Abstract:
PURPOSE:Distortions in magnetic resonance imaging (MRI) compromise spatial fidelity, potentially impacting delineation and dose calculation. We characterized 2D and 3D large field of view (FOV), sequence-independent distortion at various positions in a 1.0 T high-field open MR simulator (MR-SIM) to implement correction maps for MRI treatment planning. METHODS:A 36 × 43 × 2 cm(3) phantom with 255 known landmarks (∼1 mm(3)) was scanned using 1.0 T high-field open MR-SIM at isocenter in the transverse, sagittal, and coronal axes, and a 465 × 350 × 168 mm(3) 3D phantom was scanned by stepping in the superior-inferior direction in three overlapping positions to achieve a total 465 × 350 × 400 mm(3) sampled FOV yielding >13 800 landmarks (3D Gradient-Echo, TE/TR/α = 5.54 ms/30 ms/28°, voxel size = 1 × 1 × 2 mm(3)). A binary template (reference) was generated from a phantom schematic. An automated program converted MR images to binary via masking, thresholding, and testing for connectivity to identify landmarks. Distortion maps were generated by centroid mapping. Images were corrected via warping with inverse distortion maps, and temporal stability was assessed. RESULTS:Over the sampled FOV, non-negligible residual gradient distortions existed as close as 9.5 cm from isocenter, with a maximum distortion of 7.4 mm as close as 23 cm from isocenter. Over six months, average gradient distortions were -0.07 ± 1.10 mm and 0.10 ± 1.10 mm in the x and y directions for the transverse plane, 0.03 ± 0.64 and -0.09 ± 0.70 mm in the sagittal plane, and 0.4 ± 1.16 and 0.04 ± 0.40 mm in the coronal plane. After implementing 3D correction maps, distortions were reduced to <1 pixel width (1 mm) for all voxels up to 25 cm from magnet isocenter. CONCLUSIONS:Inherent distortion due to gradient nonlinearity was found to be non-negligible even with vendor corrections applied, and further corrections are required to obtain 1 mm accuracy for large FOVs. Statistical analysis of temporal stability shows that sequence independent distortion maps are consistent within six months of characterization.
journal_name
Med Physjournal_title
Medical physicsauthors
Price RG,Kadbi M,Kim J,Balter J,Chetty IJ,Glide-Hurst CKdoi
10.1118/1.4930245subject
Has Abstractpub_date
2015-10-01 00:00:00pages
5955-60issue
10eissn
0094-2405issn
2473-4209journal_volume
42pub_type
杂志文章相关文献
MEDICAL PHYSICS文献大全abstract::The anti-scatter grid has been widely used to reject scatter and increase the perceptibility of a low-contrast object in chest radiography; however, it also attenuated the primary x-rays, resulting in a substantial loss of information and an increased relative noise level in heavily attenuated regions. A more dose eff...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2739805
更新日期:2007-07-01 00:00:00
abstract::This paper examines statistical errors in the measurement of arterial stenoses by digital videodensitometry. Images of vessel phantoms were acquired using digital subtraction angiographic techniques with low concentrations of an iodine contrast medium and low levels of x-ray exposure. Effects of the spatial and tempor...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.595652
更新日期:1985-11-01 00:00:00
abstract::The purpose of this paper is to examine the thermal dose distribution, to configure the optimal absorbed power deposition, and to design an appropriate heating strategy for ultrasound thermal therapy. This work employs simulation programs, which are based on the transient bio-heat transfer equation and an ideal absorb...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1406516
更新日期:2001-10-01 00:00:00
abstract:PURPOSE:The first main aim of this study was to illustrate the absorbed dose rate distribution from 177Lu in sections of xenografted prostate cancer (PCa) tumors using high resolution digital autoradiography (DAR) and compare it with hypothetical identical radioactivity distributions of 90Y or 7 MeV alpha-particles. Th...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4967877
更新日期:2016-12-01 00:00:00
abstract::Significant advances have been made in recent years to improve photon dose calculation. However, accurate prediction of dose perturbation effects near the interfaces of different media, where charged particle equilibrium is not established, remain unsolved. Furthermore, changes in atomic number, which affect the multi...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.597611
更新日期:1995-07-01 00:00:00
abstract:PURPOSE:New technology has been developed to treat age-related macular degeneration (AMD) using 100 kVp pencil-beams that enter the patient through the radio-resistant sclera with a depth of interest between 1.6 and 2.6 cm. Measurement of reference and relative dose in a kilovoltage x-ray beam with a 0.42 cm diameter f...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4863482
更新日期:2014-02-01 00:00:00
abstract::The visual observer performance for detecting low-contrast patterns by employing both the receiver operating characteristic (ROC) and the 9-alternative forced choice (9-AFC) method is investigated. For each observer and each experimental technique, the results were reproducible with respect to a repeated observation. ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.596986
更新日期:1993-05-01 00:00:00
abstract::We have reported in our previous studies on the methodology, and feasibility of 4D-PET (Gated PET) acquisition, to reduce respiratory motion artifact in PET imaging of the thorax. In this study, we expand our investigation to address the problem of respiration motion in PET/CT imaging. The respiratory motion of four l...
