Position paper: recommendations for a digital mammography quality assurance program V4.0.

Abstract:

:In 2001 the ACPSEM published a position paper on quality assurance in screen film mammography which was subsequently adopted as a basis for the quality assurance programs of both the Royal Australian and New Zealand College of Radiologists (RANZCR) and of BreastScreen Australia. Since then the clinical implementation of digital mammography has been realised and it has become evident that existing screen-film protocols were not appropriate to assure the required image quality needed for reliable diagnosis or to address the new dose implications resulting from digital technology. In addition, the advantages and responsibilities inherent in teleradiology are most critical in mammography and also need to be addressed. The current document is the result of a review of current overseas practice and local experience in these areas. At this time the technology of digital imaging is undergoing significant development and there is still a lack of full international consensus about some of the detailed quality control (QC) tests that should be included in quality assurance (QA) programs. This document describes the current status in digital mammography QA and recommends test procedures that may be suitable in the Australasian environment. For completeness, this document also includes a review of the QA programs required for the various types of digital biopsy units used in mammography. In the future, international harmonisation of digital quality assurance in mammography and changes in the technology may require a review of this document. Version 2.0 represented the first of these updates and key changes related to image quality evaluation, ghost image evaluation and interpretation of signal to noise ratio measurements. In Version 3.0 some significant changes, made in light of further experience gained in testing digital mammography equipment were introduced. In Version 4.0, further changes have been made, most notably digital breast tomosynthesis (DBT) testing and QC have been addressed. Some additional testing for conventional projection imaging has been added in order that sites may have the capability to undertake dose surveys to confirm compliance with diagnostic reference levels (DRLs) that may be established at the National or State level. A key recommendation is that dosimetry calculations are now to be undertaken using the methodology of Dance et al. Some minor changes to existing facility QC tests have been made to ensure the suggested procedures align with those most recently adopted by the Royal Australian and New Zealand College of Radiologists and BreastScreen Australia. Future updates of this document may be provided as deemed necessary in electronic format on the ACPSEM's website ( https://www.acpsem.org.au/whatacpsemdoes/standards-position-papers and see also http://www.ranzcr.edu.au/quality-a-safety/radiology/practice-quality-activities/mqap ).

authors

Heggie JCP,Barnes P,Cartwright L,Diffey J,Tse J,Herley J,McLean ID,Thomson FJ,Grewal RK,Collins LT

doi

10.1007/s13246-017-0583-x

subject

Has Abstract

pub_date

2017-09-01 00:00:00

pages

491-543

issue

3

eissn

0158-9938

issn

1879-5447

pii

10.1007/s13246-017-0583-x

journal_volume

40

pub_type

杂志文章
  • A model based on the Pennes bioheat transfer equation is valid in normal brain tissue but not brain tissue suffering focal ischaemia.

    abstract::Ischaemic stroke is a major public health issue in both developed and developing nations. Hypothermia is believed to be neuroprotective in cerebral ischaemia. Conversely, elevated brain temperature is associated with poor outcome after ischaemic stroke. Mechanisms of heat exchange in normally-perfused brain are relati...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-017-0595-6

    authors: Lillicrap T,Tahtalı M,Neely A,Wang X,Bivard A,Lueck C

    更新日期:2017-12-01 00:00:00

  • A computer-assisted method for pathogenicity assessment and genetic reporting of variants stored in the Australian Inherited Retinal Disease Register.

    abstract::The assignment of pathogenicity to variants suspected of causing an inherited retinal disease and the subsequent creation of molecular genetic reports sent to clinical geneticists and ophthalmologists has traditionally been time-consuming and subject to error and ambiguity. The purpose of this paper is to describe a c...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-015-0420-z

    authors: Huynh E,De Roach J,McLaren T,Thompson J,Montgomery H,Kap C,Hoffmann L,Lamey T

    更新日期:2016-03-01 00:00:00

  • A proton therapy system in Nagoya Proton Therapy Center.

    abstract::The purpose of this paper is to describe an outline of a proton therapy system in Nagoya Proton Therapy Center (NPTC). The NPTC has a synchrotron with a linac injector and three treatment rooms: two rooms are equipped with a gantry and the other one is equipped with a fixed horizontal beamline. One gantry treatment ro...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-016-0456-8

    authors: Toshito T,Omachi C,Kibe Y,Sugai H,Hayashi K,Shibata H,Yasui K,Tanaka K,Yamamoto T,Yoshida A,Nikawa E,Asai K,Shimomura A,Okumura I,Suzuki T,Kinou H,Isoyama S,Ogino H,Iwata H,Shibamoto Y,Mizoe J

    更新日期:2016-09-01 00:00:00

  • Evaluation of EBT radiochromic film using a multiple exposure technique.

