Abstract:
PURPOSE:For stereotactic radiosurgery (SRS), accurate evaluation of dose-volume metrics for small structures is necessary. The purpose of this study was to compare the DVH metric capabilities of five commercially available SRS DVH analysis tools (Eclipse, Elements, Raystation, MIM, and Velocity). METHODS:DICOM RTdose and RTstructure set files created using MATLAB were imported and evaluated in each of the tools. Each structure set consisted of 50 randomly placed spherical targets. The dose distributions were created on a 1 mm grid using an analytic model such that the dose-volume metrics of the spheres were known. Structure sets were created for 3, 5, 7, 10, 15, and 20 mm diameter spheres. The reported structure volume, V100% [cc], and V50% [cc], and the RTOG conformity index and Paddick Gradient Index, were compared with the analytical values. RESULTS:The average difference and range across all evaluated target sizes for the reported structure volume was -4.73%[-33.2,0.2], 0.11%[-10.9, 9.5], -0.39%[-12.1, 7.0], -2.24%[-21.0, 1.3], and 1.15%[-15.1,0.8], for TPS-A through TPS-E respectively. The average difference and range for the V100%[cc] (V20Gy[cc]) was -0.4[-24.5,9.8], -2.73[-23.6, 1.1], -3.01[-23.6, 0.6], -3.79[-27.3, 1.3], and 0.26[-6.1,2.6] for TPS-A through TPS-E respectively. For V50%[cc](V10Gy[cc]) in TPS-A through TPS-E the average and ranger were -0.05[-0.8,0.4], -0.18[-1.2, 0.5], -0.44[-1.4, 0.3], -0.26[-1.8, 2.6], and 0.09[-1.4,2.7]. CONCLUSION:This study expanded on the previously published literature to quantitatively compare the DVH analysis capabilities of software commonly used for SRS plan evaluation and provides freely available and downloadable analytically derived set of ground truth DICOM dose and structure files for the use of radiotherapy clinics. The differences between systems highlight the need for standardization and/or transparency between systems, especially when evaluating plan quality for multi-institutional clinical trials.
journal_name
Med Physjournal_title
Medical physicsauthors
Stanley DN,Covington E,Liu H,Alexandrian A,Cardan R,Bridges D,Thomas EM,Fiveash JB,Popple Rdoi
10.1002/mp.14645subject
Has Abstractpub_date
2020-12-09 00:00:00eissn
0094-2405issn
2473-4209pub_type
杂志文章相关文献
MEDICAL PHYSICS文献大全abstract:PURPOSE:In this work, the fabrication, operation, and evaluation of a probe-format graphite calorimeter - herein referred to as Aerrow - as an absolute clinical dosimeter of high-energy photon beams while in the presence of a B = 1.5 T magnetic field is described. Comparable to a cylindrical ionization chamber (IC) in ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13968
更新日期:2020-03-01 00:00:00
abstract::The goal of this study was to develop a technique to distinguish benign and malignant breast lesions in secondarily digitized mammograms. A set of 51 mammograms (two views/patient) containing lesions of known pathology were evaluated using six different morphological descriptors: circularity, mu R/sigma R (where mu R ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.597707
更新日期:1996-08-01 00:00:00
abstract:PURPOSE:To develop an individually optimized contrast-enhanced (CE) 4D-computed tomography (CT) for radiotherapy simulation in pancreatic ductal adenocarcinomas (PDA). METHODS:Ten PDA patients were enrolled. Each underwent three CT scans: a 4D-CT immediately following a CE 3D-CT and an individually optimized CE 4D-CT ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4963213
更新日期:2016-10-01 00:00:00
abstract:PURPOSE:Two types of tumor hypoxia most likely exist in human cancers: Chronic hypoxia due to the paucity of blood capillaries and acute hypoxia due to temporary shutdoWn of microvasculatures or fluctuation in the red cell flux. In a recent hypoxia imaging study using 18F-FMISO PET, the authors observed variation in tr...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3213092
更新日期:2009-10-01 00:00:00
abstract::The purpose of this study was to implement the Monte Carlo method for clinical radiotherapy dose calculations. We used the EGS4/BEAM code to obtain the phase-space data for 6-20 MeV electron beams and 4, 6, and 15 MV photon beams for Varian Clinac 1800, 2100C, and 2300CD accelerators. A multiple-source model was used ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.598729
更新日期:1999-10-01 00:00:00
abstract::The validity of conclusions drawn from multi-institutional radiotherapy studies is dependent upon the degree to which participating hospitals adhere to study requirements. A major effort undertaken by the Quality Assurance Review Center (QARC) assesses whether or not patient data which are deemed evaluable actually co...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.595221
更新日期:1982-03-01 00:00:00
abstract:PURPOSE:Compton camera imaging (CCI) systems are currently under investigation for radiotherapy dose reconstruction and verification. The ability of such a system to provide real-time images during dose delivery will be limited by the computational speed of the image reconstruction algorithm. In this work, the authors ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3519873
更新日期:2011-01-01 00:00:00
abstract:PURPOSE:To determine the accuracy and test-retest repeatability of fast radiofrequency (RF) transmit measurement approaches used in Dynamic Contrast Enhanced Magnetic Resonance Imaging (DCE-MRI). Spatial variation in the transmitted RF field introduces bias and increased variance in quantitative DCE-MRI metrics includi...
