic radiosurgery (SRS) system through nondosimetric, dosimetric, and end-to-end (E2E) tests. METHODS:A comprehensive set of TPS commissioning and validation tests were developed using published guidelines. Nondosimetric validatio...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14740
更新日期:2021-01-28 00:00:00
abstract:PURPOSE:When an intensity-modulated radiation beam is delivered to a moving target, the interplay effect between dynamic beam delivery and the target motion due to miss-synchronization can cause unpredictable dose delivery. The portal dose image in electronic portal imaging device (EPID) represents radiation attenuated...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4799941
更新日期:2013-05-01 00:00:00
abstract::Heating rate (HR) patterns in cylindrical structures were studied with inhomogeneous limb phantoms. These phantoms, arm and thigh models consisting of fat, bone, and muscle material, were heated with Clini-Therm L, M, and MS applicators at 915 MHz. The thigh model is 18 cm in diameter with 2.65-cm-thick fat on the out...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.596627
更新日期:1991-11-01 00:00:00
abstract::New radiation therapy techniques such as IMRT present significant efficiency due to their highly conformal dose distributions. A consequence of the complexity of their dose distributions (high gradients, small irradiation fields, low dose distribution, ...) is the requirement for better precision quality assurance tha...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2897966
更新日期:2008-05-01 00:00:00
abstract::It has recently been shown that longitudinal aliasing can be a significant and detrimental presence in reconstructed single-slice helical computed tomography (CT) volumes. This aliasing arises because the directly measured data in helical CT are generally undersampled by a factor of at least 2 in the longitudinal dire...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1350583
更新日期:2001-03-01 00:00:00
abstract:PURPOSE:This article presents an iterative method for compensation of motion artifacts for slowly rotating computed tomography (CT) systems. Patient's motion introduces inconsistencies among projections and yields severe reconstruction artifacts for free-breathing acquisitions. Streaks and doubling of structures can ap...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3244035
更新日期:2009-11-01 00:00:00
abstract::It is well-known that the edge effect produced by phase contrast imaging results in the edge enhancement of x-ray images and thereby sharpens those images. It has recently been reported that phase contrast imaging using practical x-ray tubes with small focal spots has improved image sharpness as observed in the phase ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1992087
更新日期:2005-08-01 00:00:00
abstract:PURPOSE:To describe Acuros® CTS, a new software tool for rapidly and accurately estimating scatter in x-ray projection images by deterministically solving the linear Boltzmann transport equation (LBTE). METHODS:The LBTE describes the behavior of particles as they interact with an object across spatial, energy, and dir...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.12850
更新日期:2018-05-01 00:00:00
abstract:PURPOSE:To validate a delivery quality assurance (DQA) protocol for tomotherapy based on the measurement of the leaf open times (LOTs). In addition, to show the correlation between the mean relative LOT discrepancy and the dose deviation in the planning target volume (PTV). MATERIALS AND METHODS:We used a LOT measurem...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14348
更新日期:2020-09-01 00:00:00