    abstract::Radiochromic film is a self developing two-dimensional dosimeter system that is widely used in radiotherapy. Since its development some 40 years ago many improvements have been made; however the sensitivity can vary across the film and a non-linear dose response remains. The former can be addressed using a double expo...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-011-0067-3

    authors: Roozen K,Kron T,Haworth A,Franich R

    更新日期:2011-06-01 00:00:00

  • Echocardiographic assessment of long-term hemodynamic characteristics of mechanical mitral valve prostheses with different mitral valvular diseases.

    abstract::Mitral stenosis (MS) and mitral insufficiency (MI) have different pre-operative hemodynamic characteristics. However, it is unclear if there are differences in long-term echocardiographic characteristics of MS and MI patients after mechanical mitral valve replacement. This study is to compare long-term echocardiograph...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-016-0521-3

    authors: Zhu X,Li Q,Tang H,Xiao X

    更新日期:2017-03-01 00:00:00

  • Automatic counting of nuclear tracks.

    abstract::Two software packages designed for automatic counting nuclear tracks are tested and compared. The image processing in the first package is divided into three parts segmentation, speckle elimination and removal of ill-formed track holes. The second package is also divided into three parts but the last part is modified ...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:

    authors: Coakley K,Wong CF,Forbes K,Anh V

    更新日期:1992-06-01 00:00:00

  • Optimal selection of mother wavelet for accurate infant cry classification.

    abstract::Wavelet theory is emerging as one of the prevalent tool in signal and image processing applications. However, the most suitable mother wavelet for these applications is still a relative question mark amongst researchers. Selection of best mother wavelet through parameterization leads to better findings for the analysi...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-014-0264-y

    authors: Saraswathy J,Hariharan M,Nadarajaw T,Khairunizam W,Yaacob S

    更新日期:2014-06-01 00:00:00

  • Long term OSLD reader stability in the ACDS level one audit.

    abstract::The Australian Clinical Dosimetry Service (ACDS) has demonstrated the capacity to perform a basic dosimetry audit on all radiotherapy clinics across Australia. During the ACDS's three and a half year trial the majority of the audits were performed using optically stimulated luminescence dosimeters (OSLD) mailed to fac...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-014-0320-7

    authors: Alves AD,Lye J,Kenny J,Dunn L,Lehmann J,Cole A,Kron T,Butler D,Johnston P,Williams I

    更新日期:2015-03-01 00:00:00

  • Performance of independent dose calculation in helical tomotherapy: implementation of the MCSIM code.

    abstract::Currently, a software-based second check dose calculation for helical tomotherapy (HT) is not available. The goal of this study is to evaluate the dose calculation accuracy of the in-house software using EGS4/MCSIM Monte Carlo environment against the treatment planning system calculations. In-house software was used t...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-012-0165-x

    authors: Dzintars E,Stathakis S,Mavroidis P,Sadeghi A,Papanikolaou N

    更新日期:2012-12-01 00:00:00

  • ACPSEM ROSG Oncology-PACS and OIS working group recommendations for quality assurance.

    abstract::The Australasian College of Physical Scientists and Engineers in Medicine (ACPSEM) Radiation Oncology Specialty Group (ROSG) formed a series of working groups in 2011 to develop position papers for guidance of radiation oncology medical physics practice within the Australasian setting. These position papers are intend...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-013-0238-5

    authors: Shakeshaft J,Perez M,Tremethick L,Ceylan A,Bailey M

    更新日期:2014-03-01 00:00:00

  • Image quality improvement for medium and large field of view Elekta XVI scans.

    abstract::Cone-beam computed tomography (CBCT) has become the standard imaging technique for many sites treated with Radiotherapy. The Elekta X-ray Volumetric Imaging (XVI) system allows for the acquisition of CBCTs with three different diameters; small (27 cm), medium (41 cm) and large (50 cm) (Elekta in XVI Corrective Mainten...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-019-00817-7

    authors: Little DP

    更新日期:2019-12-01 00:00:00

  • A Monte Carlo technique to establish the water/tissue equivalence of phantom materials.

    abstract::The quality of phantom materials is crucial for accurate dosimetry in radiotherapy. A wide range of factors such as density, electron density and elemental composition can influence the radiation properties, and hence the absorbed dose, of materials. New materials can be tested by direct measurements which requires co...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:

    authors: Keall P,Kron T,Hoban P

    更新日期:1993-09-01 00:00:00

  • An investigation into lung correction for tangential breast irradiation.

