What Is the Optimal Setting for a Continuous-Flow Left Ventricular Assist Device in Severe Mitral Regurgitation?

Abstract:

:Excessive left ventricular (LV) volume unloading can affect right ventricular (RV) function by causing a leftward shift of the interventricular septum in patients with mitral regurgitation (MR) receiving left ventricular assist device (LVAD) support. Optimal settings for the LVAD should be chosen to appropriately control the MR without causing RV dysfunction. In this study, we assessed the utility of our electrocardiogram-synchronized rotational speed (RS) modulation system along with a continuous-flow LVAD in a goat model of MR. We implanted EVAHEART devices after left thoracotomy in six adult goats weighing 66.4 ± 10.7 kg. Severe MR was induced through inflation of a temporary inferior vena cava filter placed within the mitral valve. We evaluated total flow (TF; the sum of aortic flow and pump flow [PF]), RV fractional area change (RVFAC) calculated by echocardiography, left atrial pressure (LAP), LV end-diastolic pressure (LVEDP), LV end-diastolic volume (LVEDV), and LV stroke work (LVSW) with a bypass rate (PF divided by TF) of 100% under four conditions: circuit-clamp, continuous mode, co-pulse mode (increased RS during systole), and counter-pulse mode (increased RS during diastole). TF tended to be higher in the counter-pulse mode. Moreover, RVFAC was significantly higher in the counter-pulse mode than in the co-pulse mode, whereas LAP was significantly lower in all driving modes than in the circuit-clamp condition. Furthermore, LVEDP, LVEDV, and LVSW were significantly lower in the counter-pulse mode than in the circuit-clamp condition. The counter-pulse mode of our RS modulation system used with a continuous-flow LVAD may offer favorable control of MR while minimizing RV dysfunction.

journal_name

Artif Organs

journal_title

Artificial organs

authors

Naito N,Nishimura T,Takewa Y,Kishimoto S,Date K,Umeki A,Ando M,Ono M,Tatsumi E

doi

10.1111/aor.12702

subject

Has Abstract

pub_date

2016-11-01 00:00:00

pages

1039-1045

issue

11

eissn

0160-564X

issn

1525-1594

journal_volume

40

pub_type

杂志文章
  • Arterial pressure: its role in perfusion and as a criterion of extracorporeal circulation adequacy.

    abstract::Thirty-five patients undergoing open-heart surgery are reported. Cardiac index, mean arterial pressure, oxygen consumption, lactate, pyruvate, and lactate excess in the blood were measured. An inverse correlation of arterial pressure with oxygen consumption and lactate excess was found during perfusion. Comparison of ...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1982.tb01675.x

    authors: Osipov VP,Khodas MJ,Dementjeva II,Pjatnitskaja GKh

    更新日期:1982-08-01 00:00:00

  • HeartQuest ventricular assist device magnetically levitated centrifugal blood pump.

    abstract::Improvements in implantable ventricular assist device (VAD) performance will be required to obtain patient outcomes that are comparable with those of heart transplantation. The HeartQuest VAD (WorldHeart, Oakland, CA, U.S.A.) is an advanced device, with full magnetic suspension of the rotor, designed to address specif...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2006.00223.x

    authors: Bearnson GB,Jacobs GB,Kirk J,Khanwilkar PS,Nelson KE,Long JW

    更新日期:2006-05-01 00:00:00

  • Further development of hybrid functional electrical stimulation orthoses.

    abstract::In this study two aspects of hybrid functional electrical stimulation (FES) orthoses were investigated: joint motion constraints and FES control strategies. First, the effects of joint motion constraints on the gait of normal subjects were investigated using modern motion analysis systems, including electromyogram (EM...

    journal_title:Artificial organs

    pub_type: 临床试验,杂志文章

    doi:10.1111/j.1525-1594.1997.tb04648.x

    authors: Yang L,Granat MH,Paul JP,Condie DN,Rowley DI

    更新日期:1997-03-01 00:00:00

  • Resection of huge retrosternal goiter through a novel combined cervical and robot-assisted approach.

