Use of simulated annealing for optimization of alignment parameters in limited MRI acquisition volumes of the brain.

Abstract:

:Studies suggest that clinical outcomes are improved in repeat trigeminal neuralgia (TN) Gamma Knife radiosurgery if a different part of the nerve from the previous radiosurgery is treated. The MR images taken in the first and repeat radiosurgery need to be coregistered to map the first radiosurgery volume onto the second treatment planning image. We propose a fully automatic and robust three-dimensional (3-D) mutual information- (MI-) based registration method engineered by a simulated annealing (SA) optimization technique. Commonly, Powell's method and Downhill simplex (DS) method are most popular in optimizing the MI objective function in medical image registration applications. However, due to the nonconvex property of the MI function, robustness of those two methods is questionable, especially for our cases, where only 28 slices of MR T1 images were utilized. Our SA method obtained successful registration results for all the 41 patients recruited in this study. On the other hand, Powell's method and the DS method failed to provide satisfactory registration for 11 patients and 9 patients, respectively. The overlapping volume ratio (OVR) is defined to quantify the degree of the partial volume overlap between the first and second MR scan. Statistical results from a logistic regression procedure demonstrated that the probability of a success of Powell's method tends to decrease as OVR decreases. The rigid registration with Powell's or the DS method is not suitable for the TN radiosurgery application, where OVR is likely to be low. In summary, our experimental results demonstrated that the MI-based registration method with the SA optimization technique is a robust and reliable option when the number of slices in the imaging study is limited.

journal_name

Med Phys

journal_title

Medical physics

authors

Li X,Zhang P,Brisman R,Kutcher G

doi

10.1118/1.1944287

subject

Has Abstract

pub_date

2005-07-01 00:00:00

pages

2363-2370

issue

7Part1

eissn

0094-2405

issn

2473-4209

journal_volume

32

pub_type

杂志文章
  • Measurement of primary bremsstrahlung spectrum from an 8-MeV linear accelerator.

    abstract::The bremsstrahlung spectrum from an 8-MeV linear accelerator has been measured using a NaI(T1) spectrometer system. The spectrum shows a low-energy cutoff at 0.4 MeV and the maximum photon energy to be approximately 6% greater than the nominal energy. The maximum emission of energy fluence was 1.6 and 1.8 MeV for meas...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594221

    authors: Levy LB,Waggener RG,Wright AE

    更新日期:1976-05-01 00:00:00

  • A voxel-based multiscale model to simulate the radiation response of hypoxic tumors.

    abstract:PURPOSE:In radiotherapy, it is important to predict the response of tumors to irradiation prior to the treatment. This is especially important for hypoxic tumors, which are known to be highly radioresistant. Mathematical modeling based on the dose distribution, biological parameters, and medical images may help to impr...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4903298

    authors: Espinoza I,Peschke P,Karger CP

    更新日期:2015-01-01 00:00:00

  • Photon-beam subsource sensitivity to the initial electron-beam parameters.

    abstract::One limitation to the widespread implementation of Monte Carlo (MC) patient dose-calculation algorithms for radiotherapy is the lack of a general and accurate source model of the accelerator radiation source. Our aim in this work is to investigate the sensitivity of the photon-beam subsource distributions in a MC sour...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1884385

    authors: Fix MK,Keall PJ,Siebers JV

    更新日期:2005-04-01 00:00:00

  • Quantitative impact of changes in marrow cellularity, skeletal size, and bone mineral density on active marrow dosimetry based upon a reference model.

    abstract:PURPOSE:The hematopoietically active tissues of skeletal bone marrow are a prime target for computational dosimetry given potential risks of leukemia and, at higher dose levels, acute marrow toxicity. The complex three-dimensional geometry of trabecular spongiosa, however, complicates schema for dose assessment in such...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12002

    authors: Geyer AM,Schwarz BC,Hobbs RF,Sgouros G,Bolch WE

    更新日期:2017-01-01 00:00:00

  • Angular measurement of the cobalt-60 emitted radiation spectrum from a radiosurgery irradiator.

    abstract::The photon energy spectrum emanating from a Leksell Gamma Knife, Model 23004B, was measured between 0.250 and 3.5 MeV with the sources exposed. Measurements were made using a 2x2 inch NaI detector enclosed in a lead-shielded apparatus having a 1/4 inch diameter measurement aperture, which reduced the amount of radiati...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1355305

    authors: Drzymala RE,Sohn JW,Guo C,Sobotka LG,Purdy JA

    更新日期:2001-04-01 00:00:00

  • Image reconstruction algorithm for a rotating slat collimator.