journal_title:Medical physics
pub_type: 临床试验,杂志文章
doi:10.1118/1.1809778
更新日期:2004-12-01 00:00:00
abstract::Extensive information has been made available relative to the physical aspects necessary for calculation of radiation absorbed dose from radiopharmaceuticals. A similar data base for the biological factors involved in these calculations has not been documented as thoroughly. The authors present an extensive literature...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.594679
更新日期:1980-05-01 00:00:00
abstract:PURPOSE:The purpose of this study was to investigate the potential of MR-guided pulsed focused ultrasound (pFUS) for the enhancement of drug uptake in prostate tumors in vivo using doxorubicin (Dox). METHODS:An antitumor drug Dox, an orthotopic animal prostate tumor model using human prostate cancer, LNCaP cell line, ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4705346
更新日期:2012-05-01 00:00:00
abstract::A beam stop technique was used to measure the densitometric scatter fractions under three regions of a humanoid chest phantom utilizing LaOBr and Gd2O2S screens. For these receptors, the scatter fractions under the lung and retrocardiac areas were 13%-36% lower than published values for CaWO4. In the mediastinal area,...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.597041
更新日期:1993-03-01 00:00:00
abstract:PURPOSE:The aiming is getting accurate liver contour structures automatically for following adaptive radiation therapy in daily CBCT images which is very low-contrast comparing the planning CT. METHODS:Probabilistic atlas is constructed by 50 intravenous contrast planning CT images by iterative affine registration pro...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4734730
更新日期:2012-06-01 00:00:00
abstract:PURPOSE:Total and regional body composition are important indicators of health and mortality risk, but their measurement is usually restricted to controlled environments in clinical settings with expensive and specialized equipment. A method that approaches the accuracy of the current gold standard method, dual-energy ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14492
更新日期:2020-12-01 00:00:00
abstract:PURPOSE:To commission and benchmark a vendor-supplied (Varian Medical Systems) Monte Carlo phase-space data for the 6 MV flattening filter free (FFF) energy mode on a TrueBeam linear accelerator for the purpose of quality assurance of clinical volumetric modulated arc therapy (VMAT) treatment plans. A method for render...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4773883
更新日期:2013-02-01 00:00:00
abstract:PURPOSE:Accurate segmentation of lungs with severe interstitial lung disease (ILD) in thoracic computed tomography (CT) is an important and difficult task in the development of computer-aided diagnosis (CAD) systems. Therefore, we developed in this study a texture analysis-based method for accurate segmentation of lung...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3222872
更新日期:2009-10-01 00:00:00
abstract:PURPOSE:To present a noninvasive technique for directly measuring the CT bow-tie filter attenuation with a linear array x-ray detector. METHODS:A scintillator based x-ray detector of 384 pixels, 307 mm active length, and fast data acquisition (model X-Scan 0.8c4-307, Detection Technology, FI-91100 Ii, Finland) was use...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4932364
更新日期:2015-11-01 00:00:00
abstract::Atomic Energy of Canada Limited 60Co teletherapy units have adjustable collimator hinges to adapt to source diameters from 0.75 to 2.50 cm. Adjustment of the collimator hinge to the manfacturer's settings can cause a loss of radiation field flatness as high as 20% for large fields. The recommended settings appear to c...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.594170
更新日期:1975-03-01 00:00:00
abstract:PURPOSE:Automatic breast ultrasound (ABUS) imaging has become an essential tool in breast cancer diagnosis since it provides complementary information to other imaging modalities. Lesion segmentation on ABUS is a prerequisite step of breast cancer computer-aided diagnosis (CAD). This work aims to develop a deep learnin...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14569
更新日期:2020-10-30 00:00:00
abstract:PURPOSE:To use EGSnrc Monte Carlo simulations for magnetic field dosimetry to determine optimal measurement orientations, calculate beam quality conversion factors for 32 cylindrical and three parallel-plate (PP) ion chambers, evaluate the beam quality and angular dependence of these factors, and examine the magnetic f...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.12716