    abstract::A 53-year-old man was referred to our department with a history of cough, especially at night, for 3 months. He was found to have a huge retrosternal goiter. We applied a novel combined cervical and robot-assisted approach to resect this difficult goiter. Using the da Vinci robot system, the intrathoracic part of the ...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12196

    authors: Wang S,Xu S,Liu B

    更新日期:2014-05-01 00:00:00

  • The human genome efforts and the cDNA project.

    abstract::Molecular biology has been moving swiftly toward clarifying the entire genome of organisms, including human. The human genome efforts that promote this transaction are characterized by the large-scale, high throughput production of data about the structure of the genome of the molecular level and its computer-assisted...

    journal_title:Artificial organs

    pub_type: 杂志文章,评审

    doi:10.1111/j.1525-1594.1996.tb04553.x

    authors: Matsubara K

    更新日期:1996-08-01 00:00:00

  • Diabetes-induced alterations in latissimus dorsi muscle properties impair effectiveness of dynamic cardiomyoplasty in rats.

    abstract::Short-term diabetes was induced in male Wistar rats with streptozotocin injection. The effects of diabetes on latissimus dorsi (LD) muscle contractile and biochemical properties and acute cardiomyoplasty (CDM) were assessed and compared with data from 16 control rats. Isometric force, contractile properties, and fatig...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2004.47349.x

    authors: De Angelis K,Senna S,Irigoyen MC,Cestari IA

    更新日期:2004-04-01 00:00:00

  • Outcome analysis for prediction of intraaortic balloon pump support failure and long-term survival in high-risk patients undergoing mitral valve surgery.

    abstract::The outcome predictors of intra-aortic balloon pump (IABP) in patients who undergo mitral valve surgery remain unknown. This study aimed to retrospectively review valvular surgery in patients who received an IABP to identify the predictors of failure of IABP support and anticipate the necessary therapy. This retrospec...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.13669

    authors: Liang M,Wang C,Feng K,Chen G,Wang K,Wu Z

    更新日期:2020-08-01 00:00:00

  • Hemodynamic Evaluation of a Biological and Mechanical Aortic Valve Prosthesis Using Patient-Specific MRI-Based CFD.

    abstract::Modeling different treatment options before a procedure is performed is a promising approach for surgical decision making and patient care in heart valve disease. This study investigated the hemodynamic impact of different prostheses through patient-specific MRI-based CFD simulations. Ten time-resolved MRI data sets w...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12955

    authors: Hellmeier F,Nordmeyer S,Yevtushenko P,Bruening J,Berger F,Kuehne T,Goubergrits L,Kelm M

    更新日期:2018-01-01 00:00:00

  • Estimation of the native cardiac output from a rotary blood pump flow: in vitro study.

    abstract::The rotary blood pump will be an implantable left ventricular assist device (LVAD) in the near future. However, the best control method and the interrelationship between the rotary blood pump and native heart functions are unclear. An estimation was made of the native heart cardiac output from the change of an LVAD's ...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.1998.06127.x

    authors: Nakata K,Ohashi Y,Tayama E,Ohtsuka G,Takami Y,Mueller J,Glueck J,Nosé Y

    更新日期:1998-05-01 00:00:00

  • Anticoagulation with fondaparinux for hemodiafiltration in patients with heparin-induced thrombocytopenia: dose-finding study and safety evaluation.

    abstract::The optimal anticoagulation regimen for hemodialysis (HD) in patients with heparin-induced thrombocytopenia (HIT) has not been defined. Hemodiafiltration (HDF) adds a large convective component to HD, thereby changing the pharmacokinetics of most anticoagulants. Data on coagulation regimens for HDF are scant. We there...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12002

    authors: Mahieu E,Claes K,Jacquemin M,Evenepoel P,Op De Beek K,Bogaert AM,Kuypers D,Verhamme P,Meijers B

    更新日期:2013-05-01 00:00:00

  • Antimicrobial Wound Dressing Containing Silver Sulfadiazine With High Biocompatibility: In Vitro Study.