    abstract::A slat collimator in single photon emission computed tomography consists of a set of parallel slats. As the collimator spins, the detector measures a one-dimensional projection data set. A complete data set can be obtained by rotating the detector/collimator assembly around the object (patient) while the collimator sp...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1485057

    authors: Zeng GL,Gagnon D,Matthews CG,Kolthammer JA,Radachy JD,Hawkins WG

    更新日期:2002-07-01 00:00:00

  • Reduction of ring artifacts in CBCT: detection and correction of pixel gain variations in flat panel detectors.

    abstract:PURPOSE:In using flat panel detectors (FPD) for cone beam computed tomography (CBCT), pixel gain variations may lead to structured nonuniformities in projections and ring artifacts in CBCT images. Such gain variations can be caused by change in detector entrance exposure levels or beam hardening, and they are not accou...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4893278

    authors: Altunbas C,Lai CJ,Zhong Y,Shaw CC

    更新日期:2014-09-01 00:00:00

  • SU-E-T-283: Carbon Ion Therapy Innovations in Dosimetry and Dose Delivery.

    abstract:OBJECTIVE:Carbon ion particle therapy delivers sharp Bragg's peaks, and hence can produce dose delivery of very high precision and conformality. To complement this characteristic, particle therapy centers implement a number of technologies unique to these centers, and are not commonly known to the medical physics commu...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735351

    authors: Cernica G,Mun S

    更新日期:2012-06-01 00:00:00

  • Three-dimensional reconstruction of vascular trees. Theory and methodology.

    abstract::In this paper we examine the few-view reconstruction problem as it applies to imaging vascular trees. A fully automated reconstruction algorithm is described that circumvents the traditional "correspondence problem," using only notions of consistency and connectivity. It is assumed that the vascular tree is a connecte...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597704

    authors: Henri CJ,Peters TM

    更新日期:1996-02-01 00:00:00

  • Accuracy of a 3D laser∕camera surface imaging system for setup verification of the pelvic and thoracic regions in radiotherapy treatments.

    abstract:PURPOSE:The aim of this paper is to investigate the accuracy of a laser∕camera surface imaging system (Sentinel) when used on patients treated in the thorax or pelvic regions and to evaluate system employment conditions and patient setup procedures that provide more accurate results. METHODS:The system was tested on t...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4769428

    authors: Pallotta S,Simontacchi G,Marrazzo L,Ceroti M,Paiar F,Biti G,Bucciolini M

    更新日期:2013-01-01 00:00:00

  • SU-E-T-07: Edema Induced Changes in Tumor Cell Survival Fraction and Tumor Control Probability in 131Cs Permanent Prostate Implant Patients.

    abstract:PURPOSE:The purpose of this study is to estimate the effect of edema, developed during implant procedure, on tumor cell surviving fraction(SF) and tumor control probability(TCP) in the patients of prostate cancer who underwent 131 Cs permanent seed implants. METHODS:The impact of edema on SF and TCP, was calculated us...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735061

    authors: Kehwar TS,Huq MS,Smith RP

    更新日期:2012-06-01 00:00:00

  • Review and evaluation of MRI nonuniformity corrections for brain tumor response measurements.

    abstract::Current MRI nonuniformity correction techniques are reviewed and investigated. Many approaches are used to remedy this artifact, but it is not clear which method is the most appropriate in a given situation, as the applications have been with different MRI coils and different clinical applications. In this work four w...

    journal_title:Medical physics

    pub_type: 杂志文章,评审

    doi:10.1118/1.598357

    authors: Velthuizen RP,Heine JJ,Cantor AB,Lin H,Fletcher LM,Clarke LP

    更新日期:1998-09-01 00:00:00

  • A filtering method for signal equalization in region-of-interest fluoroscopy.

    abstract::A method to significantly reduce the exposure area product in fluoroscopy using a pre-patient region-of-interest (ROI) attenuator is presented. The attenuator has a thin central region and a gradually increasing thickness away from the center. It is shown that the unwanted brightening artifact caused by the attenuator...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1470205

    authors: Robert N,Komljenovic PT,Rowlands JA

    更新日期:2002-05-01 00:00:00

  • Technical Note: Transconvolution based equalization of positron energy effects for the use of 68 Ge/68 Ga phantoms in determining 18 F PET recovery.

