Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing.

Abstract:

OBJECTIVE:The Welch Allyn Pro BP 2000, an automated sphygmomanometer, was subjected to ANSI/AAMI/ISO testing. PATIENTS AND METHODS:The protocol used was the same-arm sequential procedure described in the ANSI/AAMI/ISO 2013 Standard. Eight-eight (53 adults and 35 children aged 3-12 years old) patients completed testing successfully. Arm circumferences ranged from 15 to 49.5 cm. Seven different cuff sizes were used, including two long cuffs. RESULTS:All requirements for age, sex, blood pressure (BP), and cuff sizes were fulfilled. The mean±SD was -2.8±6.37 mmHg for systolic BP for criterion 1 and -3.6±6.14 mmHg for diastolic BP. The SD for criterion 2 were 5.29 for systolic BP and 5.75 for diastolic BP. All data passed the Standard's requirements. CONCLUSION:The Pro BP 2000 uses an accurate inflationary algorithm. The time for each inflation/deflation cycle is short, thus improving patient comfort. There is also arrhythmia detection to caution the use of BP values obtained during irregular heart rhythms. The algorithm for the professional-grade Pro BP 2000 is also contained in the Home BP 1700, targeted for home use. This encourages out-of-office self-measurement. The healthcare professional can be confident that the values obtained at different sites are comparable.

journal_name

Blood Press Monit

authors

Alpert BS

doi

10.1097/MBP.0000000000000350

subject

Has Abstract

pub_date

2018-12-01 00:00:00

pages

315-317

issue

6

eissn

1359-5237

issn

1473-5725

journal_volume

23

pub_type

临床试验,杂志文章
  • Simultaneous validation of the SunTech 247 diagnostic station blood pressure measurement device according to the British Hypertension Society protocol, the International Protocol and the Association for the Advancement of Medical Instrumentation standards

    abstract:OBJECTIVE:SunTech 247 is an oscillometric blood pressure (BP) monitor manufactured by SunTech Medical Inc., Morrisville, North Carolina, USA. The aims of this study were (i) to determine its accuracy using the 1993 modified British Hypertension Society (BHS) protocol, the 2002 International Protocol (IP) and the 2002/2...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e328330c873

    authors: Polo Friz H,Facchetti R,Primitz L,Beltrame L,Galbiati V,Ricioppo A,Bombelli M,Sega R

    更新日期:2009-10-01 00:00:00

  • Effects of salt loading and potassium supplement on the circadian blood pressure profile in salt-sensitive Chinese patients.

    abstract:BACKGROUND AND OBJECTIVE:Salt-sensitive (SS) patients more frequently showed a nondipper blood pressure pattern and were associated with more serious target organ damage than non-SS patients. We aimed to investigate whether potassium supplement can improve the blunted nocturnal blood pressure fall in SS patients expose...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000276

    authors: Guo TS,Dai Y,Ren KY,Mu JJ,Ren J,Wang D,Wang Y,Chu C,Li Y,Yuan ZY

    更新日期:2017-12-01 00:00:00

  • Feasibility of oscillometric aortic pressure and stiffness assessment using the VaSera VS-1500: comparison with a common tonometric method.

    abstract:OBJECTIVES:A number of operator-independent oscillometric devices to measure hemodynamics and arterial stiffness became available recently, but some and in particular VaSera VS-1500 do not provide estimates of aortic pressures and aortic pulse wave velocity (aPWV). The aim of this work was the retrospective application...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000137

    authors: Endes S,Bachler M,Li Y,Mayer C,Hanssen H,Hametner B,Schmidt-Trucksäss A,Wassertheurer S

    更新日期:2015-10-01 00:00:00

  • Propaq Neonatal monitor used with its own single- or with Critikon twin-hose cuffs: does it matter?.

    abstract::Some oscillometric non-invasive blood pressure (NIBP) monitors use twin-hose cuffs, whereas others use single-hose cuffs. The incompatibility hinders the transfer of patients from monitors using a single-hose cuff to monitors using twin-hose cuffs and vice versa. In particular, the Propaq Encore uses a single-hose cuf...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Amoore JN

    更新日期:1997-12-01 00:00:00

  • Acute effects of moderate-intensity and high-intensity exercise on hemodynamic and autonomic reactivity to the cold pressor test in young adults with excess body weight.

    abstract:OBJECTIVE:To investigate the effects of a single session of moderate-intensity and high-intensity interval exercise in cardiovascular reactivity to the cold pressor test in young adults with excess body weight. METHODS:Twenty-two subjects with excess body weight (23 ± 2 years; 30.0 ± 3.4 kg·m) performed three sessions...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000422

    authors: Meireles K,Peçanha T,Dias ARL,Souza KA,Araújo JA,Silva JS,Rezende DAN,Santos RJ,Cambri LT,Arsa G