    abstract::Many patients all over the world suffer from acute wounds caused by traumas or burns. In most crucial cases, skin regeneration cannot be promoted spontaneously, and skin grafts are applied as the main treatment. However, this therapy has some drawbacks which motivate researchers to develop wound dressings. In this stu...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12682

    authors: Mohseni M,Shamloo A,Aghababaei Z,Vossoughi M,Moravvej H

    更新日期:2016-08-01 00:00:00

  • Glucose-insulin kinetics of a bioartificial pancreas made of an AN69 hydrogel hollow fiber containing porcine islets and implanted in diabetic mice.

    abstract::The goal of this study was to determine whether porcine islets encapsulated in hollow fibers made of AN69 copolymer can correct hyperglycemia in diabetic mice and provide normal tolerance to a glucose challenge. In vitro perifusion of hollow fibers demonstrated the rapid kinetics of insulin release in response to gluc...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.1998.05087.x

    authors: Delaunay C,Darquy S,Honiger J,Capron F,Rouault C,Reach G

    更新日期:1998-04-01 00:00:00

  • Intravascular micropump for augmented liver perfusion: first in vivo experience.

    abstract::The aim of this study was to assess the in vivo performance of a new microaxial rotary blood pump developed for long-term intraportal implantation. The pump, measuring 7 mm in diameter, has a single stage impeller and is powered by a microelectric motor. The pump was implanted into the portal vein in 13 large white pi...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.2001.06772.x

    authors: Havlík R,Kerkhoffs W,Jiao LR,Schumacher O,Reul H,Habib N

    更新日期:2001-05-01 00:00:00

  • Effectiveness of a New Single-Needle Single-Pump Dialysis System with Simultaneous Monitoring of Dialysis Dose.

    abstract::To date, single-needle (SN) hemodialysis (HD) requires a dialysis machine equipped with two blood pumps-one controlling arterial blood flow (Qb) and one controlling venous Qb. B. Braun has developed an innovative single-pump SN HD system. Therefore, usability is improved by reducing complexity. The aim of this study w...

    journal_title:Artificial organs

    pub_type: 杂志文章,随机对照试验

    doi:10.1111/aor.13149

    authors: Bieser W,Welsch M,Jakob M,Meibaum J,Rodriguez DA,Wolff H,Marcelli D,Barth C

    更新日期:2018-08-01 00:00:00

  • Circulatory loop design and components introduce artifacts impacting in vitro evaluation of ventricular assist device thrombogenicity: A call for caution.

    abstract::Mechanical circulatory support (MCS) devices continue to be hampered by thrombotic adverse events (AEs), a consequence of device-imparted supraphysiologic shear stresses, leading to shear-mediated platelet activation (SMPA). In advancing MCS devices from design to clinical use, in vitro circulatory loops containing th...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.13626

    authors: Li M,Walk R,Roka-Moiia Y,Sheriff J,Bluestein D,Barth EJ,Slepian MJ

    更新日期:2020-06-01 00:00:00

  • Magnevad status of design improvements human blood results and preliminary sheep trial.

    abstract::Magnevad II is an improved version of Magnevad I and both are fourth-generation axial flow left ventricular assist devices. A major simplification has been implemented and tested. Magnevad I used a magnetically suspended rotor requiring active axial control. Magnevad II is completely passive but otherwise similar. A s...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2005.29113.x

    authors: Goldowsky M,Lafaro R,Reed G

    更新日期:2005-10-01 00:00:00

  • Experimental study on shear stress distributions in a centrifugal blood pump.

    abstract::Wall shear stress on the pump casing cover was measured using a surface-mounted hot-film sensor. In addition, the shear stress distribution in the pump was qualitatively investigated by means of oil-film visualization. The characteristics of shear stress in the pump are discussed, including the results on the oil-film...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2007.00421.x

    authors: Mizunuma H,Nakajima R

    更新日期:2007-07-01 00:00:00

  • Renal osteodystrophy in dialysis patients: diagnosis and treatment.

