SU-E-T-162: Rapid Assessment of Cyberknife IRIS Variable Collimator Performance Using a Commercial Diode Array.

Abstract:

PURPOSE:To describe an efficient method for verifying the size and centering of the radiation field from the Cyberknife IRIS variable collimator with sub-millimeter accuracy using a general purpose commercially available diode array. METHODS:We present a technique using a conventional diode array (Sun Nuclear Profiler) with the array at an extended distance of 320 cm. The projection of the 4 mm diode spacing back to the 80 cm field definition distance gives an effective spacing of 1 mm, sufficient to confirm proper operation of the IRIS. We describe the data acquisition process and present data comparing the Profiler measurements to scanned measurements for both profile and FWHM analysis and reproducibility of the technique over repeated measurements. RESULTS:Average difference between original water scanner measurements and diode array measurements over the 12 aperture sizes) from 5 mm to 60 mm) were - 0.14 mm (range 0.03 mm to 0.83 mm). Reproducibility and centering measurements had a similar range of accuracy. CONCLUSIONS:A general purpose commercially available diode array can be used to quickly and accurately characterize the field size and centering of the Cyberknife IRIS variable collimator system with sub-millimeter accuracy subsequent to service, software recalibration, software upgrades or associated with routine QA. This technique avoids the time consuming and cumbersome water tank scanning with a diode and the difficulties associated with image based measurements (CR or radiochromic film) that require time consuming and careful calibration and choice of threshold values.

journal_name

Med Phys

journal_title

Medical physics

authors

Gibbs G,White G,Bulz G

doi

10.1118/1.4735220

subject

Has Abstract

pub_date

2012-06-01 00:00:00

pages

3740

issue

6Part12

eissn

0094-2405

issn

2473-4209

journal_volume

39

pub_type

杂志文章
  • Characterization of mesothelioma and tissues present in contrast-enhanced thoracic CT scans.

    abstract:PURPOSE:The purpose of this study was to characterize the Hounsfield unit (HU) distributions of mesothelioma and other tissues present in contrast-enhanced thoracic CT scans, to compare the HU distributions of mesothelioma, muscle, and liver by scanner and reconstruction filter/kernel combination, and to assess interpa...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3537610

    authors: Corson N,Sensakovic WF,Straus C,Starkey A,Armato SG 3rd

    更新日期:2011-02-01 00:00:00

  • Objected constrained registration and manifold learning: a new patient setup approach in image guided radiation therapy of thoracic cancer.

    abstract:PURPOSE:The management of thoracic malignancies with radiation therapy is complicated by continuous target motion. In this study, a real time motion analysis approach is proposed to improve the accuracy of patient setup. METHODS:For 11 lung cancer patients a long training fluoroscopy was acquired before the first trea...

    journal_title:Medical physics

    pub_type: 临床试验,杂志文章

    doi:10.1118/1.4794489

    authors: Chen T,Jabbour SK,Qin S,Haffty BG,Yue N

    更新日期:2013-04-01 00:00:00

  • SU-E-T-181: Development of a Single-Marker-Based Optical Tracking System to Enhance Patient Safety in Radiotherapy.

    abstract:PURPOSE:To develop an accurate and efficient single-marker-based optical tracking system to enhance patient safety and quality of care in radiotherapy. METHODS:The optical tracking system consists of a Polaris camera as well as in-house software. The camera is mounted inside the treatment room and detects the location...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735240

    authors: Yan G,Lu B,Mittauer K,Liu C,Li J

    更新日期:2012-06-01 00:00:00

  • Development of a dynamic quality assurance testing protocol for multisite clinical trial DCE-CT accreditation.

    abstract:PURPOSE:Credentialing can have an impact on whether or not a clinical trial produces useful quality data that is comparable between various institutions and scanners. With the recent increase of dynamic contrast enhanced-computed tomography (DCE-CT) usage as a companion biomarker in clinical trials, effective quality a...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4812429

    authors: Driscoll B,Keller H,Jaffray D,Coolens C

    更新日期:2013-08-01 00:00:00

  • Sparse and optimal acquisition design for diffusion MRI and beyond.

    abstract:PURPOSE:Diffusion magnetic resonance imaging (MRI) in combination with functional MRI promises a whole new vista for scientists to investigate noninvasively the structural and functional connectivity of the human brain-the human connectome, which had heretofore been out of reach. As with other imaging modalities, diffu...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3700166

    authors: Koay CG,Ozarslan E,Johnson KM,Meyerand ME

    更新日期:2012-05-01 00:00:00

  • An evaluation of ionization chambers for the relative dosimetry of kilovoltage x-ray beams.