    更新日期:2020-04-01 00:00:00

  • Determination of accuracy in neonates for non-invasive blood pressure device using an improved algorithm.

    abstract:BACKGROUND:New software (SuperSTAT(R) algorithm) with enhancements aimed at shorter determination times was developed for a non-invasive blood pressure (NIBP) device and a clinical evaluation was conducted to verify accuracy. OBJECTIVE:To determine the accuracy of the new algorithm according to ANSI/AAMI SP10-1992 and...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200204000-00006

    authors: Nelson RM,Stebor AD,Groh CM,Timoney PM,Theobald KS,Friedman BA

    更新日期:2002-04-01 00:00:00

  • Hidden errors of aneroid sphygmomanometers.

    abstract:BACKGROUND:Measurement of blood pressure remains the most commonly performed screening test in medical practice. With the likely removal of mercury sphygmomanometers from the workplace alternative devices are required. Of these the aneroid sphygmomanometer is popular both in the community and hospital setting. We inves...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200212000-00003

    authors: Waugh JJ,Gupta M,Rushbrook J,Halligan A,Shennan AH

    更新日期:2002-12-01 00:00:00

  • The importance of the sympathetic nervous system and systolic hypertension in patients with hypertension: benefits in treating patients with increased cardiovascular risk.

    abstract::The sympathetic nervous system is intimately involved in both the origin and the perpetuation of the hypertensive state, although there is some debate as to its precise contribution to the long-term hypertensive condition. Catecholamines are leading candidates to be both the pressor mechanism that initiates the rise i...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/00126097-200010002-00005

    authors: Sica DA

    更新日期:2000-01-01 00:00:00

  • Reference values for ambulatory blood pressure and self-measured blood pressure based on prospective outcome data.

    abstract::Although self-measured blood pressure and ambulatory blood pressure are widely used for research and clinical purposes, reference values are still scarcely supported by prospective outcome data. For self-measured blood pressure, values of 135 mmHg systolic and 85 mmHg diastolic (an average of two measurements in the m...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/00126097-200112000-00011

    authors: Verdecchia P

    更新日期:2001-12-01 00:00:00

  • Aortic systolic pressure derived with different calibration methods: associations to brachial systolic pressure in the general population.

    abstract:BACKGROUND:There is increasing evidence that the method of calibration directly influences the association between brachial systolic blood pressure (bSBP) and estimated aortic systolic blood pressure (aSBP) and subsequently affects prognostic and diagnostic differentiation power of the latter. OBJECTIVE:The aim of thi...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章

    doi:10.1097/MBP.0000000000000319

    authors: Wassertheurer S,Hametner B,Mayer CC,Hafez A,Negishi K,Papaioannou TG,Protogerou AD,Sharman JE,Weber T

    更新日期:2018-06-01 00:00:00

  • Brachial or wrist blood pressure in obese patients: which is the best?

    abstract:OBJECTIVE:Blood pressure (BP) measured in obese patients with a large arm circumference using a cuff of standard width may be overestimated. METHODS:We compared in this study the BP readings obtained with oscillometric devices at the left arm (OMRON HEM 705-CP) and the left wrist (OMRON R6) (Omron Medizintechnik, Mann...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e3282fd16e3

    authors: de Senarclens O,Feihl F,Giusti V,Engelberger RP,Rodieux F,Gomez P,Liaudet L,Waeber B

    更新日期:2008-06-01 00:00:00

  • The Maxi Stabil 3: validation of an aneroid device according to a modified British Hypertension Society protocol.

    abstract:BACKGROUND:Measuring blood pressure with mercury-independent alternatives is rapidly gaining attention. Mercury will be phased out of clinical use as a result of environmental, health and safety concerns and it is of vital importance that any alternative to mercury has to be of a recognized standard. We assessed the ac...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200304000-00005

    authors: Reinders A,Jones CR,Cuckson AC,Shennan AH

    更新日期:2003-04-01 00:00:00

  • Validation of the A&D UA-774 (UA-767Plus) device for self-measurement of blood pressure.

    abstract:OBJECTIVES:To determine the accuracy of the UA-774 (UA-767Plus) blood pressure monitor developed by the A&D Company. METHODS:Device evaluation was performed according to the modified British Hypertension Society (BHS) protocol released in 1993. Eighty-five subjects with characteristics outlined in the BHS protocol wer...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200408000-00008

    authors: Verdecchia P,Angeli F,Poeta F,Reboldi GP,Borgioni C,Pittavini L,Porcellati C