    abstract::A solid water breast phantom with a lung equivalent material was constructed to investigate interface effects of 6 MV tangential photon fields. The phantom was built so that Thermo-luminescence Dosimeters (TLD) could be used for comparison to a pencil beam and Equivalent Tissue-Air-Ratio (ETAR) algorithms. The investi...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:

    authors: Hill B,Brown R

    更新日期:1998-03-01 00:00:00

  • An effective approach to classify epileptic EEG signal using local neighbor gradient pattern transformation methods.

    abstract::Electroencephalographic (EEG) signal records the neuronal activity in the brain and it is used in the diagnosis of epileptic seizure activities. Human inspection of non-stationary EEG signal for diagnosing epilepsy is cumbersome, time-consuming and inaccurate. In this paper an effective automatic approach to detect ep...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-018-0697-9

    authors: Sairamya NJ,Thomas George S,Balakrishnan R,Subathra MSP

    更新日期:2018-12-01 00:00:00

  • SAR and temperature distribution in the rat head model exposed to electromagnetic field radiation by 900 MHz dipole antenna.

    abstract::Rats are often used in the electromagnetic field (EMF) exposure experiments. In the study for the effect of 900 MHz EMF exposure on learning and memory in SD rats, the specific absorption rate (SAR) and the temperature rise in the rat head are numerically evaluated. The digital anatomical model of a SD rat is reconstr...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-013-0202-4

    authors: Yang L,Hao D,Wu S,Zhong R,Zeng Y

    更新日期:2013-06-01 00:00:00

  • Evaluation of dose enhancement in presence of gold nanoparticles in eye brachytherapy by 103Pd source.

    abstract::In recent years, the use of gold nanoparticles in radiation therapy has been introduced as a new approach in radiotherapy. The aim of this study is to investigate the effect of gold nanoparticles (GNPs) in plaque brachytherapy for choroidal melanoma using Monte Carlo (MC) simulation. MCNPX code was used for simulation...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-017-0555-1

    authors: Rezaei H,Zabihzadeh M,Ghorbani M,Goli Ahmadabad F,Mostaghimi H

    更新日期:2017-09-01 00:00:00

  • Hidden pattern discovery on epileptic EEG with 1-D local binary patterns and epileptic seizures detection by grey relational analysis.

    abstract::This paper proposes a novel approach to detect epilepsy seizures by using Electroencephalography (EEG), which is one of the most common methods for the diagnosis of epilepsy, based on 1-Dimension Local Binary Pattern (1D-LBP) and grey relational analysis (GRA) methods. The main aim of this paper is to evaluate and val...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-015-0362-5

    authors: Kaya Y

    更新日期:2015-09-01 00:00:00

  • Detection of left ventricle systolic dysfunction from shape deformity.

    abstract::Coronary artery disease producing ischemic cardiomyopathy is the most frequent cause of left ventricular systolic dysfunction. Non-ischemic cardiomyopathies can also produce systolic dysfunction; they may be inherited as genetic disorders or occur sporadically. These coronary injuries have repercussions on the left ve...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/BF03178864

    authors: Manivannan J,Reddy MR,Thanikachalam S,Kumar RA

    更新日期:2005-03-01 00:00:00

  • Characterization of a Fricke dosimeter at high energy photon and electron beams used in radiotherapy.

    abstract::The dosimetric features of the Fricke dosimeter in clinical linear accelerator beams are considered. Experimental data were obtained using various nominal energies 6 and 18 MV, 12 and 15 MeV, including the (60)Co γ-ray beam. The calibration of the dosimeters was performed using the ionization chamber as a reference do...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-011-0093-1

    authors: Moussous O,Khoudri S,Benguerba M

    更新日期:2011-12-01 00:00:00

  • Depth of anaesthesia assessment based on adult electroencephalograph beta frequency band.

    abstract::This paper presents a new method to apply timing characteristics of electroencephalograph (EEG) beta frequency bands to assess the depth of anaesthesia (DoA). Firstly, the measured EEG signals are denoised and decomposed into 20 different frequency bands. The Mobility (M), permutation entropy (PE) and Lempel-Ziv compl...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-016-0459-5

    authors: Li T,Wen P

    更新日期:2016-09-01 00:00:00

  • Human brain wave activity during exposure to radiofrequency field emissions from mobile phones.

    abstract::The aim of this study was to determine whether there is an effect of mobile phone electromagnetic field emissions on the human electroencephalograph (EEG). EEG recordings from ten awake subjects were taken during exposure to radiofrequency (RF) emissions from a mobile phone positioned behind the head. Two experimental...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 临床试验,杂志文章,随机对照试验

    doi:10.1007/BF03179176

    authors: D'Costa H,Trueman G,Tang L,Abdel-rahman U,Abdel-rahman W,Ong K,Cosic I

    更新日期:2003-12-01 00:00:00

  • Non invasive measurement of myocardial motion using magnetic resonance tagging.