    abstract:PURPOSE:Avoiding measurement variability from 18 F phantom preparation by using 68 Ge/68 Ga phantoms for the determination of 18 F recovery curves (RC) in clinical quality assurance measurements and for PET/CT site qualification in multicentre clinical trials. METHODS:RCs were obtained from PET/CT measurements of seve...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12330

    authors: Prenosil GA,Hentschel M,Fürstner M,Krause T,Weitzel T,Klaeser B

    更新日期:2017-07-01 00:00:00

  • Feasibility study of patient-specific energy verification using a multilayer acrylic-disk radiation sensor.

    abstract:PURPOSE:Obtaining an integral depth-dose (IDD) curve using a recently developed acrylic-disk radiation sensor (ADRS) is time-consuming because its single structure requires point-by-point measurements in a water phantom. The goal of this study was to verify the ability of a newly designed multilayer ADRS, composed of 2...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14326

    authors: Son J,Shin D,Kim T,Park S,Rah JE

    更新日期:2020-09-01 00:00:00

  • Task-based assessment of breast tomosynthesis: effect of acquisition parameters and quantum noise.

    abstract:PURPOSE:Tomosynthesis is a promising modality for breast imaging. The appearance of the tomosynthesis reconstructed image is greatly affected by the choice of acquisition and reconstruction parameters. The purpose of this study was to investigate the limitations of tomosynthesis breast imaging due to scan parameters an...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3357288

    authors: Reiser I,Nishikawa RM

    更新日期:2010-04-01 00:00:00

  • Addendum to the AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon beams.

    abstract::An addendum to the AAPM's TG-51 protocol for the determination of absorbed dose to water in megavoltage photon beams is presented. This addendum continues the procedure laid out in TG-51 but new kQ data for photon beams, based on Monte Carlo simulations, are presented and recommendations are given to improve the accur...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4866223

    authors: McEwen M,DeWerd L,Ibbott G,Followill D,Rogers DW,Seltzer S,Seuntjens J

    更新日期:2014-04-01 00:00:00

  • Surface imaging for real-time patient respiratory function assessment in intensive care.

    abstract:PURPOSE:Monitoring of physiological parameters is a major concern in Intensive Care Units (ICU) given their role in the assessment of vital organ function. Within this context, one issue is the lack of efficient noncontact techniques for respiratory monitoring. In this paper, we present a novel noncontact solution for ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14557

    authors: Nazir S,Pateau V,Bert J,Clement JF,Fayad H,l'Her E,Visvikis D

    更新日期:2020-10-28 00:00:00

  • Sci-Thurs PM: Delivery-03: Optical imaging of microscopic radiation dose gradients using a digital microscope.

    abstract::Using superior localization and immobilization methods, stereotactic radiosurgery is capable of delivering spheres of dose as small as a few millimetres in diameter to intracranial targets. For targets abutting critical structures, the most conformal treatments minimize adverse radiation side effects and it is importa...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2965910

    authors: Keller BM,Peressotti C,Pignol JP

    更新日期:2008-07-01 00:00:00

  • SU-E-I-38: Optimizing CT Slice Width Measurements.

    abstract:PURPOSE:For almost four decades bead and wire ramps have been used in the evaluation of slice width as part of QA testing for CT scanners. While each of these approaches have been recognized and accepted as reliable, in this paper we investigate the differences, advantages and limitations of these tools. Moreover, we s...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734753

    authors: Marteinsdóttir M,Olafsdottir H,Fredriksson J,Andresson A,Healy A,Kristbjornsson A,Kristinsson S,Levy J,Goodenough D

    更新日期:2012-06-01 00:00:00

  • A boundary element approach for image-guided near-infrared absorption and scatter estimation.

    abstract::Multimodality NIR spectroscopy systems offer the possibility of region-based vascular and molecular characterization of tissue in vivo. However, computationally efficient 3D image reconstruction algorithms specific to these image-guided systems currently do not exist. Image reconstruction is often based on finite-elem...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2795832

    authors: Srinivasan S,Pogue BW,Carpenter C,Yalavarthy PK,Paulsen K

    更新日期:2007-11-01 00:00:00

  • Incorporation of treatment plan spatial and temporal dose patterns into a prostate intrafractional motion management strategy.

    abstract:PURPOSE:Periodic MV∕KV radiographs taken during volumetric modulated arc therapy (VMAT) for hypofractionated treatment provide guidance in intrafractional motion management. The choice of imaging frequency and timing are key components in delivering the desired dose while reducing associated overhead such as imaging do...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4742846

    authors: Zhang P,Hunt M,Happersett L,Cox B,Mageras G

    更新日期:2012-09-01 00:00:00

  • Mammographic imaging with a small format CCD-based digital cassette: physical characteristics of a clinical system.