更新日期:2018-02-01 00:00:00
abstract:PURPOSE:The authors discuss measurement methods and instrumentation useful for the characterization of the gray tracking performance of medical color monitors for diagnostic applications. The authors define gray tracking as the variability in the chromaticity of the gray levels in a color monitor. METHODS:The authors ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4953186
更新日期:2016-07-01 00:00:00
abstract::A doorless entry system has been designed for use in high-energy radiation therapy rooms. The doorless entry improves patient access, safety and throughput, increases the physical safety for the radiation therapists, and avoids the significant cost of heavy shielding doors. The design meets concerns relating to radiat...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.598181
更新日期:1998-02-01 00:00:00
abstract::In a practical classifier design problem, the true population is generally unknown and the available sample is finite-sized. A common approach is to use a resampling technique to estimate the performance of the classifier that will be trained with the available sample. We conducted a Monte Carlo simulation study to co...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2868757
更新日期:2008-04-01 00:00:00
abstract:PURPOSE:To characterize the radiation dose and three-dimensional (3D) imaging performance of a recently developed mobile, isocentric C-arm equipped with a flat-panel detector (FPD) for intraoperative cone-beam computed tomography (CBCT) (Cios Spin 3D, Siemens Healthineers) and to identify potential improvements in 3D i...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13983
更新日期:2020-03-01 00:00:00
abstract::Reconstruction of electron spectra of medical accelerators from measured depth dose distributions is an attractive tool for commissioning of a Monte Carlo treatment planning system. However, the reconstruction method is an inverse radiation transport problem which is poorly conditioned, in the sense it may become unst...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1461840
更新日期:2002-04-01 00:00:00
abstract:PURPOSE:The pulse line ion accelerator (PLIA) is a low-cost accelerator concept originally designed to accelerate heavy ions. Our group has been investigating the use of PLIA to accelerate light ions and believe a multi-stage PLIA could be useful for short half-life PET isotope production. The goal of this work was to ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13875
更新日期:2019-12-01 00:00:00
abstract:PURPOSE:To evaluate the image quality of virtual monochromatic images synthesized from dual-source dual-energy computed tomography (CT) in comparison with conventional polychromatic single-energy CT for the same radiation dose. METHODS:In dual-energy CT, besides the material-specific information, one may also synthesi...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3658568
更新日期:2011-12-01 00:00:00
abstract:PURPOSE:The aim of this research was to explore x-ray beam parameters with intent to optimize pediatric x-ray settings in the cardiac catheterization laboratory. This study examined the effects of peak x-ray tube voltage (kVp) and of copper (Cu) x-ray beam filtration independently on the image quality to dose balance f...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3488911
更新日期:2010-10-01 00:00:00
abstract::A restricted angular scattering model for electron penetration in dense media is presented. In the model, the Fermi-Eyges transport equation is modified through the addition of an extra term which may be interpreted as representing an apparent force opposing the scattering of electrons into wider angles. The introduct...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.596747
更新日期:1991-01-01 00:00:00
abstract:PURPOSE:Air-communicating well ionization chambers are commonly used to assess air kerma strength of sources used in brachytherapy. The signal produced is supposed to be proportional to the air density within the chamber and, therefore, a density-independent air kerma strength is obtained when the measurement is correc...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4831757
更新日期:2013-12-01 00:00:00
abstract::Techniques have been developed for reducing motion blurring artifacts by using respiratory gated computed tomography (CT) in sinogram space and quantitatively evaluating the artifact reduction. A synthetic sinogram was built from multiple scans intercepting a respiratory gating window. A gated CT image was then recons...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2074187
更新日期:2005-11-01 00:00:00