    abstract::In this work, the authors have evaluated ten different ionization chambers for the relative dosimetry of kilovoltage x-ray beams in the energy range of 50-280 kVp. Percentage depth doses in water and relative detector response (in Solid Water and in air) were measured for each of the x-ray beams studied using a number...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3183820

    authors: Hill R,Mo Z,Haque M,Baldock C

    更新日期:2009-09-01 00:00:00

  • Electromagnetic guided couch and multileaf collimator tracking on a TrueBeam accelerator.

    abstract:PURPOSE:Couch and MLC tracking are two promising methods for real-time motion compensation during radiation therapy. So far, couch and MLC tracking experiments have mainly been performed by different research groups, and no direct comparison of couch and MLC tracking of volumetric modulated arc therapy (VMAT) plans has...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4946815

    authors: Hansen R,Ravkilde T,Worm ES,Toftegaard J,Grau C,Macek K,Poulsen PR

    更新日期:2016-05-01 00:00:00

  • Characteristic image quality of a third generation dual-source MDCT scanner: Noise, resolution, and detectability.

    abstract:PURPOSE:The purpose of this work was to assess the inherent image quality characteristics of a new multidetector computed tomography system in terms of noise, resolution, and detectability index as a function of image acquisition and reconstruction for a range of clinically relevant settings. METHODS:A multisized imag...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4923172

    authors: Solomon J,Wilson J,Samei E

    更新日期:2015-08-01 00:00:00

  • The use of a diode matrix in commissioning activities for electron energies > or = 9 MeV: a feasibility study.

    abstract::The contribution of a commercially available diode matrix (MapCHECK, provided by Sun Nuclear, Melbourne, FL) for the commissioning procedures of the voxel based Monte Carlo (VMC++) algorithm for electron beams of MasterPlan treatment planning system was investigated. The attention is mainly focused on the calculation ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3081414

    authors: Borca VC,Pasquino M,Ozzello F,Tofani S

    更新日期:2009-04-01 00:00:00

  • Robust initialization of 2D-3D image registration using the projection-slice theorem and phase correlation.

    abstract:PURPOSE:The image registration literature comprises many methods for 2D-3D registration for which accuracy has been established in a variety of applications. However, clinical application is limited by a small capture range. Initial offsets outside the capture range of a registration method will not converge to a succe...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3366252

    authors: van der Bom MJ,Bartels LW,Gounis MJ,Homan R,Timmer J,Viergever MA,Pluim JP

    更新日期:2010-04-01 00:00:00

  • Technical Note: Compact three-tesla magnetic resonance imager with high-performance gradients passes ACR image quality and acoustic noise tests.

    abstract:PURPOSE:A compact, three-tesla magnetic resonance imaging (MRI) system has been developed. It features a 37 cm patient aperture, allowing the use of commercial receiver coils. Its design allows simultaneously for gradient amplitudes of 85 millitesla per meter (mT/m) sustained and 700 tesla per meter per second (T/m/s) ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4941362

    authors: Weavers PT,Shu Y,Tao S,Huston J 3rd,Lee SK,Graziani D,Mathieu JB,Trzasko JD,Foo TK,Bernstein MA

    更新日期:2016-03-01 00:00:00

  • Comparison of two independent exposure measurement techniques for clinical I-125 seeds.

    abstract::This investigation contains the results of exposure measurements on two types of clinical I-125 seeds: a model 6702 no-marker seed, and a model 6711 silver wire seed. Measurements were made locally with a parallel-plate mammography chamber, traceable to the National Bureau of Standards (NBS) calibration. The same seed...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.595784

    authors: Kubo H

    更新日期:1985-03-01 00:00:00

  • Dose planning with comparison to in vivo dosimetry for epithermal neutron irradiation of the dog brain.

    abstract::Boron neutron capture therapy (BNCT) is an experimental type of radiotherapy, presently being used to treat glioblastoma and melanoma. To improve patient safety and to determine the radiobiological characteristics of the epithermal neutron beam of Finnish BNCT facility (FiR 1) dose-response studies were carried on the...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1517048

    authors: Seppälä T,Auterinen I,Aschan C,Serén T,Benczik J,Snellman M,Huiskamp R,Ramadan UA,Kankaanranta L,Joensuu H,Savolainen S