    abstract::The physical characteristics of a clinical charge coupled device (CCD)-based imager (Senovision, GE Medical Systems, Milwaukee, WI) for small-field digital mammography have been investigated. The imager employs a MinR 2000 (Eastman Kodak Company, Rochester, NY) scintillator coupled by a 1:1 optical fiber to a front-il...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1286720

    authors: Vedantham S,Karellas A,Suryanarayanan S,Levis I,Sayag M,Kleehammer R,Heidsieck R,D'Orsi CJ

    更新日期:2000-08-01 00:00:00

  • A comparison of the physical imaging properties of Fuji ST-V, ST-VA, and ST-VN computed radiography image plates.

    abstract::The physical imaging performance of ST-V, ST-VA, and ST-VN computed radiography image plates, read with a Philips AC3 acquisition system, was investigated at 70 kVp with 0.5 mm of copper beam filtration for four different entrance air kerma values: 0.5, 2.6, 26, and 260 micro Gy. Measurements included characteristic r...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598412

    authors: Kengyelics SM,Davies AG,Cowen AR

    更新日期:1998-11-01 00:00:00

  • Metal artifact correction for x-ray computed tomography using kV and selective MV imaging.

    abstract:PURPOSE:The overall goal of this work is to improve the computed tomography (CT) image quality for patients with metal implants or fillings by completing the missing kilovoltage (kV) projection data with selectively acquired megavoltage (MV) data that do not suffer from photon starvation. When both of these imaging sys...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4901551

    authors: Wu M,Keil A,Constantin D,Star-Lack J,Zhu L,Fahrig R

    更新日期:2014-12-01 00:00:00

  • Directional scatter imaging for the stereoscopic tracking of fiducial markers in a single kV exposure.

    abstract:PURPOSE:To demonstrate, via Monte Carlo simulation, that an image obtained from the patient-generated scattered radiation forced to impinge on the detector from a known direction by means of parallel-focused grids, can be used to complement the information conveyed by the primary image, such that accurate stereoscopic ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.12712

    authors: Garnica-Garza HM

    更新日期:2018-02-01 00:00:00

  • Focal spot motion of linear accelerators and its effect on portal image analysis.

    abstract::The focal spot of a linear accelerator is often considered to have a fully stable position. In practice, however, the beam control loop of a linear accelerator needs to stabilize after the beam is turned on. As a result, some motion of the focal spot might occur during the start-up phase of irradiation. When acquiring...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1576229

    authors: Sonke JJ,Brand B,van Herk M

    更新日期:2003-06-01 00:00:00

  • Optimization of radiography applications using x-ray beams emitted by compact accelerators. Part I. Monte Carlo study of the hard x-ray spectrum.

    abstract:PURPOSE:A 3-year project called LABSYNC has been recently funded by the European Commission, with the aim of designing a radiation facility based on a compact light source, i.e., a laboratory-sized commercial synchrotron, capable of accelerating electrons up to 6 or 20 MeV. An accurate spectral description of hard x ra...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3223357

    authors: Marziani M,Taibi A,Di Domenico G,Gambaccini M

    更新日期:2009-10-01 00:00:00

  • An automated dosimetry data-acquisition and analysis system at the LAMPF pion therapy facility.

    abstract::An automated data-acquisition and analysis system has been developed for dosimetry measurements on the pion therapy beam at the Clinton P. Anderson Meson Physics Facility Biomedical Channel in Los Alamos using a PDP-11/45 computer and CAMAC interface. Initialization, test, and monitor programs allow the user to set th...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.594418

    authors: Rosen I,Smith A,Lane R,Kelsey C,Lake D,Hogstrom K,Somers J,Helland J,Kittell R,Amols H,Bradbury J,Richman C

    更新日期:1978-03-01 00:00:00

  • Comparison of three strategies in management of independent movement of the prostate and pelvic lymph nodes.

    abstract:PURPOSE:Concurrent irradiation of the prostate and pelvic lymph nodes is technically challenging due to treating one moving target and one immobile target. The purposes of this article are to propose a new management strategy and to compare this strategy to the conventional isocenter shift method and the previously pro...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3480505

    authors: Xia P,Qi P,Hwang A,Kinsey E,Pouliot J,Roach M 3rd

    更新日期:2010-09-01 00:00:00