    更新日期:2004-08-01 00:00:00

  • Blood pressure assessment during resistance exercise: comparison between auscultation and Finapres.

    abstract:OBJECTIVES:The objective of this study was to test whether the auscultatory method and Finapres provide similar readings for blood pressure response during two resistance exercise protocols of different intensities, and to gain some insight into whether auscultation can be used to monitor cardiovascular stress during r...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e32809ef9f1

    authors: Polito MD,Farinatti PT,Lira VA,Nobrega AC

    更新日期:2007-04-01 00:00:00

  • Association of vitamin D deficiency with arterial stiffness in newly diagnosed hypertension.

    abstract:OBJECTIVE:Vitamin D has beneficial effects on vascular endothelial function, blood pressure (BP) and arterial stiffness. Arterial stiffness increases in early-stage hypertensive patients and it is a strong predictor of cardiovascular morbidity and mortality. The purpose of this study was to assess the association betwe...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000497

    authors: Cakal S,Çakal B,Karaca O

    更新日期:2020-11-23 00:00:00

  • Angiotensin II receptor blockers following intravenous nicardipine administration to lower blood pressure in patients with hypertensive intracerebral hemorrhage: a prospective randomized study.

    abstract:BACKGROUND AND OBJECTIVE:In patients with hypertensive intracerebral hemorrhage (HICH), intravenous nicardipine is primarily used to lower blood pressure (BP). However, there are few studies investigating the role of oral antihypertensives administered after intravenous nicardipine to prevent BP from rising. Angiotensi...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0000000000000225

    authors: Inamasu J,Nakae S,Adachi K,Hirose Y

    更新日期:2017-02-01 00:00:00

  • The effect of crossing legs on blood pressure: a randomized single-blind cross-over study.

    abstract:BACKGROUND:Although there is a theoretical basis for crossing legs to increase blood pressure, there are no published data addressing this question. As a result guidelines for measurement of blood pressure are not consistent in recommending that patients legs should not be crossed during measurement. OBJECTIVE:To dete...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章,随机对照试验

    doi:

    authors: Peters GL,Binder SK,Campbell NR

    更新日期:1999-04-01 00:00:00

  • Validation of the A&D BP UA-651 device with a wide-range cuff for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010.

    abstract::The aim of this study was to determine the accuracy of the A&D BP UA-651 device coupled to a wide-range cuff for home blood pressure (BP) measurement according to the International Protocol of the European Society of Hypertension. The device was evaluated in 33 patients. The mean age of the patients was 56.5±15.1 year...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000100

    authors: Benetti E,Fania C,Palatini P

    更新日期:2015-06-01 00:00:00

  • A short standardized protocol for measuring the QKD interval: comparison with 24 h monitoring and reproducibility.

    abstract::BACKGROUND: Ambulatory measurement for the QKD interval is a new method for evaluation of the rigidity of large arterial trunks. Reliable indices of arterial rigidity can be distilled from the large number of measurements (usually around 100) obtained during 24 h monitoring. OBJECTIVE: To describe a shorter standardiz...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Mas D,Gosse P,Julien V V,Jarnier P,Lemetayer P,Clementy J

    更新日期:1998-08-01 00:00:00

  • Accuracy and applicability of the Terumo ES-H55 double-cuff sphygmomanometer for hospital use.

    abstract:BACKGROUND:We have developed a new blood pressure (BP)-measuring device with a double-cuff, and evaluated its accuracy and applicability for practical use. DESIGN:The double-cuff method features a small cuff placed in the centre of the compression cuff to detect the oscillation of arterial pulsation, by which algorith...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章

    doi:10.1097/00126097-200310000-00005

    authors: Tochikubo O,Nishijima K,Ohshige K,Kimura K

    更新日期:2003-10-01 00:00:00

  • Comparison of the effects of two different teaching methods in blood pressure measurement training: a randomized controlled study.

    abstract:OBJECTIVE:To compare the effects of a high-fidelity simulator and conventional teaching model used during blood pressure measurement training on students' self-confidence, self-efficacy and anxiety levels. METHODS:A randomized controlled trial design was used. The study was conducted on 59 first-year nursing students ...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0000000000000401

    authors: Terzi B,Topbaş E,Bingöl G,Mavi Aydoğdu SG

    更新日期:2019-12-01 00:00:00

  • 'Diagnostic mode' improves adherence to the home blood pressure measurement schedule.

    abstract:BACKGROUND:The accuracy of home blood pressure measurement (HBPM) depends on adherence to the measurement schedule. We investigated the number of deviations from the requested schedule using an HBPM device equipped with a diagnostic mode that only allows patients to take a fixed number of BP readings at preset times. ...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0b013e328357352a

    authors: Wessel SE,van der Hoeven NV,Cammenga M,van Montfrans GA,van den Born BJ

    更新日期:2012-10-01 00:00:00

  • Which is the correct term for blood pressure measurements taken at home?