    更新日期:2002-11-01 00:00:00

  • SU-F-BRCD-09: Total Variation (TV) Based Fast Convergent Iterative CBCT Reconstruction with GPU Acceleration.

    abstract:PURPOSE:To improve image quality and reduce imaging dose in CBCT for radiation therapy applications and to realize near real-time image reconstruction based on use of a fast convergence iterative algorithm and acceleration by multi-GPUs. METHODS:An iterative image reconstruction that sought to minimize a weighted leas...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4735747

    authors: Xu Q,Yang D,Tan J,Anastasio M

    更新日期:2012-06-01 00:00:00

  • A double mirror W/C multilayer monochromator for radiation biology applications.

    abstract::A double-mirror multilayer monochromator was developed for the purpose of irradiating live cell cultures at the Synchrotron Radiation Center, University of Wisconsin-Madison. The monochromator is designed for the soft x-ray region with photon energies between 270 and 2400 eV. Multilayer mirrors with 55 bilayers of W/C...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.598259

    authors: MacKay JF,Pearson DW,Nelms BE,DeLuca PM Jr,Gould MN,Lagally MG

    更新日期:1998-05-01 00:00:00

  • An automated procedure for the quality assurance of electron beam output and energy.

    abstract::In this article, we report on the development of a simple and accurate method for quality assurance of electron beam output and energy. Aluminum disks of thicknesses d(max) or d50 of the particular electron energy are positioned sequentially over a parallel-plate ion chamber and the ratio of the two signals is compare...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1688040

    authors: Woo MK,Videla NG

    更新日期:2004-05-01 00:00:00

  • Contrast enhancement of portal images by selective histogram equalization.

    abstract::Because of the high energy of the treatment beam, contrast of portal verification films is very poor. A simple contrast enhancement technique is described which we have labeled selective histogram equalization (SHE), to improve visualization of double-exposure portal images and this facilitate the beam verification pr...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597085

    authors: Crooks I,Fallone BG

    更新日期:1993-01-01 00:00:00

  • X-ray imaging with amorphous selenium: X-ray to charge conversion gain and avalanche multiplication gain.

    abstract::Fluoroscopy is a low dose imaging technique. As such, a very sensitive detector is required to create images of good quality. Present day flat panel active matrix read out systems introduce an amount of noise that inhibits present direct and indirect methods from producing optimal quality images at fluoroscopic exposu...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.1513157

    authors: Hunt DC,Kirby SS,Rowlands JA

    更新日期:2002-11-01 00:00:00

  • 3D modeling of effects of increased oxygenation and activity concentration in tumors treated with radionuclides and antiangiogenic drugs.

    abstract:PURPOSE:Formation of new blood vessels (angiogenesis) in response to hypoxia is a fundamental event in the process of tumor growth and metastatic dissemination. However, abnormalities in tumor neovasculature often induce increased interstitial pressure (IP) and further reduce oxygenation (pO2) of tumor cells. In radiot...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.3615164

    authors: Lagerlöf JH,Kindblom J,Bernhardt P

    更新日期:2011-08-01 00:00:00

  • Performance characteristics of transmission imaging using a uniform sheet source with parallel-hole collimation.

    abstract::Transmission imaging is receiving increasing attention in SPECT due to the need to compensate for nonuniform attenuation in cardiac-chest SPECT. The quality of a transmission image has an important effect on the measured attenuation distribution. To improve image quality, knowledge of the performance characteristics o...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.596752

    authors: Cao Z,Tsui BM

    更新日期:1992-09-01 00:00:00

  • B-spline active rays segmentation of microcalcifications in mammography.

    abstract::Accurate segmentation of microcalcifications in mammography is crucial for the quantification of morphologic properties by features incorporated in computer-aided diagnosis schemes. A novel segmentation method is proposed implementing active rays (polar-transformed active contours) on B-spline wavelet representation t...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2991286

    authors: Arikidis NS,Skiadopoulos S,Karahaliou A,Likaki E,Panayiotakis G,Costaridou L

    更新日期:2008-11-01 00:00:00

  • Active energy selective image detector for dual-energy computed radiography.

    abstract::A new energy selective detector for dual energy computed radiography has been developed that combines many of the advantages of x-ray tube voltage switching and single exposure double screen detectors. The new active detector utilizes electro-optical modulation of the response of the storage phosphor screens to allow ...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597831

    authors: Alvarez RE

    更新日期:1996-10-01 00:00:00

  • Comparison of measured and estimated maximum skin doses during CT fluoroscopy lung biopsies.