journal_title:Medical physics
pub_type: 杂志文章,多中心研究
doi:10.1002/mp.13518
更新日期:2019-06-01 00:00:00
abstract:PURPOSE:Previous studies have shown atlas-based segmentation using a single best matched (SBM) atlas subject can significantly reduce contouring time. A new multi-atlas approach has been shown to provide greater accuracy than SBM for cancer of the head and neck. The goal of this study was to evaluate the multi-atlas te...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4734942
更新日期:2012-06-01 00:00:00
abstract:PURPOSE:To present a system for robotic 4D ultrasound (US) imaging concurrent with radiotherapy beam delivery and estimate the proportion of liver stereotactic ablative body radiotherapy (SABR) cases in which robotic US image guidance can be deployed without interfering with clinically used VMAT beam configurations. M...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4964454
更新日期:2016-11-01 00:00:00
abstract:PURPOSE:The authors combine the registration of 2D beam's eye view (BEV) images and 3D planning computed tomography (CT) images, with relative, markerless tumor tracking to provide automatic absolute tracking of physician defined volumes such as the gross tumor volume (GTV). METHODS:During treatment of lung SBRT cases...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4900603
更新日期:2014-12-01 00:00:00
abstract::Convolution/superposition software has been used to produce a library of photon pencil beam dose matrices. This library of pencil beams is designed to serve as a tool for both education and investigation in the field of radiotherapy optimization. The elegance of this pencil beam model stems from its cylindrical symmet...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.598615
更新日期:1999-07-01 00:00:00
abstract::Volumetric modulated arc therapy (VMAT) has recently been used to improve the dose distribution and efficiency of treatment delivery over the standard intensity-modulated radiotherapy (IMRT) technique. This study compares the dosimetry between RapidArc plan and standard IMRT plan for head and neck cancer. Three head a...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4740174
更新日期:2012-07-01 00:00:00
abstract:PURPOSE:An original radionuclide calibrator method for activity determination is presented. The method could be used for intercomparison surveys for short half-life radioactive sources used in Nuclear Medicine, such as(99m)Tc or most positron emission tomography radiopharmaceuticals. METHODS:By evaluation of the resul...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4884035
更新日期:2014-07-01 00:00:00
abstract:PURPOSE:This article presents a new reconstruction method for 3D imaging using a multiple 360 degrees circular orbit cone beam CT system, specifically a way to combine 3D volumes reconstructed with each orbit. The main goal is to improve the noise performance in the combined image while avoiding cone beam artifacts. M...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3484058
更新日期:2010-10-01 00:00:00
abstract::In this study we address the problem of predicting the position of a moving lung tumor during respiration on the basis of external breathing signals--a technique used for beam gating, tracking, and other dynamic motion management techniques in radiation therapy. We demonstrate the use of neural network filters to corr...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2134958
更新日期:2005-12-01 00:00:00
abstract:PURPOSE:The authors have developed a quantitative calibration method for a multileaf collimator (MLC) which measures individual leaf positions relative to the MLC backup jaw on an Elekta Synergy linear accelerator. METHODS:The method utilizes a commercially available two-axis detector array (Profiler 2; Sun Nuclear Co...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.3218767
更新日期:2009-10-01 00:00:00
abstract:PURPOSE:To develop a method of using two-dimensional (2D) magnetic resonance thermometry, and three-dimensional (3D) Gaussian modeling to predict the volume, shape, and location of 1 day postoperative T1w high-intensity focused ultrasound lesions in medication refractory tremor patients; thereby facilitating a better c...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.13868
更新日期:2019-12-01 00:00:00
abstract:PURPOSE:To validate a delivery quality assurance (DQA) protocol for tomotherapy based on the measurement of the leaf open times (LOTs). In addition, to show the correlation between the mean relative LOT discrepancy and the dose deviation in the planning target volume (PTV). MATERIALS AND METHODS:We used a LOT measurem...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1002/mp.14348