    abstract::This article reviews the clinical, biological, radiological, and pathological procedures and their respective indications for the practical diagnosis of the following various histological patterns of renal osteodystrophy: osteitis fibrosa due to parathyroid hormone (PTH) hypersecretion: osteomalacia or rickets due to ...

    journal_title:Artificial organs

    pub_type: 杂志文章,评审

    doi:10.1046/j.1525-1594.1998.06198.x

    authors: Fournier A,Oprisiu R,Hottelart C,Yverneau PH,Ghazali A,Atik A,Hedri H,Said S,Sechet A,Rasolombololona M,Abighanem O,Sarraj A,El Esper N,Moriniere P,Boudailliez B,Westeel PF,Achard JM,Pruna A

    更新日期:1998-07-01 00:00:00

  • Bacterial and fungal infection in neonates undergoing venoarterial extracorporeal membrane oxygenation: an analysis of the registry data of the extracorporeal life support organization.

    abstract::A review and analysis of 5,001 neonatal venoarterial (VA) extracorporeal membrane oxygenation (ECMO) cases showed that bacterial and fungal infection occurred in 147 (2.9%) and 26 (0.6%) patients, respectively, with an overall incidence of 3.5%. Bivariate analysis was used to compare infected infants with controls, ba...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1996.tb04428.x

    authors: Douglass BH,Keenan AL,Purohit DM

    更新日期:1996-03-01 00:00:00

  • Evaluation of double filtration plasmapheresis, thermofiltration, and low-density lipoprotein adsorptive methods by crossover test in the treatment of familial hypercholesterolemia patients.

    abstract::A comparative assessment has been made regarding efficacy and safety of the double filtration plasmapheresis (DFPP), thermofiltration (TFPP), and low-density lipoprotein (LDL) adsorptive (PA) methods by making a crossover test on heterozygous familial hypercholesterolemia patients. Treatments by DFPP, TFPP (secondary ...

    journal_title:Artificial organs

    pub_type: 临床试验,杂志文章

    doi:10.1111/j.1525-1594.1996.tb04448.x

    authors: Suzuki M,Yamane S,Matsugane T,Nobuto T,Azuma N,Nishide T,Shinomiya M,Saito K,Sasaki N,Nosé Y

    更新日期:1996-04-01 00:00:00

  • Effects of scaling on centrifugal blood pumps.

    abstract::Experimental studies on the effects of scaling on the performance of centrifugal blood pumps were conducted in a closed-loop test rig. For the prototype, eight different impellers of the same outer diameter of 25 mm were tested at 1,500, 2,000, and 2,500 revolutions per minute (rpm) using blood analog as fluid medium....

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.2002.07119.x

    authors: Wong YW,Chan WK,Yu SC,Chua LP

    更新日期:2002-11-01 00:00:00

  • Quantification of the secondary flow in a radial coupled centrifugal blood pump based on particle tracking velocimetry.

    abstract::Secondary flow in the centrifugal blood pump helps to enhance the washout effect and to minimize thrombus formation. On the other hand, it has an adverse effect on pump efficiency. Excessive secondary flow may induce hemolytic effects. Understanding the secondary flow is thus important to the design of a compact, effi...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2004.29012.x

    authors: Watanabe N,Masuda T,Iida T,Kataoka H,Fujimoto T,Takatani S

    更新日期:2005-01-01 00:00:00

  • Experiencing Functional Electrical Stimulation Roots on Education, and Clinical Developments in Paraplegia and Tetraplegia With Technological Innovation.

    abstract::Cybernetics-based concepts can allow for complete independence for paralyzed individuals, including sensory motor recovery. Spinal cord injuries are responsible for a huge stress on health and a financial burden to society. This article focuses on novel procedures such as functional diagnosis for paraplegics and tetra...

    journal_title:Artificial organs

    pub_type: 杂志文章,评审

    doi:10.1111/aor.12620

    authors: Varoto R,Cliquet A Jr

    更新日期:2015-10-01 00:00:00

  • Termination of extracorporeal membrane oxygenation for cardiac support.