    abstract:PURPOSE:To measure patient-specific maximum skin dose (MSD) associated with CT fluoroscopy (CTF) lung biopsies and to compare measured MSD with the MSD estimated from phantom measurements, as well as with the CTDIvol of patient examinations. METHODS:Data from 50 patients with lung lesions who underwent a CT fluoroscop...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4884231

    authors: Zanca F,Jacobs A,Crijns W,De Wever W

    更新日期:2014-07-01 00:00:00

  • Maximum likelihood estimation of proton irradiated field and deposited dose distribution.

    abstract::In proton therapy, it is important to evaluate the field irradiated with protons and the deposited dose distribution in a patient's body. Positron emitters generated through fragmentation reactions of target nuclei can be used for this purpose. By detecting the annihilation gamma rays from the positron emitters, the a...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2712572

    authors: Inaniwa T,Kohno T,Yamagata F,Tomitani T,Sato S,Kanazawa M,Kanai T,Urakabe E

    更新日期:2007-05-01 00:00:00

  • ARPM-net: A novel CNN-based adversarial method with Markov random field enhancement for prostate and organs at risk segmentation in pelvic CT images.

    abstract:PURPOSE:The research is to develop a novel CNN-based adversarial deep learning method to improve and expedite the multi-organ semantic segmentation of CT images and to generate accurate contours on pelvic CT images. METHODS:Planning CT and structure datasets for 120 patients with intact prostate cancer were retrospect...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1002/mp.14580

    authors: Zhang Z,Zhao T,Gay H,Zhang W,Sun B

    更新日期:2020-11-05 00:00:00

  • Suitability of laser stimulated TLD arrays as patient dose monitors in high dose x-ray imaging.

    abstract::Skin entrance doses of patients undergoing interventional x-ray procedures are capable of causing skin damage and should be monitored routinely. Single TLD chips are not suitable because the location of maximum skin exposure cannot be predicted. Most photographic films are too sensitive at diagnostic x-ray energies fo...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597973

    authors: Geise RA,Schueler BA,Lien W,Jones SC

    更新日期:1997-10-01 00:00:00

  • SU-C-218-06: Evaluation of a CT-Based, Semi-Automated Lung Mass Estimation Method Under Varying Acquisition Conditions.

    abstract:PURPOSE:To evaluate a simple, semi-automated lung mass estimation method on CT scans from a variety of acquisition techniques for mass correction of MIRD dose estimates. METHODS:CT scans from ten patients undergoing stereotactic body radiation therapy treatment planning on a GE DST PET/CT scanner were analyzed retrosp...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4734654

    authors: Busse N,Erwin W,Pan T

    更新日期:2012-06-01 00:00:00

  • The stability of mechanical calibration for a kV cone beam computed tomography system integrated with linear accelerator.

    abstract::The geometric accuracy and precision of an image-guided treatment system were assessed. Image guidance is performed using an x-ray volume imaging (XVI) system integrated with a linear accelerator and treatment planning system. Using an amorphous silicon detector and x-ray tube, volumetric computed tomography images ar...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.2143141

    authors: Sharpe MB,Moseley DJ,Purdie TG,Islam M,Siewerdsen JH,Jaffray DA

    更新日期:2006-01-01 00:00:00

  • Noise, sampling, and the number of projections in cone-beam CT with a flat-panel detector.

    abstract:PURPOSE:To investigate the effect of the number of projection views on image noise in cone-beam CT (CBCT) with a flat-panel detector. METHODS:This fairly fundamental consideration in CBCT system design and operation was addressed experimentally (using a phantom presenting a uniform medium as well as statistically moti...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.4875688

    authors: Zhao Z,Gang GJ,Siewerdsen JH

    更新日期:2014-06-01 00:00:00

  • A general model of microcirculatory blood flow effects in gradient sensitized MRI.

    abstract::A general expression is derived for the NMR signal from a fluid undergoing random directional flow such as encountered within the microcirculation. The dependence of the echo amplitude on flow velocity, sample morphology, and experimental parameters are described in terms of a temporal velocity autocorrelation functio...

    journal_title:Medical physics

    pub_type: 杂志文章

    doi:10.1118/1.597170

    authors: Kennan RP,Gao JH,Zhong J,Gore JC

    更新日期:1994-04-01 00:00:00