    abstract::Magnetic resonance imaging has become a key modality within the modern radiology department. In addition to the conventional slice images, three dimensional angiographic images, and quantitative flow measurements can be obtained. More recently methods have been developed which enable one to visualize the motion of obj...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章,评审

    doi:

    authors: McKinnon GC,Fischer SE,Maier SE

    更新日期:1991-12-01 00:00:00

  • Computer model to optimise contrast in chest radiography.

    abstract::X-ray imaging of the chest is challenging because of the large differences in x-ray attenuation between the mediastinal and lung regions of the chest. This large dynamic range of chest exit exposure, along with associated large changes in the scatter component of the beam, makes film based chest imaging difficult to o...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:

    authors: McLean D

    更新日期:1993-12-01 00:00:00

  • Short tangential arcs in VMAT based breast and chest wall radiotherapy lead to conformity of the breast dose with lesser cardiac and lung doses: a prospective study of breast conservation and mastectomy patients.

    abstract::Volumetric modulated arc therapy (VMAT) is modern rotational intensity modulated therapy used for treatment of several sites. The study aimed to analyze partial tangential arc VMAT treatment planning and delivery, including analyzing the cardiac and contralateral breast doses resulting from this technique. A total of ...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-017-0558-y

    authors: Munshi A,Sarkar B,Anbazhagan S,Giri UK,Kaur H,Jassal K,Ganesh T,Kalyan Mohanti B

    更新日期:2017-09-01 00:00:00

  • An application of linear output error modelling for studying lymphocyte migration in peripheral lymphoid tissues.

    abstract::Lymphocyte recirculation between lymphatic and blood vessels and migration through tissues are essential mechanisms underlying immunological surveillance. However, the kinetics of lymphocyte migration through lymphoid tissues remains poorly understood. The present study of lymphocyte migration, based on a sheep model ...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/BF03178774

    authors: Srikusalanukul W,De Bruyne F,McCullagh P

    更新日期:2002-09-01 00:00:00

  • Effect of computed tomography number-relative electron density conversion curve on the calculation of radiotherapy dose and evaluation of Monaco radiotherapy treatment planning system.

    abstract::The accuracy of a computed tomography (CT)-relative electron density (RED) curve may have an indirect impact on the accuracy of dose calculation by a treatment planning system (TPS). This effect has not been previously quantified for input of different CT-RED curves from different CT-scan units in the Monaco TPS. This...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-019-00745-6

    authors: Hasani M,Farhood B,Ghorbani M,Naderi H,Saadatmand S,Karimkhani Zandi S,Knaup C

    更新日期:2019-06-01 00:00:00

  • A low cost ECG central station for intensive care.

    abstract::A electrocardiogram (ECG) central station has been developed for the Intensive Care Unit at Christchurch Hospital. The system allows the selection and display of four ECGs selected from seven bedside monitors in the Unit. It also provides for automatic display, hard copy and audio alarm of ECG from any monitor which h...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:

    authors: Jans R,Elder AG,Jones RD,Fright WR,Bones PJ

    更新日期:1990-03-01 00:00:00

  • Instruments to assess and measure personal and environmental radiofrequency-electromagnetic field exposures.

    abstract::Radiofrequency-electromagnetic field (RF-EMF) exposure of human populations is increasing due to the widespread use of mobile phones and other telecommunication and broadcasting technologies. There are ongoing concerns about potential short- and long-term public health consequences from RF-EMF exposures. To elucidate ...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章,评审

    doi:10.1007/s13246-015-0412-z

    authors: Bhatt CR,Redmayne M,Abramson MJ,Benke G

    更新日期:2016-03-01 00:00:00

  • Proposing feature engineering method based on deep learning and K-NNs for ECG beat classification and arrhythmia detection.

    abstract::Arrhythmia is slow, fast or irregular heartbeat. Manual ECG assessment and disease classification is an error-prone task because of vast differences in ECG morphology and difficulty in accurate identifying ECG components. Moreover, proposing a computer-aided diagnosis system for heartbeat classification can be useful ...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:10.1007/s13246-019-00814-w

    authors: Khatibi T,Rabinezhadsadatmahaleh N

    更新日期:2019-11-26 00:00:00

  • Automated ventilator testing.

    abstract::A new era has arrived for the Biomedical Engineering Department at the Royal Women's Hospital in Melbourne. We have developed a system to qualitatively test for intermittent or unconfirmed faults, associated with Bear Cub ventilators. Where previous testing has been inadequate, computer logging is now used to interfac...

    journal_title:Australasian physical & engineering sciences in medicine

    pub_type: 杂志文章

    doi:

    authors: Ghaly J,Smith AL

    更新日期:1994-06-01 00:00:00