    abstract::The determination of when to stop extracorporeal membrane oxygenation (ECMO) rests upon demonstration of the return of adequate cardiac function to support vital organs and permit subsequent recovery. In general, patients with myocardial stun will recover function within several days. Factors that limit recovery inclu...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1046/j.1525-1594.1999.06445.x

    authors: Steinhorn DM

    更新日期:1999-11-01 00:00:00

  • Technical feasibility of umbilical cannulation in midgestation lambs supported by the EXTra-uterine Environment for Neonatal Development (EXTEND).

    abstract::EXTEND (EXTra-uterine Environment for Neonatal Development) is a novel system for supporting extremely premature infants that replicates in utero conditions by maintaining a sterile fluid environment and providing gas exchange via a pumpless arteriovenous oxygenator circuit connected to the umbilical vessels. Target g...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.13524

    authors: Hornick MA,Mejaddam AY,McGovern PE,Hwang G,Han J,Peranteau WH,Partridge EA,Davey MG,Flake AW

    更新日期:2019-12-01 00:00:00

  • Influence of hydration state on plasma volume changes during ultrafiltration.

    abstract::To assess the influence of hydration state on plasma volume (PV) changes during ultrafiltration, 11 clinically normhydrated patients on maintenance hemodialysis were studied intraindividually during 2 hydration states differing by 2-15% of lean body mass (LBM). Plasma volume was measured continually during a 15-min ul...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1995.tb02352.x

    authors: Wizemann V,Leibinger A,Mueller K,Nilson A

    更新日期:1995-05-01 00:00:00

  • Nitric oxide added to the sweep gas of the oxygenator during cardiopulmonary bypass in infants: A pilot randomized controlled trial.

    abstract::Our objective was to assess the effect of nitric oxide added to the sweep gas of the oxygenator during cardiopulmonary bypass (CPB) in infants on platelet count, platelet function, clinical outcomes, and safety. A randomized, double-blinded, placebo-controlled clinical trial in infants less than a year of age undergoi...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.13788

    authors: Niebler RA,Chiang-Ching H,Daley K,Janecke R,Jobe SM,Mitchell ME,Varner C,Woods K,Scott JP

    更新日期:2020-08-01 00:00:00

  • Microbial and endotoxin contamination in water and dialysate in the central United States.

    abstract::The purified water supplies and randomly selected dialysates of 51 chronic and acute dialysis centers in the central United States were surveyed to assess the relative risks to dialysis patients from microbial and endotoxin contamination. A culture medium more sensitive than those generally employed in routine quality...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.1990.tb01601.x

    authors: Klein E,Pass T,Harding GB,Wright R,Million C

    更新日期:1990-04-01 00:00:00

  • Building a Better Neonatal Extracorporeal Life Support Circuit: Comparison of Hemodynamic Performance and Gaseous Microemboli Handling in Different Pump and Oxygenator Technologies.

    abstract::Neurologic complications during neonatal extracorporeal life support (ECLS) are associated with significant morbidity and mortality. Gaseous microemboli (GME) in the ECLS circuit may be a possible cause. Advances in neonatal circuitry may improve hemodynamic performance and GME handling leading to reduction in patient...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/aor.12908

    authors: Glass K,Trivedi P,Wang S,Woitas K,Kunselman AR,Ündar A

    更新日期:2017-04-01 00:00:00

  • Cytotoxicity of two novel formulations of calcium phosphate cements: a comparative in vitro study.

    abstract::The purpose was to evaluate the cytotoxicity of two novel formulations (alpha and beta) of calcium phosphate cements. Positive control, represented by a commercial hydroxyapatite cement, and negative control were included for comparative purposes. A continuous lineage of fibroblastic cells was used, and the effect of ...

    journal_title:Artificial organs

    pub_type: 杂志文章

    doi:10.1111/j.1525-1594.2005.29022.x

    authors: Rossa C Jr,Marcantonio E Jr,Santos LA,Boschi AO,Raddi MS

    更新日期:2005-02-01 00:00:00