    abstract::The terms 'self-blood pressure' and 'home blood pressure' are being used to describe measurements of blood pressure taken by the patients at home. However, home measurements are not always self-measurements, because these are often taken by the patients' relatives. There is little evidence on the effect of self-measur...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/00126097-200308000-00007

    authors: Stergiou GS

    更新日期:2003-08-01 00:00:00

  • Characteristics of conventional blood pressure in studies on the predictive power of ambulatory blood pressure.

    abstract:BACKGROUND:It is commonly believed that the associations of left ventricular mass and cardiovascular morbidity/mortality with blood pressure are stronger for 24-h ambulatory pressure than for conventional clinic or casual pressure. METHODS:The investigation comprised a review of relevant studies, with particular empha...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,meta分析

    doi:10.1097/00126097-200202000-00006

    authors: Fagard RH,Staessen JA

    更新日期:2002-02-01 00:00:00

  • 24 h non-invasive ambulatory blood pressure monitoring: do the number of recordings per hour and/or ways of analyzing day and night matter?

    abstract:OBJECTIVE:To analyze whether the number of blood pressure registrations per hour and/or different ways of defining day and night have any influence on the result of ambulatory blood pressure monitoring. DESIGN AND METHODS:The study used blood pressure data from 40 normotensive subjects (20 men) and 40 (20 men) treated...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200110000-00005

    authors: Enström IE,Pennert KM

    更新日期:2001-10-01 00:00:00

  • Reduction of morning blood pressure surge after treatment with nifedipine GITS at bedtime, but not upon awakening, in essential hypertension.

    abstract:OBJECTIVES:The extent of morning blood pressure (BP) surge upon wakening has been associated with increased incidence of stroke and cardiovascular mortality. This trial investigated the antihypertensive efficacy and effects on the morning BP surge of awakening versus bedtime administration of nifedipine in essential hy...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0b013e32832e0d80

    authors: Hermida RC,Ayala DE,Mojón A,Alonso I,Fernández JR

    更新日期:2009-08-01 00:00:00

  • Comparison of two automated sphygmomanometers for use in the office setting.

    abstract:OBJECTIVE:To compare readings obtained using two automated blood pressure (BP) recording devices, BpTRU and Omron 907, in a clinical setting. METHODS:Two series of 50 patients attending a hypertension unit for 24-h ambulatory BP monitoring had BP recorded either every 1 or every 2 min using BpTRU and Omron 907 devices...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e32831e314f

    authors: Myers MG,Valdivieso M,Kiss A,Tobe SW

    更新日期:2009-02-01 00:00:00

  • Morning hypertension in chronic kidney disease is sustained type, but not surge type.

    abstract:OBJECTIVE:We have shown that as renal function deteriorates, the circadian blood pressure (BP) rhythm shifts to a nondipper pattern and the duration until nocturnal BP decline [dipping time (DT)] is prolonged. We investigated whether or not morning hypertension (BP 2 h after awakening >135/85 mmHg) in chronic kidney di...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e32835026c6

    authors: Mizuno M,Fukuda M,Miura T,Wakamatsu T,Naito T,Sato R,Togawa H,Sasakawa Y,Tomonari T,Ono M,Kato Y,Ichikawa T,Shirasawa Y,Ito A,Yoshida A,Kimura G

    更新日期:2012-02-01 00:00:00

  • Sensitivity, specificity, and predictive values of a forearm blood pressure measurement method in severe obesity.

    abstract::For blood pressure assessment, it has been reported that forearm blood pressure measurement appears to be as valid as an upper-arm blood pressure measurement method in individuals with severe obesity when correlated to the intra-arterial method, considered as the gold standard. Sensitivity, specificity, and predictive...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000097

    authors: Leblanc MÈ,Cloutier L,Poirier P

    更新日期:2015-04-01 00:00:00

  • The effect of Ramadan fasting on ambulatory blood pressure in treated hypertensive patients using diuretics.

    abstract:OBJECTIVE:The influence of the Ramadan fasting on blood pressure (BP) in patients taking diuretics therapy has not been addressed specifically yet. The aim of this study was to examine the effect of long Ramadan fasting on BP with ambulatory BP monitoring in hypertensive patients using diuretics. METHODS:This prospect...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000451

    authors: Aslan S,Riza Demir A,Kahraman S,Memiç K,Avci Y,Gürbak İ,Karabulut E,Ertürk M

    更新日期:2020-08-01 00:00:00