更新日期:2020-09-01 00:00:00
abstract:PURPOSE:The application of compressed sensing (CS) technology in magnetic resonance imaging (MRI) is to accelerate the MRI scan speed by incoherent undersampling of k-space data and nonlinear iterative reconstruction of MRI images. This paper generalizes the existing rosette trajectories to configure the sampling patte...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4928152
更新日期:2015-09-01 00:00:00
abstract::Positron emission tomography (PET) has shown an increase in both sensitivity and specificity over computed tomography (CT) in lung cancer. However, motion artifacts in the 18F fluorodioxydoglucose (FDG) PET images caused by respiration persists to be an important factor in degrading PET image quality and quantificatio...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1448824
更新日期:2002-03-01 00:00:00
abstract::The purpose of this study was to investigate the effect of z overscanning on eye lens dose and effective dose received by pediatric patients undergoing head CT examinations. A pediatric patient study was carried out to obtain the exposure parameters and data regarding the eye lens position with respect to imaged volum...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2207212
更新日期:2006-07-01 00:00:00
abstract:PURPOSE:To introduce video surface imaging guidance in synchronization with 4D cone-beam CT (CBCT) scans, and in combination with respiration- gated or target-tracked dose delivery to treat mobile tumors, without collaterally damaging nearby critical structures. METHODS:The approach uses the concept that the integral ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4734688
更新日期:2012-06-01 00:00:00
abstract::Flat-panel (FP) based digital radiography systems have recently been introduced as a new and improved digital radiography technology; it is important to evaluate and compare this new technology with currently widely used conventional screen/film (SF) and computed radiography (CR) techniques. In this study, the low-con...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.1408620
更新日期:2001-11-01 00:00:00
abstract::A 4-W, laser-based system for the positional readout of thermoluminescent (TL) plates was utilized to obtain dose profiles for a 60Co radiotherapy beam. Experimentally obtained glow curves for LiF resulting from the rapid heating rates characteristic of laser heating agreed well with results predicted using a first-or...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.596580
更新日期:1990-05-01 00:00:00
abstract:PURPOSE:To obtain on-treatment volumetric patient anatomy during respiratory gated volumetric modulated arc therapy (VMAT). METHODS:On-board imaging device integrated with Linacs offers a viable tool for obtaining patient anatomy during radiation treatment delivery. In this study, the authors acquired beam-level kV im...
journal_title:Medical physics
pub_type: 信件
doi:10.1118/1.4794925
更新日期:2013-04-01 00:00:00
abstract::Proton beams offer several advantages over conventional radiation techniques for treating cancer and other diseases. These advantages might be negated if the leakage and scatter radiation from the beamline and patient are too large. Although the leakage and scatter radiation for the double scattering proton beamlines ...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.2805086
更新日期:2008-01-01 00:00:00
abstract:PURPOSE:Severe artifacts in kilovoltage-CT simulation images caused by large metallic implants can significantly degrade the conspicuity and apparent CT Hounsfield number of targets and anatomic structures, jeopardize the confidence of anatomical segmentation, and introduce inaccuracies into the radiation therapy treat...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4762814
更新日期:2012-12-01 00:00:00
abstract:PURPOSE:The purpose of this technical note is to introduce variance component analysis to the estimation of systematic and random components in setup error of radiotherapy. METHODS:Balanced data according to the one-factor random effect model were assumed. RESULTS:Analysis-of-variance (anova)-based computation was ap...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4961397
更新日期:2016-09-01 00:00:00
abstract:PURPOSE:The authors describe the integration of optical imaging with a targeted small animal irradiator device, focusing on design, instrumentation, 2D to 3D image registration, 2D targeting, and the accuracy of recovering and mapping the optical signal to a 3D surface generated from the cone-beam computed tomography (...
journal_title:Medical physics
pub_type: 杂志文章
doi:10.1118/1.4894730
更新日期:2014-10-01 00:00:00